K242536 · Tautona Group · HIF · Oct 15, 2024 · Obstetrics/Gynecology
Device Facts
Record ID
K242536
Device Name
TauTona Pneumoperitoneum Assist Device (TPAD)
Applicant
Tautona Group
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Oct 15, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TauTona Pneumoperitoneum Assist Device (TPAD) is intended for use with a Veress needle for the establishment of a pneumoperitoneum during laparoscopic procedures.
Device Story
Single-use device for laparoscopic surgery; assists Veress needle insertion. Device features pad with suction cup array, port for hospital vacuum line, vacuum control button, and handle. User applies suction to manipulate and pull tissue around insertion site; crescent shape allows device removal while Veress needle remains in patient. Operated by surgeon in clinical setting. Provides stable tissue access for pneumoperitoneum establishment. Benefits include standardized tissue manipulation during needle placement.
Clinical Evidence
Two-phase clinical study (n=30 total). Phase 1 (n=20) evaluated TPAD vs. standard of care (SOC) for Palmer's point access; Phase 2 (n=10) evaluated umbilical region access. Primary endpoints included surgeon satisfaction, procedure timing (incision to needle insertion, incision to insufflation), and patient satisfaction. Results showed no statistically significant differences between TPAD and SOC across all metrics. No severe or moderate device-related adverse events reported.
Technological Characteristics
Single-use, crescent-shaped pad with suction cup array. External hospital vacuum source. No electrical components, software, or energy sources. Biocompatibility per ISO 10993-1. Performance testing includes dimensional verification, vacuum holding force, and tensile strength.
Indications for Use
Indicated for use with a Veress needle to establish pneumoperitoneum during laparoscopic procedures in all regions of the abdominal wall, including Palmer's Point and umbilical regions.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
October 15, 2024
TauTona Group Mike Blanchette Chief Operating Officer 604 Fifth Ave Suite D Redwood City, California 94063
Re: K242536
> Trade/Device Name: TauTona Pneumoperitoneum Assist Device (TPAD) Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF Dated: August 21, 2024 Received: August 26, 2024
Dear Mike Blanchette:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242536
Device Name
TauTona Pneumoperitoneum Assist Device (TPAD)
Indications for Use (Describe)
The TauTona Pneumoperitoneum Assist Device (TPAD) is intended for use with the Veress Needle for the establishment of a pneumoperitoneum during laparoscopic procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X | Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K242536 Page 1 of 4
# 510(k) SUMMARY
# SUBMITTER INFORMATION
| Applicant: | TauTona Research and Development Group, LLC | |
|------------|--------------------------------------------------|--|
| Contact: | Mike Blanchette | |
| Phone: | 650-503-8250 | |
| Email: | mblanchette@tautonagroup.com | |
| Address: | 604 Fifth Ave, Suite D<br>Redwood City, CA 94063 | |
# CORRESPONDENT INFORMATION
| Contact: | Joshua Crist |
|----------|---------------------------------------|
| Title: | Regulatory Consultant- Medical Device |
| Firm: | Biologics Consulting Group |
#### DATE PREPARED: 10/10/24
# DEVICE INFORMATION
| Device Name: | TauTona Pneumoperitoneum Assist Device (TPAD) |
|--------------------|-----------------------------------------------|
| Common Name: | TPAD |
| Regulation Number: | 884.1730 |
| Regulation Name: | Laparoscopic insufflator |
| Product Code: | HIF |
| Regulatory Class: | Class II |
# PREDICATE DEVICE INFORMATION
| Device Name: | TauTona Pneumoperitoneum Assist Device<br>(TPAD) |
|----------------|--------------------------------------------------|
| 510(k) Number: | K233020 |
| Manufacturer: | TauTona Research and Development Group, LLC |
The predicate device has not been subject to a design related recall.
{5}------------------------------------------------
# 510(k) SUMMARY
# DEVICE DESCRIPTION
The TauTona Pneumoperitoneum Assist Device (TPAD) is a single use device used during laparoscopic surgical procedures. The device consists of a pad which contains an array of suction cups on one side; a port that attaches to a standard hospital vacuum line; a button to control vacuum to the suction cups; and a handle to allow user to manipulate the device. Suction is applied to the device to allow the user to manipulate (hold and pull on) the tissue around the Veress needle insertion site. The device is crescent in shape to allow removal of the device after use while the Veress needle remains in the patient.
# INDICATIONS FOR USE
The TauTona Pneumoperitoneum Assist Device (TPAD) is intended for use with a Veress needle for the establishment of a pneumoperitoneum during laparoscopic procedures.
# COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The table below compares the intended use and the technological characteristics of the subject device and predicate device.
| | Subject Device<br>TauTona<br>Pneumoperitoneum Assist<br>Device (TPAD) | Predicate Device<br>TauTona<br>Pneumoperitoneum Assist<br>Device (TPAD)<br>K233020 | Comparison |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | HIF | HIF | Same |
| Regulation Number | 884.1730 | 884.1730 | Same |
| Regulatory Class | II | II | Same |
| Indications for Use | The TauTona<br>Pneumoperitoneum Assist<br>Device (TPAD) is intended<br>for use with a Veress<br>needle for the establishment<br>of a pneumoperitoneum<br>during laparoscopic<br>procedures. | The TauTona<br>Pneumoperitoneum Assist<br>Device TPAD is intended for<br>use in the upper left quadrant of<br>the abdominal wall (i.e.,<br>Palmer's Point) with a Veress<br>needle for the establishment of a<br>pneumoperitoneum during<br>laparoscopic procedures. | Similar indications<br>for use. Predicate<br>TPAD location for<br>use is for Palmer's<br>Point Access. This<br>510k is to expand<br>to all regions of the<br>abdominal wall,<br>including Palmer's<br>Point and<br>umbilical regions. |
| Veress Needle Insertion | Use of negative pressure to<br>attach to skin to support Veress<br>needle insertion. Device can be | Use of negative pressure to<br>attach to skin to support Veress<br>needle insertion. Device can be<br>removed while Veress Needle<br>is in place. | Same |
#### Comparator Table for Subject and Predicate Devices Table 1:
{6}------------------------------------------------
K242536 Page 3 of 4
| | Subject Device<br>TauTona<br>Pneumoperitoneum Assist<br>Device (TPAD) | Predicate Device<br>TauTona<br>Pneumoperitoneum Assist<br>Device (TPAD) K233020 | Comparison |
|----------------------|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|------------|
| | removed while Veress Needle<br>is in place. | | |
| Vacuum Application | Array of suction cups | Array of suction cups | Same |
| Vacuum Source | External – Hospital Vacuum<br>Line | External – Hospital Vacuum<br>Line | Same |
| Software/Electronics | None | None | Same |
| Biocompatible | Biocompatibility testing in<br>accordance with ISO 10993-1 | Biocompatibility testing in<br>accordance with ISO 10993-1 | Same |
# 510(k) SUMMARY
# SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
# Biocompatibility Testing
Studies were performed and confirmed that the Subject device is biocompatible for the intended patient contact profile in accordance with ISO 10993-1.
# Electrical Safety
Not applicable. The subject device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.
#### Electromagnetic Compatibility (EMC)
Not applicable. The subject device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.
#### Software
Not applicable. The subject device contains no software.
#### Performance Testing
Performance testing meets its design requirements and performs as intended. This testing is the same as the predicate device. The performance tests include:
- Dimensional testing for vacuum line compatibility, device features ●
- Vacuum control /holding forces. ●
- Tensile strength
{7}------------------------------------------------
K242536 Page 4 of 4
#### 510(k) SUMMARY
#### Clinical Data
The clinical study was performed in two phases. The first phase included 20 patients and evaluated the TPAD vs. Standard of Care (SOC) for patients undergoing laparoscopic surgery where the Veress Needle was used to establish a pneumoperitoneum (see K233020). The majority of Veress entries were in the upper left quadrant (i.e. Palmer's point). The second phase of the study included 10 patients and focused on the umbilical region of the abdominal wall. In both phases the studies compared satisfaction, time, and skin impacts. The primary endpoint of the average surgeon questionnaire scoring was not significantly different for SOC and TPAD across all responses. The average timing measurements from the incision to Veress Needle Insertion, as well as time from incision to insufflation, were not significantly different between SOC and TPAD for both timing durations. The average patient questionnaire scoring was not significantly different for SOC and TPAD across all responses. No Severe or Moderate Adverse Events related to the TPAD were observed.
Overall, across this wide range of metrics, the use of TPAD for Veress needle insertion and insufflation was not significantly different than the use of the standard of care for Veress needle insertion and insufflation. These clinical performance data therefore support that the subject device is substantially equivalent to the predicate.
#### CONCLUSION
The subject device has the same technological characteristics and intended use as the predicate device, with a more general indications for use statement. Clinical testing conducted with the subject device demonstrates that it is as safe and effective as the predicate device to support a substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.