HemoSphere Alta Advanced Monitoring Platform (ALTAALL1/ALTCR1/ALTASR1); HemoSphere Alta- monitor pressure cable (HEMAPSC200);Acumen AFM cable- HemoSphere Alta monitor (HEMAFM100)
Applicant
Edwards Lifesciences
Product Code
DYG · Cardiovascular
Decision Date
Dec 9, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1240
Device Class
Class 2
Attributes
AI/ML, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Right Ventricular Cardiac Output
Machine learning algorithm
—
—
—
—
—
—
Indications for Use
The HemoSphere Alta Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The HemoSphere Alta Advanced Monitoring Platform is intended for use with compatible Edwards Swan- Ganz/Swan-Ganz Jr/Swan-Ganz IQ catheters and oximetry catheters. FloTrac Jr sensors. Acumen IO sensors, TruWave DPTs, ForeSight /ForeSight Jr sensors/ForeSight IQ sensor, Acumen IQ fluid meter and ClearSight/ClearSight Jr/Acumen IQ finger cuffs.
Device Story
HemoSphere Alta platform visualizes hemodynamic/volumetric data; integrates Edwards critical care technologies. Inputs: pressure waveforms (arterial, pulmonary artery, right ventricular), thermodilution signals, oximetry data (SvO2, ScvO2, StO2), and fluid meter flow rates. Transforms inputs via algorithms (e.g., RVCO, GHI, Smart Wedge, HPI, CAI, AFM) to provide continuous/intermittent hemodynamic parameters. Used in OR, ICU, ER by clinicians. Output displayed on monitor; assists in goal-directed therapy, fluid management, and hemodynamic optimization. Benefits: provides physiological insight into patient status, future instability, and fluid responsiveness; aids clinical decision-making regarding hemodynamic support.
Clinical Evidence
No new clinical trials performed. Bench testing, system integration, and mechanical testing conducted. Usability study performed per FDA human factors guidance. Clinical waveform data collected to support design and validation of the RVCO algorithm. Electrical safety and EMC compliance verified per IEC 60601 series standards.
Technological Characteristics
Programmable diagnostic computer; integrates hemodynamic monitoring modules. Materials: medical-grade plastics/metals. Connectivity: wired sensors/cables (pressure, oximetry, AFM). Energy: AC power. Software: ML-based and rule-based algorithms. Sterilization: N/A (monitor is capital equipment; sensors/cables are sterile/single-use). Standards: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-8, 62304, 62366-1, 60601-2-34, 60601-2-57, 60601-2-49, 80601-2-49.
Indications for Use
Indicated for adult and pediatric critical care patients in hospital environments requiring monitoring of cardiac output, hemodynamic parameters, venous oxygen saturation, tissue oxygen saturation, and blood pressure. Specific features indicated for surgical or non-surgical patients receiving advanced hemodynamic monitoring, including hypotension prediction, fluid management, and right ventricular pressure/cardiac output assessment. Contraindications: Refer to specific catheter/sensor labeling for target population limitations.
Regulatory Classification
Identification
A flow-directed catheter is a device that incorporates a gas-filled balloon to help direct the catheter to the desired position.
Predicate Devices
HemoSphere Alta Advanced Monitoring Platform (K232294)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 9, 2024
Edwards LifeSciences Varad Raghuwanshi Director, Regulatory Affairs One Edwards Way Irvine. California 92614
Re: K242451
Trade/Device Name: HemoSphere Alta Advanced Monitoring Platform (ALTAALL1/ALTCR1/ALTASR1); HemoSphere Alta- monitor pressure cable (HEMAPSC200); Acumen AFM cable- HemoSphere Alta monitor (HEMAFM100) Regulation Number: 21 CFR 870.1425 Regulation Name: Programable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, DQE, QAQ, MUD, DXN, DSB. FLL, QMS, QNL, QEM Dated: August 16, 2024 Received: August 16, 2024
Dear Varad Raghuwanshi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242451
#### Device Name
HemoSphere Alta Advanced Monitoring Platform (ALTAALL1/ALTCR1/ALTASR1); HemoSphere Alta- monitor pressure cable (HEMAPSC200); Acumen AFM cable- HemoSphere Alta monitor (HEMAFM100)
#### Indications for Use (Describe)
HemoSphere Alta™ Advanced Monitor with Swan-Ganz Technology
The HemoSphere Alta monitor when used with the HemoSphere Alta Swan-Ganz patient cable and Edwards Swan-Ganz catheters is indicated for use in adult and petical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed in a hospital environment. Refer to the Edwards Swan-Ganz Ir catheter indications for use statement for information on target patient population specific to the catheter being used.
The Global Hypoperfusion Index (GHI) algorithm provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability. The GHI algorithm is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring with the Swan-Ganz catheter. The GHI algorithm is considered to provide additional information regarding the patient's predicted future risk for clinical deterioration, as well as identifying patients at low risk for deterioration. The product predictions are for reference only and no therapeutic decisions should be made based solely on the GHI algorithm predictions.
When used in combination with a Swan-Ganz catheter connected to a pressure transducer, the Edwards Lifesciences Smart Wedge algorithm measures and provides pulmonary artery occlusion pressure and assesses the quality of the pulmonary artery occlusion pressurement. The Smart Wedge algorithm is indicated for use in critical care patients over 18 years of age receiving advanced hemodynamic monitoring. The Smart Wedge algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Smart Wedge algorithm parameters.
HemoSphere Alta Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere Alta monitor when used with the HemoSphere oximetry cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
HemoSphere Alta Advanced Monitor with HemoSphere Pressure Cable or HemoSphere Alta Monitor Pressure Cable
The HemoSphere Alta monitor when used with the HemoSphere Pressure Cable or HemoSphere Alta monitor Pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, FloTrac Jr, Acumen IQ, and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
The Edwards Lifesciences Acumen Hypotension Prediction Index software feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring.
{4}------------------------------------------------
The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
When used in combination with the Swan-Ganz technology connected to a compatible Swan-Ganz catheter, the Edward Lifesciences Right Ventricular Pressure (RVP) algorithm provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVP algorithm is indicated for critically ill patients over 18 years of age receiving advanced hemodynamic monitoring in the operating room (OR) and intensive care unit (ICU). The RVP algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Right Ventricular Pressure (RVP) parameters.
When used in combination with the HemoSphere Pressure Cable connected to a compatible Swan-Ganz catheter, the Right Ventricular Cardiac Output (RVCO) feature provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVCO algorithm is intended for use in surgical patients over 18 years of age that require advanced hemodynamic monitoring. The Right Ventricular Cardiac a continuous cardiac output and derived parameters.
The Cerebral Adaptive Index (CAI) Algorithm is an informational index to help assess the level of coherence or lack thereof between Mean Arterial Pressure (MAP) and the Absolute Levels of Blood Oxygenation Saturation (StO2) in patient's cerebral tissue. MAP is acquired by the HemoSphere pressure cable or HemoSphere Alta Pressure Cable and StO2 is acquired by the ForeSight oximeter cable. CAI is intended for use in patients over 18 years of age receiving advanced hemodynamic monitoring. CAI is not indicated to be used for treatment of any disease or condition and no therapeutic decisions should be made based solely on the Cerebral Adaptive Index (CAI) Algorithm.
HemoSphere Alta Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor
The Acumen assisted fluid management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy.
Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the assisted fluid management suggestions.
# Acumen IQ Fluid Meter
The Acumen IQ fluid meter is a sterile single use device that is intended to be used with the HemoSphere Alta AFM cable and AFM software feature to inform the user of the rate of flow. The device is intended to be used by qualified personnel or clinicians in a clinical setting for up to 24 hours.
HemoSphere Alta Advanced Monitor with ForeSight Oximeter Cable
The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states. The ForeSight oximeter cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight oximeter cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere Alta monitor.
• When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents ≥40 kg.
· When used with Medium Sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects ≥3 kg.
· When used with Small Sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects < 8 kg and non-cerebral use on pediatric subjects <5kg.
{5}------------------------------------------------
The Edwards Algorithm for Measurement of Blood Hemoglobin is indicated for continuously monitoring changes to hemoglobin concentration in the circulating blood of adults and transitional adolescents ≥ 40 kg receiving advanced hemodynamic monitoring using HemoSphere ForeSight Oximeter Cable and ForeSight IQ Sensors in cerebral locations.
HemoSphere Alta Advanced Monitor with ClearSight Technology
The HemoSphere Alta Monitor when used with the HemoSphere ClearSight technology, pressure controller and a compatible Edwards finger cuff are indicated for Adult and Pediatric patients of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere Alta monitor and compatible Edwards' finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acure is intended for use in surgical patients or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Edwards. The logo features a large, white letter "E" superimposed on a gray square grid. The word "Edwards" is written in a simple, sans-serif font below the square.
# 510(k) Summary – HemoSphere Alta Advanced Monitoring Platform
#### I. Submitter:
| Sponsor: | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614 | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Establishment<br>Registration<br>Number: | 2015691 | |
| Contact Person: | Varad Raghuwanshi<br>Director, Regulatory Affairs<br>One Edwards Way<br>Irvine, CA 92614<br>Telephone: (949) 756-4502<br>Fax: (949) 809-2972 | |
| Date Prepared: | December 5, 2024 | |
| II.<br>Device Information: | | |
| Platform Name<br>(Name of the Device) | HemoSphere Alta Advanced Monitoring Platform (ALTAALL1/<br>ALTCR1/ALTASR1) | |
| Trade Name: | HemoSphere Alta Advanced Cardiac Monitor<br>HemoSphere Alta Advanced Smart Recovery Monitor<br>HemoSphere Alta All-on-One monitor<br>HemoSphere Alta- Monitor Pressure Cable (HEMAPSC200)<br>HemoSphere Alta AFM Cable (HEMAFM100)<br>Right Ventricular Cardiac Output Feature | |
| Common Name: | Programmable Diagnostic/Oximetry/Ejection Fraction/ Noninvasive blood<br>pressure measurement computer/Adjunctive predictive Cardiovascular<br>Indicator/Adjunctive Open Loop Fluid Therapy Recommender | |
| Classification<br>Name: | Programmable Diagnostic Computer<br>Fiberoptic Oximeter Catheter<br>Adjunctive Predictive Cardiovascular Indicator<br>Oximeter, Tissue Saturation<br>Noninvasive blood pressure measurement system | 21 CFR 870.1425<br>21 CFR 870.1230<br>21 CFR 870.2210<br>21 CFR 870.2700<br>21 CFR 870.1130 |
| K242451 | Thermometer, Electronic, Clinical | 21 CFR 880.2910 |
| | Adjunctive Open Loop Fluid Therapy Recommender | 21 CFR 870.5600 |
| | Medium-Term Adjunctive Predictive Cardiovascular Indicator | 21 CFR 870.2210 |
| | System, Catheter or Guidewire, Steerable (Magnetic)<br>Steerable catheter control system | 21 CFR 870.1290 |
| Image: Edwards Logo | | |
21 CFR 870.2770
{7}------------------------------------------------
| Product Code<br>and Regulatory<br>Class: | DQK, | Class II |
|------------------------------------------|------|----------|
| | DQE, | Class II |
| | QAQ, | Class II |
| | MUD, | Class II |
| | DXN, | Class II |
| | DSB, | Class II |
| | FLL, | Class II |
| | QMS, | Class II |
| | QNL, | Class II |
| | QEM, | Class II |
#### III. Predicate Device
| Primary<br>Predicate Device: | - HemoSphere Alta Advanced Monitoring Platform manufactured by<br>Edwards Lifesciences, K232294, cleared on October 31st, 2023. It<br>is the base device on which the software modifications have been<br>implemented. It has been utilized for substantial equivalence in<br>terms of the graphical user interface (GUI) used, indication for use,<br>intended use, technological characteristics, cardiac output<br>algorithms, accessories, basic device functionality, new<br>configuration Pressure cable (HEMAPSC200), optional feature to<br>use pulse rate for associated derived parameter calculation, updated<br>GHI algorithm and the subject Right Ventricular Cardiac Output<br>(RVCO) Algorithm. |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Additional<br>Predicate<br>Devices: | - HemoSphere Advanced Monitoring Platform by Edwards<br>Lifesciences, K223865, cleared on June 9th, 2023, utilized for AFM<br>automated fluid tracking and the automatic zeroing of the Heart<br>reference sensor feature.<br>- Cerebral Adaptive Index (CAI) algorithm by Edwards Lifesciences,<br>K223651, cleared on May 26, 2023, utilized for the Cerebral<br>Adaptive Index (CAI) parameter.<br>- Edwards Algorithm for Measurement of Blood Hemoglobin by<br>Edwards Lifesciences, K230612, cleared on November 17th, 2023,<br>utilized for the Total Blood Hemoglobin (tHb) parameter. |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a logo for a company or brand named "Edwards". The logo features a large, stylized letter "E" in white, set against a gray, grid-like background. The word "Edwards" is written in a simple, sans-serif font below the graphic.
- Smart Wedge algorithm by Edwards Lifesciences, K230579 cleared on August 17th, 2023, utilized for the Smart Wedge functionality.
- Acumen Hypotension Prediction Index (HPI) algorithm by -Edwards Lifesciences, K230057, cleared on June 8th, 2023, utilized for the HPI Adjustable MAP (Hypotension threshold) feature.
- -Swan-Ganz Catheters, FloTrac sensors, ClearSight finger cuffs, HemoSphere Advanced Monitoring Platform by Edwards Lifesciences, K231248, cleared on September 21, 2023, for Pediatric patients > 12 years of age Indication expansion when used with Swan-Ganz catheters, FloTrac sensors, ClearSight finger cuffs.
#### IV. Device Description
Device The HemoSphere Alta Advanced Monitoring Platform is Edwards' next-Description: generation platform that provides a means to interact with and visualize hemodynamic and volumetric data on a screen. It incorporates a comprehensive view of patient hemodynamic parameters with an intuitive and easy user interface. The HemoSphere Alta Advanced Monitoring Platform is designed to provide monitoring of cardiac flow with various core technologies coupled with other technologies-based features such as Algorithms and Interactions. It integrates Edwards existing Critical Care technologies into a unified platform.
> The Right Ventricular Cardiac Output (RVCO) feature is a machinelearning algorithm that calculates and displays continuous cardiac output (CORV) from the right ventricle using as inputs the right ventricular pressure waveform and derived right ventricular pressure parameters such as SYSRVF, DIARVP, MRVP, RVEDP, PRRV and Max RV dP/dt from the existing Right Ventricular Pressure (RVP) algorithm and if available, intermittent cardiac output (iCO).
#### V. Indications for Use:
# HemoSphere Alta Advanced Monitor with Swan-Ganz Technology
The HemoSphere Alta monitor when used with the HemoSphere Alta Swan-Ganz patient cable and Edwards Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Edwards. The logo features a large, white letter "E" superimposed on a gray square. Behind the "E", there are faint white circles and lines, creating a subtle geometric pattern. Below the square, the word "Edwards" is written in a simple, sans-serif font.
blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter and Swan-Ganz Jr catheter indications for use statement for information on target patient population specific to the catheter being used.
The Global Hypoperfusion Index (GHI) algorithm provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability. The GHI algorithm is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring with the Swan-Ganz catheter. The GHI algorithm is considered to provide additional information regarding the patient's predicted future risk for clinical deterioration, as well as identifying patients at low risk for deterioration. The product predictions are for reference only and no therapeutic decisions should be made based solely on the GHI algorithm predictions.
When used in combination with a Swan-Ganz catheter connected to a pressure cable and pressure transducer, the Edwards Lifesciences Smart Wedge algorithm measures and provides pulmonary artery occlusion pressure and assesses the quality of the pulmonary artery occlusion pressure measurement. The Smart Wedge algorithm is indicated for use in critical care patients over 18 years of age receiving advanced hemodynamic monitoring. The Smart Wedge algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Smart Wedge algorithm parameters.
# HemoSphere Alta Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere Alta monitor when used with the HemoSphere oximetry cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows a logo for Edwards. The logo features a large, white letter "E" superimposed on a gray square grid. A faint circular design is behind the letter "E". The word "Edwards" is written in a serif font below the logo.
# HemoSphere Alta Advanced Monitor with HemoSphere Pressure Cable or HemoSphere Alta Monitor Pressure Cable
The HemoSphere Alta monitor when used with the HemoSphere Pressure Cable or HemoSphere Alta monitor Pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment.
The Edwards Lifesciences Acumen Hypotension Prediction Index software feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
When used in combination with the Swan-Ganz technology connected to a compatible Swan-Ganz catheter, the Edward Lifesciences Right Ventricular Pressure (RVP) algorithm provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVP algorithm is indicated for critically ill patients over 18 years of age receiving advanced hemodynamic monitoring in the operating room (OR) and intensive care unit (ICU). The RVP algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Right Ventricular Pressure (RVP) parameters.
When used in combination with the HemoSphere Pressure Cable connected to a compatible Swan-Ganz catheter, the Right Ventricular Cardiac Output (RVCO) feature provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVCO algorithm is intended for use in surgical or non-surgical patients over 18 years of age that require advanced hemodynamic monitoring. The Right Ventricular Cardiac output provides a continuous cardiac output and derived parameters.
The Cerebral Adaptive Index (CAI) Algorithm is an informational index to help assess the level of coherence or lack thereof between Mean Arterial Pressure (MAP) and the Absolute Levels of Blood Oxygenation Saturation (StO2) in patient's cerebral tissue. MAP is acquired by the HemoSphere
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the text "K242451". The text is in a bold, sans-serif font. The text is black and the background is white.
Image /page/11/Picture/1 description: The image shows a logo for Edwards. The logo features a large, white letter "E" superimposed on a gray square grid. Below the grid, the word "Edwards" is written in a simple, sans-serif font.
pressure cable or HemoSphere Alta Pressure Cable and StO2 is acquired by the ForeSight oximeter cable. CAI is intended for use in patients over 18 years of age receiving advanced hemodynamic monitoring. CAI is not indicated to be used for treatment of any disease or condition and no therapeutic decisions should be made based solely on the Cerebral Adaptive Index (CAI) Algorithm.
# HemoSphere Alta Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor
The Acumen assisted fluid management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions
## Acumen IQ Fluid Meter
The Acumen IQ fluid meter is a sterile single use device that is intended to be used with the HemoSphere Alta AFM cable and AFM software feature to inform the user of the rate of flow. The device is intended to be used by qualified personnel or clinicians in a clinical setting for up to 24 hours.
should be made based solely on the assisted fluid management suggestions.
# HemoSphere Alta Advanced Monitor with ForeSight Oximeter Cable
The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states. The ForeSight oximeter cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight oximeter cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere Alta monitor.
- When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents ≥40 kg.
- . When used with Medium Sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects ≥3 kg.
- . When used with Small Sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects <8 kg and noncerebral use on pediatric subjects <5kg.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows a logo for a company named Edwards. The logo features a large, white letter "E" superimposed on a gray square with a grid pattern. Behind the "E", there are faint white circles. Below the square, the word "Edwards" is written in a serif font.
The Edwards Algorithm for Measurement of Blood Hemoglobin is indicated for continuously monitoring changes to hemoglobin concentration in the circulating blood of adults and transitional adolescents > 40 kg receiving advanced hemodynamic monitoring using HemoSphere ForeSight Oximeter Cable and ForeSight IQ Sensors in cerebral locations.
## HemoSphere Alta Advanced Monitor with ClearSight Technology
The HemoSphere Alta Monitor when used with the HemoSphere ClearSight technology, pressure controller and a compatible Edwards finger cuff are indicated for Adult and Pediatric patients of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere Alta monitor and compatible Edwards' finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
#### Intended Use: The HemoSphere Alta Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting.
The HemoSphere Alta Advanced Monitoring Platform is intended for use with compatible Edwards Swan- Ganz/Swan-Ganz Jr/Swan-Ganz IQ catheters and oximetry catheters. FloTrac Jr sensors. Acumen IO sensors, TruWave DPTs, ForeSight /ForeSight Jr sensors/ForeSight IO sensor, Acumen IQ fluid meter and ClearSight/ClearSight Jr/Acumen IQ finger cuffs.
A comprehensive list of parameters available while monitoring with the HemoSphere Alta Advanced Monitor and a connected Swan-Ganz patient
{13}------------------------------------------------

Edwards
cable are listed below. Only iCO, iCI, iSVR, and iSVRI are available to the pediatric patient population.
| Parameter | Description | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------|------------------------|-----------------------------------------------------------|
| CO | continuous cardiac output | Adult only | Operating Room,<br>Intensive Care Unit,<br>Emergency Room |
| sCO | STAT cardiac output | | |
| CPO | Cardiac Power Output | | |
| CI | continuous cardiac index | | |
| sCI | STAT cardiac index | | |
| EDV | right ventricular end diastolic volume | | |
| sEDV | STAT right ventricular end diastolic<br>volume | | |
| EDVI | right ventricular end diastolic volume<br>index | | |
| sEDVI | STAT right ventricular end diastolic<br>volume index | | |
| HRavg | averaged heart rate | | |
| LVSWI | left ventricular stroke work index | | |
| PVR | pulmonary vascular resistance | | |
| PVRI | pulmonary vascular resistance index | | |
| RVEF | right ventricular ejection fraction | | |
| sRVEF | STAT right ventricular ejection fraction | | |
| RVSWI | right ventricular stroke work index | | |
| SV | stroke volume | | |
| SVI | stroke volume index | | |
| SVR | systemic vascular resistance | | |
| SVRI | systemic vascular resistance index | | |
| iCO | intermittent cardiac output | Adult and<br>Pediatric | |
| iCI | intermittent cardiac index | | |
| iSVR | intermittent systemic vascular resistance | | |
| iSVRI | intermittent systemic vascular resistance<br>index | | |
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere Alta Advanced Monitor and a connected HemoSphere oximetry cable are as listed below:
| Parameter | Description | Patient Population | Hospital Environment |
|-----------|----------------------------------|---------------------|-----------------------------------------|
| SvO2 | Mixed Venous Oxygen Saturation | Adult and Pediatric | Operating Room,<br>Intensive Care Unit, |
| ScvO2 | Central Venous Oxygen Saturation | | Emergency Room |
A comprehensive list of parameters available while monitoring with the HemoSphere Alta monitor and both a connected HemoSphere Alta Swan-Ganz patient cable and Oximetry cable are listed below:
| Parameter | Description | Patient<br>Population | Hospital<br>Environment |
|-----------|---------------------------------------------------------------|------------------------|-------------------------|
| DO2 | Oxygen Delivery | | |
| DO2I | Oxygen Delivery Indexed | Adult and<br>Pediatric | Operating Room, |
| VO2 | Oxygen Consumption | | Intensive Care Unit, |
| VO2e | Estimated Oxygen Consumption when<br>ScvO2 is being monitored | | Emergency Room |
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image shows the logo for Edwards. The logo is a square with a white letter E inside of it. There are white circles behind the letter E. The word "Edwards" is written in black below the square.
| VO2I | Oxygen Consumption Index | | |
|-------|---------------------------------------------------------------------|------------|--|
| VO2Ie | Estimated Oxygen Consumption Index<br>when ScvO2 is being monitored | | |
| GHI | global hypoperfusion index | Adult only | |
A comprehensive list of parameters available while monitoring with the HemoSphere Alta monitor and both a connected HemoSphere Alta Swan-Ganz patient cable and pressure cable are listed below:
| Parameter | Description | Patient<br>Population | Hospital<br>Environment |
|----------------------------------------------------------------------------------------------------------|-------------------------------------------|-----------------------|-----------------------------------------|
| CO20s | 20-second cardiac output | | |
| CI20s | 20-second cardiac index | | |
| SV20s | 20-second stroke volume | | |
| SVI20s | 20-second stroke volume index | | |
| PAOP | Pulmonary artery occlusion pressure2 | | |
| RVEDP | Right Ventricular End Diastolic Pressure | Adult only | Operating room,<br>intensive care unit, |
| RV dP/dt | Maximal Right Ventricular Systolic slope2 | | emergency room |
| SYSRVP | Systolic Right Ventricular Pressure | | |
| DIARVP | Right Ventricular Diastolic Pressure2 | | |
| MRVP | Mean Right Ventricular Pressure2 | | |
| CORV | Right ventricular cardiac output2 | | |
| CIRV | Right ventricular cardiac index2 | | |
| CPORV | Right ventricular cardiac power output2 | | |
| CPIRV | Right ventricular cardiac power index2 | | |
| SVRV | Right Ventricular Stroke Volume2 | | |
| SVIRV | Right Ventricular Stroke Volume Index2 | | |
| PRRVP | Right Ventricular Pulse Rate2 | | |
| CORV | Right Ventricular Cardiac Output2 | | |
| 1 20-second flow parameters are only available if the 20s flow parameter feature is enabled. | | | |
| Please Contact your local Edwards representative for more information on enabling this advanced feature. | | | |
| 2 RVP and RVCO parameters are available when using a Swan-Ganz IQ catheter.
A comprehensive list of parameters available while monitoring with the HemoSphere Alta monitor and a connected HemoSphere pressure cable or HemoSphere Alta pressure Cable are listed below:
| Parameter | Description | Patient<br>Population | Hospital<br>Environment |
|-----------|----------------------------------------------------------|---------------------------------------------|-----------------------------------------------------------|
| CO/<br>CI | Continuous Cardiac Output1/ Continuous<br>Cardiac Index1 | | |
| CVP | Central Venous Pressure | | |
| CPO | Cardiac Power output | | |
| CPI | Cardiac Power index | | |
| DIAART | Systemic arterial diastolic blood<br>pressure | Adult and<br>Pediatric ≥ 12<br>years of age | Operating Room,<br>Intensive Care Unit,<br>Emergency Room |
| DIAPAP | pulmonary artery diastolic blood<br>pressure | | |
| | Systolic slope2 | | |
| Eadyn | Dynamic Arterial Elastance2 | | |
| MAP | Mean Arterial Pressure | | |
| MPAP | Mean Pulmonary Arterial Pressure | | |
{15}------------------------------------------------
Image /page/15/Picture/1 description: The image shows a logo with the letter 'E' in white, superimposed on a gray square with a grid pattern. A partial circle design is behind the 'E'. Below the square, the word 'Edwards' is written in a simple, sans-serif font.
| PPV | pulse pressure variation1 | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|-------------------------------------|
| PR | Pulse rate | |
| SV/ | Stroke Volume1/ | |
| SVI | Stroke Volume Index1 | |
| SVR/ | Systemic Vascular Resistance1/ Systemic | |
| SVRI | Vascular Resistance1 Index | |
| SVV | Stroke Volume Variation1 | |
| SYSART | Systolic Blood Pressure | |
| SYSPAP | Systolic pulmonary artery blood pressure | |
| RVEDP | Right Ventricular End Diastolic Pressure | Adult Only |
| RV dP/dt | Maximal right ventricular systolic slope | Operating Room, Intensive Care Unit |
| SYSRVP | Systolic Right Ventricular Pressure | |
| DIARVP | Right Ventricular Diastolic Pressure | |
| MRVP | Mean Right Ventricular Pressure | |
| PRRVP | Right Ventricular Pulse Rate | |
| HPI | Acumen Hypotension Prediction Index | |
| 1FloTrac parameters are available when using a FloTrac/Acumen IQ sensor and if the FloTrac feature is enabled 2HPI parameters are available when using an Acumen IQ sensor and if the HPI feature is activated. | | |
A list of Acumen Assisted Fluid Management (AFM) outputs available for surgical patients ≥18 years of age while monitoring with the HemoSphere Alta monitor and a connected HemoSphere pressure cable are listed below:
| AFM output | Patient Population | Hospital<br>Environment |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-------------------------|
| Fluid Bolus Suggested | | |
| Test Bolus Suggested | | |
| Fluid Not Suggested | | |
| Suggestions Suspended | | |
| Bolus In Progress... | ≥18 years of age only | Operating room |
| Bolus Complete | | |
| Bolus Complete; Analyzing<br>Hemodynamic Response | | |
| Tracked Case Vol. | | |
| Flow Rate | | |
| Bolus Volume | | |
| AFM outputs are available when using an Acumen IQ sensor and if the AFM feature is activated.<br>Flow rate and Bolus Volume are visible when using automatic fluid tracking mode. | | |
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere Alta monitor and both a connected HemoSphere pressure cable and oximetry cable are listed below:
| Parameter | Description | Patient Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|--------------------|-----------------------------------------------------------|
| DO2 | Oxygen Delivery | Adult only | Operating Room,<br>Intensive Care Unit,<br>Emergency Room |
| DO2I | Oxygen Delivery Indexed | | |
| VO2 | Oxygen Consumption | | |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | |
| VO2I | Oxygen Consumption Index | | |
| VO2Ie | Estimated Oxygen Consumption<br>Index when ScvO2 is being<br>monitored | | |
{16}------------------------------------------------
Image /page/16/Picture/1 description: The image shows a logo with the letter 'E' in white, set against a gray square background divided into a grid pattern. Behind the 'E', there are faint circular lines, adding a subtle design element. Below the square, the word 'Edwards' is written in a simple, sans-serif font, completing the logo.
Tissue oxygen saturation, StO2, can be monitored with the HemoSphere Alta monitor and a connected ForeSight oximeter cable as listed below.
| Parameter | Description | Patient Population | Hospital<br>Environment |
|-------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|-----------------------------------------------|-----------------------------------------------------------|
| StO2 | Tissue oxygen saturation | | |
| ∆ctHb | relative change in total<br>hemoglobin | Adult and Pediatric | Operating Room,<br>Intensive Care Unit,<br>Emergency Room |
| tHb | Total Hemoglobin | Adults and transitional<br>adolescents ≥40 kg | |
| Total hemoglobin (tHb) is available when monitoring using a HemoSphere ForeSight Oximeter<br>Cable and two Foresight IQ Sensors in cerebral locations | | | |
comprehensive list of parameters available for adult populations while monitoring with the HemoSphere Alta monitor and both a connected HemoSphere pressure cable or HemoSphere Alta Pressure Cable and a connected ForeSight oximeter cable are listed below:
| Parameter | Description | Patient<br>Population | Hospital Environment |
|-----------|--------------------------|-----------------------|-----------------------------------------------------------|
| CAI | Cerebral Adaptive Index1 | Adult only | Operating Room, Intensive<br>Care Unit, Emergency<br>Room |
A comprehensive list of parameters available while monitoring with the HemoSphere Alta Advanced Monitor and a connected Pressure controller are listed below:
| Parameter | Description | Patient<br>Population | Hospital<br>Environment |
|-----------|-----------------------------------------|-----------------------|-----------------------------------|
| CO | Continuous Cardiac Output | | |
| CPO | Cardiac Power Output | | |
| CI | Continuous Cardiac Index | | |
| DIA | arterial diastolic bloodpressure | | |
| SYS | Systolic Blood Pressure | | |
| MAP | Mean Arterial Pressure | | |
| PPV | pulse pressure variation | | |
| PR | Pulse rate | | |
| SV | Stroke Volume | Adult and | |
| SVI | Stroke Volume Index | pediatric ≥ 12 | |
| SVR | Systemic Vascular Resistance | | Operating Room,<br>Intensive Care |
| SVRI | Systemic Vascular Resistance Index | | Unit, Emergency |
| SVV | Stroke Volume Variation | | Room |
| dP/dt | systolic slope | Adult only | |
| Eadyn | Dynamic Arterial Elastance1 | | |
| HPI | Acumen Hypotension Prediction<br>Index¹ | | |
I HPI parameters are available when using an Acumen IQ finger cuff and heart reference sensor (HRS)
NOTE: CO/CI and SV/SVI are measured using a reconstructed brachial arterial waveform. All other monitored parameters use a reconstructed radial arterial waveform. SVR/SVRI are derived from CO/CI and MAP along with an entered or monitored CVP value.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for Edwards. The logo is a gray square with a white grid pattern. Inside the square is a large, white letter "E". The word "Edwards" is written in black below the square.
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere advanced monitor and both a connected pressure controller and oximetry cable are listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|---------------------------------------------------------------------------|-----------------------|----------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | HemoSphere<br>ClearSight<br>Module and<br>HemoSphere<br>Oximetry<br>Cable | Adult only | Operating<br>Room,<br>Intensive<br>Care Unit |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption<br>Index when ScvO2 is being<br>monitored | | | |
#### VI. Comparison of Technological Characteristics with the Predicate Devices:
The subject and predicate devices are based on the following same technological elements:
- I Hardware: The subject device uses the same platform as the predicate (K232294, cleared October 31st, 2023). Additionally, the updated hardware configurations for the cables: HemoSphere Alta Pressure Cable and HemoSphere Alta AFM Cable remain identical in terms of the functionality, intended use and indication to their respective predicate devices.
- I Indication and Intended Use: The subject device has the same intended use and indication as the predicates.
- Technological characteristics: The new features/algorithm included in the subject device have the same technological characteristics as their respective predicate devices.
- I Accessories/Components: The subject and the predicate device both use the same accessories and components, Pressure controller, Heart Reference Sensor, Acumen fluid meter.
The following technological differences exist between the subject and predicate devices:
#### K242451
{18}------------------------------------------------
Image /page/18/Picture/1 description: The image shows the logo for Edwards. The logo features a large, white letter "E" superimposed on a gray square. Behind the "E", there are faint white circles and lines, creating a subtle geometric pattern. Below the square, the word "Edwards" is written in a simple, sans-serif font.
- Hardware: The subject device includes new hardware configuration of the configurations for the cables: HemoSphere Alta Pressure Cable and HemoSphere Alta AFM Cable. The updated configuration includes the modified connectors and updated design for user convenience.
- I Software: The subject device includes an updated software to support the new integrated algorithms and features.
- Graphical User Interface (GUI): The subject device includes an updated graphical user interface to accommodate new integrated algorithms and features.
- I New Algorithm: The subject HemoSphere Alta Advanced Monitoring Platform includes a new Right Ventricular Cardiac Output (RVCO) algorithm, which uses the existing technologies, RVP and iCO algorithms of the HemoSphere Alta Advanced monitoring platform device (K232294, cleared October 31, 2023).
A brief description of the key highlights of the subject HemoSphere Alta™ Advanced Monitoring Platform and a comparative overview with the predicate HemoSphere Advanced Monitoring Platform is provided below:
- I New features and/or algorithms for the HemoSphere Alta Advanced Monitoring Platform (original K232294, cleared October 31st, 2023):
- I Cerebral Adaptive Index (CAI):
The existing Cerebral Adaptive Index (CAI) Algorithm (K223651, cleared May 26th, 2023) has been integrated into the HemoSphere Alta Advanced Monitoring.
- Edwards Algorithm for Measurement of Blood Hemoglobin (tHb):
The existing Edwards Algorithm for Measurement of Blood Hemoglobin (K230612, cleared Nov 17th, 2023) has been integrated into the HemoSphere Alta Advanced Monitoring.
- Smart Wedge Algorithm:
{19}------------------------------------------------
Image /page/19/Picture/1 description: The image shows a logo with the letter 'E' in white, set against a gray square background. The letter 'E' is stylized and has a geometric design with circles and lines. Below the square is the word 'Edwards' in a serif font.
The Smart Wedge Algorithm (K230579, cleared August 18th, 2023) has been integrated into the HemoSphere Alta Advanced Monitoring.
- I Right Ventricular Cardiac Output (RVCO) Algorithm:
The RVCO algorithm is a new algorithm that uses the right ventricular pressure waveform to calculate continuous cardiac output. The right ventricular cardiac output (CORV) parameter provides an additional option for the measurement of cardiac output and trends it continuously for use in the surgical and nonsurgical setting. The algorithm is based on the physiological relationship between flow and pressure in the pulmonary cardiovascular system, and it was trained with large-scale data to capture this relationship between pulmonary flow and right ventricular pressure.
- HemoSphere Alta ™ monitor Pressure cable ■ (HEMAPSC200):
A New configuration of the existing HemoSphere Pressure Cable (HEMPSC100) is being introduced to provide improved user interface.
- Modification to the existing elements of the HemoSphere Alta Advanced Monitoring Platform:
- Acumen Assisted Fluid Management (AFM) – Automatic bolus tracking with Acumen AFM cable:
The Acumen Assisted Fluid Management (AFM) feature has been updated to allow for an automated fluid tracking mode cleared in K223865 on June 9th, 2023, for HemoSphere Advanced Monitoring Platform. Similar to the predicate device, this automated fluid tracking mode for the AFM software feature is achieved via a modified Acumen AFM cable and the existing AFM fluid meter.
Additionally, for user convenience, the automatic bolus tracking feature can now also be used independently in "Fluid Meter Only" mode to track fluid administration. In this mode, users can track fluid delivery via an existing Acumen fluid meter instead of manually tracking and logging. During this mode, the AFM algorithm will be disabled, and the user will not get any bolus suggestions.
{20}------------------------------------------------
Image /page/20/Picture/1 description: The image shows a logo for Edwards. The logo features a gray square with a white letter "E" inside. The letter "E" is stylized with circles around it. Below the square, the word "Edwards" is written in a serif font.
- Modification to the existing Acumen Hypotension Prediction Index (HPI) feature:
The existing HPI feature for HemoSphere Alta Advanced Monitoring Platform has been modifi…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.