← Product Code [SBF](/productcode/SBF) · K242569

# Mixed Reality Spine Navigation (K242569)

_Brainlab AG · SBF · May 16, 2025 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K242569

## Device Facts

- **Applicant:** Brainlab AG
- **Product Code:** [SBF](/productcode/SBF.md)
- **Decision Date:** May 16, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Software as a Medical Device

## Indications for Use

Mixed Reality Spine Navigation is an accessory to the Spine & Trauma Navigation Medical Device. The software is intended to display 2D navigation screens as well as a floating 3D virtual representation and stereotaxic information of tracked instruments on a virtual display using a Mixed Reality headset. The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed stereotaxic information.

## Device Story

Accessory to Spine & Trauma Navigation system; utilizes Magic Leap 2 MedTech optical see-through head-mounted display. Inputs navigation data (patient registration, instrument tracking) via Wi-Fi from Brainlab IGS platforms (Curve, Kick 2, Buzz). Transforms data into 2D navigation screens and 3D virtual anatomical models rendered stereoscopically in headset. Used in operating room by surgeons; provides floating virtual content overlaid on real-world view. Enables hands-free gesture control for viewing navigation data during pedicle screw placement. Output assists surgeon in visualizing instrument trajectory relative to patient anatomy; enhances spatial awareness during surgery. Does not replace primary navigation screen; intended for use in conjunction with stereotaxic information displayed on IGS platform.

## Clinical Evidence

No clinical testing was required. Evidence consists of software verification and validation (Enhanced Documentation level), hardware safety/EMC testing (IEC 60601-1, IEC 60601-1-2), biocompatibility (ISO 10993-1), cleaning validation, optical property testing (IEC 63145 series), and usability testing in a simulated OR environment. Network performance benchmarking confirmed bandwidth requirements.

## Technological Characteristics

Optical see-through head-mounted display (Magic Leap 2 MedTech); Wi-Fi connectivity for data transfer; optical tracking; gesture-based user interface. Software: Mixed Reality Spine Navigation 1.0. Materials: Biocompatible components per ISO 10993-1. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 63145-20-10/20-20/22-10 (eyewear displays).

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Spine & Trauma Navigation ([K221618](/device/K221618.md))

## Reference Devices

- CommandEP Data Manager PC (DPC02) ([K223290](/device/K223290.md))
- CommandEP HMD (HMD02) ([K223290](/device/K223290.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

May 16, 2025

Brainlab AG
Esther Moreno Garcia
Manager Regulatory Affairs
Olof-Palme-Str.9
Munich, 81829
Germany

Re: K242569
Trade/Device Name: Mixed Reality Spine Navigation
Regulation Number: 21 CFR 882.4560
Regulation Name: Augmented Reality Stereotaxic Instrument
Regulatory Class: Class II
Product Code: SBF
Dated: August 28, 2024
Received: May 12, 2025

Dear Esther Moreno Garcia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}

2K242569 - Esther Moreno Garcia
Page

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

3K242569 - Esther Moreno Garcia
Page

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K242569  |   |
|  Device Name Mixed Reality Spine Navigation  |   |
|  Indications for Use (Describe) Mixed Reality Spine Navigation is an accessory to the Spine & Trauma Navigation Medical Device. The software is intended to display 2D navigation screens as well as a floating 3D virtual representation and stereotaxic information of tracked instruments on a virtual display using a Mixed Reality headset. The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed stereotaxic information.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

BRAINLAB

# 510(k) Summary - K242569

May 15, 2025

General Information

|  Manufacturer | Brainlab AG; Olof-Palme-Str.9, 81829, Munich, Germany  |   |
| --- | --- | --- |
|  Establishment Registration | 8043933  |   |
|  Trade Name | Mixed Reality Spine Navigation  |   |
|  Classification Name | Augmented Reality Stereotaxic Instrument  |   |
|  Product Code | SBF  |   |
|  Regulation Number | 882.4560  |   |
|  Regulatory Class | II  |   |
|  Panel | Orthopedic  |   |
|  Predicate Device | K221618 - Spine & Trauma Navigation  |   |
|  Reference Device | K223290, CommandEP Data Manager PC (DPC02), CommandEP HMD (HMD02)  |   |
|  Contact Information  |   |   |
|  Primary Contact |   | Alternate Contact  |
|  Esther Moreno Garcia
Manager Regulatory Affairs
Phone: +49 89 99 15 68 0
Email: regulatory.affairs@brainlab.com |   | Chiara Cunico
Phone: +49 89 99 15 68 0
Fax: +49 89 99 15 68 5033
Email: chiara.cunico@brainlab.com  |

1. Indications for Use

Mixed Reality Spine Navigation is an accessory to the Spine &amp; Trauma Navigation Medical Device. The software is intended to display 2D navigation screens as well as a floating 3D virtual representation and stereotactic information of tracked instruments on a virtual display using a Mixed Reality headset. The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed stereotaxic information.

2. Device Description

The Subject device Mixed Reality Spine Navigation is an accessory to the Spine &amp; Trauma Navigation System. It consists of the software Mixed Reality Spine Navigation 1.0 and the mixed reality headset Magic Leap 2 Medtech (ML2). The software application allows the display in the mixed reality headset of 2D navigation views and a 3D hovering model of the patient anatomy, including stereotactic information of tracked instruments to support the surgeon during pedicle screw placement procedures.

The software Mixed Reality Spine Navigation is installed and running on the Magic Leap 2 MedTech glasses and can only be used in combination with a Brainlab Image Guided Surgery (IGS) platform (Curve Navigation 17700, Kick 2 Navigation Station or Buzz Navigation Ceiling-Mounted), where the Spine &amp; Trauma Navigation software is running. All required navigation

K242569, Page 1 of 8

{5}

BRAINLAB

data, such as the patient registration, is transferred over Wi-Fi from the IGS platform to the Magic Leap 2 MedTech. Based on these data, 2D views and a 3D model are rendered by the computing unit of the Magic Leap 2 MedTech and displayed stereoscopically in the headset. Thus, navigation information displayed on the screen(s) of the IGS platform can be simultaneously displayed on Magic Leap 2 MedTech during the surgery.

Magic Leap 2 MedTech is an optical see-through head-mounted display: images are projected on semi-transparent optical layers, giving the surgeons the possibility to have virtual content displayed around the patient while having a view of the real world. If needed, corrective lenses can be attached to the headset.

# 3. Substantial Equivalence

For the Substantial Equivalence determination, comparison of the Subject Device features with the following predicate device and reference device was carried out:

K221618 - Spine &amp; Trauma Navigation (product code: OLO)

K223290, CommandEP Data Manager PC (DPC02), CommandEP HMD (HMD02) (product code: LLZ)

The predicate device was chosen since both the subject device and the predicate are used together and have similar technological characteristics and use case. The reference device was chosen for comparison basis of the Magic Leap 2 Medtech glasses.

The main difference compared to the predicate device is the use of the Magic Leap 2 Medtech headset allowing to view 2D and 3D views in the proximity of the patient's anatomy.

K242569, Page 2 of 8

{6}

BRAINLAB

|  Feature | Predicate Device K221618 | Reference Device K223290 | Subject Device  |
| --- | --- | --- | --- |
|  Indications for use | Spine & Trauma Navigation is intended as an intraoperative image-guided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or Intraoperative 2D or 3D image data.

Spine & Trauma Navigation enables computer assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments.

The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or | N/A | Mixed Reality Spine Navigation is an accessory to the Spine & Trauma Navigation Medical Device. The software is intended to display 2D navigation screens as well as a floating 3D virtual representation and stereotactic information of tracked instruments on a virtual display using a Mixed Reality headset.

Similar to predicate device. The subject device provides another way of displaying navigation information for spine screw placement procedures.  |

K242569, Page 3 of 8

{7}

BRAINLAB

|  Feature | Predicate Device K221618 | Reference Device K223290 | Subject Device  |
| --- | --- | --- | --- |
|   | 2D fluoroscopic image) and/or an image data based model of the anatomy. |  |   |
|  Intended use environment | The navigation system shall be used in an operating room / suite.
The navigation system is not intended to be used inside an MR environment. | N/A | The intended use environment is the operating room. The device is not intended to be used in the MR environment. Pre- and/or intraoperative MR imaging is not part of the clinical workflows for which the device is used.
Same as predicate device.  |
|  Software version | Spine & Trauma Navigation 2.0 | N/A | Mixed Reality Spine Navigation 1.0
Spine & Trauma Navigation 2.0  |
|  Mixed Reality Headset | N/A, no Magic leap 2 available | CommandEP HMD (Magic Leap 2 glasses):
- Includes headset, compute pack and controller.
- Possible to connect simultaneously up to five headsets
- Provided non-sterile | Magic Leap 2 Medtech:
- Includes headset, compute pack and controller.
Additionally it comes with a backpack for carrying the compute pack under the sterile gown and possibility of prescription lenses.
- Possible to connect simultaneously up to four headsets
- Provided non-sterile  |

K242569, Page 4 of 8

{8}

BRAINLAB

|  Feature | Predicate Device K221618 | Reference Device K223290 | Subject Device  |
| --- | --- | --- | --- |
|   |  |  | Similar to reference device.  |
|  Date transfer between mixed reality headset and platform | N/A, no Magic leap 2 available | Encrypted wireless transfer (Wi-Fi) | Encrypted wireless transfer (Wi-Fi). Identical to reference device.
Same as reference device.  |
|  User interaction | Touch screen of IGS platform and remote control | CommandEP HMD (Magic Leap 2 MedTech:
Hands-free gaze-dwell controls (the use of the controller is not required). | IGS platforms:
Touch screen and Remote Control

Magic Leap 2 MedTech:
Gesture controls (the use of the controller is not required).

Same as predicate for interaction with IGS platform and similar to reference device. User interface and gesture controls have been validated during usability testing.  |
|  Localization technique | Optical tracking | N/A | Optical tracking. Same as predicate device.  |
|  Navigation accuracy | • Mean Positional Error of the placed instrument's tip ≤ 2 mm
• Mean Angular Error of the placed instrument's axis ≤ 2° | N/A | • Mean Positional Error of the placed instrument's tip ≤ 2 mm
• Mean Angular Error of the placed instrument's axis ≤ 2°
Same as predicate device.  |

K242569, Page 5 of 8

{9}

BRAINLAB

|  Feature | Predicate Device K221618 | Reference Device K223290 | Subject Device  |
| --- | --- | --- | --- |
|  View features | Inline views for navigation, 3Ds views, Probe 's eye view, autopilot and DDR view. | N/A | All views displayed in the IGS platforms screen of the predicate can also be displayed when using the subject device. With the Magic Leap 2 Medtech, two additional views (virtual displays) are available: 2D floating display with Probe 's eye and trajectory displays and a 3D floating model.

Similar to predicate device, same views available (except for DRR and autopilot views) and additional floating views in the proximity of the patient anatomy.  |
|  Intra-operative planning of implants | Views and features to enable intraoperative planning of implants | N/A | Spine &Trauma Navigation software is also running in the IGS platform when using Mixed Reality Spine Navigation, thus same views and features as predicate device for intraoperative planning of screws. In addition, planned screws and trajectories can also be displayed in 2D and 3D views in the ML2 Medtech.  |
|  GUI technology | HTML5 | N/A | Mixed Reality Toolkit  |

K242569, Page 6 of 8

{10}

BRAINLAB

|  Feature | Predicate Device K221618 | Reference Device K223290 | Subject Device  |
| --- | --- | --- | --- |
|   |  |  | Similar to predicate device. Same GUI style implemented with a different technology.  |
|  IGS platforms | Curve Navigation 17700
Kick 2 Navigation Station (several models)
Buzz Navigation (Ceiling-Mounted)
Curve 1.1/1.2 Navigation Station | N/A | Curve Navigation 17700 (several models)
Kick 2 Navigation Station (several models)
Buzz Navigation (Ceiling-Mounted)
Curve 1.1/1.2 Navigation Station

An additional Curve Navigation platform #17701 has been introduced, remaining platforms are the same as predicate device.  |

K242569, Page 7 of 8

{11}

BRAINLAB

## 4. Performance Data

The following testing was conducted on the Subject Device to establish substantial equivalence with the predicate device. Test methods are equivalent to the predicate and reference devices.

### Software Verification and Validation Testing

Software verification and validation testing has been conducted and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions". These include successful implementation of product specifications, incremental testing for different release candidates, testing of risk control measures, compatibility testing or cybersecurity tests. The documentation submitted is for Enhanced Documentation level.

### Hardware verification

- For basic safety and EMC, testing was provided according to IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION and IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION.
- Biocompatibility evaluation was carried out in accordance with ISO 10993-1: 2018.
- Cleaning validation in accordance with FDA’s guidance “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling”.
- Testing in accordance with IEC 63145-20-10:2019 / 63145-20-20:2019 / 63145-22-10:2020, applicable to Eyewear displays, was provided in order to assess optical properties such as luminance or chromaticity.

### Bench Tests

- Usability testing was provided to validate the pedicle screw placement workflow using Mixed Reality Spine Navigation in a simulated OR environment.
- Network performance benchmarking has been performed to ensure that Mixed Reality Spine Navigation is working as intended with minimum bandwidth requirements.

No clinical testing was required for the subject device.

## 5. Conclusion

The comparison of the Subject Device with the predicate device shows that Mixed Reality Spine Navigation (1.0) has similar functionality, intended use and technological characteristics as the predicate device. Based on the comparison to the predicate and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate device.

K242569, Page 8 of 8

---

**Source:** [https://fda-staging.innolitics.com/device/K242569](https://fda-staging.innolitics.com/device/K242569)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
