The Falcon HST is an EEG and respiratory signal recorder. The device is intended for use by adult patients in the home or clinical environment, under the direction of a qualified healthcare practitioner, to aid in the diagnosis of sleep disorders.
Device Story
Falcon HST is an EEG and respiratory signal recorder worn on the patient's chest. It collects physiological data including EEG, EOG, ECG, and respiratory signals via inductive plethysmography bands, electrodes, pressure transducers for nasal airflow, and solid-state accelerometers for body position. Pulse oximetry is integrated via external module. The device features a touch-screen LCD and internal memory for data storage. Data is transferred to a PC via USB-C for review and analysis using Profusion Sleep software (v5.1+). The system is operated by patients in home or clinical environments under physician direction. Physicians use the software output to aid in the diagnosis of respiratory and cardiac-related sleep disorders. The device benefits patients by enabling ambulatory sleep monitoring, reducing the need for in-lab polysomnography.
Clinical Evidence
Bench testing compared Falcon HST against Zmachine Synergy for EEG, respiratory effort, nasal airflow, body position, and pulse oximetry, showing high agreement. Safety/performance testing conducted per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-26, and IEC 60601-2-47. Clinical performance testing validated Profusion PSG software 5.1 against gold-standard polysomnography data for calculating Apnea Hypopnea Index (AHI) and sleep staging (N1, N2, N3, REM, Wake).
Indicated for adult patients in home or clinical settings to aid in the diagnosis of sleep disorders, under the direction of a qualified healthcare practitioner.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
February 20, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Compumedics Limited % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Boulevard Warren, New Jersey 07059
Re: K242447
Trade/Device Name: Falcon HST Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLV, OMC, MNR Dated: December 20, 2024 Received: December 20, 2024
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Jay R. Gupta -S
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
Device Name
Falcon HST
| Indications for Use (Describe) |
|--------------------------------|
|--------------------------------|
The Falcon HST is an EEG and respiratory signal recorder. The device is intended for use by adult patients in the home or clinical environment, under the direction of a qualified healthcare practitioner, to aid in the diagnosis of sleep disorders.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Compumedics. The logo features a blue circle with a stylized "m" inside of it. To the right of the circle is the word "COMPUMEDICS" in blue, and below that is the tagline "Defining Life's Signals" also in blue.
# AJ231 Falcon HST 510(k) Summary
#### Device Name: Falcon HST
The following 510(k) summary is being submitted in accordance with 21 CFR 807.92.
## Submission Details
#### Applicant Information:
| Applicant Information: | |
|------------------------|--------------------------------------------------------------|
| Submitter: | Compumedics Limited |
| Address: | 30-40 Flockhart Street, Abbotsford 3067, Victoria, Australia |
| Phone number: | +61 (0) 3 8420 7300 |
| Fax number: | +61 (0) 3 8420 7399 |
| Contact person: | John Joseph |
| Address: | 30-40 Flockhart Street, Abbotsford 3067, Victoria, Australia |
| Phone number: | +61 (0) 3 8420 7362 |
| Fax number: | + 61 (0) 3 8420 7399 |
| Email: | jjoseph@compumedics.com.au |
Submission Prepared: November 25, 2024
## Subject Device Information:
| Trade name: | Falcon HST |
|-----------------------|-----------------------|
| Common name: | Falcon HST |
| Primary Product code: | OLV, OMC, MNR |
| Classification Names: | Electroencephalograph |
| Panel: | Neurological Devices |
| Device class: | II |
| Regulation numbers: | 21 CFR 882.1400 |
## Predicate Device Information
### Primary Predicate Device:
| Trade Name: | Zmachine Synergy |
|-----------------------|-----------------------------------------|
| Manufacturer: | Consolidated Research of Richmond, Inc. |
| 510(k) number: | K172986 |
| 510(k) Decision Date: | 19th Dec 2017 |
| Classification: | Electroencephalograph |
| Product code: | OLV, OMC, MNR |
| Device class: | II |
| Regulation numbers: | 21 CFR 882.1400 |
#### Secondary Predicate Device:
| Trade Name: | Zmachine DT-100 |
|-----------------------|----------------------------------------|
| Manufacturer: | Consolidated Research of Richmond, Inc |
| 510(k) number: | K101830 |
| 510(k) Decision Date: | 31st Mar 2011 |
| Classification: | Electroencephalograph |
| Product code: | OLV, OMC |
| Device class: | II |
| Regulation numbers: | 21 CFR 882.1400 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Compumedics. The logo consists of a blue circle with a stylized "m" inside, followed by the word "COMPUMEDICS" in blue, all caps. Below the company name is the tagline "Defining Life's Signals" also in blue.
## Intended Use / Indications For Use
The Falcon HST is an EEG and respiratory signal recorder. The device is intended for use by adult patients in the home or clinical environment, under the direction of a qualified healthcare practitioner, to aid in the diagnosis of sleep disorders.
## Device Description
The Falcon HST comprises hardware and software which provide separate parameters for recording, review, and analysis of collected and stored physiological parameters, including EEG, EOG, ECG and respiratory signals, which are then used as an aid in the diagnosis of respiratory and/or cardiac related sleep disorders by qualified physicians.
The Falcon system consists of the main unit and the charging cradle. The main unit is a small device that is worn on the patient's chest over clothing. It is equipped with a touch-screen LCD and contains various channel inputs such as for the inductive plethysmography bands, and electrodes. The Falcon charging cradle is used to charge the device, as well as provide a USB interface for transferring study data to the PC.
## Technology overview
#### Electroencephalograph (EEG)
Both Falcon HST and Zmachine Synergy use the same EEG hardware to acquire an EEG channel characteristic.
#### Respiratory effort
Respiratory Effort is sensed by the Falcon HST using a thoracic and abdominal effort belt, however Zmachine Synergy uses only a thoracic effort belt to record chest expansion and contraction during inhalation and exhalation (i.e. respiratory effort).
#### Respiratory nasal airflow
Both Falcon HST and Zmachine Synergy use the similar pressure transducer to acquire the respiratory nasal airflow signals.
#### Body position
Both the Falcon HST and Zmachine Synergy use solid-state accelerometer to acquire the body position signal.
#### Blood oxygen saturation and pulse
Both the Falcon HST and Zmachine Synergy use the similar pulse oximeter hardware and data output modes to acquire the blood oxygen saturation and pulse signals.
#### PC Software
The Falcon HST device is compatible with Profusion Sleep software v5.1 or greater, while the predicate device Zmachine DT-100 (K101830) is compatible with their analysis software. The main differences involve the addition of Falcon HST file format support and data retrieval.
Profusion PSG software 5.1 is identical to the Profusion PSG software which was separately cleared as a part of K072201 and K093223, and as a result, required software documentation is provided regarding this software.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Compumedics. The logo consists of a blue circle with a cursive "m" inside of it, followed by the word "COMPUMEDICS" in blue. Below the word "COMPUMEDICS" is the phrase "Defining Life's Signals" in a smaller font.
The differences in software versions do not constitute a substantial change as they are not changes to underlying software specifications and are instead driver-level changes necessary for data conversion from the Falcon HST device. The system has been subject to functional testing with full coverage of data flow paths including USB communication which represents the predominant change between versions. Additionally, the Profusion Sleep interface has been used as part of the system setup for IEC 60601-1-2 and IEC 80601-2-26 tests as these include tests relating to signal display and accuracy of signal reproduction, especially in a noisy environment.
## Substantial Equivalence
This document represents the discussion of the substantial equivalence of the Falcon HST with respect to its primary predicate device the Zmachine Synergy (K172986). To discuss the substantial equivalence of Profusion Sleep 5.1 software feature support to Falcon HST, another predicate device Zmachine DT-100 (K101830) is included as Zmachine Synergy (K172986) only focuses on hardware device application. Additionally, this document provides a comparison of intended use, as well as a comprehensive evaluation of technological characteristics. Available performance data is summarised to further support safety and effectiveness and substantial equivalence with respect to its predicate devices.
The Zmachine Synergy device was launched in 2017 by Consolidated Research of Richmond, Inc. It has been sold in respective markets since then. The intended use of the Falcon HST is the same as the Zmachine Synergy, as is the channels recorded and the basic use and technology.
The main differences are that the Falcon HST is a single unit with colour touch screen rather than a single button configuration in the Zmachine Synergy. The Falcon HST has a charging cradle unlike Zmachine Synergy which has a USB charging cable.
Based on the Indications for Use, results of technological, performance & safety testing and a comparison to the predicate devices, it is concluded that the Falcon HST system with Profusion Sleep 5.1 feature support is substantially equivalent to the Zmachine Synergy and Zmachine DT-100, and there are no new issues identified in terms of safety or effectiveness with respect to its intended use.
The below table contains a multi-parameter comparison between Falcon HST system with Profusion Sleep 5.1 support feature, the primary predicate device the Zmachine Synergy and the predicate device Zmachine DT-100. This table is included in AJ056 Falcon Substantial Equivalence Discussion with further analysis of differences where applicable.
| Characteristic | Falcon HST | Zmachine Synergy (Primary<br>predicate device) | Zmachine DT-100<br>(Predicate device) |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Usage Information | | | |
| Manufacturer | Compumedics Ltd. | Consolidated Research of<br>Richmond, Inc. | Consolidated Research of<br>Richmond, Inc. |
| 510(k) Number | | K172986 | K101830 |
| Classification<br>regulation | 21 CFR 882.1400 | 21 CFR 882.1400 | 21 CFR 882.1400 |
| Product Code | OLV, OMC, MNR | OLV, OMC, MNR | OLV and OMC |
| Indications for<br>use | The Falcon HST is an EEG and<br>respiratory signal recorder.<br>The device is intended for use<br>by adult patients in the home<br>or clinical environment,<br>under the direction of a<br>qualified healthcare<br>practitioner, to aid in the<br>diagnosis of sleep disorders. | The Zmachine Synergy is an EEG<br>and respiratory signal recorder.<br>The device is intended for use by<br>adult patients in the home or<br>clinical environment, under the<br>direction of a qualified healthcare<br>practitioner, to aid in the<br>diagnosis of sleep disorders. | The CRI Zmachine is a single-<br>channel, EEG acquisition and<br>analysis system, designed for<br>use in the home or clinical<br>environments. This device is<br>intended to be used by qualified<br>healthcare practitioners to<br>monitor the wake and sleep<br>states of adult patients and as<br>an adjunct to their diagnosis of<br>sleep disorders. |
| Type of Use | Prescription-Use Only | Prescription-Use Only | Prescription-Use Only |
| Intended<br>Environment | Home or clinical environment | Home or clinical environment | Home or clinical environment |
| Patient<br>Population | Adults | Adults | Adults |
| Acquired<br>Channels | Airflow<br>Pulse Oximeter<br>Respiratory Effort<br>Body Position<br>EEG | Airflow<br>Pulse Oximeter<br>Respiratory Effort<br>Body Position<br>EEG | EEG |
| Sensor<br>technology | Airflow: Solid state pressure<br>transducer (±6 mbar range)<br>Pulse Oximetry: Nonin 3150<br>Wrist-Ox2, BLE module.<br>Respiratory Effort: Thoracic<br>effort belt based on<br>respiratory inductance<br>plethysmography (RIP).<br>Body position: Solid state<br>accelerometer, multiple<br>differential EXG channels. | Airflow: Solid state pressure<br>transducer (±6 mbar range)<br>Pulse Oximetry: Nonin XPOD LP<br>module.<br>Respiratory Effort: Thoracic effort<br>belt based on respiratory<br>inductance plethysmography<br>(RIP).<br>Body position: Solid state<br>accelerometer, multiple EEG<br>channels. | Zmachine DT-100 EEG<br>technology |
| Sensor<br>technology | LED indicators. A single<br>colour LED is provided to<br>indicate unit status. This LED<br>is turned on or off (either<br>continuously or flashing)<br>under control of the<br>firmware. The displayable<br>colour is blue. | LED indicators. A full-color LED<br>indicator is located beside each<br>connector (airflow, effort,<br>oximeter,<br>and EEG) to indicate correct or<br>incorrect hookup/operation.<br>Another<br>full-color LED is located under the<br>main power button to indicate<br>system status. | LCD display. A full color,<br>320×240 pixel, LCD display<br>with an LED backlight is used to<br>present all system information. |
| Power Source | Internally powered using li-<br>ion rechargeable battery | Internally powered using li-ion<br>rechargeable battery | Internally powered using li-ion<br>rechargeable battery |
| Internal<br>memory/ data<br>Storage | Storage on Falcon HST<br>onboard memory | Fixed microSD card | Removeable microSD card |
| Communication<br>Interface | USB-C | USB | USB card reader |
| Access to<br>recorded data | Recorded data is stored in the<br>device. When the device is<br>connected to PC via USB cable<br>the device provides access to<br>its internal memory. | Recorded data is stored in the<br>device. When the device is<br>connected to PC via USB cable the<br>device provides access to its<br>internal memory. | Recorded data is stored in the<br>device. The microSD card is<br>removed and connected to PC<br>via USB card reader to access<br>internal memory. |
| Recorded data<br>format | Each channel of recorded data<br>is stored in an individual file<br>of the proprietary data<br>format. | Each channel of recorded data is<br>stored in an individual file of the<br>GSC2 data format. | Each channel of recorded data<br>is stored in an individual file of<br>the binary data format. |
| Device<br>dimensions | 92 x 78 x 11 mm | 120 x 61 x 24 mm | 116 x 68 x 21 mm |
| Type of Device | EEG-based sleep monitor | EEG-based sleep monitor | EEG-based sleep monitor |
| No. of EEG<br>Channels | 4 | 1 | 1 |
| Electrode<br>Placement | Mastoid | Mastoid | Mastoid |
| Analyze EEG<br>Data in Real<br>time | Yes | n/a | Yes |
| Sleep<br>Classification | Wake and Sleep | Wake and Sleep | Wake and Sleep |
| EEG Analysis<br>Methodology | | | |
| Proprietary adaptive<br>algorithm using time and<br>frequency domain features | n/a | Proprietary adaptive algorithm<br>using time and frequency<br>domain features | |
| Can Provide<br>Information to<br>Assist the<br>Physician in<br>Diagnosis of<br>Sleep<br>Disorders? | n/a | Yes | |
| Can Provide<br>Information to<br>Assist the<br>Physician in the<br>Application of<br>Treatment | n/a | Yes | |
| Provide<br>Information to<br>Assist Physician<br>in the<br>Evaluation of<br>Treatment<br>Efficacy? | n/a | Yes | |
| Calculates<br>Summary Sleep<br>Statistics? | n/a | Yes | |
| Battery<br>Powered? | Yes | Yes | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Compumedics. The logo features a blue circle with a stylized "m" inside, followed by the word "COMPUMEDICS" in blue, sans-serif font. Below the company name is the tagline "Defining Life's Signals" also in blue.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Compumedics. The logo consists of a blue circle with a stylized "m" inside, followed by the word "COMPUMEDICS" in blue, and the tagline "Defining Life's Signals" below. The logo is simple and professional, and the blue color gives it a sense of trust and reliability.
The table above shows that there are no significant differences between Zmachine Synergy and the predicate devices that adversely affect product safety and effectiveness.
## Testing Summary
Design and verification activities have been performed on the Falcon HST as a result of the risk analysis and product requirements. External tests have been completed for electrical safety (IEC 60601-1:2005 (Third Edition) + COR1:2006 + COR2:2007 + A1:2012 + A2:2020), EMC (IEC 60601-1-2:2014 + A1:2020, EN 60601-1-2:2015+A1:2021), and mechanical and environmental requirements (IEC 60601-1-11:2015+A1:2020). In addition, side-by-side bench comparison testing summarized in the table below, was performed in which a Falcon HST was compared against a Zmachine Synergy (K172986).
| Test | Test Method Summary | Results |
|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EEG Input<br>Circuit | EEG is sensed by the Falcon HST and<br>Zmachine Synergy using the identical EEG<br>amplifier and analog-to-digital conversion<br>hardware. As such, the acquired signals<br>from both systems are expected to be<br>substantially equivalent. | The EEG characteristics were found to be<br>in high agreement with the design limits<br>for all points of comparison. As such, the<br>EEG recording capabilities were found to<br>be substantially equivalent. |
| | To test, an arbitrary waveform generator<br>was connected to the EEG input of a Falcon<br>HST and Zmachine Synergy device. | |
| | Signals were generated by connecting this<br>arbitrary waveform generator to the<br>systems electrodes. Tests were not | |
| | performed simultaneously, as waveforms | |
| | were able to be replicated for each test using the same waveform generator. | |
| Respiratory<br>Effort | Respiratory Effort is sensed by the Falcon HST using a thoracic and abdominal effort belt, however Zmachine Synergy uses only a thoracic effort belt to record chest expansion and contraction during inhalation and exhalation (i.e. respiratory effort).<br><br>To test, the Falcon HST and Zmachine Synergy were subject to simulated inhalation and exhalation. As such, the acquired signals from both systems are expected to be substantially equivalent. | The acquired data from each system was analyzed in order to compare the Respiratory Effort readings.<br><br>Both units produced similar readings. As such, the Respiratory Effort characteristics were found to be substantially equivalent. |
| Respiratory<br>Airflow | Respiratory Airflow is sensed by the Falcon HST and Zmachine Synergy using the similar pressure transducer hardware. As such, the acquired signals from both systems are expected to be substantially equivalent.<br><br>Testing required use of nasal cannula in the same setting with the same breathing rate for both products.<br><br>The acquired signals from both systems are expected to be substantially equivalent. | The acquired data from each system was analyzed in order to compare the Respiratory Airflow readings.<br><br>Both units produced similar readings. As such, the Respiratory Airflow characteristics were found to be substantially equivalent. |
| Body<br>Position | Body Position is sensed by the Falcon HST and Zmachine Synergy using an accelerometer module.<br><br>As both systems are measuring the angle of the system with respect to gravity, and both systems are worn by the patient in the same configuration, the acquired signal from both systems are expected to be substantially equivalent if it can be demonstrated that the Falcon HST reports angle with regard to gravity appropriately against an angular reference. To test, the Falcon HST and Zmachine Synergy devices are rotated through 360 degrees against an angular reference. | The acquired data from the Falcon HST and Zmachine Synergy was analyzed in order to compare the appropriate body position readings.<br><br>As such, the Body Position recording capabilities were found to be substantially equivalent. |
| Pulse<br>Oximetry | Pulse Oximetry is sensed by the Falcon HST and Zmachine Synergy using identical pulse oximeter hardware and similar data output modes. The device used was a NOnin Medical 3150 WristOx, being FDA approved as per FDA 510K Nonin Medical 3150 WristOx – K102350 As such, the acquired signals from both are expected to be substantially equivalent. Testing was performed by attaching the pulse oximeter | The acquired data from each system was analysed in order to compare the Pulse Oximeter readings.<br><br>The heart rate and oxygen saturation readings were found to be in high agreement when comparing the two systems together. As such, the Pulse Oximeter recording capabilities were found to be substantially equivalent. |
| to the index finger and obtaining a reading<br>from the device one at a time, under the<br>same conditions. | Both units produced similar readings. As<br>such, the Pulse Oximeter recording<br>capabilities were found to be<br>substantially equivalent. | |
| As such, the acquired signals from both<br>systems are expected to be substantially<br>equivalent. | | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Compumedics. The logo consists of a blue circle with a stylized "m" inside, followed by the word "COMPUMEDICS" in blue. Below the word "COMPUMEDICS" is the tagline "Defining Life's Signals" also in blue.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Compumedics. The logo consists of a blue circle with a stylized "m" inside, followed by the word "COMPUMEDICS" in blue, and the tagline "Defining Life's Signals" below. The logo is simple and professional, and the blue color gives it a sense of trust and reliability.
Additionally, following performance test was performed:
- Electroencephalograph safety and performance testing as per IEC 80601-2-26:2019, Testing against the particular standard for electroencephalograph includes validation against multiple essential performance requirements such as accuracy of amplitude and rate of variation signal reproduction, input dynamic range and differential offset voltage, input noise, frequency response, and common mode rejection ratio.
- . Ambulatory Electrocardiography systems safety and essential performance testing as per IEC 60601-2-47:2012.
- . Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems IEC 62133-2:2017/AMD1:2021
- . Bench testing against Falcon functional requirements to ensure that performance meets hardware and software design specifications including functionality substantially equivalent to the Zmachine Synergy predicate device.
All tests passed with results equivalent to the Zmachine Synergy and Zmachine DT-100 and did not raise additional concerns of safety and effectiveness. Clinical performance testing was provided to validate that the performance of the Profusion PSG software 5.1 produces substantially equivalent results for calculating the apnea hypopnea index (AHI) and sleep staging (N1, N2, N3, REM and Wake) when compared to expert review of gold standard polysomnography data.
## Conclusion
Compumedics Ltd. considers the Falcon HST to be substantially equivalent to its primary predicate device the Zmachine Synergy and predicate device the Zmachine DT-100. The indications for use, technological characteristics, and underlying principles of operation are the same. There are no questions of safety and effectiveness, and substantial equivalence is supported by extensive performance testing data.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.