K242288 · Medical Brands Laboratories B.V. · GEH · Jan 21, 2025 · General, Plastic Surgery
Device Facts
Record ID
K242288
Device Name
Advanced Cryo Wart Remover
Applicant
Medical Brands Laboratories B.V.
Product Code
GEH · General, Plastic Surgery
Decision Date
Jan 21, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Advanced Cryo Wart Remover is designed as an over-the-counter (OTC) self-care device indicated for adults and children from the age of four years and older for the treatment of common and plantar warts.
Device Story
Advanced Cryo Wart Remover is an OTC cryotherapy device for home use by laypersons. It consists of a 15 mL pressurized aerosol canister containing a freezing agent (1,1-Difluoroethane, Propane, and Isopentane) and a metal applicator tip. To operate, the user holds the device downward and presses side buttons five times to charge the tip. The cold tip is applied to common warts on hands, tops of feet, and toes for 20 seconds, or to plantar warts on calloused skin for 40 seconds. The device provides localized cryosurgery to remove warts. It is non-sterile and reusable. The device benefits patients by providing a safe, intuitive, and effective self-care option for wart removal without requiring professional clinical intervention.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing and human factors studies. Bench testing included biocompatibility (ISO 10993-5, 10, 17, 18), shelf life (ASTM D3090-72), transportation (ASTM D4169), and comparative thermal performance. Human factors studies (n=10 for usability, n=20 for self-assessment) confirmed safe operation, accurate wart identification, and labeling comprehension (Flesch Reading Ease 61.22).
Technological Characteristics
Cryosurgical device; 15 mL pressurized aerosol canister; freezing agent: 1,1-Difluoroethane, Propane, Isopentane; metal applicator tip; non-sterile; reusable; manual activation via side buttons; shelf life 36 months; biocompatibility per ISO 10993; thermal performance testing.
Indications for Use
Indicated for adults and children aged 4+ for the treatment of common and plantar warts.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration" in a stacked format.
January 21, 2025
Medical Brands Laboratories B.V. Pieternella Bouter Chief Commercial Officer (CCO) Weesperstraat 114 Diemen, 1112AP Netherlands
Re: K242288
Trade/Device Name: Advanced Cryo Wart Remover Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: August 2, 2024 Received: December 20, 2024
Dear Pieternella Bouter:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long H. Chen--C Digitally signed by Long H.Chen-S
Date: 2025.01.21 11:14:35 -05'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242288
Device Name
## Advanced Cryo Wart Remover
Indications for Use (Describe)
The Advanced Cryo Wart Remover is designed as an over-the-counter (OTC) self-care device indicated for adults and children from the age of four years and older for the treatment of common and plantar warts.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for medicalbrands. The logo consists of the word "medicalbrands" in a sans-serif font, with a shield-shaped icon above it. The shield has a grid pattern inside and a "TM" symbol in the upper right corner.
# 807.92(a)(1)
| 1. | Submission Sponsor: | Medical Brands Laboratories B.V.<br>Weesperstraat 114,<br>1112 AP, Diemen,<br>The Netherlands<br>+31 (0)20 345 53 30<br>Drs. P. Bouter |
|----|---------------------|----------------------------------------------------------------------------------------------------------------------------------------|
|----|---------------------|----------------------------------------------------------------------------------------------------------------------------------------|
- Date Prepared: 2. Up-dated January 2025 (v5.0)
# 807.92(a)(2)
#### Device Identification: ന്
| Trade/Proprietary Name: | Advanced Cryo Wart Remover |
|-------------------------|-------------------------------------------------------------------------------|
| Common/Usual Name: | Cryogenic OTC Pen |
| 510 (k) number: | K242288 |
| Classification Name: | Layperson wart removal cryogenic device |
| Regulation Number: | 21 CFR 878.4350 |
| Product Code: | GEH, Unit, Cryosurgical, Accessories |
| Class: | II |
| Classification Panel: | General and Plastic Surgery (878.4350 - Cryosurgical unit<br>and accessories) |
# 807.92(a)(3)
#### 4. Legally marketed Primary predicate device:
| Device Name: | Wartie® |
|-------------------|----------------------|
| Proprietary Name: | Wartie® Wart Remover |
| 510 (k) number: | K140314 |
| Manufacturer: | YouMedical BV |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for medicalbrands. The logo consists of the word "medicalbrands" in a sans-serif font, with a shield-shaped icon above it. The shield icon has a dashed line pattern inside and a "TM" symbol in the upper right corner.
> Requlation Number 21 CFR 878.4350
Product Code GEH, Unit, Cryosurgical, Accessories
The Wartie® Wart Remover is used for treatment of warts by providing an over the counter cryosurqery product (for the treatment of warts), to be used at home.
# 807.92(a)(4)
### 5. Description of the Subject Device
The Advanced Cryo Wart Remover is an over-the-counter (OTC) cryotherapy device designed for the treatment of common and plantar warts. The device includes:
- A 15 mL / 0.51 oz pressurized aerosol canister containing a freezing agent (1,1-Difluoroethane, Propane, and Isopentane).
- . A metal tip that applies the cryogen directly to the wart
- Detailed instructions, including illustrated descriptions. ●
To activate the product, the lay user must, while holding the device facing downward, press the side buttons firmly five times in a row. Apply the cold metal tip for 20 seconds on warts located on the hands, tops of feet, and toes. Apply for 40 seconds on plantar warts in areas with thicker or calloused skin.
The device is non-sterile, reusable, and intended for home use by laypersons. No special training or sterilization is required for operation.
## 807.92(a)(5)
#### Indication for use and Intended Users of Subject Device 6.
The Advanced Cryo Wart Remover is an over-the-counter (OTC) device indicated for adults and children from the age of four years and older for the treatment of common and plantar warts.
## 807.92(a)(5)
#### Indication for use Comparison 7.
The subject device has the same indications for use as the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for medicalbrands. The logo consists of the word "medicalbrands" in a sans-serif font, with a shield-shaped icon above it. The shield icon has a grid pattern inside and a trademark symbol on the top right.
# 807.92(a)(6)
### Technological Comparison 8.
The subject device has the same technological characteristics as the predicate devices. It shares the same intended use, mechanism of action, and technology as the predicate devices. Whereby the subject device share the following similar characteristics:
- Gas composition of cryogen; Mixture of DMPE Dimethyl ether, Propane, Isobutane versus 1,1 ● Difluoroethane, Propane, Isopentane)
# 807.92(b)
### 9. Non-clinical Performance Data
To demonstrate safety and effectiveness of the Advanced Cryo Wart Remover and to show substantial equivalence to the predicate device, the following non-clinical tests were performed.
- Biocompatibility Testing per ISO 10993-1 Supports biocompatibility of the device. ●
- 。 ISO 10993-5 (cytotoxicity)
- 。 ISO 10993-10 (irritation)
- · ISO 10993-10 (sensitization)
- · ISO 10993-17 (toxicological risk assessment (TRA))
- · ISO 10993-18 (chemical characterization)
Advanced Cryo Wart Remover passed the testing in accordance with:
- Shelf Life Testing per ASTM D3090-72 Supports shelf life of 36 months.
- · Shelf life testing was based on device stability and functionality testing.
### Transportation Testing per ASTM D4169
### Bench Testing - Supports device performance.
- Advanced Cryo Wart Remover was subjected to comparative thermal performance testing that included thermoelectric testing to determine freeze temperature extremes, freeze temperature plateau, and freeze duration. The test provided data on both the temperature profile at the target surface and on the tissue freezing capacity of the device. The results of the tests demonstrated comparable performance to the predicate devices.
### Label and IFU Comprehension Study - Supports device self-selection & labeling comprehension.
- A Flesch test was performed on the proposed labeling to ensure that the instructions were considered 'readable' to lay consumers. The test generated a Flesch Reading Ease score of 61.22, Flesch-Kincaid Grade Level of 7.26, and a ReadablePro rating of 'A'. According to the report created, the text has a Reach score of 100%, meaning that the text is readable for 100% of the addressable audience, equating to approximately 85% of the general public.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for medicalbrands. The logo consists of the word "medicalbrands" in a sans-serif font, with a shield-shaped icon above it. The shield is outlined and has dashed lines inside, creating a grid-like pattern. There is a trademark symbol to the right of the shield.
### Human Factors Study - Supports safe usability of the device for lay-person use.
- A usability study compliant with EN ISO 8317:2004 was conducted to evaluate the design and safe use of the Advanced Cryo Wart Remover by lay consumers. The study assessed the device's usability, closure system, and instructions for use. Ten participants aged 50-70 found the device easy to use, citing its intuitive design, easy grip, and straightforward application. All users successfully operated the opening mechanism, requiring no changes to the device or instructions.
- A secondary usability study evaluated user interpretation time (one or two seconds) to ensure accurate device charging. Ten participants tested three methods (counting, phrase articulation, word articulation) five times each. Results showed that phrase and word articulation provided more consistent timing than counting. The study confirmed that the device design and Instructions for Use enable safe and effective use.
- · A self assessment study including among 20 participants, 95% (19/20) accurately identified images of common and plantar warts.. All participants correctly confirmed their eligibility and understood the device's purpose and target users (common and plantar warts on hands and feet for ages 4+).
## 10. Statement of Substantial Equivalence
Based on the information presented, it is concluded that the Advanced Cryo Wart Remover is substantially equivalent to the primary predicate device Wart Remover and the second primary predicate device Dr. Scholl's Skin Tag Remover.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.