OverCEM SA Universal; OverCEM SA Translucent; OverCEM SA Opaque; OverCEM Ti-Abutment; OverCEM Cer-Abutment
K242366 · Overfibers S.R.L. · EMA · Jan 10, 2025 · Dental
Device Facts
Record ID
K242366
Device Name
OverCEM SA Universal; OverCEM SA Translucent; OverCEM SA Opaque; OverCEM Ti-Abutment; OverCEM Cer-Abutment
Applicant
Overfibers S.R.L.
Product Code
EMA · Dental
Decision Date
Jan 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Indications for Use
1. Definitive cementation of inlays, onlays, crowns and bridges, in ceramic or zirconia, composite or metal: 2-3 unit Maryland bridges and 3-unit inlay/onlay bridges (to be avoided for patients with bruxism or periodontitis). 2. Definitive cementation of endocanal posts. 3. Definitive cementation of ceramic, zirconia, composite or metal restorations on implant abutments and natural teeth. 4. Definitive cementation of zirconium oxide superstructures for two-unit abutments
Device Story
OverCEM SA is a dual-curing (self-curing and light-curing), self-adhesive resin luting cement. It consists of two pastes (A and B) contained in a one-body syringe, delivered via an automixing tip. Used by dental professionals in clinical settings to bond restorative materials (ceramic, zirconia, composite, metal) and endocanal posts to natural teeth or implant abutments. The device functions through polymerization of methacrylates initiated by chemical and light-activated catalysts. It provides a stable, strong connection between the restoration and the tooth/implant structure. The clinician applies the cement directly to the restoration or prepared cavity, removes excess, and cures using a light-curing unit. The device benefits patients by providing a reliable, radio-opaque, and color-stable luting solution for various dental prosthetics.
Clinical Evidence
No clinical testing performed. Evidence is based on bench testing including diametral tensile strength, film thickness, compression strength, working/setting time, flexural strength, water sorption/solubility, color stability, radio-opacity, and adhesive bond strength (ISO 29022:2013). Biocompatibility evaluated per ISO 10993-1, including cytotoxicity (ISO 10993-5), skin sensitization, and intracutaneous reactivity (ISO 10993-10).
Technological Characteristics
Dual-curing, self-adhesive resin luting cement. Materials: Methacrylates. Form factor: One-body syringe with automixing tips. Energy source: Light-curing unit (>1000mW/cm²). Standards: ISO 4049 (polymer-based restorative materials), ISO 13116 (radio-opacity), ISO 7491 (color stability), ISO 29022 (shear bond strength), ISO 10993 (biocompatibility). Connectivity: None. Sterilization: Not applicable (non-sterile).
Indications for Use
Indicated for definitive cementation of ceramic, zirconia, composite, or metal inlays, onlays, crowns, bridges (including 2-3 unit Maryland and 3-unit inlay/onlay bridges), endocanal posts, and zirconium oxide superstructures on natural teeth or implant abutments. Contraindicated for patients with bruxism or periodontitis regarding specific bridge types.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 10, 2025
Overfibers S.r.l. Castrenze Genovese Company Representative Via Malatesta.7 Imola (BO), 40026 ITALY
Re: K242366
Trade/Device Name: OverCEM SA Universal; OverCEM SA Translucent; OverCEM SA Opaque; OverCEM Ti-Abutment; OverCEM Cer-Abutment Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: Class II Product Code: EMA Dated: August 9, 2024 Received: December 11, 2024
Dear Castrenze Genovese:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAELE. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K242366
Device Name OverCEM SA Universal; OverCEM SA Translucent; OverCEM SA Opaque; OverCEM Ti-Abutment; OverCEM Cer-Abutment Indications for Use (Describe) 1. Definitive cementation of inlays, onlays, crowns and bridges, in ceramic or zirconia, composite or metal: 2-3 unit Maryland bridges and 3-unit inlav/onlay bridges (to be avoided for patients with bruxism or periodontitis). 2. Definitive cementation of endocanal posts. 3. Definitive cementation of ceramic, zirconia, composite or metal restorations on implant abutments
and natural teeth.
4. Definitive cementation of zirconium oxide superstructures for two-unit abutments
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Overfibers. The word "over" is in black, and the word "fibers" is in black with a blue swoosh above it. Below the word "fibers" is the phrase "improve your performance" in red.
## 510(k) Summary - K242366
#### 1. Contact Details:
| Applicant Name : | Overfibers S.r.l. |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Applicant Address: | Via Malatesta, 7 Imola (BO) 40026, Italy |
| Applicant Contact Telephone: | +39 345 5002938 |
| Applicant Contact: | Castrenze Genovese |
| Applicant Contact Email: | ezio.genovese@overfibers.com |
| Date Prepared: | 11/12/2024 |
| 2. Device Name: | |
| Device Trade Name: | OverCEM SA Universal;<br>OverCEM SA Translucent;<br>OverCEM SA Opaque;<br>OverCEM Ti-Abutment;<br>OverCEM Cer-Abutment. |
| Common Name: | Dental cement |
| Classification Name: | Cement, Dental |
| Regulation Number: | 872.3275 |
| Product Code(s): | EMA |
#### 3. Legally Marketed Predicate Devices
| Product | Applicant | 510(k) No. | Code No | Predicate | Decision Date |
|-----------|-----------------|------------|---------|-----------|---------------|
| G-CEM ONE | GC America Inc. | K200798 | EMA | Primary | 11/24/2020 |
#### 4. Device Description Summary
OverCEM is dual-curing, self-adhesive resin luting cement. The components consists of Paste A and B, which are filled in a one-body syringe. Both pastes are automixed with a mixing tip. The operating principle of the device is based on two reactions, one self-curing. OverCEM is designed to strongly and stably connect reconstruction materials and endocanal posts to patients' teeth or implants.
Image /page/4/Picture/8 description: The image shows three syringes of OverCEM SA Dual-Curing Self-Adhesive Resin Cement. The top syringe is labeled "Opaque", the middle syringe is labeled "Translucent", and the bottom syringe is labeled "Universal". Two blue caps with applicators are shown next to the top syringe.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Overfibers. The logo has the word "overfibers" in bold, black letters. There is a blue swoosh above the word "improve your performance" which is in smaller, red letters. The logo is simple and modern.
#### 5. Intended Use/Indications for Use
- 1. Definitive cementation of inlays, onlays, crowns and bridges, in ceramic or ziroonia, composite or metal; 2-3 unit Maryland bridges and 3-unit inlay/onlay bridges (to be avoided for patients with bruxism or periodontitis).
- 2. Definitive cementation of endocanal posts.
- Definitive cementation of ceramic, zirconia, composite or metal restorations on implant abutments and က natural teeth.
- 4. Definitive cementation of zirconium oxide superstructures for two-unit abutments
## 6. Package:
#### 1. Refill OverCEM SA
OverCEM SA syringe (Universal/Translucent/Opaque) 9,5 g (1), T-Mixer regular (8), T-Mixer wide (5) and endo-tip (5).
#### 2. Refill OverCEM Abutment
OverCEM Abutment syringe (Ti-Abutment) 9.5 g (1), T-Mixer reqular (13).
#### 3. Starter Kit OverCEM SA
- OverCEM SA syringe (Universal+Translucent+Opaque) 9,5 g (3), T-Mixer regular (24), T-Mixer wide ● (15) and endo-tip (15);
- OverCEM SA syringe (Universal) 9,5 g (3), T-Mixer regular (24), T-Mixer wide (15) and endo-tip (15); ●
- OverCEM SA syringe (Translucent) 9,5 g (3), T-Mixer regular (24), T-Mixer wide (15) and endo-tip ● (15):
- OverCEM SA syringe (Opaque) 9,5 g (3), T-Mixer regular (24), T-Mixer wide (15), and endo-tip (15). ●
## 4. Starter Kit OverCEM Abutment:
- OverCEM Abutment syringe (Ti-Abutment) 9,5 g (3), T-Mixer regular (30); ●
- OverCEM Abutment syringe (Cer-Abutment) 9,5 q (3), T-Mixer reqular (30). ●
#### 7. Shelf life and storage condition
- Shelf life 2 year: ●
- . Recommended for optimal performance, store at temperature of 2-25°C (36.0-77.0°F). Avoid exposure of the cement to direct bright light or humidity during cement application to avoid setting time shortening.
#### 8. Performance Bench Tests
The device has been tested according to ISO 4049:2009 - Dentistry - Polymer-based restorative materials, ISO 13116:2015 - Dentistry - Test Method for Determining Radio-Opacity of Materials, ISO 7491:2000 -Dental materials — Determination of colour stability and ISO 29022:2013 - Dentistry — Notchededge shear bond strength test.
Performance testing includes:
- Diametral tensile strength
- Film thickness
- Compression strength ●
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Overfibers. The word "over" is in black, and the word "fibers" is in black. There is a blue swoosh above the word "fibers". Below the two words is the phrase "improve your performance".
- Working time
- Setting time ●
- Flexural strength ●
- Water sorption
- Solubility
- Color stability ●
- Radio-Opacity ●
- . Adhesive bond strength
## 9. Non-Clinical Performance Testing
A biocompatibility assessment was completed according to ISO 10993-1:2021, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. With reference to the ISO 10993-1 standard, medical devices belonging to the OverCEM family are classified as: externally communicating, tissue/bone/dentin contact devices for a permanent contact duration (>30 days).
The tests performed are:
- . In vitro cytotoxicity based on the criteria of the protocol of ISO 10993-5:2009
- Skin sensitisation based on the criteria of the protocol of ISO 10993-10:2010 ●
- Intracutaneous reactivity based on the criteria of the protocol of ISO 10993-10:2010 ●
In conclusion, biocompatibility of OverCEM is acceptable device from the biological evaluation result.
#### 10. Clinical Performance Testing
No clinical testing has been performed on this device.
## 11. Comparison of Technology
The applicant and predicate device have the same polymerisation methods.
The raw materials used by the predicate device are substantially equivalent to those of the applicant. The applicant and predicate device are the same in function, and similar in composition and intended use. This supports that the compatibility of the applicant device is substantially equivalent to the predicate devices.
#### 12. Differences
The OverCEM device, unlike the predicate device, is a self-adhesive resin luting cement with dual-curing (selfcuring and light-curing), without the use of a second component (the predicate device needs the combination G-CEM ONE and G-CEM ONE ADHESIVE ENHANCING PRIMER for self-curing).
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Overfibers. The word "overfibers" is written in a bold, sans-serif font, with the "over" portion in a lighter color than the "fibers" portion. A blue swoosh underlines the word. Below the word, the phrase "improve your performance" is written in a smaller, sans-serif font.
| | Applicant Device (K242366) | Primary predicate (K200798) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Trade name | OverCEM SA Universal | G-CEM ONE |
| | OverCEM SA Translucent | |
| | OverCEM SA Opaque | |
| | OverCEM Ti-Abutment | |
| | OverCEM Cer-Abutment | |
| Manufacturer | Overfibers S.r.l. | GC Corporation |
| Prescription/over-the-counter use | Prescription Use | Prescription Use |
| Product category | Dual-curing, self-adhesive resin luting cement. | Self-adhesive resin cement |
| Paste/paste ratio | Paste A / Paste B = 1.0 (w/w) | Paste A / Paste B = 1.3 / 1.0 (w/w) |
| Indication for Use | 1. Definitive cementation of inlays, onlays, crowns and bridges, in | 1. Cementation of all types of all ceramic, resin and metal-based inlays, |
| | ceramic or zirconia, composite or metal; 2-3 unit Maryland bridges | onlays, crowns and bridges. |
| | and 3-unit inlay/onlay bridges (to be avoided for patients with | 2. Cementation of metal, ceramic, fiber posts, and cast post and cores. |
| | bruxism or periodontitis). | 3. Cementation of all ceramic and composite veneers. |
| | 2. Definitive cementation of endocanal posts. | 4. Final cementation of crowns and bridges on implant abutments. |
| | 3. Definitive cementation of ceramic, zirconia, composite or metal | |
| | restorations on implant abutments and natural teeth. | |
| | 4. Definitive cementation of zirconium oxide superstructures for | |
| | two-unit abutments | |
| Product description | The components consists of Paste A and B, which are filled in a | The components consist of Paste A and B, which are filled in a one-body |
| | one-body syringe. Both pastes are automixed with a mixing tip. | syringe. Both pastes are automixed with a mixing tip and directly applied |
| | The operating principle of the device is based on two reactions, | to restorations or the prepared cavity. |
| | one self-curing and one light-curing. OverCEM is designed to | |
| | strongly and stably connect reconstruction materials and | |
| | endocanal posts to patients' teeth or implants. | |
| Instruction for use | 1. Tooth preparation; | 1. Tooth preparation |
| | 2. Preparation of the root canal; | 2. Application of G-CEM ONE ADHESIVE ENHANCING PRIMER |
| | 3. Restoration preparation; | (Optional) |
| | 4. Dispensing; | 3. Restoration preparation |
| | 5. Cementation; | 4. Dispensing |
| | 6. Excess cement removal; | 5. Cementation |
| | 7. Final checks; | 6. Excess cement removal |
| | 8. Final polishing. | 7. Final set |
| | | 8. Final polishing |
| Curing specification | Light cure using a light curing unit: | Light cure using a light curing unit. |
| | 20-30 sec (>1000mW/cm² lamp) | 10 sec. (High Power LED Light) (>1200mW/cm2) |
| | | 20 sec. (Halogen/LED) (700 mW/cm2) |
| | Self-curing: | Self-curing: G-CEM ONE can be combined with the dedicated Adhesive |
| | 35°C/95°F: | Enhancing Primer (AEP). This innovative Primer induces what we call |
| | Working time: 01:30 | "touch cure mechanism” for an accelerated and very effective dark-cure |
| | Setting time: 04:00 | (self-cure) reaction of G-CEM ONE. |
| | 23°C/73.4°F: | |
| | Working time: 03:00 | |
| | Setting time: 06:00 | |
| Comparison of technology | Methacrylates contained in Paste A polymerize by polymerization | Methacrylates contained in Paste A polymerize by polymerization |
| | initiators contained in Paste A and Paste B. In addition, they also | initiators contained in Paste A and Paste B. In addition, they also |
| | polymerize by light irradiation with photo polymerization initiators | polymerize by light irradiation with photo polymerization initiators |
| | contained in Paste A. | contained in Paste A. |
| | For self-curing, working and setting times are influenced by | Furthermore, using G-CEM ONE ADHESIVE ENHANCING PRIMER for |
| | storage temperature, environment and oral temperature. The | the cavity and abutment tooth, surface modification and cement |
| | curing of OverCEM SA slows down considerably at room | polymerization are promoted and hardened with the cement. |
| | temperature and accelerates in the oral cavity. OverCEM SA is a | Methacrylates contained in this material are very hydrophobic and set |
| | dual material and therefore sensitive to daylight and artificial light. | material is stable. Therefore, the ingredients in the cured material are |
| | The working time is reduced if it is used under a white light | difficult to elute in water. |
| | operating lamp: reduce the light intensity to avoid unwanted | |
| | polymerisation. | Flexural Strength [MPa]: 150 |
| | Methacrylates contained in this material are very hydrophobic and | Film Thickness [µm]: 5 |
| | set material is stable. Therefore, the ingredients in the cured | Setting Time [s]: 240 |
| | material are difficult to elute in water. | Working Time [s]: 165 |
| | | Water sorption [µm/mm³]: 0.0000025 |
| | Flexural Strength [MPa]: 102.25 | Water solubility [µm/mm³]: 0.0000016 |
| | Film Thickness [µm]: 23.10 | Color stability: Excellent colour stability |
| | Setting Time [s]: 240 declared in the IFU (200.40 test result) | |
| | Working Time [s]: 90 | Regarding the adhesion strength, the following results are obtained for |
| | Water sorption [µm/mm³]: .45 | different substrates: |
| | Water solubility [µm/mm³]: 0.883 | Dentin [MPa]: 9.24 ± 2.06 |
| | Color stability: Excellent overall behaviour in colour stability tests | Enamel [MPa]: 26.45 ± 7.72 |
| | | Zirconia ceramic [MPa]: 32.45 ± 3.09 |
| Regarding the adhesion strength, the following results are obtained for different substrates:<br>• Dentin [MPa]: $12.95 \pm 2.22$<br>• Enamel [MPa]: $29.62 \pm 7.75$<br>• Zirconia ceramic [MPa]: $44.21 \pm 6.53$<br>• Glass Ceramic [MPa]: $31.40 \pm 5.68$<br>• Composite [MPa]: $31.27 \pm 3.51$<br>• Metal [MPa]: $34.84 \pm 4.97$ | • Glass Ceramic [MPa]: $19.55 \pm 5.06$<br>• Composite [MPa]: $22.85 \pm 4.77$<br>• Metal [MPa]: $29.34 \pm 4.71$<br>There is no information on the radio-opacity of the device.<br><br>The data on the properties of the device were taken from the documentation that the manufacturer made available on its website and from the scientific literature.<br><br>Data for adhesion strength were obtained from a test performed by Overfibers S.r.l according to ISO 29022:2013. | |
| The OverCEM devices are radio-opaque. | | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Overfibers. The word "overfibers" is written in a bold, sans-serif font, with the "over" in a smaller font size than "fibers". A blue swoosh underlines the word "overfibers". Below the logo, the tagline "improve your performance" is written in a smaller, sans-serif font.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.