The Autoclavable Cassettes are used in healthcare facilities to organize surgical and prosthetic instruments during sterilization, storage, and transport during and after surgical uses. The Autoclavable Cassettes are not intended to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated sterilization wrap to maintain the sterility of the enclosed devices.
Device Story
Autoclavable Cassettes store, organize, and protect surgical and prosthetic instruments during sterilization, storage, and transport. Device consists of a PPSU lid, base, and tray with silicone rubber grommets to secure instruments. Perforated design allows moist heat penetration during steam sterilization. Healthcare providers place instruments into the cassette, secure the lid, and wrap the unit in an FDA-cleared sterilization wrap. The cassette is then processed via pre-vacuum steam sterilization (132°C for 4 minutes, 30 minutes drying). The device facilitates instrument positioning and organization during surgical procedures. Benefits include organized instrument management and protection of instruments during the sterilization cycle and subsequent storage.
Clinical Evidence
Bench testing only. Validations performed per ANSI/AAMI ST98:2022 (manual cleaning), ANSI/AAMI/ISO 17665-1:2006 (sterilization/drying time), and ISO 10993-1:2018 (biocompatibility). Simulated use testing confirmed device integrity over 100 cleaning/sterilization cycles. Packaging validated per ASTM D4169-23e1. All tests passed.
Technological Characteristics
Materials: Polyphenylsulfone (PPSU), silicone rubber, stainless steel (hinge pin). Design: Perforated plastic tray/lid with silicone grommets. Sterilization: Pre-vacuum steam (132°C, 4 min, 30 min dry). Reusable up to 100 cycles. Biocompatibility: ISO 10993-1/5 compliant (non-cytotoxic).
Indications for Use
Indicated for use in healthcare facilities to organize surgical and prosthetic instruments during sterilization, storage, and transport. Intended for use with a validated sterilization wrap to maintain sterility of enclosed devices.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 21, 2025
Prismatik Dentalcraft, Inc. Haiyan Wei Principal Regulatory Affairs Specialist 2144 Michelson Drive Irvine, California 92612
Re: K242564
Trade/Device Name: Autoclavable Cassette (Glidewell HT™ Implant Guided Surgical Kit, Glidewell HTTM Implant Surgical Kit, Glidewell™ Prosthetic Kit) Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: February 18, 2025 Received: February 18, 2025
Dear Haiyan Wei:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen A. Digitally signed by
Stephen A. Anisko -S
Anisko -S Date: 2025.03.21
14:49:15 -04'00'
for: Christopher K. Dugard, M.S. Division Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
### 510(k) Number (if known) K242564
### Device Name
Autoclavable Cassette (Glidewell HTM Implant Guidewell HT™ Implant Surgical Kit, Glidewell™ Prosthetic Kit)
### Indications for Use (Describe)
The Autoclavable Cassettes are used in healthcare facilities to organize surgical and prosthetic instruments during sterilization, storage, and transport during and after surgical uses. The Autoclavable Cassettes are not intended to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated sterilization wrap to maintain the sterility of the enclosed devices.
Sterilization parameters:
Pre-vacuum steam: 132°C (270°F) for 4 minutes with 30 minutes drying time.
Sterilization validations for the worst-case Autoclavable Cassette and instruments. The Autoclavable Cassettes are validated for a maximum load of 641 grams for Glided Surgical Kit, 476 grams for Glidewell HTTM Implant Surgical Kit. and 192 grams for Glidewell™ Prosthetic Kit.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. On the left side of the logo is a triangle with a color gradient that starts with pink at the top, transitions to blue in the middle, and ends with yellow at the bottom. To the right of the triangle is the company name, "PRISMATIK DENTALCRAFT, INC.", written in a blue sans-serif font.
## 510(k) Summary K242564
## Submission Date: February 18, 2025
#### I. Contact Details
Applicant Name: Prismatik Dentalcraft, Inc.
Address: 2144 Michelson Drive, Irvine, CA 92612 United States
Applicant Contact: Ms. Haiyan Wei, Principal Regulatory Affairs Specialist
Email: Haiyan.Wei@glidewelldental.com
Phone: 949-399-1984
Secondary Contact: Ms. So Hyun Park, Regulatory Affairs Manager
Email: So.Park(@glidewelldental.com
Phone: 949-863-5479
#### II. Device Name
Device Trade Name: Autoclavable Cassette (Glidewell HTTM Implant Guided Surgical
Kit, Glidewell HTTM Implant Surgical Kit, Glidewell™ Prosthetic Kit)
Common Name: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories
Classification Name: Sterilization Wrap
Regulation Number: 21 CFR 880.6850
Regulatory Class: Class II
Product Code: KCT
#### III. Predicate Device
Predicate 510(k) #: K180791
Predicate Trade Name: Straumann® BLX Surgical Cassette
Classification Name: Sterilization Wrap
Regulation Number: 21 CFR 880.6850
Regulatory Class: Class II
Product Code: KCT
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.
#### Device Description IV.
The Autoclavable Cassettes are designed to store, organize, and protect various surgical and prosthetic instruments that are sterilized by the healthcare provider. The Autoclavable Cassettes are composed of three main components: a lid, a base, and a tray. The components are made of polyphenylsulfone PPSU. The tray has silicone rubber grommets to hold the instruments in place. The lid latches to the base holds the tray, and all three components are designed to secure the kit into a single unit. The cassette protects the interior components during sterilization and subsequent storage. The kits are perforated to allow for penetration of moist heat during steam sterilization and require the use of an FDA-cleared sterilization wrap to maintain sterility. To facilitate the surgical procedure and the correct use and the positioning of the instrument, the tray has instrument descriptions on the surface.
Surgical and prosthetic instruments of the Glidewell Implant System are intended to be placed in the Autoclavable Cassette. These surgical and prosthetic instruments which are the contents of the sterilization cassette are all Class 1 exempt as described in 21 CFR 872.3980 (Endosseous dental implant accessories).
| Device<br>Model Name | Cassette<br>Size | Max No. of<br>Instruments | Weight of<br>tray empty<br>(g) | Weight of<br>tray full<br>(g) | Vent to<br>Volume<br>Ratio |
|----------------------------------------------------|----------------------------|---------------------------|--------------------------------|-------------------------------|-------------------------------------------|
| Glidewell<br>HT™ Implant<br>Guided<br>Surgical Kit | 212mm x<br>173mm x<br>64mm | 63 | 484 | 641 | $3.11 X10^{-3}$<br>(mm-1)<br>0.079 (in-1) |
| Glidewell<br>HT™ Implant<br>Surgical Kit | 199mm x<br>174mm x<br>60mm | 52 | 407 | 476 | $6.09 X10^{-4}$<br>(mm-1)<br>0.016 (in-1) |
| Glidewell™<br>Prosthetic Kit | 96mm x<br>74mm x<br>60mm | 11 | 163 | 192 | $8.50 X10^{-4}$<br>(mm-1)<br>0.022 (in-1) |
The Autoclavable Cassettes will be marketed in the following variations with accompanied instruments:
#### V. Indications for Use
The Autoclavable Cassettes are used in healthcare facilities to organize surgical and prosthetic instruments during sterilization, storage, and transport during and after surgical uses. The Autoclavable Cassettes are not intended to maintain sterility; they are intended
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a colorful, prismatic design. To the right of the triangle is the company name, "PRISMATIK DENTALCRAFT, INC." in blue text.
to be used in conjunction with a legally marketed, validated sterilization wrap to maintain the sterility of the enclosed devices.
Sterilization parameters:
Pre-vacuum steam: 132°C (270° F) for 4 minutes with 30 minutes drying time.
Sterilization validations for the worst-case Autoclavable Cassette included cassette and instruments. The Autoclavable Cassettes were validated for a maximum load of 641 grams for Glidewell HTM Implant Guided Surgical Kit, 476 grams for Glidewell HTTM Implant Surgical Kit, and 192 grams for Glidewell™ Prosthetic Kit.
#### VI. Technological Comparison
The subject device and predicate device share the following characteristics:
- Same intended use ●
- . Same design
- Same principal of operation .
- Same sterilization method
- Reusable ●
The subject device is technologically different from the predicate device as follows:
- Materials (same materials except medium size cassette which has a lid hinge pin ● made of stainless steel)
- Cassette size/weight ●
- Vent-to-Volume ratio .
- . Drying time
The technological characteristics of the subject device are compared to the predicate device in Table 1 below. Any noted differences in the technological characteristics between the subject device and the predicate device do not raise new concerns of safety and effectiveness.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a triangle on the left side, which is divided into several sections with different colors. To the right of the triangle is the text "PRISMATIK DENTALCRAFT, INC." The text is in a sans-serif font and is blue.
| Characteristics | Subject Device<br>(K242564) | Predicate Device<br>(K180791) | Comparison |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Autoclavable Cassette (Glidewell HTTTM Implant<br>Guided Surgical Kit, Glidewell HTTTM Implant<br>Surgical Kit, GlidewellTM Prosthetic Kit) | Straumann® BLX Surgical<br>Cassette | N/A |
| Manufacturer | Prismatik Dentalcraft, Inc. | Straumann USA, LLC | N/A |
| Regulatory Classification | | | |
| Regulation<br>Number | 21 CFR 880.6850 | 21 CFR 880.6850 | Same |
| Classification<br>Name | Sterilization Wrap | Sterilization Wrap | Same |
| Classification | II | II | Same |
| Product Code | KCT | KCT | Same |
| Intended Use/Indication For Use | | | |
| Intended Use/<br>Indications for<br>Use | The Autoclavable Cassettes are<br>used in healthcare facilities to<br>organize surgical and prosthetic<br>instruments during sterilization,<br>storage, and transport during<br>and after surgical uses. The<br>Autoclavable Cassettes are not<br>intended to maintain sterility;<br>they are intended to be used in<br>conjunction with a legally<br>marketed, validated sterilization<br>wrap to maintain the sterility of<br>the enclosed devices.<br><br>Sterilization parameters:<br>Pre-vacuum steam: 132°C<br>(270° F) for 4 minutes with 30<br>minutes drying time.<br><br>Sterilization validations for the<br>worst-case Autoclavable<br>Cassette included cassette and<br>instruments. The Autoclavable<br>Cassettes were validated for a<br>maximum load of 641 grams for<br>Glidewell HTTM Implant | The Straumann® BLX Cassette<br>is used in healthcare facilities to<br>organize, enclose, cleaning,<br>sterilize, transport, and store<br>medical devices between<br>surgical uses. The BLX<br>Cassette is not intended to<br>maintain sterility; it is intended<br>to be used in conjunction with a<br>legally marketed, validated<br>sterilization wrap.<br><br>The BLX Cassette has been<br>validated for a maximum load<br>of 300 grams, including cassette<br>and instruments.<br><br>Sterilization parameters: Pre-<br>vacuum steam: 132°C (270° F)<br>for 4 minutes with 20 minutes<br>drying time. | Same; except<br>for minor<br>differences in<br>the device<br>trade name,<br>and device<br>size/weight. |
| Characteristics | Subject Device<br>(K242564) | Predicate Device<br>(K180791) | Comparison |
| Technological Characteristics | | | |
| | Guided Surgical Kit, 476 grams<br>for Glidewell HT™ Implant<br>Surgical Kit, and 192 grams for<br>Glidewell™ Prosthetic Kit. | | |
| Design | Plastic tray and lid | Plastic tray and lid | Same |
| Material | Polyphenylsulfone (Radel<br>5000)<br>Silicon<br>Stainless steel (lid hinge pin for<br>medium size only) | Polyphenylsulfone (Radel<br>5000)<br>Silicon | Same; except<br>for the medium<br>size of the<br>subject device,<br>which has a lid<br>hinge pin made<br>of stainless<br>steel. |
| Material<br>Compatible with<br>Sterilization<br>Method | Yes | Yes | Same |
| Perforated | Yes; Allow moist heat (steam)<br>penetration to achieve<br>sterilization. | Yes; Allow moist heat (steam)<br>penetration to achieve<br>sterilization. | Same |
| Enclosed Feature | Silicone grommets | Silicone grommets | Same |
| Reusable | Yes, up to 100x | Yes, up to 100x | Same |
| Sterilization<br>Method | Moist heat (steam) | Moist heat (steam) | Same |
| Cycles | Pre-vacuum | Pre-vacuum | Same |
| Parameters | Pre-vacuum:<br>132°C (270° F) for 4 minutes<br>with 30 minutes drying time. | Pre-vacuum:<br>132°C (270° F) for 4 minutes<br>with 20 minutes drying time. | Similar; more<br>drying time<br>required for the<br>subject device. |
| Sterile barrier | FDA cleared sterilization pouch | FDA cleared sterilization pouch | Same |
| Biocompatibility | The biocompatibility<br>assessment was performed per<br>ISO 10993-1 and testing was<br>performed using methods<br>described in ISO 10993-5. The<br>results indicate the subject<br>devices are non-cytotoxic. | The biocompatibility<br>assessment was performed per<br>ISO 10993-1 and testing was<br>performed using methods<br>described in ISO 10993-5. The<br>results indicate the subject<br>devices are non-cytotoxic. | Same |
# Table 1: Comparison Table Between Subject Device and Predicate Device
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into sections of different colors, including pink, red, yellow, green, and blue. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a color wheel design. To the right of the triangle is the company name in blue text. The words "PRISMATIK" and "DENTALCRAFT, INC." are stacked on top of each other.
#### Summary of Non-Clinical Testing VII.
The performance during multiple reprocessing steps for Autoclavable Cassette, recommended in the labeling, was validated according to applicable as recommendations in the FDA guidance document - Guidance for Industry and Food and Drug Administration Staff: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015. The worst-case cassettes were tested for performance based on critical impact factors including weight, vent to volume ratio, and worst-case instruments to be contained. The summary of performance testing is provided in Table 2 below.
| Type of Testing | Test Purpose | Acceptance Criteria | Test Result |
|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Manual Cleaning<br>Validation<br>(ANSI/AAMI<br>ST98:2022) | Evaluate and validate<br>the manual cleaning of<br>the subject device | Visual Inspection: No<br>visible soil<br>Hemoglobin Test: <2.2<br>$\mu$ g/cm²<br>Protein Test: <6.4<br>$\mu$ g/cm² | Pass |
| Sterilization<br>Validation<br>(ANSI/AAMI/ISO<br>17665-1:2006/<br>(R)2013) | Validate sterilization<br>cycle and drying time of<br>the subject devices | SAL ≤10-6 using the<br>biological indicator<br>(BI) overkill method | Pass |
| Reprocessing of<br>Medical Devices in a<br>Health Care Setting | Life Cycle (simulated<br>use) testing after 100<br>cleaning and<br>sterilization cycles | Test samples must<br>withstand 100 cycles of use<br>(cleaning,<br>sterilization, and functional<br>tests) without physical<br>damage and compromising<br>function. | Pass |
| Biocompatibility<br>(ISO 10993-1:2018) | Chemical<br>Characterization and<br>Cytotoxicity | Cellular vitality for the<br>extract of the sample<br>>70%, which is considered<br>non-cytotoxic. | Pass |
| Packaging Validation<br>(ASTM D4169-23e1) | Packaging performance<br>testing | Packaging must withstand<br>the distribution<br>environment | Pass |
Table 2. Performance Testing Summary
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into sections of different colors, including pink, blue, and yellow. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.
#### VIII. Conclusion
Based on the intended use, technological characteristics, and non-clinical performance data, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, cleared under K180791, Class II (21 CFR 880.6850), product code KCT.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.