ArgenZ MTZ zirconia can be used for the production of full-contour and substructure restorations up to and including a full arch. All blanks are processed by dental laboratories or by dental professionals
Device Story
ArgenZ MTZ is a multi-translucent, multi-layer yttria-stabilized zirconia disc; contains oxides of Hafnium, Aluminum, and coloring oxides (iron, erbium, cobalt, manganese). Device is a non-sterile, single-use blank processed by dental laboratories or professionals using CAD/CAM milling systems to fabricate dental restorations. Material composition varies by layer (5Y zirconia top, 4Y zirconia bottom, with intervening layers of both); color intensity and yttria content vary from incisal to dentine. Output is a custom-milled dental restoration. Benefits include aesthetic, durable, full-contour or substructure dental prosthetics.
Clinical Evidence
Bench testing only. No clinical or animal data provided. Device performance evaluated against ISO 6872:2015 (Dentistry – Ceramic materials) requirements, demonstrating flexural strength ≥ 800 MPa and chemical solubility < 100 µg/cm². Biocompatibility evaluated per ISO 10993-1, including cytotoxicity testing per ISO 10993-5:2009.
Technological Characteristics
Yttria-stabilized zirconia (4 mole and 5 mole percent). Disc form factor (10mm to 30mm thickness). Complies with ISO 6872:2015 (Type II, Class 5). Flexural strength ≥ 800 MPa; chemical solubility < 100 µg/cm²; radioactivity ≤ 1.0 Bq·g⁻¹. Non-sterile. Processed via CAD/CAM milling.
Indications for Use
Indicated for the production of full-contour and substructure dental restorations, including full-arch restorations, for use by dental laboratories or dental professionals.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 24, 2024
Argen Corporation % Patsy Trisler Owner/Regulatory Consultant Trisler Consulting (dba) 306 Turnberry Ct Lebanon, Indiana 46052
Re: K242458
Trade/Device Name: ArgenZ MTZ Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: August 19, 2024 Received: August 19, 2024
Dear Patsy Trisler:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K242458
Device Name
ArgenZ MTZ
Indications for Use (Describe)
ArgenZ MTZ zirconia can be used for the production of full-contour and substructure restorations up to and including a full arch.
All blanks are processed by dental laboratories or by dental professionals
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) Number: | K242458 |
|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER Name: | The Argen Corporation |
| Address: | 5855 Oberlin Drive<br>San Diego, CA 92121-3706 |
| Contact Person: | Paul Cascone |
| Phone: | 858.455.7900 |
| Date Prepared: | October 17, 2024 |
| DEVICE Trade<br>Name: | ArgenZ MTZ |
| Common Name: | Dental Zirconia Ceramics |
| Classification Name | Porcelain powder for clinical use |
| Number | 21 CFR 872.6660 |
| Product Code | EIH |
| Regulatory Class | 2 |
| Review Panel | Dental |
| PRIMARY<br>PREDICATE: | K222626, Dental Zirconia Ceramic, Aidite (Qinhuangdao)<br>Technology Co., Ltd |
| REFERENCE<br>DEVICES: | K150919, ArgenZ Esthetic Plus, The Argen Corporation |
| | K190079, ArgenZ HT+, The Argen Corporation |
| DEVICE<br>DESCRIPTION | The ArgenZ MTZ (referring to multi-translucent, multi-layer)<br>zirconia consist of yttria-stabilized pre-shaded 4 mole and 5<br>mole percent zirconia. |
| | They are manufactured (pressed and sintered) into disc forms<br>and offered multiple thicknesses and multiple shades. They<br>contain oxides of Hafnium and Aluminum and minute amounts<br>of coloring oxides of iron, erbium, cobalt and manganese. |
| | The discs are single use, non-sterile devices and used for<br>fabricating dental restorations, using CAD/CAM systems that<br>can accommodate the disc sizes. |
| SUMMARY OF<br>TECHNOLOGICAL<br>CHARACTERISTICS | The zirconia material complies with ISO 6872:2015 as a type II,<br>meeting the requirements for a class 5 material, with:<br>• flexural strength above 800 MPa and<br>• chemical solubility of less than 100 µg/cm². |
| | The physical device is a disc with thicknesses that vary from 10<br>mm to 30mm. Within a given thickness the top layer is made of<br>5 Y Zirconia, the bottom layer from 4 Y zirconia with the<br>intervening layers different combinations of 5Y and 4Y. |
| | The color intensity increases going from incisal to dentine and<br>the yttria content decreases from incisal to dentine. |
| INDICATIONS FOR<br>USE STATEMENT | ArgenZ MTZ zirconia can be used for the production of full-<br>contour and substructure restorations up to and including a full<br>arch.<br>All blanks are processed by dental laboratories or by dental<br>professionals. |
| SAFETY TESTING | The device's biocompatibility was evaluated according to ISO<br>10993, Biological evaluation of medical devices – Part 1:<br>Evaluation and testing within a risk management system. |
| | According to the Biological Risk Assessment, Cytotoxicity<br>testing (per ISO 10993-5:2009) was conducted. |
| PERFORMANCE<br>TESTING | Bench testing, performed according to ISO 6872:2015<br>Dentistry – Ceramic materials, showed the ArgenZ MTZ meets<br>the requirements specified in the standard. The test<br>measurements are in the SE table below. |
| | In vivo Animal and Human Clinical performance testing were not<br>provided for this submission. |
| COMPARISON TO<br>THE PREDICATE<br>DEVICE<br>and<br>SUBSTANTIAL<br>EQUIVALENCE<br>CONCLUSION | The information and data provided in this 510(k) establish that<br>the ArgenZ MTZ zirconia device is substantially equivalent to<br>the Aidite primary predicate device's Multilayer 3D-pro model in<br>the intended use, design, principle of operation, and technology<br>including the use of the same zirconia materials. |
| | Further ArgenZ MTZ's material is a combination of the two<br>reference devices, Argen's K150919 (consisting of 5 mole<br>percent zirconium) and K190079 (consisting of 4 mole percent<br>zirconium); and ArgenZ MTZ is manufactured using the same<br>processes. |
| | The differences in the Indications for Use statement between<br>ArgenZ MTZ and the primary predicate does not change the<br>Intended Use – both are zirconia blanks for use in making<br>dental restorations with CAD/CAM technologies. |
| | For complete presentation of the similarities and differences,<br>see the following SE Comparison table. |
{5}------------------------------------------------
{6}------------------------------------------------
## Substantial Equivalence Comparison Table
| Comparison item | Proposed Device (K242458) | Predicate Device (K222626) | Comment |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Manufacturer | The Argen Corporation | Aidite (Qinhuangdao) Technology<br>Co., Ltd. | None |
| Product Name | ArgenZ MTZ | Dental Zirconia Ceramic | None |
| Product Code | EIH | EIH | Same |
| Regulation<br>Number | 21 CFR § 872.6660 | 21 CFR § 872.6660 | Same |
| Classification | Class II | Class II | Same |
| Prescription Use | Yes | Yes | Same |
| Indications for<br>Use | ArgenZ MTZ zirconia can be used<br>for the production of full-contour<br>and substructure restorations up<br>to and including a full arch.<br>All blanks are processed by<br>dental laboratories or by dental<br>professionals. | Dental Zirconia Ceramic are used<br>for dental restorations using<br>different CAD/CAM or manual<br>milling machines.<br>All blanks are processed through<br>dental laboratories or by dental<br>professionals. | Same Intended<br>Use<br>Similar<br>Indications |
| Class (per<br>ISO 6872:<br>2015) | Class 5 | Class 5: Multilayer-3D pro | Same |
| Composition | Based on yttria-stabilized zirconia<br>5Y zirconia on top of 4Y zirconia | Based on yttria-stabilized zirconia<br>5Y zirconia on top of 4Y zirconia | Same |
| Color | Color | Color | Similar |
| Intended User | Professional dental technicians | Professional dental technicians | Same |
| Single Use | Yes | Yes | Same |
| Sterile | Non-sterile | Non-sterile | Same |
| Physical<br>Properties | Conform to ISO 6872:2015 | Conform to ISO 6872:2015 | Same |
| Uniformity | Uniform | Uniform | Same |
| Freedom from<br>extraneous<br>materials | Free from extraneous materials | Free from extraneous materials | Same |
| Radioactivity | ≤ 1.0 Bq·g⁻¹ | ≤ 1.0 Bq·g⁻¹ | Same |
| Flexural strength | ≥ 800 MPa | Multilayer-3Dpro: ≥800 MPa | Similar |
| Chemical<br>solubility | < 100 µg/cm² | < 100 µg/cm² | Same |
| Linear thermal<br>expansion<br>coefficient | (9.9)×10⁻⁶ K⁻¹ | Multilayer-3D pro:<br>(10.4±0.5)×10⁻⁶ K⁻¹ | Similar |
| Biocompatibility | Conforms to ISO 10993-1 | Conforms to ISO 7405:2018 | Similar |
| Labeling | Complies with 21 CFR 801 | Complies with 21 CFR 801 | Same |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.