K242672 · Imperative Care, Inc. · NRY · Jan 14, 2025 · Cardiovascular
Device Facts
Record ID
K242672
Device Name
Zoom System
Applicant
Imperative Care, Inc.
Product Code
NRY · Cardiovascular
Decision Date
Jan 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Zoom System, when used with the Zoom Aspiration Pump (or equivalent vacuum pump), is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of last known well. Patients who are ineligible for intravenous thrombolytic drug therapy or who have not responded to thrombolytic drug therapy are candidates for treatment. The Zoom Aspiration Tubing and the Zoom POD Aspiration Tubing are intended to connect the Zoom (71, 55, 45, 35) Catheter and the TracStar LDP Large Distal Platform, the Zoom 88 Large Distal Platform, or the Zoom 88 Large Distal Platform Support to the Zoom Canister of the Zoom Aspiration Pump (or equivalent vacuum pump) and to allow the user to control the fluid flow.
Device Story
Zoom System is a mechanical neurothrombectomy device for acute ischemic stroke. Components include Zoom reperfusion catheters (71, 55, 45, 35), Large Distal Platform (LDP) catheters (Zoom 88 LDP, Zoom 88 LDP Support, TracStar LDP), aspiration tubing (with or without POD clot filter), and aspiration pump. System uses vacuum aspiration to remove thrombus from intracranial arteries. Operated by physicians in clinical/hospital settings. Catheters are single-lumen, braid/coil-reinforced, variable stiffness with hydrophilic coating. Physician navigates catheters to occlusion site via guidewire; applies vacuum via pump to aspirate clot. Output is restored blood flow (reperfusion). Benefits include rapid revascularization, improved clinical outcomes, and reduced stroke-related disability.
Clinical Evidence
Prospective, multi-center, single-arm pivotal trial (Imperative Trial, NCT04129125) of 260 subjects. Primary effectiveness: mTICI ≥ 2b in ≤ 3 passes achieved in 84% (177/211) of analyzed cohort (p<0.001 vs PG). Primary safety: sICH rate 0.9% (2/211), meeting PG ≤ 6.0%. Secondary endpoints: 90-day all-cause mortality 12.3%, device-related SAE 1.4%, procedure-related vessel injury 0.5%.
Technological Characteristics
Catheters: single-lumen, braid/coil-reinforced, variable stiffness, radiopaque markers, hydrophilic coating. Materials: medical-grade plastics/metals. Sterilization: Ethylene Oxide (EO), SAL 10^-6. Connectivity: manual luer connections to aspiration pump. Aspiration tubing: hollow polymer tube with external pinch clamp; Zoom POD includes in-line clot filter. Compliant with ISO 10555-1 (catheter performance) and ISO 11135 (sterilization).
Indications for Use
Indicated for revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (ICA, MCA M1/M2, basilar, vertebral arteries) within 8 hours of last known well. Candidates include patients ineligible for or non-responsive to IV thrombolytic therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
January 14, 2025
Imperative Care, Inc. Teri Nguven Lead Regulatory Affairs Specialist 1359 Dell Avenue Campbell, California 95008
Re: K242672
Trade/Device Name: Zoom System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: December 16, 2024 Received: December 16, 2024
Dear Teri Nguyen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242672
Device Name Zoom System
### Indications for Use (Describe)
The Zoom System, when used with the Zoom Aspiration Pump (or equivalent vacuum pump), is indicated for use in the revascularization of patients with acute secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of last known well.
Patients who are ineligible for intravenous thrombolytic drug therapy or who have not responded to thrombolytic drug therapy are candidates for treatment.
The Zoom Aspiration Tubing and the Zoom POD Aspiration Tubing are intended to connect the Zoom (71, 55, 45, 35) Catheter and the TracStar LDP Large Distal Platform, the Zoom 88 Large Distal Platform, or the Zoom 88 Large Distal Platform Support to the Zoom Canister of the Zoom Aspiration Pump (or equivalent vacuum pump) and to allow the user to control the fluid flow.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
Imperative Care, Inc.
1359 Dell Avenue Campbell, CA 95008
Teri Nguyen
408-833-4415
January 13, 2025
#### I. SUBMITTER
Submitter's Name: Address:
Contact Person: Telephone: Email: Date of Preparation:
#### II. DEVICE
Proprietary Name: Common/Usual Name: Classification Name: Regulatory Class: Product Code: Regulation
Zoom™ System Catheter, Thrombus Retriever Percutaneous Catheter II NRY 21 CFR 870.1250
vtnguyen@imperativecare.com
### III. PREDICATE DEVICE
Proprietary Name:
Common/Usual Name: Classification Name: Product Code: Regulation: Manufacturer: 510(k) #:
Proprietary Name:
Common/Usual Name: Classification Name: Product Code: Regulation: Manufacturer: 510(k) #:
ZOOM 71 Reperfusion Catheter; ZOOM Aspiration Tubing Catheter, Thrombus Retriever Percutaneous Catheter NRY 21 CFR 870.1250 Imperative Care, Inc. K211476
ZOOM (71, 55, 45, 35) Reperfusion Catheters; ZOOM Aspiration Tubing Catheter, Thrombus Retriever Percutaneous Catheter NRY 21 CFR 870.1250 Imperative Care, Inc. K210996
{5}------------------------------------------------
#### IV. REFERENCE DEVICE
Proprietary Names:
Common/Usual Name: Classification Name: Product Codes: Regulation: Manufacturer: 510(k) #:
TracStar™ LDP Large Distal Platform™; Zoom™ 88 Large Distal Platform™: Zoom™ 88 Large Distal Platform™ Support Catheter, Percutaneous, Neurovasculature Percutaneous Catheter OJP, DOY 21 CFR 870.1250 Imperative Care, Inc. K240948
### V. DEVICE DESCRIPTION
The Imperative Care Zoom System consists of the following devices:
- Zoom Catheters .
- o Zoom™ (71, 55, 45, 35) Catheters
- Large Distal Platform Catheters (LDP Catheters) ●
- Zoom™ 88 Large Distal Platform™ (Zoom 88 LDP) o
- Zoom™ 88 Large Distal Platform™ Support (Zoom 88 LDP Support) O
- TracStar™ LDP Large Distal Platform™ (TracStar LDP) O
- Zoom Aspiration Tubing and Zoom POD Aspiration Tubing
- Zoom Aspiration Pump ●
The Zoom Catheters and the LDP Catheters are intended to be used as a system in conjunction with the Zoom Aspiration Tubing or Zoom POD Aspiration Tubing and the Zoom Aspiration Pump (or equivalent vacuum pump) to aspirate thrombus in patients with acute ischemic stroke.
The Zoom Catheters and LDP Catheters are single lumen, braid and coil reinforced, variable stiffness catheters with a radiopaque marker and a lubricious hydrophilic coating on the distal portion of the catheter. The catheters have a luer hub on the proximal end.
Dimensions for each catheter are included on the individual device label. The Zoom Catheters are compatible with 0.014" - 0.018" guidewires. The LDP Catheters are compatible with 0.038" or smaller guidewires. An additional support catheter may be used
{6}------------------------------------------------
to assist in accessing the target vasculature. The Zoom 45, 55, and 71 Catheters are compatible with 6F guide sheaths with a minimum inner diameter of 0.088". The Zoom 35 Catheter is compatible with 5F guide sheaths with a minimum inner diameter of 0.068". The LDP Catheters are compatible with 8F or greater introducer sheaths with a minimum inner diameter of 0.115". The Zoom 71 Catheter is compatible with 5F microcatheters or intermediate catheters with a maximum outer diameter of 0.065". The Zoom 45 and 55 Catheters are compatible with 2.4F microcatheters or intermediate catheters with a maximum outer diameter of 0.031". The Zoom 35 Catheter is compatible with 1.4F microcatheters or intermediate catheters with a maximum outer diameter of 0.018". The LDP Catheters are compatible with 6F microcatheters or intermediate catheters with a maximum outer diameter of 0.083".
All catheters are packaged with an accessory rotating hemostasis valve (RHV). The RHV is intended to be attached to the proximal hub of the catheter and used to control hemostasis during use with other devices.
The Zoom Aspiration Tubing and the Zoom POD Aspiration Tubing (Zoom POD) are comprised of a hollow cylindrical tube which is bonded to a standard luer fitting that connects to the Zoom Catheter and the LDP Catheter and a slip fit connector that connects to the canister on the aspiration pump. The Zoom Aspiration Tubing and Zoom POD Aspiration Tubing are made of common medical grade polymers.
In addition to the accessories discussed above, the adjunctive devices and supplies listed below could be used with the Zoom System.
- Guidewires .
- Support/Diagnostic Catheters ●
- Introducer Sheaths .
- Aspiration Pump* .
- Capable of achieving pressure between -20inHg to max vacuum (-29.9 inHg)
- 0 Airflow rating of 0 23 LPM
{7}------------------------------------------------
## 0 IEC 60601-1 Compliant
* Imperative Care offers the Zoom Aspiration Pump which meets the indicated criteria.
### INDICATIONS FOR USE VI.
The Zoom System, when used with the Zoom Aspiration Pump (or equivalent vacuum pump), is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of last known well.
Patients who are ineligible for intravenous thrombolytic drug therapy or who have not responded to thrombolytic drug therapy are candidates for treatment.
The Zoom Aspiration Tubing and the Zoom POD Aspiration Tubing are intended to connect the Zoom (71, 55, 45, 35) Catheter and the TracStar LDP Large Distal Platform, the Zoom 88 Large Distal Platform, or the Zoom 88 Large Distal Platform Support to the Zoom Canister of the Zoom Aspiration Pump (or equivalent vacuum pump) and to allow the user to control the fluid flow.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH VII. PREDICATE AND REFERENCE DEVICES
The predicate devices are the Imperative Care ZOOM (71, 55, 45, 35) Reperfusion Catheters (ZOOM Reperfusion Catheters) and ZOOM Aspiration Tubing cleared under K211476 and K210996. The predicate and subject devices share the same intended use and basic technological characteristics, and equivalent performance characteristics, demonstrated through bench, and clinical testing. Table 1 provides a comparison of the subject Zoom System Catheters and the predicate ZOOM Reperfusion Catheters. Table 2 provides a comparison of the subject Zoom Aspiration Tubing and the Zoom POD Aspiration Tubing and the predicate ZOOM Aspiration Tubing.
{8}------------------------------------------------
| Device<br>Attribute | Predicate Device | Subject Device | Reference Device |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Product<br>Classification | Class II<br>Product Code: NRY<br>21 CFR 870.1250 | Same as Predicate | Class II<br>Product Code: QJP, DQY<br>21 CFR 870.1250 |
| Product Name | ZOOM 71 Reperfusion Catheter<br>ZOOM 55 Reperfusion Catheter<br>ZOOM 45 Reperfusion Catheter<br>ZOOM 35 Reperfusion Catheter | Zoom 71 Catheter<br>Zoom 55 Catheter<br>Zoom 45 Catheter<br>Zoom 35 Catheter<br>TracStar LDP<br>Zoom 88 LDP<br>Zoom 88 LDP Support | TracStar LDP<br>Zoom 88 LDP<br>Zoom 88 LDP Support |
| 510(k) Number | K211476, K210996 | K242672 | K240948 |
| Manufacturer | Imperative Care, Inc. | Same | Same |
| Indications for<br>Use | The ZOOM Reperfusion Catheters,<br>with the ZOOM Aspiration Tubing<br>and ZOOM Aspiration Pump (or<br>equivalent vacuum pump), are<br>indicated for use in the<br>revascularization of patients with<br>acute ischemic stroke secondary to<br>intracranial large vessel occlusive<br>disease (within the internal carotid,<br>middle cerebral – M1 and M2<br>segments, basilar, and vertebral<br>arteries) within 8 hours of symptom<br>onset.<br>Patients who are ineligible for<br>intravenous tissue plasminogen<br>activator (IV t-PA) or who fail IV t-<br>PA therapy are candidates for<br>treatment. | The Zoom System, when used with the<br>Zoom Aspiration Pump (or equivalent<br>vacuum pump), is indicated for use in the<br>revascularization of patients with acute<br>ischemic stroke secondary to intracranial<br>large vessel occlusive disease (within the<br>internal carotid, middle cerebral - M1 and<br>M2 segments, basilar, and vertebral arteries)<br>within 8 hours of last known well.<br>Patients who are ineligible for intravenous<br>thrombolytic drug therapy or who have not<br>responded to thrombolytic drug therapy are<br>candidates for treatment. | The TracStar LDP Large Distal Platform is<br>indicated for the introduction of<br>interventional devices into the peripheral,<br>coronary, and neuro vasculature.<br>The Zoom 88 Large Distal Platform and<br>Zoom 88 Large Distal Platform Support are<br>indicated for the introduction of<br>interventional devices into the peripheral,<br>coronary, and neuro vasculature. |
| Device<br>Attribute | Predicate Device | Subject Device | Reference Device |
| Condition<br>Supplied | Sterile, Single Use Only | Same | Same |
| Sterilization<br>Method | Ethylene Oxide (EO), SAL 10-6<br>One (1) EO Cycle | Same | Same<br>Up to Two (2) EO Cycles |
| Inner Diameter<br>(Distal) | 0.035"-0.071" | Same as Predicate and Reference Devices | 0.088" |
| Outer Diameter<br>(Distal) | 0.051"-0.083" | Same as Predicate and Reference Devices | 0.106" |
| Inner Diameter<br>(Proximal) | 0.047"-0.071" | Same as Predicate and Reference Devices | 0.088" |
| Outer Diameter<br>(Proximal) | 0.061"-0.083" | Same as Predicate and Reference Devices | 0.110" |
| Effective<br>Working Length | 137 cm – 160 cm | Same as Predicate and Reference Devices | 80 cm – 110 cm |
| Tip Design | Beveled distal edge, soft, flexible,<br>atraumatic tip | Same | Same |
| Materials | Commonly used medical grade<br>plastics and metals with hydrophilic<br>coating | Same | Same |
| Packaged<br>Accessories | RHV | Same | Same |
| Packaging<br>Configuration | The catheter is placed in a protective<br>polyethylene tube and then mounted,<br>along with the RHV accessory, onto a<br>polyethylene packaging card.<br><br>The packaging card is inserted into a<br>Tyvek® pouch which is then sealed.<br><br>The sealed pouch and Instructions for<br>Use (IFU) are placed in a shelf carton | Same | Same |
| Device<br>Attribute | Predicate Device | Subject Device | Reference Device |
| Aspiration<br>Pump | Zoom Aspiration Pump | Same as Predicate Device | N/A - Not indicated for reperfusion |
| Shelf Life | 12 Months (1 Year) | Same as Reference Device | 3 Years |
# Table 1: Comparison of Technological Characteristics with Predicate and Reference Devices – Zoom System Catheters
{9}------------------------------------------------
{10}------------------------------------------------
# Table 2: Comparison of Technological Characteristics with Predicate Device – Zoom and Zoom POD Aspiration Tubing
| Device Attribute | Predicate Device | Subject Device |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | ZOOM Aspiration Tubing | Zoom Aspiration Tubing<br>Zoom POD Aspiration Tubing |
| Indications for Use | The ZOOM Aspiration Tubing is intended to connect the<br>ZOOM Reperfusion Catheter to the ZOOM Canister of the<br>ZOOM Aspiration Pump and to allow the user to control<br>the fluid flow. | The Zoom Aspiration Tubing and the Zoom POD Aspiration<br>Tubing are intended to connect the Zoom (71, 55, 45, 35) Catheter<br>and the TracStar LDP Large Distal Platform, the Zoom 88 Large<br>Distal Platform, or the Zoom 88 Large Distal Platform Support to<br>the Zoom Canister of the Zoom Aspiration Pump (or equivalent<br>vacuum pump) and to allow the user to control the fluid flow. |
| Condition Supplied | Sterile and Single Use | Same |
| Sterilization Method | EO, SAL 10-6 | Same |
| Tubing Inner Diameter<br>(ID) | 0.110" | Same |
| Working Length | 104'' | Zoom Aspiration Tubing: 104"<br>Zoom POD Aspiration Tubing: 117" |
| Flow Control<br>Mechanism | External Pinch Clamp | Same |
| Clot Filter | N/A – does not have a clot filter | Zoom Aspiration Tubing: N/A<br>Zoom POD Aspiration Tubing: in-line clot filter |
| Packaging<br>Configuration | The Zoom Aspiration Tubing is coiled and placed inside a<br>labeled pouch and sealed. The sealed pouch and IFU are<br>placed in a labeled shelf carton box. | Same |
| Shelf Life | 43 Months | Same |
{11}------------------------------------------------
#### VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
The tissue contacting materials, including raw materials, manufacturing processes and manufacturing aids, of the subject Zoom Catheters are similar to the predicate ZOOM Reperfusion Catheters (K211476 and K210996). Original testing on the predicate devices apply to the subject Zoom Catheters and additional biocompatibility testing was not required.
The subject LDP Catheters are identical to the reference LDP Catheters (K240948). There are no changes to the patient contacting materials, including raw materials, manufacturing processes, and manufacturing aids, of the subject LDP Catheters. Therefore, the original testing on the reference devices apply to the subject LDP Catheters and additional biocompatibility testing was not required.
The changes made to the Zoom Aspiration Tubing to create the Zoom POD Aspiration Tubing (Zoom POD) were evaluated for the Zoom POD in accordance with the FDA guidance document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"," and International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process." Supporting biocompatibility testing (cytotoxicity, sensitization, and irritation / intracutaneous reactivity) was completed to confirm that the addition of the in-line clot filter component (collection pod) did not result in any significant changes to the biocompatibility of the Zoom POD Aspiration Tubing.
{12}------------------------------------------------
# Non-clinical Bench Testing
A summary of the evaluated design and performance specifications is presented in Table 3 and Table 4. Performance specifications and test methods were based primarily on catheter performance standard ISO 10555-1. The test results were reviewed and found to demonstrate that the differences between the subject devices and predicate or reference devices do not significantly impact any performance parameters that would negatively affect the safety or effectiveness of the subject devices.
| Test Attribute | Specification | Results |
|----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Delivery, Compatibility,<br>and Retraction<br>(Trackability) | The catheter shall be able to be delivered,<br>deployed, and retracted per the IFU within a<br>simulated neurovascular model without<br>incurring any damage to the catheter. | Pass |
| Tip Flexibility | The catheter distal tip flexibility shall be<br>comparable to the predicate. | Pass |
| Visual Inspection | The catheter shall meet visual inspection<br>criteria.<br>The printing on the strain relief must be legible. | Pass |
| Dimensional (Distal ID,<br>Proximal ID, Distal Outer<br>Diameter (OD), Proximal<br>OD) | All defined catheter dimensions are within the<br>specified tolerances. | Pass |
| Catheter Bond Strength | The catheter shall have sufficient bond strength<br>to remain intact throughout a procedure. | Pass |
| Freedom from Leakage -<br>Positive Pressure | The catheter must remain leak free under<br>specified test conditions. | Pass |
| Freedom from Leakage -<br>Negative Pressure | The catheter must remain leak free under<br>specified test conditions. | Pass |
| Dynamic Burst | The catheter must withstand pressures under<br>dynamic flow conditions. | Pass |
| Static Burst | The catheter shall meet criteria for static burst<br>pressure testing. | Pass |
| Test Attribute | Specification | Results |
| Catheter Torque Strength | With the catheter tip constrained from<br>movement, the proximal end was rotated until<br>failure. The catheter shall not be damaged when<br>rotated at least two (2) full rotations (720<br>degrees). | Pass |
| Kink Resistance | There shall be no kinking of the catheter shaft<br>around respective clinically relevant minimum<br>bend radii in distal tip, medial, and proximal<br>locations. | Pass |
| Flexibility | The catheters need to have acceptable flexure<br>values for tracking in the vasculature. | Pass |
| Proximal Shaft Stiffness | The stiffness of the proximal shaft was<br>evaluated to ensure comparable stiffness to the<br>predicate. | Pass |
| Delivery Force | The catheters shall not be too stiff or require<br>excessive force to safely navigate and track to<br>the target neurovasculature. | Pass |
| Compatibility with other<br>Devices (External) | The catheters shall be able to be delivered<br>through the minimum introducer sheath or guide<br>catheter size indicated in the product labeling. | Pass |
| Guidewire Compatibility | The catheters shall be able to be delivered over<br>the maximum size guidewire indicated in the<br>product labeling. | Pass |
| Microcatheter/Intermediate<br>Catheter Compatibility | The catheters shall be able to accommodate a<br>microcatheter/intermediate catheter up to the<br>maximum size indicated in the product labeling. | Pass |
| Luer Compatibility | Devices and RHV accessory shall be compatible<br>with standard syringe luer fittings per ISO<br>80369-7. | Pass |
| Accessory Compatibility | Devices shall be compatible with an RHV. | Pass |
| Coating - Particulate | The amount of particulate matter generated<br>during simulated use testing shall be determined<br>and compared to competitive products and<br>techniques. | Pass |
| Coating - Lubricity,<br>Durability and Integrity | Coating must be lubricious with a specified<br>average pull force. There were no coating<br>anomalies or significant wear observed post<br>simulated use. | Pass |
| Clot Retrieval | The catheters will be able to be delivered in a<br>clinically relevant vascular model and<br>effectively aspirate clots when used with the<br>aspiration pump and aspiration tubing set when<br>placed at appropriate locations based on the size<br>of the catheter. | Pass |
| Test Attribute | Specification | Results |
| Lumen Integrity | The catheter lumen shall not collapse under<br>vacuum. | Pass |
| Vacuum Force at Catheter<br>Tip | The vacuum force at the tip of the catheter<br>should be comparable to the vacuum force at the<br>tip of the predicate devices. | Pass |
| Flowrate - Positive<br>(forward) Pressure | The catheter lumen shall allow for a minimum<br>flowrate comparable to comparator products. | Pass |
| Flowrate - Negative<br>(Vacuum) Pressure | The minimum flowrate under a vacuum shall<br>be similar to or greater than comparator<br>devices. | Pass |
Table 3: Tests and Performance Specifications for Zoom System Catheters
{13}------------------------------------------------
{14}------------------------------------------------
# Table 4: Tests and Performance Specifications for Zoom System Aspiration Tubing
| Test Attribute | Specification | Results |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Vacuum Force at Catheter<br>Tip | The vacuum force delivered by the aspiration<br>tubing to the tip of the catheter should be<br>comparable to the vacuum force delivered by<br>the predicate aspiration tubing. | Pass |
| Connector Compatibility | The aspiration tubing connectors shall securely<br>connect to the pump canister lid and standard<br>luer fittings. | Pass |
| Lumen Collapse Test | The tubing lumen shall not collapse under<br>vacuum. | Pass |
| Flow Control<br>Functionality | The flow control mechanism shall allow users to<br>start and stop flow multiple times when the<br>connected pump is running at maximum<br>vacuum. | Pass |
| Freedom From Leakage | The vacuum pressure delivered at the tip of the<br>aspiration tubing shall be consistent with the<br>pressure generated by the pump. | Pass |
| Tensile Strength | The bonds between the tubing and connectors<br>shall be sufficiently strong to ensure the tubing<br>remains intact during use. | Pass |
| Clot Filter Functionality<br>(Zoom POD only) | The clot filter should be able to be opened and<br>closed without causing leak. | Pass |
| Aspiration Flowrate | Minimum aspiration catheter flowrate shall be<br>achieved with the Zoom Aspiration Tubing and<br>Zoom POD Aspiration Tubing. | Pass |
{15}------------------------------------------------
## Sterilization
The Zoom System devices (Zoom Catheters, LDP Catheters, Zoom Aspiration Tubing, Zoom POD Aspiration Tubing) are sterilized using validated EO processes with a sterility assurance level (SAL) of 106. The sterilization processes were validated per the overkill method in accordance with ISO 11135, "Sterilization of Health-Care Products - Ethylene Oxide - Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices."
## Shelf Life and Packaging
Accelerated aging testing based on ASTM F1980 was previously conducted to verify packaged device performance of the subject LDP Catheters and Zoom Aspiration Tubing. Accelerated aging testing based on ASTM F1980 was conducted to verify packaged device performance of the subject Zoom Catheters and Zoom POD Aspiration Tubing. A minimum shelf life was established based on these testing and is indicated by the expiration date provided on the product labeling. Verification testing of the subject devices followed the same test methods as the predicate devices.
Packaging and sterile barrier integrity following a transportation challenge has been verified for the packaging configurations used for the Zoom Catheters, LDP Catheters, and Zoom Aspiration Tubing within the Zoom System. Aging testing has also been performed that supports the sterile barrier integrity following aging.
### Animal Studies
A series of subacute and chronic animal studies were conducted to evaluate the safe and effective use of the Zoom Catheters and LDP Catheters with the Imperative Care Zoom Aspiration Pump and Zoom Aspiration Tubing in a porcine model. Testing included worst case positioning of the catheter in a "wedged" position, retrieval of various types of clots, and evaluation of various aspiration pressures. The Penumbra System was used as a control device in all completed studies. The studies concluded that:
{16}------------------------------------------------
- All vessels treated with the Zoom Catheters and the LDP Catheters were patent, . with no perforation, dissection or thrombosis following completion of the procedure.
- The subacute and chronic vascular response was histologically comparable between . the Zoom Catheters/LDP Catheters and control Penumbra System Reperfusion Catheter.
- The subacute and chronic pathological findings were comparable between the . Zoom Catheters/LDP Catheters and control Penumbra System Reperfusion Catheter.
## Clinical Study
The Imperative Care Zoom Catheters and the LDP Catheters were studied in the Imperative Trial, a prospective, multi-center, open-label, single-arm pivotal clinical investigation designed to assess the safety and effectiveness of the Zoom System in subjects diagnosed with acute ischemic stroke secondary to large vessel occlusions and undergoing a stroke mechanical neurothrombectomy procedure within 8 hours of last known well. A total of 260 evaluable subjects were enrolled across 26 clinical sites in the United States. Further details on this study can be accessed at ClinicalTrials.gov using the identifier NCT04129125.
| | Intervention/<br>Treatment<br>(Day 0) | 24-Hour<br>Assessment<br>(Day 1) | Discharge/Day<br>7 Assessment<br>(Day 3-7) | 90-Day<br>Follow Up<br>Assessment<br>(Day 90) | Total Withdrawal,<br>LTF, and Deaths |
|--------------------------------------|---------------------------------------|----------------------------------|--------------------------------------------|-----------------------------------------------|--------------------------------------|
| Evaluable<br>Subjects | 260 | 260 | 259 | 253 | |
| Withdrawal | 0 | 0 | 1 | 4 | 5 |
| Lost to<br>Follow-up<br>(LTF) | N/A | 0 | 0 | 5 | 5 |
| Death | 0 | 1* | 5 | 27 | 33 |
| Subjects<br>Completing<br>Each Visit | 260 | 259 | 253 | 217 | |
Reconciliation of Evaluable Subjects
One subject death occurred 2 days post-procedure, but declining status prevented the 24-hour visit from being completed.
{17}------------------------------------------------
### Imperative Trial Mechanical Neurothrombectomy Procedure
Mechanical neurothrombectomy was performed using the Zoom System under general anesthesia or conscious sedation, at the discretion of the investigator. The Zoom Catheters and the LDP Catheters were the primary device. An LDP Catheter was employed to access the vasculature in all cases, while its use to aspirate was according to the treating physicians' preference. Aspiration was applied through an LDP Catheter alone or in conjunction with a Zoom (71, 55, 45, or 35) Catheter either proximal to the clot or at the clot in 92% (238/260) of subjects and in 52% (124/238) of these subjects, aspiration was applied through an LDP Catheter alone or in conjunction with a Zoom (71, 55, 45, or 35) Catheter at the clot. If multiple revascularization attempts were required, the Zoom System was recommended to be used for at least the first three attempts before considering switching to any non-Zoom System device. If the modified thrombolysis in cerebral infarction (mTICI) score ≥ 2b revascularization was not achieved after three attempts with the study device, the procedure was considered a treatment failure for the…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.