K242485 · Breas Medical AB · MOD · May 20, 2025 · Anesthesiology
Device Facts
Record ID
K242485
Device Name
EveryWare
Applicant
Breas Medical AB
Product Code
MOD · Anesthesiology
Decision Date
May 20, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
EveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. EveryWare is intended to be used in hospital, institutional, provider, and home care settings.
Device Story
EveryWare is a cloud-based medical device data management platform. It collects treatment data from compatible Breas respiratory devices via a cellular modem (iLink) connected via USB or other data transfer methods (SD card/file upload). Data is transmitted to a secure cloud-hosted system. Authorized healthcare representatives (physicians, clinicians, DME providers) access the platform via a web browser to view patient data, generate reports, and remotely adjust prescription or performance settings on non-life support therapy devices. The device does not automate diagnosis or therapy, nor does it alter the intended use of connected ventilators. It facilitates clinical follow-up and management of therapy settings, potentially improving patient care efficiency by allowing remote adjustments and monitoring.
Clinical Evidence
No clinical data. Bench testing only, including software verification and validation, cybersecurity assessment, and human factors/usability engineering in accordance with FDA guidance and consensus standards (IEC 62304, IEC 60601-1, IEC 60601-1-2).
Technological Characteristics
Web-based graphical user interface; cloud-hosted data storage; cellular modem (iLink) connectivity via USB. Supports non-life support therapy devices (product codes MNS, MNT, NOU, CBK). Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 62304 (software lifecycle).
Indications for Use
Indicated for patients prescribed compatible non-life support therapy devices. Used by healthcare representatives (physicians, nurses, respiratory therapists, physiotherapists, technicians, DME/HCP/HMP providers) in hospital, institutional, and home care settings to manage patient data and adjust prescription/performance settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Respironics Inc. Care Orchestrator (Sapphire) (K152356)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 20, 2025
Breas Medical AB
Ivan Liljegren
Senior Vice President, Global QA/RA
Foretagsvagen 1
Molnlycke, 45333
Sweden
Re: K242485
Trade/Device Name: EveryWare
Regulation Number: 21 CFR 868.5895
Regulation Name: Continuous ventilator
Regulatory Class: Class II
Product Code: MOD, MNS, MNT, NOU, CBK
Dated: April 14, 2025
Received: April 16, 2025
Dear Ivan Liljegren:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242485 - Ivan Liljegren
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K242485 - Ivan Liljegren
Page 3
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
Submission Number (if known)
K242485
Device Name
EveryWare
Indications for Use (Describe)
EveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. EveryWare is intended to be used in hospital, institutional, provider, and home care settings.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) SUMMARY - K242485
Breas Medical
EveryWare
510(k) Owner
Breas Medical AB
Foretagsvagen 1
Molnlycke, Vastra Gotaland SWEDEN 435 33
Telephone: +46 31868800
Submission Correspondent
Ivan Liljegren
Sr. Vice President, Global QARA
Breas Medical AB
Foretagsvagen 1, Molnlycke,
Vastra Gotalands Lan (Se-14), Sweden 453 33
Date Prepared:
May 15, 2025
Trade Names of Device
EveryWare
Common or Usual Name
(Accessory to) Continuous Ventilator
Classification Name
Regulation: 21 C.F.R. §868.5895 Continuous Ventilator, Class II
Primary Predicate Device(s)
ResMed Ltd. AirView cleared in K151901
Secondary Predicate Device(s)
Respironics Inc. Care Orchestrator (Sapphire) cleared in K152356
Device Description
EveryWare is a platform that gathers treatment information from Breas respiratory devices in a secure cloud-based system. The data from ventilator is sent over a cellular modem through a cellular network to a cloud-hosted secure data storage system. The authenticated users access this cloud hosted system from a browser in their computer.
EveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by
Page 1 of 7
{5}
healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. EveryWare is intended to be used in hospital, institutional, provider, and home care settings. EveryWare securely connects compatible medical devices located at the point of patient care to the cloud, and provides authorized healthcare representatives the means to manage patient and device information and settings. EveryWare does NOT alter the intended use of connected medical devices or provide functions to automate diagnosis or therapy.
## Indications for Use
EveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. EveryWare is intended to be used in hospital, institutional, provider, and home care settings.
## Indications for Use Comparison
The Indications for Use for EveryWare fall fully within the indications for use of the primary predicate device and the secondary predicate device.
## Technological Characteristics Compared to Predicate
Similar to ResMed AirView and Respironics Care Orchestrator, EveryWare is a medical device data management system comprised of a web-based graphical user interface and a component for wireless connection to the compatible therapy devices. Compatible therapy devices for EveryWare only include devices belonging to the MNS, MNT, NOU and CBK product codes, whereas AirView and Care Orchestrator are also compatible with therapy devices that are in BZD product code.
Aside from supporting clinicians to manage data on patients who are prescribed with compatible therapy devices in accordance with the intended use of these therapy devices, both EveryWare, AirView and Care Orchestrator allow settings in the compatible non-life support therapy devices to be changed remotely. EveryWare does not allow ventilation modes to be changed remotely.
The wireless connection of the therapy devices to AirView is through a cellular modem that can either be embedded in the therapy device, or provided by means of a separate connectivity module. The physical component for wireless connection of the therapy devices to Care Orchestrator is embedded within the therapy device itself, and can either be a bluetooth or cellular modem wireless connection. EveryWare uses a cellular modem called iLink that needs to be connected via USB to the therapy device.
The following table presents similarities and differences with the predicates. There are no additional technological features that the EveryWare has which are not already cleared in AirView and/or Care Orchestrator.
Page 2 of 7
{6}
| Characteristic | Breas Medical AB, EveryWare (subject device) | ResMed Ltd. AirView (K151901) | Respironics, Inc Care Orchestrator (K152356) | Discussion of Differences |
| --- | --- | --- | --- | --- |
| Intended Use | EveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. | AirView is a web based solution for healthcare specialists intended to assist in the diagnosis of sleep disordered breathing in adult patients through analysis of data recorded by an AirView compatible home sleep device. Transfer and display machine and therapeutic information that has been transmitted remotely from the patient's therapy device. It is intended to support the standard follow - up care of patients that have been prescribed a compatible ResMed therapy device. AirView also provides remote settings capabilities for non-life support devices only. | Sapphire [Care Orchestrator] is intended to support clinicians by tracking data on patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. Sapphire [Care Orchestrator] provides remote patient data collection & viewing and is intended to be used by healthcare representatives (e.g.,Physicians, Clinicians, Durable Medical Equipment providers) in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. | Similar |
| Environment of Use | EveryWare is intended to be used in hospital, institutional provider, and home care settings. | Healthcare professional's computer; Clinical users, home care providers (HCPs) and other healthcare specialists. | Sapphire [Care Orchestrator] is intended to be used in hospital, institutional provider, and home care settings. | Similar |
{7}
| Characteristic | Breas Medical AB, EveryWare (subject device) | ResMed Ltd. AirView (K151901) | Respironics, Inc Care Orchestrator (K152356) | Discussion of Differences |
| --- | --- | --- | --- | --- |
| User Population | EveryWare is intended to be used by the following 3 user groups:- Prescribing physicians, nurses, respiratory therapists, physiotherapists and technicians-Durable Medical Equipment (DME)/Homecare providers (HCPs)/Home Medical Equipment providers (HMPs) including managers and administrators | Clinical users, home care providers (HCPs) and other healthcare specialists | Healthcare representatives (e.g., Physicians, Clinicians, Durable Medical Equipment providers) and patients | Similar |
| Application Type | Web-based application | Web-based application | Web-based application | Same |
| Product Codes Supported by Remote Prescription Change Functionality | MNS MNT | MNS BZD | MNS MNT BZD | Similar to K152356 with respect to compatible ventilators supported. |
| Product Codes NOT SUPPORTED by Remote Prescription Change Functionality | NOU CBK | CBK MNR | NOU CBK | Similar |
| Data Transfer Technology | Wireless (cellular modem) SD Card File upload/Internet | Wireless (cellular modem) SD Card File upload/Internet USB Stick/Internet | Wireless (Bluetooth or cellular modem) SD Card/Internet | Similar |
{8}
| Characteristic | Breas Medical AB, EveryWare (subject device) | ResMed Ltd. AirView (K151901) | Respironics, Inc Care Orchestrator (K152356) | Discussion of Differences |
| --- | --- | --- | --- | --- |
| Functionality | Centralized database Patient Management Display therapy data Generate reports Settings management for non-life supporting devices | Centralized database Patient Management Display therapy data Generate reports Settings management for non-life supporting devices | Centralized database Patient Management Display therapy data Generate reports Settings management for non-life supporting devices | Same |
| Therapy Settings Management (for non-life support devices), including: | Performance Settings Prescription | Performance Settings Prescription Mode | Performance Settings Prescription Mode | Similar. The absence of Therapy Mode Change does not raise new questions of safety and effectiveness. |
| - Therapy Mode Change | No | Yes | Yes | Different The absence of Therapy Mode Change does not raise new questions of safety and effectiveness. |
| - Target Volume Setting | Yes | Yes | Yes | Same |
| - IPAP & EPAP Settings | Yes | Yes | Yes | Same |
| - Breath Rate and Insp. Time Setting | Yes | Yes | Yes | Same |
| - Rise Time & Ramp Pressure Setting | Yes | Yes | Yes | Same |
| - Humidifier Setting | No | Yes | Yes | Different The absence of Humidifier Setting does not raise new questions of safety and effectiveness. |
Page 5 of 7
{9}
Page 6 of 7
| Characteristic | Breas Medical AB, EveryWare (subject device) | ResMed Ltd. AirView (K151901) | Respironics, Inc Care Orchestrator (K152356) | Discussion of Differences |
| --- | --- | --- | --- | --- |
| - High Minute Volume, Disconnection & Apnea alarm | Yes | No | Yes | Similar to K152356 |
| - Mask Resistance & Mask Resistance Lock | No | No | Yes | Similar to K151901. |
| - View Optional Screen | No | No | Yes | Similar to K151901. |
| - Tubing Type & Tubing Type Lock | No | No | Yes | Similar to K151901. |
| - Automated Airway Management | No | No | Yes | Similar to K151901. |
| Reports | Detailed report (includes compliance information) | Detailed report (includes compliance information) | Detailed report (includes compliance information) | Same |
| Labeling | Online help file within the application | Online help file within the application | Online help file within the application | Same |
| Viewing of data | Interactive viewing of data based on user selected date range | Interactive viewing of data based on user selected date range | Interactive viewing of data based on user selected date range | Same |
## Non-clinical Testing
EveryWare was designed and subjected to performance testing in accordance with all applicable requirements, specifications, guidance documents and recognized consensus standards:
- Content of Premarket Submissions for Device Software Functions. Guidance for Industry and Food and Drug Administration Staff (June 14, 2023)
- Off-the-Shelf Software Use in Medical Devices. Guidance for Industry and Food and Drug Administration Staff (August 11, 2023)
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submission. Guidance for Industry and Food and Drug Administration Staff (September 27, 2023)
{10}
- Applying Human Factors and Usability Engineering to Medical Devices. Guidance for Industry and Food and Drug Administration Staff (February 3, 2016)
- IEC 60601-1 Edition 3.2:2020
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Edition 4.1:2020
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 62304 Edition 1.1 2015
Medical device software - Software life cycle processes
EveryWare, by itself, is not intended for diagnosis and treatment of disease or medical condition. EveryWare does not change the performance specification of the connected therapeutic devices. Therefore, no further performance data is required to demonstrate substantial equivalence.
## Conclusion
EveryWare is substantially equivalent to the AirView and Care Orchestrator (Sapphire), as the devices share a common intended use and technological characteristics as demonstrated through performance testing.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.