← Product Code [NAY](/productcode/NAY) · K242318

# da Vinci SP Surgical System (SP1098) (K242318)

_Intuitive Surgical, Inc. · NAY · Nov 26, 2024 · Gastroenterology, Urology · SESU_

**Canonical URL:** https://fda-staging.innolitics.com/device/K242318

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [NAY](/productcode/NAY.md)
- **Decision Date:** Nov 26, 2024
- **Decision:** SESU
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K242318 · Nov 26, 2024 | da Vinci SP Surgical System (SP1098) | Intuitive Surgical, Inc. | Published clinical literature (2010-2024) on multiport robotic-assisted colorectal surgery; Systematic review of clinical cohorts | The sponsor conducted a systematic literature review according to PRISMA methods to compare clinical outcomes of the da Vinci SP system against pooled data from published literature on multiport robotic systems. This evidence was used to demonstrate that the SP system does not raise different questions of safety or effectiveness for colorectal procedures. | Systematic Literature Review; Colorectal Surgery; Comparative Effectiveness; Clinical Outcomes |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Systematic Literature Review (2010-2024); Systematic literature review and meta-analysis comparison; Study Period: 2010 - March 2024 | Patients undergoing LAR/TME and right colectomy procedures; Sample Size: N=11,340 to N=48,348 (pooled literature cohorts); Number of Sites: Multiple (international) | Multiport robotic systems (da Vinci Xi/Multiport) | Operative time, length of stay, estimated blood loss, adverse event rates, conversion rate, mortality, readmission, reoperation |

## Indications for Use

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical da Vinci SP Instruments during urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

## Device Story

The da Vinci SP Surgical System is a computer-controlled robotic platform for minimally invasive surgery. It comprises a Surgeon Console, Vision Cart, and Patient Cart. The surgeon operates from the console using hand controls and foot pedals while viewing a 3D image of the anatomy. The Patient Cart, positioned at the bedside, features an instrument arm with four drives that manipulate up to three instruments and a camera through a single port. The system enables precise tissue manipulation, including grasping, cutting, dissection, ligation, electrocautery, and suturing. It is used by trained physicians in an operating room to perform complex procedures, including colorectal surgeries like LAR/TME and colectomies. The system provides enhanced visualization and instrument control, potentially reducing surgical trauma and improving patient outcomes compared to traditional approaches.

## Clinical Evidence

Prospective, multicenter, single-arm clinical study (IDE) of 60 subjects (40 US, 20 South Korea) undergoing colorectal procedures (29 LAR/TME, 28 right colectomy, 3 other). Primary endpoints included safety and performance. LAR/TME cohort: 0% conversion rate, 0% device-related SAE rate, 10.3% major AE rate. Right colectomy cohort: 0% conversion rate, 0% device-related SAE rate, 0% major AE rate. Comparison to multiport robotic literature (N=11,340 for LAR/TME; N=10,907 for right colectomy) showed 95% CI overlap for all clinical parameters, confirming substantial equivalence.

## Technological Characteristics

Computer-controlled robotic surgical system consisting of Surgeon Console, Vision Cart, and Patient Cart. Features 3D visualization, four-drive instrument arm for single-port access, and EndoWrist SP instruments. Connectivity via networking hardware in Vision Cart. Software-controlled movement. No changes to materials, sterilization, or electrical/mechanical performance from predicate.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- da Vinci SP Surgical System, Model SP1098, and da Vinci SP Instruments and Accessories ([K240502](/device/K240502.md))

## Reference Devices

- da Vinci Xi Surgical System, Model IS4000, EndoWrist Instruments, and Accessories ([K171632](/device/K171632.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 26, 2024

Intuitive Surgical, Inc. Mike Yramategui Fellow Regulatory Engineer 1020 Kifer Road Sunnyvale, California 94086

Re: K242318

Trade/Device Name: da Vinci SP Surgical System (SP1098) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NAY Dated: November 1, 2024 Received: November 1, 2024

Dear Mike Yramategui:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

The OHT4: Office of Surgical and Infection Control Devices has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the device's labeling:

{1}------------------------------------------------

The safety and effectiveness of this device for use in the performance of general laparoscopic surgery procedures have not been established. This device is only intended to be used for single port urological, colorectal, and general thoracoscopic surgical procedures and for transoral otolaryngology surgical procedures in the oropharynx for benign tumors and malignant tumors classified as T1 and T2 with the da Vinci EndoWrist SP Instruments and the da Vinci SP Surgical System (SP1098).

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). 

{2}------------------------------------------------

The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

BLETA
VUNIQI -S
Digitally signed by
BLETA VUNIQI -S
Date: 2024.11.26
15:28:57 -05'00'

For. Binita Ashar, M.D., M.B.A., F.A.C.S. Director OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K242318

Device Name

da Vinci SP Surgical System (SP1098)

Indications for Use (Describe)

da Vinci SP Surgical System, Model SP1098:

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical da Vinci SP Instruments during urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

# da Vinci SP Instruments:

Intuitive Surgical da Vinci SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/ pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary [21 CFR § 807.92(c)]

#### SUBMITTER INFORMATION I.

| Submitter: | Intuitive Surgical, Inc.            |
|------------|-------------------------------------|
|            | 1266 Kifer Road                     |
|            | Sunnyvale, CA 94086                 |
| Contact:   | Mike Yramategui                     |
|            | Fellow Regulatory Engineer          |
|            | Phone Number: 408-523-2145          |
|            | Fax Number: 408-523-8907            |
|            | Email: Mike.Yramategui@intusurg.com |

Date Summary Prepared: November 15, 2024

#### SUBMITTER INFORMATION II.

| Trade Name:          | da Vinci SP® Surgical System, Model SP1098,<br>da Vinci SP® Instruments, and Accessories |
|----------------------|------------------------------------------------------------------------------------------|
| Common Name:         | System, Surgical, Computer Controlled Instrument                                         |
| Classification Name: | Endoscope and Accessories (21 CFR §876.1500)                                             |
| Regulatory Class:    | Class II                                                                                 |
| Product Code:        | NAY (System, Surgical, Computer Controlled Instrument)                                   |
| Submission Type:     | Traditional 510(k)                                                                       |

# III. PREDICATE DEVICE INFORMATION

| Predicate Device: | da Vinci SP Surgical System, Model SP1098,<br>da Vinci SP Instruments, and Accessories (K240502) |
|-------------------|--------------------------------------------------------------------------------------------------|
| Reference Device: | da Vinci Xi Surgical System, Model IS4000,<br>EndoWrist Instruments, and Accessories (K171632)   |

{5}------------------------------------------------

#### IV. DEVICE DESCRIPTION

The da Vinci SP Surgical System is designed to enable complex surgery using a minimally invasive approach. The system consists of a Surgeon Console, a Vision Cart, and a Patient Cart and is used with a camera, instruments, and accessories.

The surgeon seated at the Surgeon Console controls all movement of the instruments and camera by using two hand controls and a set of foot pedals. The surgeon views the camera image on a three-dimensional (3D) viewer, which provides a view of patient anatomy and instrumentation, along with icons and other user interface features.

The Vision Cart includes supporting electronic equipment, such as the camera light source, video and image processing, and the networking hardware. The Vision Cart also has a touchscreen to view the camera image and adjust system settings.

The Patient Cart is the operative component of the da Vinci SP Surgical System. Its primary function is to support the positioning of the surgical port and to manipulate the surgical instruments and camera. The Patient Cart is positioned at the operating room and contains an instrument arm that is positioned with respect to the target patient anatomy. The instrument arm contains four instrument drives that hold up to three surgical instruments and the camera. The patient-side assistant installs and removes the camera and instruments intra-operatively.

This 510(k) is for a labeling modification only, to add "colorectal surgical procedures" to the indications, and to add "Low anterior resection / total mesorectal excision (LAR/TME), Colectomy (Right, Left, Transverse, Total, Hemi), Sigmoidectomy, and Abdominoperineal Resection (APR)" as new representative, specific procedures in the Professional Instructions for Use.

{6}------------------------------------------------

#### INDICATIONS FOR USE

#### da Vinci SP Surgical System, Model SP1098:

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System. Model SP1098) is intended to assist in the accurate control of Intuitive Surgical EndoWrist SP Instruments during urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach, and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

#### da Vinci SP Instruments:

Intuitive Surgical da Vinci SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach, and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional

#### COMPARISON OF INTENDED USE, INDICATIONS FOR USE AND V. TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The da Vinci SP Surgical System, Model SP1098 and da Vinci SP Instruments and Accessories are unchanged from the predicate device in terms of intended use, design, performance, and technological characteristics. The labeling has been changed to add "colorectal surgical procedures" to the indication statement and to add "Low anterior resection / total mesorectal excision (LAR/TME), Colectomy (Right, Left, Transverse, Total, Hemi), Sigmoidectomy, and Abdominoperineal Resection (APR)" as new representative, specific procedures in the Professional Instructions for Use.

{7}------------------------------------------------

#### VI. PERFORMANCE DATA

The addition of colorectal surgical procedures to the SP1098 Indications does not change any of the safety or performance requirements that were previously verified and / or validated for the SP1098 regarding cleaning, sterilization, packaging, shelf life, biocompatibility, software, cybersecurity, electrosurgical performance, electromagnetic compatibility, electrical safety, mechanical and electrical performance, reliability, or human factors for use in urologic (K173906), transoral surgery (K182371) or general thoracoscopic surgical procedures (K240502).

#### Cadaver and Animal Performance Testing

Comparative animal and cadaver testing was conducted to demonstrate the equivalence of the SP1098 System to the multiport da Vinci Xi system, Model IS4000 for performing colorectal surgical procedures. Cadavers were used to demonstrate device performance for anatomical access and reach. Live animals were used to assess safety and performance in cases where live tissue model was appropriate. These models replicate factors experienced during normal clinical use, including operating on perfused organs, normal tissue manipulation, and ensuring that appropriate hemostasis is achieved and maintained.

Three (3) independent practicing surgeons participated in a pre-clinical study using the SP1098 system to perform a set of colorectal surgical procedures. Each surgeon performed two (2) procedures in a cadaver model and one (1) procedure in a porcine model, for a total of 9 surgical procedures performed per system. Success criteria for each procedure are listed below (Table 1). In addition, surgeons completed questionnaires that evaluated their ability to perform surgical tasks of the SP1098 and the multiport da Vinci Xi system, which is cleared for use in colorectal surgical procedures (K171632) and serves as a reference device for this 510(k).

{8}------------------------------------------------

# Table 1

| Surgical<br>Procedure                                       | Subject | Procedure Completion Acceptance Criteria                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|-------------------------------------------------------------|---------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Low Anterior<br>Resection / Total<br>Mesorectal<br>Excision | Cadaver | Sufficient colon mobilization to ensure tension-free anastomosis Complete mesorectal dissection (e.g., intact mesorectal fascial envelope, circumferential resection, distal dissection to levator ani) Critical anatomy identified (e.g., gonadal vessels, inferior mesenteric vessels, ureter, hypogastric nerves, pelvic nerves) Appropriate vessel ligation (e.g., inferior mesenteric or superior rectal vessels) Anastomosis complete and deemed surgically acceptable upon visual inspection Acceptable leak test (i.e., no air bubbles observed during leak test) |
| Right<br>Hemicolectomy                                      | Cadaver | Sufficient colon mobilization to ensure tension-free anastomosis Critical anatomy identified (e.g., gonadal vessels, ureter, ileocolic vessels, colic vessels, duodenum) Appropriate vessel ligation (e.g., right colic, ileocolic, right branch of middle colic vessels) Anastomosis complete and deemed surgically acceptable upon visual inspection Acceptable leak test (i.e., no air bubbles observed during leak test)                                                                                                                                              |
| Low Anterior<br>Resection / Total<br>Mesorectal<br>Excision | Porcine | Sufficient colon mobilization to ensure tension- free anastomosis Anastomosis complete and deemed surgically acceptable upon visual inspection Acceptable leak test (i.e. no air bubbles observed during leak test) Hemostasis maintained                                                                                                                                                                                                                                                                                                                                 |

{9}------------------------------------------------

# Clinical Study

Intuitive conducted a prospective, multicenter, single-arm clinical study under an Investigational Device Exemption (IDE) to confirm the clinical safety and performance of the da Vinci SP Surgical System. Instruments and Accessories in colorectal surgical procedures. This study also confirms that use of the da Vinci SP Surgical System in colorectal surgical procedures does not raise different questions of safety or effectiveness. The study included 60 subjects (40 subjects in the United States and 20 subjects in South Korea) at eight (8) institutions in the United States and two (2) institutions in South Korea. The da Vinci SP Surgical System was used to perform the following procedures in this study:

- Low Anterior Resection with/without Total Mesorectal Excision (LAR/TME) (N = 29) .
- Right Colectomy (N = 28) ●
- Other colorectal procedures (N = 3) .
	- Sigmoid Colectomy (N = 2) o
	- Transverse Colectomy (N=1) O

# LAR/TME Results

Twenty-nine (29) subjects were enrolled in the study for LAR/TME with nineteen (19) subjects enrolled in the U.S. and ten (10) subjects enrolled in South Korea. Of these 29 subjects, 26 (89.7%) were malignant indications and 3 (10.3%) were benign indications. The mean BMI observed in the LAR/TME cohort was 25.7, with BMI in U.S. population was slightly higher (26.3) than the South Korean population (24.4).

All LAR/TME procedures in the study were completed using a da Vinci SP Surgical System without conversion to open or an alternative approach. Therefore, the conversion rate is 0% in the study for in the LAR/TME cohort.

There were no intraoperative adverse events and no unanticipated device-related adverse effects (UADE) reported for the LAR/TME cohort. There were eight (8) serious adverse events (SAE) reported, all deemed related to the colorectal procedure only and not related to the device. Of those eight (8) SAEs, three (3) were classified as major (all within Clavien-Dindo III classification). "Major" is defined by Clavien-Dindo Classification Grade IIIa or higher. Table 2 summarizes the study results and Table 3 lists the serious adverse events and major complications reported in the LAR/TME cohort.

{10}------------------------------------------------

| Clinical Study Parameter                                                                                                                                                                    | K242318       |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Number of Subjects, n                                                                                                                                                                       | 29            |
| Operative time (minutes), mean ± SE                                                                                                                                                         | 273.6 ± 14.84 |
| Estimated blood loss (mL), mean ± SE                                                                                                                                                        | 75.2 ± 16.12  |
| Blood transfusion rate, %                                                                                                                                                                   | 0             |
| Conversion rate, %                                                                                                                                                                          | 0             |
| Serious adverse event rate                                                                                                                                                                  | 17.2%         |
| Device-related serious adverse event rate, %                                                                                                                                                | 0             |
| Intraoperative adverse event rate, %                                                                                                                                                        | 0             |
| Subjects with major AE<br>(Clavien-Dindo Grade III/IV/V, n (%))a                                                                                                                            | 3 (10.3%)     |
| Length of hospital stay (days)b, mean ± SE                                                                                                                                                  | 4.6 ± 3.69    |
| Rate of positive surgical margin, %, n= 7                                                                                                                                                   | 0             |
| Readmission rate, %c                                                                                                                                                                        | 4 (14.3%)     |
| Reoperation rate, %                                                                                                                                                                         | 0             |
| Mortality rate%                                                                                                                                                                             | 0             |
| Notes:                                                                                                                                                                                      |               |
| a. There were no grade IV or V events.                                                                                                                                                      |               |
| b. Mean length of hospital stay in the U.S. was 2.9 days and the mean length of<br>hospital stay in South Korea was 7.7 days.                                                               |               |
| c. 4 subjects had unplanned re-admissions related to a SAE (3 classified as major and<br>1 classified as minor). None of the unpanned re-admissions were related to the da Vinci SP system. |               |

Table 2. Results Summary (LAR/TME)

| Table 3. Postoperative Serious Adverse Events and Major Complications (LAR/TME) |                   |
|---------------------------------------------------------------------------------|-------------------|
|                                                                                 |                   |
| Type                                                                            | Total study N= 29 |
| Subjects experiencing SAEs                                                      | 5                 |
| Subject Experiencing Major AE Complications                                     | 3                 |
| Total Number of SAEs by Minor/Major Classification                              | 8                 |
| Minor (All Clavien-Dindo Grade I)                                               |                   |
| Acute Renal Failure                                                             | 1                 |
| Dehydration                                                                     | 1                 |
| High Ileostomy Output                                                           | 1                 |
| Ileus                                                                           | 2                 |
| Major (within Clavien-Dindo Grade III)                                          |                   |
| Anastomotic Dehiscence                                                          | 1                 |
| Bowel Obstruction                                                               | 1                 |
| Hematoma of Presacral Space                                                     | 1                 |

{11}------------------------------------------------

### Right Colectomy Results

Twenty-eight (28) subjects were enrolled in the study for right colectomy, with eighteen (18) subjects enrolled in the U.S. and ten (10) subjects enrolled in South Korea. Of these 28 subjects, 22 (78.6%) were malignant indications and 6 (21.4%) was for a benign indication. The mean BMI observed in the Right Colectomy cohort was 25.8, with BMI in U.S. population was slightly higher (26.9) than the South Korean population (23.8), there was no statistical significance between the two regions.

All right colectomy procedures in the study were completed using a da Vinci SP Surgical System without conversion to open or an alternative approach, therefore, the conversion rate is 0% in the study for in the right colectomy cohort.

There was one (1) intraoperative adverse event (corneal abrasion) deemed not related to the colorectal procedure and/or device. There were no unanticipated device-related adverse effects (UADE). Two (2) Subjects experienced one (1) serious adverse event each, all relate to the colorectal procedure only and none classified as major AE complications (all Clavien-Dindo Grade I). Table 4 summarizes the study results and Table 5 lists the serious adverse events and major complications reported in the right colectomy cohort.

| Clinical Study Parameter                                                                                                      | K242318       |
|-------------------------------------------------------------------------------------------------------------------------------|---------------|
| Number of Subjects, n                                                                                                         | 28            |
| Operative time (min), mean ± SE                                                                                               | 208.2 ± 11.43 |
| Estimated blood loss (mL), mean ± SE                                                                                          | 37.0 ± 8.4    |
| Transfusion rate, %                                                                                                           | 0             |
| Conversion to open rate, %                                                                                                    | 0             |
| Serious adverse event rate, %                                                                                                 | 7.1           |
| Device-related serious adverse event rate, %                                                                                  | 0             |
| Intraoperative adverse event, %                                                                                               | 3.6%          |
| Subjects with major adverse events, AE<br>(Clavien-Dindo Grade III/IV/V), n (%)                                               | 0             |
| Length of hospital stay (days), mean ± SE a                                                                                   | 3.5 ± 0.38    |
| Rate of positive surgical margins, n (%)                                                                                      | 0             |
| Readmission rate, n (%)                                                                                                       | 0             |
| Reoperation rate%                                                                                                             | 0             |
| Mortality rate, %                                                                                                             | 0             |
| Notes:                                                                                                                        |               |
| a. Mean length of hospital stay in the U.S. was 2.3 days and the mean<br>length of hospital stay in South Korea was 5.8 days. |               |

Table 4. Results Summary (Right Colectomy)

{12}------------------------------------------------

| Type                                        | Total study N=19 |
|---------------------------------------------|------------------|
| Subjects experiencing SAEs                  | 2                |
| Subject Experiencing Major AE Complications | 0                |
| Total Number of SAEs                        | 2                |
| Minor (Clavien-Dindo Grade I)               |                  |
| Acute Renal Failure                         | 1                |
| Ileus                                       | 1                |

#### Table 5. Postoperative Adverse Events (Right Colectomy)

# Comparison of da Vinci SP Study Results to Literature on Multiport Roboticassisted Colorectal Surgery

Results of this study of da Vinci SP were compared to recently published clinical literature between 2010 and March 2024 on multiport robotic systems for LAR/TME and right colectomy procedures. Systematic literature searches were done for LAR/TME, and right colectomy published literature in this time period, were conducted according to PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) methods. Table 6 summarizes the comparison in LAR/TME. Table 7 summarizes the comparison in right colectomy. Each parameter from the literature is shown based on the pooled sample size from multiple publications and 95% confidence interval based on estimated weighted average and weighted standard deviation from multiple publications. Table 8 provides a bibliography of the literature on multiport articles used for the comparison.

Comparison of the results from the da Vinci SP Surgical System to the published literature demonstrate that the 95% confidence intervals (CI) for all 12 clinical parameters in the SP study overlap the 95% Cls calculated from the published clinical literature on da Vinci Multiport systems for both LAR/TME and right colectomy. This comparison demonstrates that the da Vinci SP Surgical System is as safe and effective as the predicate device and does not raise different questions of safety or effectiveness.

{13}------------------------------------------------

| Clinical Study Parameter              | K242318          | da Vinci Multi-port<br>Literature |
|---------------------------------------|------------------|-----------------------------------|
| Operative time                        |                  |                                   |
| Sample Size                           | N = 29           | N = 11340                         |
| Mean ± SE                             | 273.6 ± 14.84    | 282.3 ± 12.38                     |
| 95% CI                                | [243.18, 303.99] | [256.43, 308.10]                  |
| Length of stay                        |                  |                                   |
| Sample Size                           | N = 29           | N = 26500                         |
| Mean ± SE                             | 4.6 ± 0.69       | 7.7 ± 0.79                        |
| 95% CI                                | [3.15, 5.96]     | [6.12, 9.37]                      |
| Estimated blood loss                  |                  |                                   |
| Sample Size                           | N = 29           | N = 5055                          |
| Mean ± SE                             | 75.2 ± 16.12     | 63.0 ± 14.66                      |
| 95% CI                                | [42.22, 108.26]  | [31.29, 94.63]                    |
| Intra-operative adverse event         |                  |                                   |
| Sample Size                           | N = 29           | N = 1739                          |
| Proportion                            | 0%               | 5.8%                              |
| 95% CI                                | [0%, 11.94%]     | [2.95%, 8.72%]                    |
| Major adverse event complication rate |                  |                                   |
| (Clavien-Dindo III, IV, V) a          |                  |                                   |
| Sample Size                           | N = 29           | N = 9928                          |
| Proportion                            | 10.3%            | 8.1%                              |
| 95% CI                                | [2.19%, 27.35%]  | [6.47%, 9.82%]                    |
| Anastomotic complication rate b       |                  |                                   |
| Sample size                           | N = 29           | N=17204                           |
| Proportion                            | 10.3%            | 6.4%                              |
| 95% CI                                | [2.19%, 27.35%]  | [4.75%, 7.97%]                    |
| Transfusion rate                      |                  |                                   |
| Sample Size                           | N = 29           | N = 10332                         |
| Proportion                            | 0%               | 2.3%                              |
| 95% CI                                | [0%, 11.94%]     | [1.35%, 3.22%]                    |
| Conversion rate                       |                  |                                   |
| Sample Size                           | N = 29           | N = 48348                         |
| Proportion                            | 0%               | 6.2%                              |
| 95% CI                                | [0%, 11.94%]     | [5.39%, 6.92%]                    |
| Mortality rate c                      |                  |                                   |
| Sample Size                           | N = 28           | N = 36974                         |
| Proportion                            | 0%               | 0.6%                              |
| 95% CI                                | [0%, 12.34%]     | [0.41%, 0.72%]                    |
| Readmission rate c                    |                  |                                   |
| Sample Size                           | N = 28           | N = 30954                         |
| Proportion                            | 14.3%            | 7.1%                              |
| 95% CI                                | [4.03%, 32.67%]  | [4.96%, 9.14%]                    |
| Reoperation rate c                    |                  |                                   |
| Sample Size                           | N = 28           | N = 8107                          |
| Proportion                            | 0%               | 5.5%                              |
| 95% CI                                | [0%, 12.34%]     | [4.37%, 6.59%]                    |

# Table 6. da Vinci SP LAR/TME versus da Vinci Multiport Literature Comparison

{14}------------------------------------------------

| Notes: | a. There were no grade IV or V events.<br>b. Anastomoses were created using handheld, third-party laparoscopic staplers or were hand-sewn.<br>c. One subject missed their Day 42 visit due to a SAE. |
|--------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

# Table 7. da Vinci SP Right Colectomy versus da Vinci Multiport Literature Comparison

| Clinical Parameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K242318          | da Vinci Multi-port<br>Literature |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|-----------------------------------|
| Operative time                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 10907                         |
| Mean ± SE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 208.2 ± 11.43    | 285.5 ± 26.92                     |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [184.76, 231.67] | [210.74, 360.25]                  |
| Length of stay (days)                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 11164                         |
| Mean ± SE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 3.5 ± 0.38       | 5.1 ± 0.28                        |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [2.75, 4.32]     | [4.37, 5.81]                      |
| Estimated blood loss                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 394                           |
| Mean ± SE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 37.0 ± 8.40      | 98.7 ± 19.65                      |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [19.73, 54.20]   | [0, 348.40]                       |
| Intra-operative adverse event rate                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 520                           |
| Proportion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 3.6%             | 2.8%                              |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [0.09%, 18.35%]  | [0%, 29.23%]                      |
| Major adverse event rate<br>(Clavien-Dindo III, IV, V)                                                                                                                                                                                                                                                                                                                                                                                                                                            |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 3203                          |
| Proportion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0%               | 6.7%                              |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [0%, 12.34%]     | [5.55%, 7.93%]                    |
| Anastomotic complication rate ª                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                  |                                   |
| Sample size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N=28             | N=14926                           |
| Proportion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0%               | 4.2%                              |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [0%, 12.34%]     | [2.53%, 5.79%]                    |
| Transfusion rate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 2883                          |
| Proportion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0%               | 7.3%                              |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [0%, 12.34%]     | [4.47%, 10.11%]                   |
| Conversion rate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 25893                         |
| Proportion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0%               | 6.0%                              |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [0%, 12.34%]     | [5.12%, 6.88%]                    |
| Mortality rate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 11684                         |
| Proportion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0%               | 0.6%                              |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [0%, 12.34%]     | [0.33%, 0.89%]                    |
| Readmission rate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                  |                                   |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 11236                         |
| Proportion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0%               | 6.8%                              |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [0%, 12.34%]     | [5.53%, 8.01%]                    |
| Reoperation rate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                  |                                   |
| Sample Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N = 28           | N = 9382                          |
| Proportion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0%               | 5.8%                              |
| 95% CI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | [0%, 12.34%]     | [4.96%, 6.56%]                    |
| The 95% confidence intervals for 9 discrete variables from the literature data are based on pooled<br>proportions and exact method. The weighted averages and 95% confidence intervals for 3 continuous<br>variables of the 12 variables from the literature data are based on weighted averages and weighted<br>standard deviations where the sample means and standard deviations for each publication are taken as<br>reported or estimated using quantiles as reported (McGrath et al. 2020). |                  |                                   |
| Notes: a. Third-party handheld laparoscopic staplers were used in the creation of the<br>anastomosis.                                                                                                                                                                                                                                                                                                                                                                                             |                  |                                   |

{15}------------------------------------------------

# VII. CONCLUSION

The Performance test data demonstrates that the ability of the subject device (SP1098) to perform colorectal surgical procedures, LAR/TME and right colectomy. The Clinical Study results confirm that there are no different issues of safety or effectiveness as compared to the multiport da Vinci system when performing colorectal surgical procedures, LAR/TME and right colectomy.

Thus, these labeling changes to the da Vinci SP Surgical System are substantially equivalent to the cleared predicate device.

{16}------------------------------------------------

#### Table 8. Bibliography of Clinical Literature on da Vinci Multiport LAR/TME (n=55)

- 1. Arron, M. N. N. N. G. Greijdanus, R. P. G. Ten Broek, J. W. T. Dekker, F. van Workum, H. van Goor, P. J. Tanis and J. H. W. de Wilt (2021). "Trends in risk factors of anastomotic leakage after colorectal cancer surgery (2011-2019): A Dutch population-based study." Colorectal Dis 23(12): 3251-3261.
- 2. Arron, M. N. N., N. G. Greijdanus, S. Bastiaans, P. A. J. Vissers, R. H. A. Verhoeven, R. P. G. ten Broek, H. M. W. Verheul, P. J. Tanis, H. van Goor and J. H. W. de Wilt (2022). "Long-Term Oncological Outcomes After Colorectal Anastomotic Leakage A Retrospective Dutch Population-based Study." Annals of Surgery 276(5): 882-889.
- Bedrikovetski, S., N. N. Dudi-Venkata, H. M. Kroon, J. W. Moore, R. A. Hunter and T. 3. Sammour (2020). "Outcomes of Minimally Invasive Versus Open Proctectomy for Rectal Cancer: A Propensity-Matched Analysis of Bi-National Colorectal Cancer Audit Data." Dis Colon Rectum 63(6): 778-787.
- Bertani, E., A. Chiappa, R. Biffi, P. P. Bianchi, D. Radice, V. Branchi, E. Cenderelli, I. 4. Vetrano, S. Cenciarelli and B. Andreoni (2011). "Assessing appropriateness for elective colorectal cancer surgery: clinical, oncological, and quality-of-life short-term outcomes employing different treatment approaches." Int J Colorectal Dis 26(10): 1317-1327.
- 5. Bhangu, A. and g. European Society of Coloproctology collaborating (2018). "An international multicentre prospective audit of elective rectal cancer surgery; operative approach versus outcome, including transanal total mesorectal excision (TaTME)." Colorectal Dis 20 Suppl 6: 33-46.
- Bliznakova, K., N. Kolev, A. Zlatarov, T. Kalinov and T. Georgiev (2023). "Feasibility and 6. Safety of Robotic-Assisted Surgery for Rectal Cancer: Short-Term Outcomes of a Pilot Study with da Vinci Xi Platform During COVID-19." Chirurgia (Bucur) 118(1): 27-38.
- 7. Burghgraef, T. A., J. C. Hol. M. L. Rutgers, R. Crolla, A. A. W. van Geloven, R. Hompes, J. W. A. Leijtens, F. Polat, A. Pronk, A. B. Smits, J. B. Tuynman, E. G. G. Verdaasdonk, P. M. Verheijen, C. Sietses and E. C. J. Consten (2022). "Laparoscopic Versus Robot-Assisted Versus Transanal Low Anterior Resection: 3-Year Oncologic Results for a Population-Based Cohort in Experienced Centers." Ann Surg Oncol 29(3): 1910-1920.
- Cheong, J. Y., C. J. Young and C. Byrne (2021). "Does the body mass index impact lymph 8. node yield for colorectal cancer resection, and does operative approach influence this: a review of bi-national colorectal cancer audit database." ANZ Journal of Surgery 91(12): 2707-2713.
- 9. Corrigan, N., H. Marshall, J. Croft, J. Copeland, D. Jayne and J. Brown (2018). "Exploring and adjusting for potential learning effects in ROLARR: a randomised controlled trial comparing robotic-assisted vs. standard laparoscopic surgery for rectal cancer resection." Trials 19(1): 339.
- 10. Debakey, Y., A. Zaghloul, A. Farag, A. Mahmoud and I. Elattar (2018). "Robotic-Assisted versus Conventional Laparoscopic Approach for Rectal Cancer Surgery, First Egyptian Academic Center Experience, RCT." Minimally Invasive Surgery 2018: 5836562.

{17}------------------------------------------------

- 11. Del Gutierrez Delgado. M. P., S. Mera Velasco. J. D. Turino Luque. I. Gonzalez Poveda. M. Ruiz Lopez and J. Santoyo Santoyo (2022). "Outcomes of robotic-assisted vs conventional laparoscopic surgery among patients undergoing resection for rectal cancer: an observational single hospital study of 300 cases." J Robot Surg 16(1): 179-187.
- 12. Ellebæk, M. B., S. K. Perdawood, S. Steenstrup, S. Khalaf, J. Kundal, S. Möller, J. C. Bang, J. Støvring and N. Qvist (2023). "Early versus late reversal of diverting loop ileostomy in rectal cancer surgery: a multicentre randomized controlled trial." Scientific Reports 13(1).
- 13. Emile, S. H., N. Horesh, M. R. Freund, Z. Garoufalia, R. Gefen, E. Silva-Alvarenga, G. Dasilva and S. D. Wexner (2022). "Outcomes of laparoscopic versus robotic-assisted resection of T4 rectal cancer: propensity score-matched analysis of a national cancer database." The British journal of surgery 110(2): 242-250.
- 14. Farah, E., A. A. Abreu, B. Rail, J. Salgado, G. Karagkounis, H. J. Zeh, 3rd and P. M. Polanco (2023). "Perioperative outcomes of robotic and laparoscopic surgery for colorectal cancer: a propensity score-matched analysis." World J Surg Oncol 21(1): 272.
- 15. Feng, Q., W. Tang, Z. Zhang, Y. Wei, L. Ren, W. Chang, D. Zhu, F. Liang, G. He and J. Xu (2022). "Robotic versus laparoscopic abdominoperineal resections for low rectal cancer: A single-center randomized controlled trial." Journal of Surgical Oncology 126(8): 1481-1493.
- 16. Feng, Q., W. Yuan, T. Li, B. Tang, B. Jia, Y. Zhou, W. Zhang, R. Zhao, C. Zhang, L. Cheng, X. Zhang, F. Liang, G. He, Y. Wei and J. Xu (2022). "Robotic versus laparoscopic surgery for middle and low rectal cancer (REAL): short-term outcomes of a multicentre randomised controlled trial." The lancet. Gastroenterology & hepatology 7(11): 991-1004.
- 17. Giesen, L. J. X., J. W. T. Dekker, M. Verseveld, R. M. P. H. Crolla, G. P. van der Schelling, C. Verhoef and P. B. Olthof (2023). "Implementation of robotic rectal cancer surgery: a cross-sectional nationwide study." Surgical Endoscopy 37(2): 912-920.
- 18. Gilmore, B., M. A. Adam, K. Rhodin, M. C. Turner, B. Ezekian, C. R. Mantyh and J. Migaly (2021). "Evolution of minimally invasive surgery for rectal cancer: update from the national cancer database." Surgical Endoscopy 35(1): 275-290.
- 19. Glencer, A. C., J. A. Lin, K. Trang, A. Greenberg, K. S. Kirkwood, M. A. Adam and A. Sarin (2022). "Assessing the role of robotic proctectomy in obese patients: a contemporary NSQIP analysis." J Robot Surg 16(6): 1391-1399.
- 20. Horsey, M. L., S. A. Parascandola, A. D. Sparks, S. Hota, M. Ng and V. Obias (2022). "The impact of surgical approach on short- and long-term outcomes after rectal cancer resection in elderly patients: a national cancer database propensity score matched comparison of robotic, laparoscopic, and open approaches." Surgical Endoscopy 36(2): 1269-1277.
- 21. Horvath, P., C. Steidle, C. Yurttas, I. Baur, A. Königsrainer and I. Königsrainer (2023). "Possible Advantages of Minimal-Invasive Approaches in Rectal Cancer Surgery: A Nationwide Analysis." Journal of Clinical Medicine 12(14).
- 22. Hu. K. Y., R. Wu, A. Szabo, T. J. Ridolfi, K. A. Ludwig and C. Y. Peterson (2020). "Laparoscopic Versus Robotic Proctectomy Outcomes: An ACS-NSQIP Analysis." J Surg Res 255: 495-501.

{18}------------------------------------------------

- 23. Hyde, L. Z., O. Baser, S. Mehendale, D. Guo, M. Shah and R. P. Kiran (2019). "Impact of surgical approach on short-term oncological outcomes and recovery following low anterior resection for rectal cancer." Colorectal disease : the official journal of the Association of Coloproctology of Great Britain and Ireland 21(8): 932-942.
- 24. Jayne, D., A. Pigazzi, H. Marshall, J. Croft, N. Corrigan, J. Copeland, P. Quirke, N. West, T. Rautio, N. Thomassen, H. Tilney, M. Gudgeon, P. P. Bianchi, R. Edlin, C. Hulme and J. Brown (2017). "Effect of Robotic-Assisted vs Conventional Laparoscopic Surgery on Risk of Conversion to Open Laparotomy Among Patients Undergoing Resection for Rectal Cancer: The ROLARR Randomized Clinical Trial." JAMA - Journal of the American Medical Association 318(16): 1569-1580.
- 25. Kethman, W. C., K. E. Bingmer, A. Ofshteyn, R. Charles, S. L. Stein, D. Dietz and E. Steinhagen (2022). "Effects of surgical approach on short- and long-term outcomes in earlystage rectal cancer: a multicenter, propensity score-weighted cohort study." Surg Endosc 36(8): 5833-5839.
- 26. Kim, J. C., S. B. Lim, Y. S. Yoon, I. J. Park, C. W. Kim and C. N. Kim (2014). "Completely abdominal intersphincteric resection for lower rectal cancer: feasibility and comparison of robot-assisted and open surgery." Surg Endosc 28(9): 2734-2744.
- 27. Kim, J. C., S. S. Yang, T. Y. Jang, J. Y. Kwak, M. J. Yun and S. B. Lim (2012). "Open versus robot-assisted sphincter-saving operations in rectal cancer patients: techniques and comparison of outcomes between groups of 100 matched patients." Int J Med Robot 8(4): 468-475.
- 28. Kim, J. Y., N. K. Kim, K. Y. Lee, H. Hur, B. S. Min and J. H. Kim (2012). "A comparative study of voiding and sexual function after total mesorectal excision with autonomic nerve preservation for rectal cancer: laparoscopic versus robotic surgery." Annals of Surgical Oncology 19(8): 2485-2493.
- 29. Kim, M. J., S. C. Park, J. W. Park, H. J. Chang, D. Y. Kim, B. H. Nam, D. K. Sohn and J. H. Oh (2018). "Robot-assisted Versus Laparoscopic Surgery for Rectal Cancer: A Phase II Open Label Prospective Randomized Controlled Trial." Ann Surg 267(2): 243-251.
- 30. Law, W. L. and D. C. C. Foo (2017). "Comparison of short-term and oncologic outcomes of robotic and laparoscopic resection for mid- and distal rectal cancer." Surg Endosc 31(7): 2798-2807.
- 31. Lei, Y., J. Jiang, S. Zhu, B. Yi and J. Li (2022). "Comparison of the short-term efficacy of two types of robotic total mesorectal excision for rectal cancer." Tech Coloproctol 26(1): 19-28.
- 32. Matsuyama, T., H. Endo, H. Yamamoto, I. Takemasa, K. Uehara, T. Hanai, H. Miyata, T. Kimura, H. Hasegawa, Y. Kakeji, M. Inomata, Y. Kitagawa and Y. Kinugasa (2021). "Outcomes of robot-assisted versus conventional laparoscopic low anterior resection in patients with rectal cancer: propensity-matched analysis of the National Clinical Database in Japan." BJS Open 5(5).

{19}------------------------------------------------

- 33. Mizoguchi, M., M. Kizuki, N. Iwata, M. Tokunaga, K. Fushimi, Y. Kinugasa and T. Fujiwara (2023). "Comparison of short-term outcomes between robot-assisted and laparoscopic rectal surgery for rectal cancer: A propensity score-matched analysis using the Japanese Nationwide diagnosis procedure combination database." Annals of Gastroenterological Surgery 7(6): 955-967.
- 34. Myrseth, E., L. S. Nymo, P. F. Gjessing, H. Korner, J. T. Kvaloy and S. Norderval (2022). "Lower conversion rate with robotic assisted rectal resections compared with conventional laparoscopy; a national cohort study." Surgical Endoscopy 36(5): 3574-3584.
- 35. Ose, I. and S. K.…

---

**Source:** [https://fda-staging.innolitics.com/device/K242318](https://fda-staging.innolitics.com/device/K242318)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).