CT Collaboration Live is a solution with on-console chat, audio and video call, screen sharing and remote access functionality to facilitate collaboration between CT users and between CT user and Philips Clinical Application Specialist.
Device Story
CT Collaboration Live is a software application integrated into Philips CT 5300 systems; enables two-way communication (text, voice, video, screen sharing) between CT system operator and remote user on Windows device. Facilitates peer-to-peer consultation, training, and remote system operation. Remote user can control CT system parameters via virtual control panel/touch screen only when granted access by on-site CT technologist; access limited to one remote user per session and revocable at any time. Output includes live image feed and communication interface for clinical guidance. Used in clinical settings to support remote expert consultation and training; images reviewed are not for diagnostic use.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical software verification testing, including design verification against Software Requirements Specification (SRS) and risk analysis/mitigation testing.
Technological Characteristics
Software application integrated into Philips CT systems. Enables remote image viewing, real-time communication, and remote system control. Connectivity via Windows-based laptop/desktop. Remote access requires technologist authorization. Software-based; no specific hardware materials or energy sources defined beyond host CT system. Verification testing confirms compliance with international/FDA-recognized consensus standards.
Indications for Use
Indicated for remote communication with CT console (via chat, call, video call, screensharing and remote access) by a qualified remote clinical user for consultation, guidance, support, and training in real time. Remote access must be granted by the CT technologist operating the system. Remote access is only available for Philips CT systems supporting CT Collaboration Live connectivity capabilities. Images reviewed remotely are not for diagnostic use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are displayed side-by-side.
November 18, 2024
Philips Healthcare (Suzhou) Co., Ltd % Yael Curtz Regulatory Affairs Team Lead No. 258, Zhongyuan Road, Suzhou Industrial Park Suzhou Jiangsu, 215024 CHINA
Re: K242329
Trade/Device Name: CT Collaboration Live Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ, JAK Dated: October 30, 2024 Received: October 30, 2024
### Dear Yael Curtz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K242329
Device Name
CT Collaboration Live (728290)
#### Indications for Use (Describe)
CT Collaboration Live is indicated for remote communication with CT console (via chat, call, video call, screen sharing and remote access) by a qualified remote clinical user for consultation, guidance, support, and training in real time. Remote access must be granted by the CT technologist operating the system. Remote access is only available for Philips CT systems supporting CT Collaboration Live connectivity capabilities. Images reviewed remotely are not for diagnostic use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
# 510(k) Summary
| This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92 | | |
|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Date Prepared: | August 5, 2024 | |
| Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd.<br>No. 258, Zhongyuan Road, Suzhou Industrial Park,<br>Suzhou Jiangsu, CHINA, 215024<br>Establishment Registration Number: 3009529630 | |
| Primary Contact Person: | Yael Curtz<br>Regulatory Affairs Team Lead<br>Phone: (972) 54-7282285<br>E-mail: yael.curtz@philips.com | |
| Secondary Contact<br>Person: | Carmit Shmuel<br>Regulatory Affairs Manager and Site Lead<br>Phone: (972) 54-2109054<br>E-mail: Carmit.shmuel@philips.com | |
| Device: | Trade Name: | CT Collaboration Live |
| | Common name: | Picture archiving and communications system |
| | Classification Name: | System, Image Processing, Radiological<br>Picture archiving and communications system<br>(PACS) |
| | Classification Regulation: | 21 CFR 892.2050<br>21 CFR 892.1750 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Primary Product Code:<br>Secondary Product Code: | LLZ<br>JAK |
| Primary Predicate<br>Device: | Trade Name: | Collaboration Live |
| | Manufacturer: | Philips Ultrasound, Inc. |
| | 510(k) Clearance: | K200179 (Predicate) |
| | Classification Name: | System, Image Processing, Radiological<br>Picture archiving and communications system<br>(PACS) |
| | Classification Regulation: | 21 CFR 892.2050<br>21 CFR 892.1550<br>21 CFR 892.1560 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | LLZ |
| | Secondary Product Code: | JYN, JYO |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the primary focus of the image and is presented against a white background.
- Device Description: The proposed CT Collaboration Live is a software application integrated in Philips Computed Tomography (CT) X-Ray CT 5300 Systems. CT Collaboration Live enables two-way communication of text, voice, image, and video information between a CT system operator and a remote user on a Windows device. CT Collaboration Live facilitates: 1) peer-to-peer consultation and training, 2) system-level remote sharing & operation, 3) access to live image feed and 4) remote expert user(s) and physician consultation. CT Collaboration Live functionality includes a remote-control feature in which the CT system operator may grant a qualified remote user control of the CT system parameters via a virtual control panel and virtual touch screen.
- Intended Use: CT Collaboration Live is a solution with on-console chat, audio and video call, screen sharing and remote access functionality to facilitate collaboration between CT users and between CT user and Philips Clinical Application Specialist.
Contraindications are not known.
- Indications for Use: CT Collaboration Live is indicated for remote communication with CT console (via chat, call, video call, screensharing and remote access) by a qualified remote clinical user for consultation, guidance, support, and training in real time. Remote access must be granted by the CT technologist operating the system. Remote access is only available for Philips CT systems supporting CT Collaboration Live connectivity capabilities. Images reviewed remotely are not for diagnostic use.
- Technological The proposed CT Collaboration Live has the same fundamental design Characteristics characteristics as the predicate Collaboration Live (K200179). CT Collaboration Live is a software application that enables remote image viewing and real-time communication between the system operator and a remote user (on a laptop or desktop computer) for the purposes of service, training/education, and peer-topeer collaboration. CT Collaboration Live also contains a remote-control feature similar to Collaboration Live, in which a qualified remote user can control the system parameters when granted access by the system operator. Like Collaboration Live, remote control in CT Collaboration Live can be granted to only one remote user at a time and may be revoked at any time by the system operator. Based on the information provided above, the CT Collaboration Live is
considered substantially equivalent to the primary currently marketed and predicate device Collaboration Live (K200179) in terms of fundamental design technology.
The following table 1 lists the technological characteristics differences for the proposed CT Collaboration Live:
| Table 1 – Technological Characteristics Comparison | | | |
|----------------------------------------------------|---------------------------------------------------------------------------------|-----------------------------------------|------------------------------------------------------|
| Design<br>Feature | Cleared to<br>Market<br>Predicate<br>Device:<br>Collaboration<br>Live (K200179) | Proposed<br>Device: CT<br>Collaboration | Conclusion |
| Regulation<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Identical, therefore<br>substantially<br>equivalent. |
{6}------------------------------------------------
| Regulation<br>Name | System, Image<br>processing,<br>Radiological-<br>Picture<br>Archiving and<br>Communications<br>System (PACS) | System, Image<br>processing,<br>Radiological-<br>Picture<br>Archiving and<br>Communications<br>System (PACS) | Identical, therefore<br>substantially<br>equivalent. |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>Code | LLZ | LLZ | Identical, therefore<br>substantially<br>equivalent. |
| Features | Remote service<br>support,<br>Remote clinical<br>training and<br>education,<br>Remote peer to<br>peer<br>collaboration<br>(non-diagnostic) | Peer to peer<br>consultation and<br>training,<br>System level<br>remote sharing<br>and operation,<br>Access to live<br>image feed,<br>Remote expert<br>user(s) and<br>physician<br>consultation | Features were<br>elaborated for<br>clarity. Equivalent,<br>no impact to safety<br>and effectiveness,<br>therefore<br>substantially<br>equivalent. |
| Remote<br>Access<br>Control | Granted by<br>technologist at<br>system | Granted by<br>technologist at<br>system | Identical, therefore<br>substantially<br>equivalent. |
| Remote<br>privileges | Secure<br>credentials | Secure<br>credentials | Identical, therefore<br>substantially<br>equivalent. |
| Number of<br>remote<br>users per<br>session | 1 | 1 | Identical, therefore<br>substantially<br>equivalent. |
| Diagnostic<br>use | Remote images<br>are not for<br>diagnostic use | Remote images<br>are not for<br>diagnostic use | Identical, therefore<br>substantially<br>equivalent. |
Based on the above, the proposed CT Collaboration Live is considered substantially equivalent to the predicate device Collaboration Live (K200179), in terms of technological characteristics.
Summary of Non-Clinical Non-clinical performance software verification testing has been performed on Performance Data: the proposed CT Collaboration Live.
> All tests were used to support substantial equivalence of the proposed CT Collaboration Live and to demonstrate that CT Collaboration Live:
> • Complies with international and FDA-recognized consensus standards and/or FDA device Specific guidance document, and;
- Meets the acceptance criteria and is adequate for its intended use.
Therefore, CT Collaboration Live is substantially equivalent to the predicate device Collaboration Live (K200179) in terms of safety and effectiveness.
Design Verification planning and testing against Software Requirements Specification (SRS), was conducted. Software verification activities demonstrate
{7}------------------------------------------------
that the CT Collaboration Live software application meets the design input requirements. Software requirements may be verified by manual test, automated test, inspection/analysis, or any combination of the three. Design verification also includes risk analysis/risk mitigation testing.
The traceability between the requirements and the test protocols are described in the Traceability Matrix. The Traceability Matrix also shows the overall test results per requirement. The summary and conclusion of results are provided in the System Verification Test Report.
#### Summary of Clinical The proposed CT Collaboration Live did not require a clinical study since Performance Data: substantial equivalence to the predicate device Collaboration Live (K200179) was demonstrated with the following attributes:
- Indication for use; ●
- Technological characteristics;
- Non-clinical performance testing; and
- Safety and effectiveness.
Substantial Equivalence The proposed CT Collaboration Live is substantially equivalent to the predicate device Collaboration Live (K200179) in terms of indications for use, Conclusion: technological characteristics and safety and effectiveness.
> Additionally, substantial equivalence was demonstrated by verification tests provided in this 510(k) premarket notification. These tests demonstrate that CT Collaboration Live complies with the requirements specified in the international and FDA-recognized consensus standards and is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.