K242631 · Smith & Nephew Inc., Endoscopy Div. · OWY · Nov 26, 2024 · General, Plastic Surgery
Device Facts
Record ID
K242631
Device Name
REGENETEN™ Bioinductive Implant
Applicant
Smith & Nephew Inc., Endoscopy Div.
Product Code
OWY · General, Plastic Surgery
Decision Date
Nov 26, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The REGENETEN Bioinductive Implant is indicated for the management and protection of tendon or extra-articular ligament injuries in which there has been no substantial loss of tendon or ligament tissue. The REGENETEN Bioinductive Implant Delivery System is indicated for the arthroscopic delivery of the REGENETEN Bioinductive Implant.
Device Story
REGENETEN Bioinductive Implant is a bioabsorbable, porous collagen sheet used for management and protection of tendon or extra-articular ligament injuries. Provided sterile and single-use, it is hydrated and applied arthroscopically via a delivery system to provide a collagen layer between injured tissue and surrounding structures. Used by surgeons in clinical settings, the implant acts as a scaffold to support healing and neo-tissue formation. It does not require design or material changes for its expanded indication; clinical benefit is supported by animal studies demonstrating increased ligament thickness and safe integration compared to sutured repair alone.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (staple pull-out, hydrothermal transition, endotoxin, mechanical integrity) and an in vivo ovine bilateral animal study. The animal study compared sutured repair (control) vs. sutured repair with implant (treatment), demonstrating increased ligament thickness and safe tissue formation without new safety issues.
Technological Characteristics
Bioabsorbable collagen sheet; porous, pliable, nonfriable. Provided sterile, non-pyrogenic, dual sterile seal. Delivered arthroscopically. Identical to predicate (K222501) in material composition and physical performance.
Indications for Use
Indicated for patients with tendon or extra-articular ligament injuries without substantial tissue loss. Bioinductivity demonstrated for tendon only.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 26, 2024
Smith & Nephew Inc., Endoscopy Div. Lacev Klungseth Pr. Regulatory Affairs Specialist 150 Minuteman Road Andover, Massachusetts 01810
Re: K242631
Trade/Device Name: REGENETEN™ Bioinductive Implant Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWY, ORQ Dated: August 28, 2024 Received: September 3, 2024
Dear Lacey Klungseth:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## CHRISTOPHER FERREIRA -S
Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K242631
Device Name REGENETEN™ Bioinductive Implant
#### Indications for Use (Describe)
The REGENETEN Bioinductive* Implant is indicated for the management and protection of tendon or extra-articular ligament injuries in which there has been no substantial loss of tendon or ligament tissue. * Bioinductivity has been demonstrated for tendon only.
The REGENETEN Bioinductive Implant Delivery System is indicated for the arthroscopic delivery of the REGENETEN Bioinductive Implant.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Smith & Nephew, Inc. Endoscopy 150 Minuteman Road Andover, MA 01810 Massachusetts, USA
T: +1 978 749 1000 T: +1 800 343 8386 (USA toll free) www.smith-nephew.com
# Smith Nephew
K242631
### Attachment 15:
## 510(k) Summary According to 21 CFR 807.92
| Date Prepared: | November 8, 2024 | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------|--|
| Submitter Name & Address: | Smith and Nephew, Inc.<br>Endoscopy<br>150 Minuteman Road<br>Andover, MA 01810 | |
| FDA Establishment Owner/Operator Number: | 1020279 | |
| FDA Establishment Registration Number: | 3003604053 | |
| Submitter / Primary Contact: | Lacey Klungseth<br>Pr. Regulatory Affairs Specialist<br>(612) 443-2734<br>Lacey.Klungseth@Smith-Nephwe.com | |
| Alternate Contact: | Jenna Horsley<br>Regulatory Affairs Director<br>(617) 218-7921<br>Jenna.Horsley@smith-nephew.com | |
| Device Trade / Proprietary Name: | REGENETEN™ Bioinductive Implant | |
| Device Common Name: | Surgical mesh | |
| Device Classification Name: | Mesh, Surgical, Collagen, Orthopaedics | |
| Regulation Medical Specialty: | Office of General & Plastic Surgery | |
| Review Panel(s): | Office of General & Plastic Surgery | |
| Product Code(s): | OWY (Bioinductive Implant),<br>ORQ (Bioinductive Implant Delivery Instrument) | |
| Regulation Numbers: | 21 CFR 878.3300 | |
| Submission Type: | Traditional 510(k) | |
| Device Class: | Class II | |
| Device Predicate: | REGENETEN™ Bioinductive Implant (K222501) | |
{5}------------------------------------------------
Smith & Nephew, Inc. Endoscopy 150 Minuteman Road Andover MA 018 Massachusetts, US
T: +1 978 749 1000 +1 800 343 8386 (USA toll free) h-nenhew com
Image /page/5/Picture/2 description: The image shows the Smith+Nephew logo in orange. Below the logo is the text "K242631" in black. The logo is a recognizable brand mark.
#### Device Description:
The REGENETEN™ Bioinductive Implant is a bioabsorbable implant device that provides a layer of collagen over injured tendons or ligaments. The implant is designed to provide a layer of collagen between a tendon or ligament and the surrounding tissue. After hydration, the implant is an easy-tohandle, pliable, nonfriable, porous collagen sheet. The REGENETEN Bioinductive Implant is provided sterile, non-pyrogenic, for single-use only, in a variety of sizes, in a dual sterile seal.
#### Proposed Intended Use
The REGENETEN™ Bioinductive* Implant is a medical device intended for the management and protection of soft tissue injuries, including tendon and extra-articular liqament.
#### Proposed Indications for Use
The REGENETEN™ Bioinductive Implant is indicated for the management and protection of tendon or extra-articular ligament injuries in which there has been no substantial loss of tendon or ligament tissue.
* Bioinductivity has been demonstrated for tendon only.
The REGENETEN™ Bioinductive Implant Delivery System is indicated for the arthroscopic delivery of the REGENETEN™ Bioinductive Implant.
#### Summary of Technological Characteristics Compared to Predicate Device:
The subject device (REGENETEN Bioinductive Implant) and the predicate device (REGENETEN Bioinductive Implant, K222501), are identical devices, having the same physical performance characteristics, packaging, and material composition. Past safety, bench testing, and animal study of the predicate device are directly applicable to the subject device. Previous verification testing completed is directly applicable to the subject device, including staple pull-out testing, hydrothermal transition temperature, endotoxin testing, and mechanical integrity.
The subject device has identical technological characteristics as the predicate device with the addition of the expanded indication. The expanded indications for use did not require any change in the device design, material composition, and packaging and this technological difference is supported by, valid, reproducible performance testing.
#### Biocompatibility:
The subject device (REGENETEN Bioinductive Implant) and the predicate device (REGENETEN Bioinductive Implant, K222501) are physically identical devices; therefore, additional biocompatibility testing is not required to support this submission.
#### Performance Data (Bench Testing):
The Regeneten Bioinductive Implant is identical to the predicate device (REGENETEN Bioinductive Implant, K222501) and performance specifications are identical. Past safety, verification, and mechanical characterization tests of the predicate device are directly applicable to the subject device, including staple pull-out testing, hydrothermal transition temperature, endotoxin testing, and mechanical integrity.
{6}------------------------------------------------
Smith & Nephew, Inc. Endoscopy 150 Minuteman Road Andover. MA 0181 Massachusetts, USA
T: +1 978 749 1000 +1 800 343 8386 (USA toll free) w.smith-nenhew.com
Image /page/6/Picture/2 description: The image contains the Smith & Nephew logo in orange. Below the logo is the text "K242631" in black. The logo is a stylized version of the company name, with a plus sign incorporated into the "N" of Nephew.
#### Performance Data (Animal Testing):
Smith and Nephew conducted an animal study with multiple timepoints evaluating the REGENETEN Bioinductive Implant in an in vivo ovine bilateral model with a primary histology endpoint to demonstrate ligament-like tissue formation post-capsulotomy.
A direct comparative analysis between sutured repair (control group) and sutured repair with Bioinductive Implant (treatment group) evaluated healing progression, formation of neo-tissue, and resemblance to native tissue. Results show the Bioinductive Implant results in increased ligament thickness and that the Bioinductive Implant can be safely implanted for ligament repair without raising any new issues of safety and effectiveness.
Bioinductivity has been demonstrated for tendon only.
#### Compliance to Standards:
This submission does not claim compliance to any new FDA recognized consensus standards.
#### Substantial Equivalence and Conclusion:
No design modifications were completed regarding the expanded indications for use and the subject device (REGENETEN Bioinductive Implant) is identical to the predicate device (REGENETEN Bioinductive Implant, K222501) in terms of material composition, design, packaging, sterilization, and biocompatibility. The only difference is the addition of ligament tissue to the indications for use.
Expansion of indications for use to include management and protection of extra-articular ligament injuries does not impact the intended surgical purpose of the device, nor does it affect the safety and efficacy of the device relative to the predicate. Both the subject and predicate devices have the same intended surgical purpose.
This difference is substantiated by reviewing the results of the animal study presented in this 510(k). Based on the comparative analysis, the REGENETEN Bioinductive Implant has been shown to be substantially equivalent to its predicate, the REGENETEN Bioinductive Implant (K222501). They are the same device: the technological characteristics, packaging, and sterilization parameters are identical. Only the indications for use have changed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.