The VersaTap™ Suture Anchor is intended to be used for soft tissue fixation during general orthopedic surgery. The VersaTap™ Suture Anchor is intended for use in arthroscopic or open surgical approaches for fixation of soft tissue and ligaments to bone/healthy tissue during tendon and ligament repairs, during procedures such as: Shoulder: Rotator cuff repair, biceps tenodesis, SLAP repair, Bankart repair Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Elbow: Tennis elbow repair Knee: Medial and lateral collateral ligament repair. Wrist: Scapholunate ligament reconstruction. The VersaPEEK™ Suture Anchor is intended to be used for soft tissue fixation during general orthopedic surgery. The VersaPEEK™ Suture Anchor is intended for use in arthroscopic or open surgical approaches for fixation of soft tissue and ligaments to bone/healthy tissue during tendon and ligament repairs, during procedures such as: Shoulder: Rotator cuff repair, biceps tenodesis, SLAP repair, Bankart repair Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Elbow: Tennis elbow repair Knee: Medial and lateral collateral ligament repair. Wrist: Scapholunate ligament reconstruction. The MicroTi™ Suture Anchor is intended to be used for soft tissue fixation during general orthopedic surgery. The MicroTi™ Suture Anchor is intended for use in arthroscopic or open surgical approaches for fixation of soft tissue and ligaments to bone/healthy tissue during tendon and ligament repairs, during procedures such as: Elbow: Ulnar/Medial Collateral Ligament Repair Foot/Ankle: Achilles Tendon Repair, Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Mid-foot Reconstruction, Hallux Valgus Reconstruction, Metatarsal Ligament Repair. Hand/Wrist: Scapholunate Ligament Reconstruction and Ulnar/ Radial Collateral Ligament Reconstruction. The VersaLat™ Ti Suture Anchor is intended to use for fixation of soft tissue and ligaments to bone/ healthy tissue during tendon and ligament repairs, during procedures such as: Shoulder: Rotator Cuff Repair, Biceps Tenodesis Elbow: Biceps Tendon Reattachment, Ulnar/Radial Collateral Ligament Reconstruction. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, and Iliotibial Band Tenodesis Foot/Ankle: Lateral stabilization, medial stabilization, Achilles tendon repair, mid-foot reconstruction, hallux valgus reconstruction, metatarsal ligament repair, metatarsal tendon repair, bunionectomy, and digital tendon transfers. The VersaLat™ Suture Anchor is intended to use for fixation of soft tissue and ligaments to bone/ healthy tissue during tendon and ligament repairs, during procedures such as: Shoulder: Rotator Cuff Repair, Biceps Tenodesis Elbow: Biceps Tendon Reattachment, Ulnar/Radial Collateral Ligament Reconstruction. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, and Illiotibial Band Tenodesis. Foot/Ankle: Lateral stabilization, medial stabilization, Achilles tendon repair, mid-foot reconstruction, hallux valgus reconstruction, metatarsal ligament repair, metatarsal tendon repair, bunionectomy, and digital tendon transfers.
Device Story
Suture anchors for soft tissue-to-bone fixation in orthopedic surgery; used in arthroscopic or open procedures. Devices include self-tapping screw-in anchors (VersaTap, MicroTi) and knotless suture anchors (VersaLat). Materials include PEEK and Titanium; designs vary by diameter and length to accommodate specific joint sizes. Operated by orthopedic surgeons. Anchors provide mechanical fixation of soft tissue/ligaments to bone; facilitate tendon/ligament repair. Benefits include secure fixation, reduced post-operative imaging artifacts (PEEK models), and flexibility in suture/tape management (knotless models).
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing included insertion torque, static pullout, and justification for dynamic pullout (fatigue), component interconnection, and corrosion testing per FDA guidance.
Technological Characteristics
Materials: PEEK and Titanium. Design: Threaded bone fixation fasteners; self-tapping or knotless configurations. Dimensions: Varying diameters and lengths. Sterilization: Sterile. Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring soft tissue-to-bone fixation during general orthopedic surgery, including arthroscopic or open procedures for tendon and ligament repairs in the shoulder, elbow, knee, wrist, and foot/ankle.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
August 29, 2024
Ortho-Design (Pty) Ltd Dian Peach Managing Director 17 Dely Road Hazelwood, Pretoria 0081 South Africa
Re: K242296
VersaTap™ Suture Anchor (ADP050); VersaPEEK™ Suture Anchor (APP002); Trade/Device Name: MicroTi™ Suture Anchor (ATP003-M); VersaLat™ Ti Suture Anchor (ADP021-TI); VersaLatTM Suture Anchor (ADP021-PK) Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: May 3, 2024
Dear Dian Peach:
Received: August 2, 2024
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combinationproducts/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-reportmedical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuingeducation/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumereducation-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Christopher Ferreira -S
Christopher Ferreira, MS Assistant Director, Biomedical Engineer DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K242296
Device Name
VersaTap TM Suture Anchor (ADP050); VersaPEEK™ Suture Anchor (APP002); MicroTi™ Suture Anchor (ATP003-M); VersaLat™ Ti Suture Anchor (ADP021-TI); VersaLat™ Suture Anchor (ADP021-PK)
Indications for Use (Describe)
#### VersaTap™ Suture Anchor
The VersaTap™ Suture Anchor is intended to be used for soft tissue fixation during general orthopedic surgery. The VersaTap™ Suture Anchor is intended for use in arthroscopic or open surgical approaches for fixation of soft tissue and ligaments to bone/healthy tissue during tendon and ligament repairs, during procedures such as:
- · Shoulder: Rotator cuff repair, biceps tenodesis, SLAP repair, Bankart repair
- · Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus
- · Elbow: Tennis elbow repair
- · Knee: Medial and lateral collateral ligament repair.
- · Wrist: Scapholunate ligament reconstruction
#### VersaPEEK™ Suture Anchor
The VersaPEEK™ Suture Anchor is intended to be used for soft tissue fixation during general orthopedic surgery. The VersaPEEK™ Suture Anchor is intended for use in arthroscopic or open surgical approaches for fixation of soft tissue and ligaments to bone/healthy tissue during tendon and ligament repairs, during procedures such as:
- · Shoulder: Rotator cuff repair, biceps tenodesis, SLAP repair, Bankart repair
- · Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus
- · Elbow: Tennis elbow repair
- · Knee: Medial and lateral collateral ligament repair.
- · Wrist: Scapholunate ligament reconstruction
#### MicroTi™ Suture Anchor
The MicroTi™ Suture Anchor is intended to be used for soft tissue fixation during general orthopedic surgery. The MicroTi™ Suture Anchor is intended for use in arthroscopic or open surgical approaches for fixation of soft tissue and ligaments to bone/healthy tissue during tendon and ligament repairs, during procedures such as:
- · Elbow: Ulnar/Medial Collateral Ligament Repair
· Foot/Ankle: Achilles Tendon Repair, Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Mid-foot Reconstruction, Hallux Valgus Reconstruction, Metatarsal Ligament Repair.
· Hand/Wrist: Scapholunate Ligament Reconstruction and Ulnar/ Radial Collateral Ligament Reconstruction.
{3}------------------------------------------------
VersaLat™ Ti Suture Anchor
The VersaLat™ Ti Suture Anchor is intended to use for fixation of soft tissue and ligaments to bone/ healthy tissue during tendon and ligament repairs, during procedures such as:
· Shoulder: Rotator Cuff Repair, Biceps Tenodesis
· Elbow: Biceps Tendon Reattachment, Ulnar/Radial Collateral Ligament Reconstruction.
· Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, and Iliotibial Band Tenodesis
· Foot/Ankle: Lateral stabilization, medial stabilization, Achilles tendon repair, mid-foot reconstruction, hallux valgus reconstruction, metatarsal ligament repair, metatarsal tendon repair, bunionectomy, and digital tendon transfers.
VersaLat™ Suture Anchor
The VersaLat™ Suture Anchor is intended to use for fixation of soft tissue and ligaments to bone/ healthy tissue during tendon and ligament repairs, during procedures such as:
- · Shoulder: Rotator Cuff Repair, Biceps Tenodesis
- · Elbow: Biceps Tendon Reattachment, Ulnar/Radial Collateral Ligament Reconstruction.
· Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, and Illiotibial Band Tenodesis.
· Foot/Ankle: Lateral stabilization, medial stabilization, Achilles tendon repair, mid-foot reconstruction, hallux valgus reconstruction, metatarsal ligament repair, metatarsal tendon repair, bunionectomy, and digital tendon transfers.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
ver-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
The following information is provided in accordance with 21CFR 807.92 for the Premarket 510(k) Summary:
| Submitter Information: | Ortho-Design (Pty) Ltd<br>17 Dely Road<br>Hazelwood, Pretoria, 0081 South<br>Africa |
|-----------------------------------------|-------------------------------------------------------------------------------------|
| Date: | 26 August 2024 |
| Contact Person:<br>Telephone:<br>Email: | Dian Peach, Managing Director<br>(+27) 12 807 1902<br>dianpeach@ortho-design.co.za |
Table 1: Subject devices and their respective submission K212381 predicate devices.
| K212381 | The enclosed submission |
|-------------------------|---------------------------------------|
| VersaTap™ Suture Anchor | VersaTap™ Suture Anchor (Device A) |
| | VersaPEEK™ Suture Anchor (Device B) |
| MiniTi™ Suture Anchor | MicroTi™ Suture Anchor (Device C) |
| VersaLat™ Suture Anchor | VersaLat™ Ti Suture Anchor (Device D) |
| | VersaLat™ Suture Anchor (Device E) |
Table 2: DEVICE A – VersaTap™ Suture Anchor
| | Subject Device | Predicate Device<br>(Cleared in K212381) |
|---------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | VersaTap™ Suture Anchor | VersaTap™ Suture Anchor<br>(Cleared in K212381) |
| Regulation Description: | Smooth or threaded metallic<br>bone fixation fastener. | Smooth or threaded metallic<br>bone fixation fastener. |
| Classification Name: | Fastener, Fixation,<br>Nondegradable, Soft Tissue | Fastener, Fixation,<br>Nondegradable, Soft Tissue |
| Regulation Number: | 21 CFR 888.3040 | 21 CFR 888.3040 |
| Device Classification: | Class II | Class II |
| Product Code: | MBI | MBI |
| Reference Device(s): | KINSA RC Suture Anchor<br>(K070908), manufactured by<br>Smith & Nephew | K061665; K173788;<br>K063453, K130274, K070758 |
| Device Description: | The VersaTap™ Suture<br>Anchor is a self-tapping suture<br>anchor. This anchor is<br>designed to combine the<br>advantages of both PEEK and<br>Titanium. The titanium tip<br>makes the product self-tapping,<br>whilst the majority of the<br>anchor is manufactured from<br>PEEK to minimize post-<br>operative imaging effects. | The VersaTap™ Suture<br>Anchor is a self-tapping<br>suture anchor mostly used as<br>a medial row anchor in<br>rotator cuff repair surgery.<br>This anchor is ingeniously<br>designed to combine the<br>advantage of both PEEK and<br>Titanium. The titanium tip<br>allows for self-tapping<br>ability whilst the majority of<br>the anchor is manufactured<br>from PEEK which minimizes<br>post-operative imaging<br>effects. |
| Indications for Use: | The VersaTap™ Suture Anchor<br>is intended to be used for soft<br>tissue fixation during general<br>orthopedic surgery. The<br>VersaTap™ Suture Anchor is<br>intended for use in arthroscopic<br>or open surgical approaches for<br>fixation of soft tissue and<br>ligaments to bone/healthy<br>tissue during tendon and<br>ligament repairs, during<br>procedures such as:<br>• Shoulder: Rotator cuff<br>repair, biceps tenodesis,<br>SLAP repair, Bankart<br>repair<br>• Foot/Ankle: Lateral<br>Stabilization, Medial<br>Stabilization, Achilles<br>Tendon Repair, Hallux<br>Valgus<br>• Elbow: Tennis elbow repair<br>• Knee: Medial and lateral<br>collateral ligament repair.<br>• Wrist: Scapholunate<br>ligament reconstruction. | The VersaTap™ Suture<br>Anchor is intended to be<br>used for soft tissue fixation<br>during general orthopedic<br>surgery. The VersaTap™<br>Suture Anchor is intended for<br>use in arthroscopic or open<br>surgical approaches for<br>fixation of soft tissue and<br>ligaments to bone/healthy<br>tissue during tendon and<br>ligament repairs, during<br>procedures such as:<br>• Shoulder: Rotator cuff<br>repair, biceps tenodesis,<br>SLAP repair, Bankart<br>repair<br>• Foot/Ankle: Lateral<br>Stabilization, Medial<br>Stabilization, Achilles<br>Tendon Repair, Hallux<br>Valgus<br>• Elbow: Tennis elbow<br>repair<br>• Knee: Medial and lateral<br>collateral ligament repair.<br>• Wrist: Scapholunate<br>ligament reconstruction |
| Relationship to the<br>Device: | The subject and predicate devices (cleared in K212381)<br>have the same intended use, materials,<br>packaging, and similar design features. | function. |
| Technological<br>Differences: | The subject and predicate devices (cleared in K212381) have similar engineering designs with minor dimensional changes. The subject device has a reduced diameter and a reduced length compared to the predicate device (cleared in K212381). | |
| Clinical Characteristics: | The subject and predicate devices (cleared in K212381) are intended to be used for soft tissue fixation during general orthopedic surgery. | |
| Technical Equivalence: | The subject and predicate devices (cleared in K212381) have similar designs, development methods, principles of operations and performance. | |
| Mechanical Testing and<br>Disclosure of<br>Performance and Safety<br>Deviation: | Substantial equivalence is supported by the results of mechanical testing including insertion torque and static pullout testing, and by justification for dynamic pullout (fatigue), component interconnection, and corrosion testing according to the FDA Guidance: 'Bone Anchors - Premarket Notification (510(k)) Submissions'. Ortho-Design's suture anchors support safety and effectiveness and can be considered substantially equivalent to the predicate device (cleared in K212381). | |
| Biocompatibility: | The subject and predicate devices (cleared in K212381) are manufactured using the same approved suppliers, manufacturing facilities, manufacturing processes, chemicals, materials, and cleaning processes. All manufacturing is conducted according to the processes defined by ISO 13485 and ISO 9001. All materials, processes, and cleaning agents used for the subject device have been used in the previously cleared device(s). Therefore biocompatibility of the subject device is substantially equivalent to the predicate device.<br><br>The subject devices have been tested to be non-pyrogenic. | |
| Quality Control<br>Measures | For both the subject and predicate device (cleared in K212381)<br>the methods, thresholds and criteria of the following are<br>identical:<br>Sterility Self-Life Packaging Pyrogenicity / Endotoxin Testing | |
| Conclusion: | The subject device demonstrates substantial equivalence by<br>non-clinical testing consisting of mechanical, biocompatibility,<br>sterility, shelf-life, packaging and pyrogenicity tests and<br>adheres to the minimum requirements set out in FDA's<br>guidance document for Bone Anchors – Premarket<br>Notification (510(k)) Submissions. | |
| | Subject Device | Predicate Device<br>(Cleared in K212381) |
| Device Trade Name: | VersaPEEK™ Suture Anchor | VersaTap™ Suture Anchor<br>(Cleared in K212381) |
| Regulation Description: | Smooth or threaded metallic<br>bone fixation fastener. | Smooth or threaded metallic<br>bone fixation fastener. |
| Classification Name: | Fastener, Fixation,<br>Nondegradable, Soft Tissue | Fastener, Fixation,<br>Nondegradable, Soft Tissue |
| Regulation Number: | 21 CFR 888.3040 | 21 CFR 888.3040 |
| Device Classification: | Class II | Class II |
| Product Code: | MBI | MBI |
| Reference Device(s): | KINSA RC Suture Anchor<br>(K070908), manufactured by<br>Smith & Nephew | K061665; K173788;<br>K063453, K130274, K070758 |
| Device Description: | The VersaPEEK™ Suture<br>Anchor is manufactured<br>completely from PEEK. This<br>material composition<br>minimizes post-operative<br>imaging effects. The suture<br>anchor is designed for ultimate<br>mechanical properties (pullout<br>strength, tensile strength, etc).<br>The VersaPEEK™ Suture<br>Anchor also has a variety of<br>suture/suture tape<br>configurations. | The VersaTap™ Suture<br>Anchor is a self-tapping<br>suture anchor mostly used as<br>a medial row anchor in<br>rotator cuff repair surgery.<br>This anchor is ingeniously<br>designed to combine the<br>advantage of both PEEK and<br>Titanium. The titanium tip<br>allows for self-tapping<br>ability whilst the majority of<br>the anchor is manufactured<br>from PEEK which minimizes<br>post-operative imaging<br>effects. |
| Indications for Use: | The VersaPEEK™ Suture<br>Anchor is intended to be used<br>for soft tissue fixation during<br>general orthopedic surgery.<br>The VersaPEEK™ Suture<br>Anchor is intended for use in<br>arthroscopic or open surgical<br>approaches for fixation of soft<br>tissue and ligaments to<br>bone/healthy tissue during<br>tendon and ligament repairs,<br>during procedures such as:<br>• Shoulder: Rotator cuff<br>repair, biceps tenodesis,<br>SLAP repair, Bankart | The VersaTap™ Suture<br>Anchor is intended to be<br>used for soft tissue fixation<br>during general orthopedic<br>surgery. The VersaTap™<br>Suture Anchor is intended for<br>use in arthroscopic or open<br>surgical approaches for<br>fixation of soft tissue and<br>ligaments to bone/healthy<br>tissue during tendon and<br>ligament repairs, during<br>procedures such as:<br>• Shoulder: Rotator cuff<br>repair, biceps tenodesis, |
| | Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Elbow: Tennis elbow repair. Knee: Medial and lateral collateral ligament repair. Wrist: Scapholunate ligament reconstruction. *The above-mentioned indications for use are identical to those of the predicate device, VersaTap™ Suture Anchor, cleared in submission K212381. | SLAP repair, Bankart repair Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Elbow: Tennis elbow repair Knee: Medial and lateral collateral ligament repair. Wrist: Scapholunate ligament reconstruction |
| Relationship to the<br>Device: | The subject and predicate devices (cleared in K212381) have the same intended use, materials, function, packaging, and similar design features. | |
| Technological<br>Differences: | The subject and predicate devices (cleared in K212381) have similar engineering designs and minor dimensional changes. The subject device has a reduced diameter compared to the predicate device (cleared in K212381). The subject device is also composed fully composed of PEEK whereas the predicate device (cleared in K212381) is composed of a PEEK body and a titanium tip. | |
| Clinical Characteristics: | The subject and predicate device (cleared in K212381) are intended to be used for soft tissue fixation during general orthopedic surgery. | |
| Technical Equivalence: | The subject and predicate devices (cleared in K212381) have similar designs, development methods, principles of operations and performance. | |
| Mechanical Testing and<br>Disclosure of<br>Performance and Safety<br>Deviation: | Substantial equivalence is supported by the results of mechanical testing including insertion torque and static pullout testing, and by justification for dynamic pullout (fatigue), component interconnection, and corrosion testing according to the FDA Guidance: 'Bone Anchors - Premarket Notification (510(k)) Submissions'. Ortho-Design's suture anchors support safety and effectiveness and can be considered substantially equivalent to the predicate device (cleared in K212381). | |
| Biocompatibility: | The subject and predicate devices (cleared in K212381) are manufactured using the same approved suppliers, manufacturing facilities, manufacturing processes, chemicals, materials, and cleaning processes. All manufacturing is conducted according to the processes defined by ISO 13485 and ISO 9001. All materials, processes, and cleaning agents used for the subject device have been used in the previously cleared device(s). Therefore biocompatibility of the subject device is substantially equivalent to the predicate device.<br>The subject devices have been tested to be non-pyrogenic. | |
| Quality Control<br>Measures | For both the subject and predicate device (cleared in K212381) the methods, thresholds and criteria of the following are identical:<br>• Sterility<br>• Self-Life<br>• Packaging<br>• Pyrogenicity / Endotoxin Testing | |
| Conclusion: | The subject device demonstrates substantial equivalence by non-clinical testing consisting of mechanical, biocompatibility, sterility, shelf-life, packaging and pyrogenicity tests and adheres to the minimum requirements set out in FDA's guidance document for Bone Anchors – Premarket Notification (510(k)) Submissions. | |
| | Subject Device | Predicate Device<br>(Cleared in K212381) |
| Device Trade Name: | MicroTiTM Suture Anchor | MiniTiTM Suture Anchor<br>(Cleared in K212381) |
| Regulation Description: | Smooth or threaded metallic<br>bone fixation fastener. | Smooth or threaded metallic<br>bone fixation fastener. |
| Classification Name: | Fastener, Fixation,<br>Nondegradable, Soft Tissue | Fastener, Fixation,<br>Nondegradable, Soft Tissue |
| Regulation Number: | 21 CFR 888.3040 | 21 CFR 888.3040 |
| Device Classification: | Class II | Class II |
| Product Code: | MBI…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.