The PPTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation. When the patient is in the seated position using the chair, the System is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the head, neck and thorax.
Device Story
PPTS is a proton therapy system using a synchrotron-based accelerator to produce a proton beam (70-250 MeV) for radiation treatment. The system utilizes pencil beam scanning to deliver prescribed doses to localized tumors. The P-ARTIS patient positioning system, featuring a 6-degree-of-freedom robotic arm, supports patients in either seated (chair) or supine (couch) positions. The system integrates CT (3D) and X-ray (2D) imaging for patient registration and position verification at the isocenter. A control subsystem (TSM) synchronizes beam delivery with hospital oncology information systems and PACS, while a safety subsystem manages interlocks and monitors beam parameters. Used in clinical radiation oncology settings, the device is operated by trained medical professionals. The output—a precisely steered proton beam—allows clinicians to target tumors while sparing surrounding healthy tissue, potentially improving treatment outcomes for patients with localized malignancies.
Clinical Evidence
Bench testing only. Performance verified through mechanical testing of the positioning system, beam performance evaluation (dose shape, rate, spot positioning), safety interface/interlock testing, and integration validation with oncology information systems and treatment planning software. Compliance with consensus standards (IEC 60601 series, AAPM TG-224) confirmed.
Technological Characteristics
Synchrotron-based proton accelerator; pencil beam scanning nozzle; 70-250 MeV energy range; 6-degree-of-freedom robotic patient positioning (P-ARTIS); integrated CT and X-ray imaging; steel and concrete shielding. Connectivity via TSM to hospital oncology information systems and PACS. Software-based control and safety interlocks. Compliant with IEC 60601-1, 60601-1-2, 60601-2-44, 60601-1-3, 60601-1-8, 60601-2-54, 60601-2-68, and AAPM TG-224.
Indications for Use
Indicated for patients with localized tumors and other conditions susceptible to radiation therapy. When using the chair, indicated for tumors in the head, neck, and thorax.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
P-Cure Proton Beam Therapy System (PPBTS) (K221996)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 12, 2025
P-Cure, Ltd.
% Lina Kontos
Regulatory Counsel
Hogan Lovells US LLP
555 Thirteenth Street NW
Washington, District of Columbia 20004
Re: K242418
Trade/Device Name: P-Cure Proton Therapy System (PPTS)
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: Class II
Product Code: LHN
Dated: September 19, 2024
Received: September 19, 2024
Dear Lina Kontos:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242419 - Lina Kontos
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K242419 - Lina Kontos
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K242418
Device Name
P-Cure Proton Therapy System (PPTS)
Indications for Use (Describe)
The PPTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation.
When the patient is in the seated position using the chair, the System is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the head, neck and thorax.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K242418
Page 1 of 5
510(k) SUMMARY
P-Cure, Ltd.'s P-Cure Proton Therapy System (PTTS)
Submitter:
P-Cure, Ltd.
14 Carmel St.
Shilat Industrial Zone, 7318800, Israel
Phone: +972 54 488 1399
Contact Person: Ori Lubin
Date Prepared: May 9, 2025
Name of Device: P-Cure Proton Therapy System (PTTS)
Common or Usual Name: Proton beam therapy systems
Classification Name: Medical Charged-Particle Radiation Therapy System, 21 CFR 892.5050
Regulatory Class: Class II
Product Code: LHN
Predicate Device:
P-Cure Proton Beam Therapy System (PPBTS) (K221996)
Reference Device:
ProTom Radiance 330 (K191521)
Intended Use / Indications for Use:
The PPTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation.
When the patient is in the seated position using the chair, the System is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the head, neck and thorax.
Technological Characteristics:
The P-CURE Proton Therapy System (PPTS) is comprised of four main subsystems that function in tandem to generate the desired dose level and distribution at the target site:
- Beam production system (Synchrotron based accelerator)
- Injector – produces and delivers protons to the synchrotron
{5}
K242418
Page 2 of 5
- Synchrotron ring – accelerates the proton beam in circular orbit (within the ring) to the desired energy level
- Extraction system – extracts the beam from the ring to the beam delivery subsystem
- Beam delivery system for a single fixed beam treatment room. Steers and monitors the extracted proton pencil beam from the synchrotron to the desired treatment location (Nozzle).
- Patient Positioning System (P-ARTIS). Mechanically orients the patient (seated or on supine); provides independent means of patient registration using CT (3D) and X-ray (2D)
- CT system (P-ARTIS CT)
- Robotic arm and chair/couch (6 Degree of freedom Couch) (P-ARTIS PPS)
- X-ray system (P-ARTIS XR)
- Positioning Software (P-ART)
- Control and Safety Systems
- Control Subsystem (TSM). Synchronizes the various subsystem actions and connects with hospital oncology information systems and PACS.
- Safety Subsystem. Includes hardware and software means to ensure safe system operation for patient and personnel. It includes subsystem interlocks, treatment beam parameters monitoring, and others.
# Summary of Technological Characteristics and Comparison to Predicate Device:
The PPTS, the predicate device, and the reference device all share the same fundamental principles of operation, namely that they accelerate a proton beam to the same energy range by using a synchrotron and identically deliver the prescribed treatment dose to the patient by a pencil beam scanning technique. The PPTS uses the same image guided and robotic patient positioning system (P-ARTIS) to move the patient ensuring delivery of the beam to the target area, as prescribed.
While the previously cleared PPBTS positioned patients in a seated position alone, the PPTS can additionally position the patient in a supine position for treatment using an off-the-shelf couch, similar to the identified reference device, ProTom Radiance 330. The PPTS, PPBTS, and ProTom Radiance 330 all use the same robotic arm for mounting the patient support.
In all three systems, the simulation CT dataset are acquired in the treatment patient orientation. For a "seated" planning, the CT dataset is acquired by using in room vertical CT (performed identically for planning for PPBTS and PPTS). For a "supine" planning, the CT simulation data is generated at standard simulation settings of radiation oncology departments (performed identically for planning for ProTom Radiance 330 and PPTS). In all three systems, orthogonal x-ray imaging is used for image guided position verification at isocenter.
A table comparing the key features of the subject, predicate, and reference devices is provided below.
{6}
K242418
Page 3 of 5
| Device Characteristic | P-Cure Proton Therapy System (PPTS) | P-Cure Proton Beam Therapy System (PPBTS)
(Predicate Device) | Radiance 330 Proton Beam Therapy System
(Reference Device) |
| --- | --- | --- | --- |
| 510(k) Number | K242418 | K221996 | K191521 |
| Intended Use/Indications for Use | The PPTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation.
When using the chair, the System is indicated to treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the head, neck and thorax. | The P-CURE system is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation in the head, neck and thorax. | The ProTom Radiance 330 is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation |
| Accelerator | Synchrotron | Synchrotron | Synchrotron |
| Particle | Protons | Protons | Protons |
| Variable | Yes; 70-250 MeV | Yes; 70-250 MeV | Yes; 70-250 MeV |
| Cycle Time | Variable | Variable | Variable |
| Beam Transport | Bending and focusing magnets | Bending and focusing magnets | Bending and focusing magnets |
| Nozzles | Pencil Beam Scanning | Pencil Beam Scanning | Pencil Beam Scanning |
| Beam Range in Patient (Tissue Depth) | 3 cm to 38 cm clinical use | 3 cm to 38 cm clinical use | 3 cm to 38 cm clinical use |
| Range Verifier | Optional | Optional | Optional |
| Control and Safety System | Software controls accelerator, beam transport and delivery, sets operational parameters, monitors systems and provides alerts regarding error conditions. | Software controls accelerator, beam transport and delivery, sets operational parameters, monitors systems and provides alerts regarding error conditions. | Software controls accelerator, beam transport and delivery, sets operational parameters, monitors systems and provides alerts regarding error conditions. |
| Mechanical Beam Stops | Yes | Yes | Yes |
| Beam Intensities | Continuously variable over range | Continuously variable over range | Continuously variable over range |
{7}
K242418
Page 4 of 5
| Shielding | Steel and Concrete | Steel and Concrete | Steel and Concrete |
| --- | --- | --- | --- |
| Gantry Rotating Angle | N/A. Fixed beam | N/A. Fixed beam | Rotating gantry and Fixed beam |
| Patient Positioner | Yes – couch or chair mounted on a robotic patient positioning device | Yes – chair mounted on a robotic patient positioning device | Yes – couch mounted on a robotic patient positioning device |
| Patient Positioner | 6 degree of freedom robotic positioning device. | 6 degree of freedom robotic positioning device. | 6 degree of freedom robotic positioning device. |
| Maximum Load | Chair: 180 kg
Couch: 235 kg | Chair: 180 kg | Unknown |
| Imaging | Yes – Diagnostic quality CT (seated configuration) and orthogonal x-ray system | Yes – Diagnostic quality CT and orthogonal x-ray system | Yes – couch mounted CBCT and x-ray system |
| Treatment Room | Horizontal fixed gantry with a robotic chair or a couch | Horizontal fixed gantry with a robotic chair | Rotating gantry with a robotic couch
Horizontal fixed gantry with a robotic couch |
## Performance Data:
The Company performed testing as follows:
- Mechanical testing to verify performance of the positioning system
- Beam performance testing to evaluate beam dose shape, beam dose, dose rate, dose monitoring, and spot positioning
- Safety interface testing to verify collision sensors, mechanical interlocks
- Simulation and validation testing to verify integration with oncology information systems
- Validation testing to verify integration with positioning system and treatment planning system
- Testing to support repeatability and reproducibility of patient positioning and immobilization
Product safety and essential performance electrical testing was conducted based upon the following consensus standards: IEC 60601-1, IEC 60601-1-2, IEC TR 60601-4-2, EN 606601-2-44, IEC 60601-1-3, IEC 60601-1-8, IEC 60601-2-54, IEC 60601-1-64, IEC 60601-2-68, IEC 62667, and AAPM TG-224.
Software was documented and validated per FDA's Guidance Document “Content of Premarket Submissions for Device Software Functions,” and per IEC
In all instances, the PTTS functioned as intended and met its specifications. Testing demonstrated substantial equivalence in terms of performance and safety to the predicate.
{8}
K242418
Page 5 of 5
# Substantial Equivalence Conclusion:
The PPTS has the same intended use and substantially similar indications for use and technological characteristics as the predicate, PPBTS (K221996). The minor differences in indications for use statements and technological characteristics between the PPTS and the predicate device do not raise new or different questions of safety or effectiveness. The completed verification and validation data demonstrate that the device meets its specifications with the commercially available patient positioning configuration, which are comparable to those of the reference device. Thus, the PPTS system can be found substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.