K242512 · Philips Medical Systems Nederland B.V. · LLZ · Oct 24, 2024 · Radiology
Device Facts
Record ID
K242512
Device Name
ROCC Console
Applicant
Philips Medical Systems Nederland B.V.
Product Code
LLZ · Radiology
Decision Date
Oct 24, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ROCC Console is intended for remote scanning, support, review, monitoring and standardization of imaging protocols of medical imaging devices. It is a multi-vendor solution allowing view-only or full access control to connected devices. ROCC Console is also intended for training of medical personnel working on connected medical imaging devices.
Device Story
ROCC Console is a cloud-based software system enabling remote expert radiology users to connect with onsite technologists operating MR/CT scanners. It facilitates real-time operational support, guidance, and protocol management. The system captures KVM (Keyboard, Video, Mouse) signals from the scanner console, transmitting them via proprietary apps or VNC/KVM switches to a remote expert. The expert views the scanner interface and provides input to assist the onsite technologist, who retains ultimate control and can terminate access. The system supports up to three concurrent connections. It is used in healthcare facilities or remote office/home settings. By enabling remote expertise, the device improves access to specialized support, standardizes imaging protocols across sites, and provides training for medical personnel, potentially enhancing scan quality and operational efficiency.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through non-clinical performance testing, specifically software verification and validation (V&V) activities. All pre-specified acceptance criteria were met.
Technological Characteristics
Software-only device; multi-vendor compatibility. Supports MR and CT modalities. Connectivity via proprietary KVM software (Console Transmitter/Receiver Apps), VNC, or off-the-shelf HW/SW KVM switches over TCP/HTTPS/WebRTC-UDP. Operates on Windows-based desktop systems and web browsers. Features user authorization and authentication. Designed for remote session management and protocol harmonization.
Indications for Use
Indicated for healthcare professionals, including expert radiology users and onsite technologists, for remote scanning, support, review, monitoring, and protocol standardization of MR and CT imaging devices. Intended for use in clinical environments, including fixed and mobile (truck) settings. No patient-specific contraindications listed.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 24, 2024
Philips Medical Systems Nederland B.V. % Rob Vingerhoed Regulatory Affairs Manager Veenpluis 6 5684 Pc BEST, NETHERLANDS
Re: K242512
Trade/Device Name: ROCC Console Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: August 23, 2024 Received: August 23, 2024
Dear Rob Vingerhoed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
#### K242512
Device Name
ROCC Console
### Indications for Use (Describe)
ROCC Console is intended for remote scanning, support, review, monitoring and standardization of imaging protocols of medical imaging devices. It is a multi-vendor solution allowing view-only or full access control to connected devices. ROCC Console is also intended for training of medical personnel working on connected medical imaging devices.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary
Prepared on: 2024-10-23
| Contact Details | | 21 CFR 807.92(a)(1) | |
|----------------------------------------------------------|------------------------------------------------|---------------------|--|
| Applicant Name | Philips Medical Systems Nederland B.V. | | |
| Applicant Address | Veenpluis 6 5684 PC Best Netherlands | | |
| Applicant Contact Telephone | +31 622632404 | | |
| Applicant Contact | Mr. Rob Vingerhoed | | |
| Applicant Contact Email | rob.vingerhoed@philips.com | | |
| Correspondent Name | Philips Medical Systems Nederland B.V. | | |
| Correspondent Address | Veenpluis 6 5684 PC Best Netherlands | | |
| Correspondent Contact Telephone | +31 622632404 | | |
| Correspondent Contact | Mr. Rob Vingerhoed | | |
| Correspondent Contact Email | rob.vingerhoed@philips.com | | |
| Device Name | | 21 CFR 807.92(a)(2) | |
| Device Trade Name | ROCC Console | | |
| Common Name | Medical image management and processing system | | |
| Classification Name | Medical image management and processing system | | |
| Regulation Number | 892.2050 | | |
| Product Code(s) | LLZ | | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | |
| Predicate # | K232744 | | |
| Predicate Trade Name (Primary Predicate is listed first) | syngo Virtual Cockpit (VB10A) | | |
| Product Code | LLZ | | |
| Device Description Summary | | 21 CFR 807.92(a)(4) | |
This premarket notification addresses the Philips ROCC Console, a Medical Image Management and Processing System (MIMPS).
ROCC Console is a software device, which is part of the ROCC (Radiology Operations Command Center) system, a cloud-based solution which can connect expert radiology users located at a remote command center with onsite technologists operating an MR/CT scanner on-demand, while enabling real-time operational support/guidance for the scanner console. This ROCC system allows healthcare professionals to share expertise, even when they are not physically present in the same location. Expert users are working from remote office locations within a healthcare facility or from a home office.
ROCC Console is a multiple function device product. The medical device functions include the ability to console of a connected MR/CT scanner from a remote location, and a Protocol Management mode, which can be used to manage updates / harmonize scan protocols across connected MR/CT scanners.
{5}------------------------------------------------
ROCC Console is vendor neutral. It is intended to be used in healthcare environments including imaging modalities (MR or CT) with digital output (DP, DVI, and HDMI) which are either fixed to one location or are in mobile (truck) environments where medical imaging services are provided.
ROCC Console supports three connection methods to exchange the KVM (Keyboard, Video and Mouse) information between the expert user and the technologist. The proprietary software KVM connection includes two applications: a "Console Transmitter App" for use by the technologist on the scanner side, and a "Console Receiver App" for use by the remotely located expert user. The Console Transmitter App is a native Windows application installed on a desktop system, which sends the console video stream over the internet to the Console Receiver App. The Console Receiver App is a web-application that can run on a web browser, and receives and displays the video stream to the remotely located expert user edits the scanner console through mouse and keyboard entries, it sends these keyboard and mouse entries back to the Console Transmitter App, which subsequently sends the scanner console. Additionally, ROCC Console supports off-the-shelf HW/SW KVM switch or VNC. ROCC Console can be used with 3 MR/CT console connections concurrently.
When a patient is to be scanned, the on-site technologist must remain present at all times and retains full control over the MR/CT scanner console and can terminate the editing authorization at any time. For CT connections, full access is limited to what is available in the software associated with the modality workspace and is not applicable to physical switches controlling the radiation exposure.
The Protocol Management mode allows access to the MR/CT scanner console without a technologist on site. It is only intended for protocol management and harmonization and not for any remote imaging support.
# Intended Use/Indications for Use
ROCC Console is intended for remote scanning, support, review, monitoring and standardization of imaging protocols of medical imaging devices. It is a multi-vendor solution allowing view-only or full access control to connected devices. ROCC Console is also intended for training of medical personnel working on connected medical imaging devices.
# Indications for Use Comparison
The intended use / indications for use statements are near identical, except for a few minor differences in applied terminology (e.g., view-only vs read-only). As these differences have essentially the same meaning, these statements are substantially equivalent.
The description of the intended clinical users also differs in applied terminology. The description for the predicate device only refers to the user profiles as used in the accompanying Instructions for Use, whereas the description for the ROCC Console also describes the professional qualifications of the intended users. As both devices are only intended to be used by healthcare professionals with similar qualifications, and not by patients or lay persons, these statements are substantially equivalent.
The description of the intended use environment for ROCC Console is more specific, but substantially equivalent with the one for the predicate device. Both devices support flexibility in the location of use.
Therefore, it can be concluded that the intended use / indications for use for both the proposed ROCC Console are the same.
## Technological Comparison
### 21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
The technological characteristics are identical for the following aspects
- · Design (Software only)
- Multi-vendor compatibility
- · Remote use
· Remote accessibility (expert user can remotely access the supported scanner and/or only view/read the scanner console
- Length of the remote sessions
- Connectivity requirements
- · User interface functions
- · Operator interface
- · Operating system
- User authorization and authentication
- Image acquisition
- · Image processing, reporting and archiving.
The following technological characteristics are considered substantially equivalent:
• Multi-modality / Hardware: ROCC Console supports MR and CT modality which is a subset of what the predicate device supports. Because it is a subset, it is regarded substantially equivalent.
• Delay / latency: the specified maximum latency of ROCC console is slightly higher than the predicate maximum
{6}------------------------------------------------
latency of ROCC Console allows the performance and safety requirements of ROCC Console to be met. The difference with it's predicate is not significant.
• Connection protocols: both devices use VNC over TCP. ROCC Console also uses HW KVM over TCP for one of it's connection methods. This means the TCP protocol is used by both devices use HTTPS connection protocols even though the data being exchanged is different, the use of the protocol is identical. The predicate device uses different terminology for the same connection protocol (UDP versus WebRTC-UDP for ROCC Console). Therefore this is substantially equivalent. · Lastly: the ROCC system communication features offered are substantially equivalent.
Substantial equivalence was demonstrated with non-clinical performance testing. V&V) activities were performed on the proposed ROCC Console and demonstrated that the predetermined acceptance criteria were successfully met and no different questions in terms of safety and effectiveness were raised. The non-clinical performance tests provided in this 510(k) submission demonstrate that the proposed ROCC Console is as safe and effective as the predicate device, the syngo Virtual Cockpit (VB10A) (K232744).
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Please see software verification and validation documents to support the performance of ROCC Console.
Not applicable.
All pre-specified performance metrics have been met as demonstrated within the software verification and validation. The ROCC Console performs as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.