K242515 · Shanghai United Imaging Healthcare Co., Ltd. · IZL · Oct 10, 2024 · Radiology
Device Facts
Record ID
K242515
Device Name
uDR 380i Pro
Applicant
Shanghai United Imaging Healthcare Co., Ltd.
Product Code
IZL · Radiology
Decision Date
Oct 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
uDR 380i Pro is a mobile digital radiography device intended to acquire X-ray images of the human anatomy for medical diagnosis. uDR 380i Pro can be used on both adult and pediatric patient by a qualified and trained operator. This device is not intended for mammography.
Device Story
Mobile digital radiography system; used in emergency rooms, ICUs, and wards. Consists of lifting column, telescopic cantilever, X-ray generator, tube, collimator, and wireless flat panel detectors. Operator (qualified/trained) positions device; acquires X-ray images; processes via workstation. Output: digital X-ray images for clinical diagnosis. Benefits: enables imaging for patients unable to be moved to traditional radiology rooms. Modifications include integration of specific Canon CXDI flat panel detectors and associated control software.
Clinical Evidence
Clinical image evaluation performed using the proposed device with CXDI-710C detector. A board-certified radiologist reviewed sample images of head, chest, abdomen, spine, pelvis, and extremities. Images were determined to be of sufficient quality for clinical diagnosis.
Technological Characteristics
Mobile X-ray system; 32kW/50kW generator; 40-150kV range. Detectors: 14"x17"/11"x14" Cesium Iodide scintillator on Amorphous Silicon. Pixel size 125μm. Standards: ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54, ISO 10993-5/10. Connectivity: DICOM. Software: Image acquisition, viewing, measurement, annotation, post-processing, and virtual grid.
Indications for Use
Indicated for acquiring X-ray images of human anatomy for medical diagnosis in adult and pediatric patients. Not intended for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 10, 2024
Shanghai United Imaging Healthcare Co.,Ltd. % Xin Gao Regulatory Affairs Manager No.2258 Chengbei Rd. Jiading District Shanghai, 201807 CHINA
Re: K242515
Trade/Device Name: uDR 380i Pro Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL Dated: August 21, 2024 Received: August 23, 2024
Dear Xin Gao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device
{2}------------------------------------------------
advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K242515
Device Name uDR 380i Pro
#### Indications for Use (Describe)
uDR 380i Pro is a mobile digital radiography device intended to acquire X-ray images of the human anatomy for medical diagnosis. uDR 380i Pro can be used on both adult and pediatric patient by a qualified and trained operator. This device is not intended for mammography.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark gray.
#### 510 (k) SUMMARY
K242515
## 1. Date of Preparation:
August 23, 2024
#### 2. Sponsor Identification
### Shanghai United Imaging Healthcare Co.,Ltd.
No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Establishment Registration Number: 3011015597
Contact Person: Xin GAO Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com
#### 3. Identification of Proposed Device
Trade Name: uDR 380i Pro Common Name: Mobile Digital Medical X-ray Imaging System Model(s): uDR 380i Pro
#### Regulatory Information
Classification Name: Mobile X-Ray System Classification: II Product Code: IZL Regulation Number: 21 CFR 892.1720 Review Panel: Radiology
#### 4. Identification of Predicate Device(s)
Predicate Device
Trade Name: uDR 380i Pro 510(k) Number: K222339 Device Name: Mobile Digital Medical X-ray Imaging System Manufacturer:Shanghai United Imaging Healthcare Co.,Ltd.
#### Regulatory Information
Classification Name: Mobile X-Ray System Classification: II Product Code: IZL Regulation Number: 21 CFR 892.1720 Review Panel: Radiology
# Page 1 of 6
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" symbol, which is dark blue and has a white vertical line running through the center. The logo is simple and modern in design.
#### 5. Device Description
uDR 380i Pro is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. It can be moved to different environments for an examination, like emergency room. ICU and ward. It mainly consists of a lifting column - telescopic cantilever frame system, system motion assembly, X-ray System (high voltage generator, x-ray tube, collimator and wireless flat panel detectors which have been cleared in K230175), power supply system and software for acquiring and processing the clinical images.
uDR 380i Pro is intended to acquire X-ray images for both adult and pediatric, especially for person who may not be able to be moved to a traditional RAD room. uDR 380i Pro have been previously cleared by FDA via K222339.
The system offers:
- A 14"×17" or 11"×14" flat panel detector
- A high-power, 32 kW or 50kW generator
- A maneuverable drive system
- X-ray tube-collimator assembly with flexible movement
- . Storage for detectors and supplies
- . Image Acquisition Workstation with touchscreen user interface
The modifications performed on the uDR 380i Pro (K222339) in this submission are due to the addition of flat panel detectors, including CXDI-710C and CXDI-810C, and CXDI Control Software NE which were cleared in K230175.
The device software is unchanged from the predicate device, except for the addition of CXDI Control Software NE.
The modifications, which do not affect the intended use or alter the fundamental scientific technology of the device.
#### 6. Indications for use
uDR 380i Pro is a mobile digital radiography device intended to acquire X-ray images of the human anatomy for medical diagnosis. uDR 380i Pro can be used on both adult and pediatric patient by a qualified and trained operator. This device is not intended for mammography.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in bold, sans-serif font. To the right of the words is a stylized "U" symbol. The symbol is a dark gray color and has a white line running vertically through the center.
#### 7. Comparison of Technological Characteristics with the Predicate Devices
A comparison between the technological characteristics of proposed and predicate devices is provided as below.
| Table 1 Comparison of Technology Characteristics | | | |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| ITEM | Predicate Device:<br>uDR 380i Pro<br>K222339 | Proposed Device:<br>uDR 380i Pro | Remark |
| Product Code | IZL | IZL | Same |
| Regulation<br>No. | 21 CFR 892.1720 | 21 CFR 892.1720 | Same |
| Class | II | II | Same |
| Indications<br>Use | uDR 380i Pro is a mobile<br>digital radiography device<br>intended to acquire X-ray<br>images of the human<br>anatomy for medical<br>diagnosis. uDR 380i Pro<br>can be used on both adult<br>and pediatric patient by a<br>qualified and trained<br>operator. This device is not<br>intended for<br>mammography. | uDR 380i Pro is a mobile<br>digital radiography device<br>intended to acquire X-ray<br>images of the human<br>anatomy for medical<br>diagnosis. uDR 380i Pro can<br>be used on both adult and<br>pediatric patient by a<br>qualified and trained<br>operator. This device is not<br>intended for mammography. | Same |
| Specifications | High Voltage Generator | | |
| Maximum<br>Output Power | 32kw/ 50kW | 32kw/ 50kW | Same |
| kV Range | 40~150kV | 40~150kV | Same |
| mA Range | 10~400mA/ 10~560mA | 10~400mA/ 10~560mA | Same |
| mAs Range | 0.1-630 mAs | 0.1-630 mAs | Same |
| X-ray Tube | | | |
| Anode Heat<br>Content | 300KHU | 300kHU | Same |
| Focus Size | 0.6/ 1.2 | 0.6/1.2 | Same |
| Anode Target<br>Angle | 14° | 14° | Same |
| Collimator | | | |
| Maximum<br>Light Field | 43cm × 43cm | 43cm × 43cm | Same |
| Adjustment<br>Method | Manual | Manual | Same |
| Flat Panel Detector | | | |
| Detector Size | 14" × 17"/ 11" × 14" | 14" × 17"/ 11" × 14" | Same |
| Scintillator<br>Material | Cesium Iodide | Cesium Iodide | Same |
| Semiconducto<br>r Material | Amorphous Silicon | Amorphous Silicon | Same |
| Pixel Size | 125μm | 125μm | Same |
| DQE | Typical:<br>58% @3uGy,0.5lp/mm | Typical:<br>0.58±10% @3uGy,0.5lp/mm<br>(for AR-C3543W&AR-<br>B2735W)<br>Typical:0.58±10% @2.5uGy<br>,0.5lp/mm (for CXDI-710C<br>& CXDI-810C) | Note 1 |
| MTF | Typical: 63% @ 1lp/mm<br>Typical: 35% @2lp/mm | Typical: 63% @ 1lp/mm<br>Typical: 35% @2lp/mm | Same |
| Anti-scatter Grid | | | |
| Grid size | 356mm × 445mm | 356mm × 445mm | Same |
| Grid ratio | 8:1 | 8:1 | Same |
| Telescoping Column | | | |
| X-ray Tube<br>Assembly<br>RVA | -315 $^{\circ}$ ~+315 $^{\circ}$ | -315 $^{\circ}$ ~+315 $^{\circ}$ | Same |
| X-Ray Tube<br>Assembly<br>Tilting Range | -30 $^{\circ}$ ~+90 $^{\circ}$ | -30 $^{\circ}$ ~+90 $^{\circ}$ | Same |
| | Battery | | |
| Battery<br>Capacity | 2.40kWh | 2.40kWh | Same |
| Image Acquisition Workstation | | | |
| Display Size | 19" | 19" | Same |
| Disk Size | 500GB | ≥500GB | Note 2 |
| Software function | | | |
| Image<br>Export/Import | Yes | Yes | Same |
| Image<br>Viewing | Yes | Yes | Same |
| Image<br>Measurement | Yes | Yes | Same |
| Image<br>Annotation | Yes | Yes | Same |
| Image Post-<br>processing | Yes | Yes | Same |
| Virtual grid | Yes | Yes | Same |
| Accessory | | | |
| Badge Reader | Yes | Yes | Same |
| <b>Safety</b> | | | |
| Electrical Safety | Comply with ANSI/AAMI ES 60601-1 | Comply with ANSI/AAMI ES 60601-1 | Same |
| EMC | Comply with IEC60601-1-2 | Comply with IEC60601-1-2 | Same |
| Biocompatibility | Comply with ISO10993-5 and ISO10993-10 | Comply with ISO10993-5 and ISO10993-10 | Same |
Table 1 Comparison of Technology Characteristics
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is in bold, dark blue font and is stacked on top of each other. To the right of the text is a dark blue emblem that resembles the letter "U" with a horizontal line through the middle.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is bisected by a vertical line and has a horizontal line across the middle. The logo is simple and modern.
| Justification | |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Note ID | Justification |
| Note 1 | DQE Performance is similar. When operated under the intended use, it did not raise new safety and effectiveness concerns. |
| Note 2 | The disk size of the proposed device is actually a range value, which is only a descriptive update, however the disk size can satisfy its intended use. So it did not raise new safety and effectiveness concerns. |
#### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### Non-Clinical Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications as it is Substantially Equivalent (SE) to the predicate device. UNITED IMAGING HEALTHCARE claims conformance to the following standards and guidance:
- ANSI/AAMI ES60601-1:2005& A1:2012 & A2:2021 Medical electric for basic > safety and essential performance.
- IEC 60601-1-2: 2014+AMD1: 2020,Edition 4.1,Medical electrical > equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- A IEC 60601-1-3: 2008+AMD1: 2013+AMD2: 2021,Edition 2.2, Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment.
- A IEC 60601-2-54: 2022, Edition 2.0, Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy.
- > IEC 60601-1-6:2010+AMD1:2013+AMD2: 2020, Editon 3.2, Medical electrical
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for United Imaging. The text "UNITED IMAGING" is in bold, dark blue font. To the right of the text is a stylized dark blue symbol that resembles the letter "U" with a white horizontal line through the middle.
equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- > IEC 62366-1: 2015+AMD1: 2020, Edition 1.1, Medical devices - Part 1: Application of usability engineering to medical devices
- A IEC 62304: 2006+AMD1:2015, Edition 1.1, Medical device software - Software life cycle processes.
- A IEC TS 60601 -4-2:2024 Edition 1.0, Medical electrical equipment - Part 4-2: Guidance a and interpretation- Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
#### Biocompatibility
- A ISO 10993-5: 2009, Edition 3.0, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
- A ISO 10993-10: 2010, Edition 3.0, Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization.
#### Software
- > NEMA PS 3.1-3.20(2022): Digital Imaging and Communications in Medicine (DICOM)
- > Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- A Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
#### Clinical Image Evaluation
The clinical image evaluation was performed under the proposed device with CXDI-710C. Sample image of Head, chest, abdomen, spine, pelvis, upper extremity and lower extremity were provided with a board certified radiologist to evaluate the image quality in this submission. Each image was reviewed with a statement indicating that image quality are sufficient for clinical diagnosis.
#### Summarv
The features described in this premarket submission are supported with the results of the testing mentioned above, the uDR 380i Pro was found to have a safety and effectiveness profile that is similar to the predicate device.
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device has same indications for use, similar performance, equivalence safety and effectiveness as the predicate device.
The differences above between the proposed device and predicate device do not affect the indications for use, technology characteristics, safety and effectiveness. And no issues are raised regarding to safety and effectiveness.
The proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.