K242308 · Zhejiang Zhonglei Medical Technology Co. , Ltd. · ITI · Nov 27, 2024 · Physical Medicine
Device Facts
Record ID
K242308
Device Name
Power wheelchair
Applicant
Zhejiang Zhonglei Medical Technology Co. , Ltd.
Product Code
ITI · Physical Medicine
Decision Date
Nov 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Motor-driven power wheelchair (Model N6102) for indoor/outdoor mobility. Input: user operation via joystick controller. Transformation: controller processes joystick input to drive dual DC brushless motors on rear wheels; system includes automatic electromagnetic braking. Output: wheelchair movement (forward, backward, turning). Used by disabled or elderly patients; operated by the user. Benefits: provides independent mobility for individuals with limited walking ability. Compact, foldable carbon fiber frame; detachable seat cushion; flip-back armrests. Max load 136kg.
Clinical Evidence
Bench testing only. Compliance with ISO 7176 series (stability, braking, dimensions, obstacle climbing, battery performance) and ISO 10993 (biocompatibility) demonstrated. No clinical data provided.
Technological Characteristics
Carbon fiber frame; solid PU tires; DC brushless motors (24V, 150W); lithium-ion battery (24V, 10Ah). Joystick-controlled dual-drive system. Automatic electromagnetic braking. Foldable design. Biocompatibility per ISO 10993-5, -10, -23. EMC per ISO 7176-21 and IEC 60601-1-2. Performance validated per ISO 7176 series.
Indications for Use
Indicated for disabled or elderly persons limited to a seated position requiring mobility assistance for indoor and outdoor use.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
November 27, 2024
Zhejiang Zhonglei Medical Technology Co. Ltd. Ariel Xiang Consultant Shanghai Sungo Management Consulting Co. Ltd. 14th Floor, 1500# Century Avenue Shanghai, Shanghai 200122 China
Re: K242308
Trade/Device Name: Power wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: October 30, 2024 Received: October 30, 2024
#### Dear Ariel Xiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page
2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Heather L. Dean -S
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K242308
Device Name
Power wheelchair
Indications for Use (Describe)
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(K) Summary
#### Document Prepared Date: 2024/10//30
#### A. Applicant
Name : Zhejiang Zhonglei Medical Technology Co.Ltd Address : 1st floor, 2nd floor, No. 66, Dayuan Street, Nanmingshan Street, Liandu District, Lishui City, Zhejiang Province, China. Contact Person: Chengwei Ying Tel: +86 13958430301 Mail:594506983@gq.com
Submission Correspondent: Primary contact: Ms. Ariel Xiang Title: Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email:shouqiu.xiang@sungoglobal.com
Secondary contact: Mr. Raymond Luo Title: Technical Director Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com
## B. Device
Trade Name: Power wheelchair Common Name: Powered wheelchair Models: N6102
Regulatory Information Classification Name: Powered Wheelchair Classification: Class II. Product code: ITI Regulation Number: 890.3860
{5}------------------------------------------------
Review Panel: Physical Medicine
#### C. Predicate device:
510Knumber: K230964 Device Name:Power Wheelchair Manufacturer:Zhejiang Innuovo Rehabilitation Devices Co.,Ltd
#### Indications for use of the device D.
The Power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
### E. Device Description
This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design. The whole wheelchair can be folded by one button and it can be easily carried or rolled after folding. The seat cushion is detachable. The armrest can be flipped backward, which is convenient for the elderly to move. Users can drive the wheelchair by themselves through the control device.
The wheelchair uses lithium batteries as its power source. The controls the drive left/right motor to realize the wheelchair forward, backward and turn functions.
The frame of the device is carbon fiber. The front wheels are driven wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels. The rear wheels are driving wheels to control the speed and direction. The wheels are Solid PU tires. When in use, the operator drives the motor of the rear wheel by operating the joystick to achieve the rear wheels movement.
The DC brushless motor and brake system are fixed on the rear wheels.
The max loading of the device is 136KG. Only for one person sit.
## F. Comparison with predicate Device
| Table 1 General Comparison |
|----------------------------|
|----------------------------|
| Elements of Comparison | Subject Device | Predicate Device<br>(K230964) | Remark |
|------------------------|-------------------|-------------------------------|--------|
| Manufacturer | Zhejiang Zhonglei | Zhejiang Innuovo | -- |
{6}------------------------------------------------
| | Medical<br>Technology Co.Ltd | Rehabilitation<br>Devices Co.,Ltd | | Driving system | Direct drive on the<br>rear<br>wheels | Direct drive on the<br>rear<br>wheels | S.E |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|----------------------------------------------|----------------------------------------------|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual name | Power Wheelchair | Power Wheelchair | -- | Brake system | Automatic<br>electromagnetic<br>brake system | Automatic<br>electromagnetic<br>brake system | S.E |
| Model(s) | N6102 | N5909 | -- | Braking distance | 780mm | ≤1.5 m | Minor difference<br>on braking<br>distance will not<br>cause different<br>performance.<br>Shorter distance<br>for braking will be<br>more safety. |
| Classification | Class II | Class II | Same | Maximum safe<br>operationalincline<br>degree | 9 ° | 9 ° | S.E |
| Classification regulation | 21 CFR890.3860 | 21 CFR890.3860 | Same | Armrest | Carbon fibers | T700 | Different material<br>used for parts in<br>contact with user,<br>which such<br>differences will not<br>impact the safety<br>and effectiveness<br>of the subject<br>device as<br>biocompatibility<br>tests are carried<br>out according to<br>ISO 10993 series. |
| Product code | ITI | ITI | Same | Main frame material | Carbon fiber<br>material | Carbon fiber<br>material | S.E |
| Indications for use | It is a motor<br>driven, indoor and<br>outdoor<br>transportation<br>vehicle with the<br>intended use to<br>provide mobility to<br>a disabled or<br>elderly person<br>limited to a seated<br>position. | It is a motor<br>driven, indoor and<br>outdoor<br>transportation<br>vehicle with the<br>intended use to<br>provide mobility to<br>a disabled or<br>elderly person<br>limited to a seated<br>position. | S.E. | Back cushion | Polyester fiber | Polyester fabric | S.E |
| Intended user | disabled or elderly<br>person limited to a<br>seated position. | disabled or elderly<br>person limited to a<br>seated position. | S.E. | Seat cushion | Polyester fiber | Polyester fiber | S.E |
| Use condition | indoor and outdoor<br>use | indoor and outdoor<br>use | S.E | Overall Dimension<br>(length*width*height) | 950*610*960mm | 900*580*860mm | Minor difference<br>on wheelchair<br>dimension will not |
| Number of wheels | 4,including two<br>front wheels and<br>two rear Wheels | 4,including two<br>front wheels and<br>two rear Wheels | S.E | | | | |
| Function of wheels | Front<br>wheels:driv<br>en wheels<br>suitable for<br>rotation,<br>acceleration,<br>retrograde Rear<br>wheels: driving<br>wheels to control<br>the speed and<br>direction | Front<br>wheels:driv<br>en wheels<br>suitable for<br>rotation,<br>acceleration,<br>retrograde Rear<br>wheels: driving<br>wheels to control<br>the speed and<br>direction | S.E | | | | |
| Movement control | By Joystick control | By Joystick control | S.E | | | | |
{7}------------------------------------------------
{8}------------------------------------------------
| (length*width*height) | | | cause different<br>performance. All<br>safety and<br>performance have<br>been validated<br>with the maximum<br>rated weight<br>dummy.<br>Analysis |
|---------------------------------------------------------------|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Front wheel size/type | 7" x1.78"/ PU<br>Solid tire<br>(180mm<br>x45mm) | 7" x 1.75"/PU Solid<br>tire | Minor difference<br>on dimension of<br>wheels will not<br>cause different<br>performance. |
| Rear wheel size/type | 9" x 2"/PU Solid<br>tire<br>(230mm<br>x50mm) | 8.5" x 1.8"/ PU<br>Solid tire | Minor difference<br>on dimension of<br>wheels will not<br>cause different<br>performance. |
| Max speed forward | 1.54 m/s (5.5km/h) | Up to 6 km/h (1.6<br>m/s), adjustable | Minor difference<br>on Max speed<br>forward of wheels<br>will not cause<br>different<br>performance. |
| Max loading weight | 136kg (≈300lbs) | 110kg (≈250lbs) | Difference on<br>loading weight will<br>not cause different<br>performance. All<br>safety and<br>performance have<br>been validated<br>with the maximum<br>loading weight. |
| Maximum distance of<br>travel on the fully charged<br>battery | 16.1km | 10 km | It is caused by the<br>size of the wheel,<br>will not cause<br>different<br>performance, The<br>further away the<br>better |
| Battery | | | |
| | li-ion battery pack<br>24V,10Ah, | li-ion battery pack<br>rechargeable, 24 VDC 10Ah | S.E |
| Battery charger | Off-board charger<br>Input: 100-240V,<br>50/60Hz, 1.5A,<br>Output: 24 Vdc, 2A; | Off-board charger<br>Input: 100-240V,<br>50/60Hz, 1.5A,<br>Output: 24 Vdc, 2A; | S.E |
| Motor | Brushless Eco-Drive,<br>24V,150W, 2pcs | Brushless DC motor; 24VDC; 150W; 2pcs | S.E |
| Electronic controller | Brushless dual-drive rocker controller | Brushless dual-drive rocker controller | S.E |
| Turning Radius | 900mm | 900 mm | S.E |
| Maximum obstacle climbing | 25mm | 40 mm | Minor difference on obstacle climbing will not cause new safety and effectiveness concerns. |
{9}------------------------------------------------
# Table 2 safety comparison
| Item | Proposed Device | Predicate Device | Results |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Biocompatibility | All user directly contacting<br>materials are compliance<br>with<br>ISO10993-5 ,ISO10993-10<br>,ISO10993-23 requirements. | All user directly contacting<br>materials are compliance with<br>ISO10993-5 and ISO10993-10<br>requirements. | S.E. |
| EMC | ISO7176-21 and IEC 60601-<br>1-2 | ISO7176-21 and IEC 60601-1-<br>2 | S.E. |
| Performance | ISO7176 series | ISO7176 series | S.E. |
| Label and<br>labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | S.E. |
| Item | Proposed Device | Predicate Device | Results |
| ISO7176-1 | The Static stability has been<br>determined after the testing<br>according to the ISO 7176-1,<br>and test results meet its<br>design specification. | The Static stability has been<br>determined after the testing<br>according to the ISO 7176-1,<br>and test results meet its<br>design specification. | S.E. |
| ISO7176-2 | The dynamic stability has<br>been determined after the<br>testing according to the ISO<br>7176-2, and test results meet<br>its design specification. | The dynamic stability has<br>been determined<br>after the testing according to<br>the ISO 7176-2, and test<br>results meet its design<br>specification. | S.E. |
| ISO7176-3 | The effectiveness of brakes<br>has been determined after the<br>testing according to the ISO<br>7176-3, and test results meet<br>its design specification. | The effectiveness of brakes<br>has been<br>determined after the testing<br>according to the ISO 7176-3,<br>and test results meet its<br>design specification. | S.E. |
| ISO7176-4 | The theoretical distance range<br>has been determined after<br>the testing according to the<br>ISO 7176-4, and test results<br>meet its design specification. | The theoretical distance range<br>has been determined after<br>the testing according to the<br>ISO 7176-4, and test results<br>meet its design specification. | S.E. |
| ISO7176-5 | The dimensions, mass has<br>been determined after the<br>testing according to the ISO<br>7176- 5. | The dimensions, mass has<br>been determined after the<br>testing according to the ISO<br>7176- 5. | S.E. |
| ISO7176-6 | The dimensions, mass has<br>been determined after the<br>testing according to the ISO<br>7176- 5. | The dimensions, mass has<br>been determined after the<br>testing according to the ISO<br>7176- 5. | S.E. |
| ISO7176-7 | The seating and wheel<br>dimensions has been<br>determined after the testing<br>according to the ISO 7176-7. | The seating and wheel<br>dimensions has been<br>determined after the testing<br>according to the ISO 7176-7. | S.E. |
| ISO7176-8 | All test results meet the<br>requirements in Clause 4 of<br>ISO 7176-8. | All test results meet the<br>requirements in Clause 4 of<br>ISO 7176-8. | S.E. |
| ISO7176-9 | The test results shown that<br>the device under tests could<br>continue to function according<br>to manufacturer's specification<br>after being subjected to each<br>of the testsspecified in Clause<br>8 of ISO 7176-9. | The test results shown that the<br>device under tests could<br>continue to function according<br>to manufacturer's specification<br>after being subjected to each<br>of the tests specified in Clause<br>8 of ISO 7176-9. | S.E. |
| ISO7176-10 | The obstacle-climbing ability<br>of device has been<br>determined after the testing<br>according to the ISO 7176-10. | The obstacle-climbing ability<br>of device has been determined<br>after the testing according to<br>the ISO 7176-10. | S.E. |
| ISO7176-11 | The test dummies used in the<br>testing of ISO 7176 series are<br>meet the requirements of ISO<br>7176-11. | The test dummies used in the<br>testing of ISO 7176 series are<br>meet the requirements of ISO<br>7176-11. | S.E. |
| ISO7176-13 | The coefficient of friction of<br>test surfaces has been<br>determined, which could be<br>used in other 7176 series tests<br>involved. | The coefficient of friction of<br>test surfaces has been<br>determined, which could be<br>used in other 7176 series tests<br>involved. | S.E. |
| ISO7176-14 | All test results meet the<br>requirements in Clause 7, 8, 9,<br>10, 11, 12, 13, 14, 15, 17 of<br>ISO 7176-14. | All test results meet the<br>requirements in Clause 7, 8, 9,<br>10, 11, 12, 13, 14, 15, 17 of<br>ISO 7176-14. | S.E. |
| ISO7176-15 | The test results shown that<br>information | The test results shown that<br>information | S.E. |
| | disclosure, documentation<br>and labelling of device meet<br>the requirements of ISO7176-<br>15. | disclosure, documentation and<br>labelling of device meet the<br>requirements of ISO<br>7176-15. | |
| ISO16840-<br>10:2021 | The performance of resistance<br>to<br>ignition meet the requirements | The performance of resistance<br>to<br>ignition meet the requirements | S.E. |
| | of ISO 16840-10:2021. | of ISO 16840-10:2021. | |
| ISO 7176-<br>21/IEC60601-<br>1-2 | The EMC performance results<br>meet the requirements of ISO<br>IEC60601-1-2. | The EMC performance results<br>meet the requirements of ISO<br>7176-21&IEC60601-1-2. | S.E. |
| ISO7176-25 | The performance of batteries<br>and charger of device meet<br>the Requirements in Clause 5<br>and 6 of ISO 7176-25. | The performance of batteries<br>and charger of device meet<br>the Requirements in Clause<br>5and 6 of ISO 7176-25. | S.E. |
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
# G. Substantial Equivalence Discussion
The subject device complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2001, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2008, ISO7176-13:1989, ISO 7176-14:2008, ISO 7176-15:1996, ISO 16840-10: 2021, ISO7176-21:2009, ISO 7176-22:2014, ISO 7176-25:2013, IEC 60601-1-2: 2014.
The intended uses for both devices are the same. Mainframes of two devices are folded by way of front and rear close.
The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14:2008. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3:2012.
Turning radius and Maximum obstacle climbing are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2:2001, ISO 7176-10:2008.
The flame retardant test of the seat cushion/backrest of the subject device is conducted according to ISO 16840-10. Therefore, both devices are assured to be under the same safety level.
The all rest different between the subject device and predicate device can proved the safety or effectiveness by the ISO 7176 series support reports.
Different material used for parts in contact with user, which such differences will not impact the safety and effectiveness of the subject device as biocompatibility tests are carried out according to ISO 10993 series.
In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
{13}------------------------------------------------
# H. Substantially Equivalency Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices, K230964 Power Wheelchair from Zhejiang Innuovo Rehabilitation Devices Co., Ltd.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.