CaRi-Plaque is intended to provide an optimized non-invasive application to analyze coronary anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images. CaRi-Plaque is a web-based image processing application. It is a non-invasive diagnostic reading software intended for use as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses. CaRi-Plaque is intended for use by internal operators who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology may be present that is near or smaller than the scanning resolution. The analysis results produced by the software and provided to the Healthcare Professional are not intended to replace the skill and judgment of a qualified medical practitioner. The analysis results should be reviewed with other clinical information which may include but is not limited to: The patient's original CT images, clinical history, symptoms, clinical risk factors, results of other diagnostic tests, and the clinical judgement of appropriately qualified Healthcare Professionals.
Device Story
CaRi-Plaque is a web-based software application for quantitative/qualitative analysis of coronary artery plaque and stenosis using CCTA DICOM data. Trained operators at regional data centers use the software to perform semi-automatic segmentation of vessels, plaque, and lumen; the process combines 3D threshold-based algorithms with manual editing tools. The device outputs a report containing quantitative metrics: Plaque Burden, Calcified/Noncalcified/Low-Density Noncalcified Plaque Volumes, Remodeling Index, and Maximum Stenosis. Reports are provided to healthcare professionals (e.g., cardiologists, radiologists) to assist in assessing coronary disease severity and treatment planning. The device does not replace clinical judgment; clinicians may request re-analysis. Benefits include non-invasive characterization of coronary plaque, aiding clinical decision-making.
Clinical Evidence
Clinical validation study included 117 subjects (85 men, 32 women, ages 27-85) across 4 sites. Performance compared device measurements to expert-reader ground truth. Primary endpoints (Pearson Correlation Coefficients) were: Maximum Stenosis (0.930), Total Plaque Volume (0.985), Calcified Plaque Volume (0.999), Noncalcified Plaque Volume (0.977), Plaque Burden (0.885), and Low Density Noncalcified Plaque Volume (0.817). Remodeling Index assessed via Cohen's Kappa (0.42). Subgroup analysis by vessel, sex, and age showed no performance bias. Inter-operator agreement and repeatability were confirmed.
Technological Characteristics
Web-based software application; client-side Google Chrome interface. DICOM 3 compliant. Employs semi-automatic, threshold-based segmentation algorithms with manual editing capabilities. Provides 2D/3D visualization (MPR, curved MPR, straightened view). Quantitative outputs include plaque volumes, burden, and stenosis metrics. Software verification/validation performed. No specific hardware materials or energy sources; standalone software-only device.
Indications for Use
Indicated for use by trained operators to analyze coronary anatomy and pathology from CT angiographic images to determine the presence and extent of coronary plaques and luminal stenoses in adult patients. Results are intended for review by qualified medical practitioners in conjunction with other clinical information to aid in determining treatment paths.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
February 20, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Caristo Diagnostics Ltd. % John Smith Partner Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW WASHINGTON DC 20004
Re: K242240
Trade/Device Name: CaRi-Plaque Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: July 30, 2024 Received: January 16, 2025
Dear John Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pm.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and
{1}------------------------------------------------
21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/unique-deviceidentification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-how-report-medicaldevice-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuingeducation/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Zhkke
, for
Jessica Lamb Assistant Director DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
510(k) Number (if known) K242240
Device Name CaRi-Plaque
#### Indications for Use (Describe)
CaRi-Plaque is intended to provide an optimized non-invasive application to analyze coronary anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images.
CaRi-Plaque is a web-based image processing application. It is a non-invasive diagnostic reading software intended for use as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses.
CaRi-Plaque is intended for use by internal operators who have been appropriately trained in the software's functions, capabilities and limitations.
Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology may be present that is near or smaller than the scanning resolution.
The analysis results produced by the software and provided to the Healthcare Professional are not intended to replace the skill and judgment of a qualified medical practitioner. The analysis results should be reviewed with other clinical information which may include but is not limited to: The patient's original CT images, clinical history, symptoms, clinical risk factors, results of other diagnostic tests, and the clinical judgement of appropriately qualified Healthcare Professionals.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Caristo diagnostics. The logo consists of the word "Caristo" in a bold, sans-serif font, with the word "diagnostics" in a smaller font below it. To the left of the word "Caristo" is a stylized graphic consisting of three concentric arcs in yellow, orange, and red. The arcs are arranged in a semi-circle shape, with the open end facing to the right.
510(k) SUMMARY Caristo's CaRi-Plaque
#### Submitter and 510(k) Owner
Caristo Diagnostics New Barclay House 234 Botley Road Oxford, OX2 0HP United Kingdom
#### Official Correspondent
John Smith Partner, Hogan Lovells US LLP 555 Thirteenth Street, NW Washington DC 20004 Telephone: +1 202 637 3638 E-mail: john.smith@hoganlovells.com
| Date of Preparation: | February 11, 2025 |
|----------------------|-------------------------------------------------------|
| Name of Device: | CaRi-Plaque |
| Classification Name: | Radiological Image Processing System |
| Regulatory Class: | Class II |
| Product Code: | LLZ |
| Regulation: | 21 CFR §892.2050 |
| Predicate Device: | Autoplaque 3.0 (K212758, Cedars-Sinai Medical Center) |
| Reference Device: | CaRiCloud v1.0 (K200274, Caristo Diagnostics) |
#### Device Description
CaRi-Plaque v1.0 ("CaRi-Plaque," the subject device) is a web-based software-only application for the quantitative and qualitative clinical analysis of previously acquired CCTA DICOM data for the purpose of characterizing and quantifying plaque formation and stenosis in coronary arteries. CaRi-Plaque aids healthcare professionals trained in cardiovascular health and patient care (including but not limited to Cardiologists, Radiologists and others) by describing the physical characteristics of coronary plaque volume and cross- sectional area, determined using a combination of 3D image thresholding computerized algorithms and manual editing tools to provide automated quantification and characterization of coronary atherosclerotic plaque and stenosis. The processing of CT scan data is performed by trained operators and the resulting CaRi-Plaque Report is provided to the healthcare professional to enable them to assess the extent and severity of coronary disease.
The CaRi-Plaque report includes visual representations of each vessel and associated quantitative outputs. These quantitative outputs include: Plaque Burden, Calcified Plaque (CP) Volume, Total Plaque
{4}------------------------------------------------
Volume, Noncalcified Plaque (NCP) Volume, Low Density Noncalcified Plaque (LD-NCP) Volume, Remodeling Index, and Maximum Stenosis. CaRi-Plaque does not replace standard clinical practice or clinical decisionmaking, and the results of the CaRi-Plaque analysis are to be used in the context of other patient information by the healthcare professional. The healthcare professional may request a re-analysis of the CT scan data if they do not agree with the report analysis.
# Intended Use / Indications for Use
CaRi-Plaque is intended to provide an optimized non-invasive application to analyze coronary anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images.
CaRi-Plaque is a web-based image processing application. It is a non-invasive diagnostic reading software intended for use as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses.
CaRi-Plaque is intended for use by internal operators who have been appropriately trained in the software's functions, capabilities and limitations.
Users should be aware that certain views make use of interpolated data. This is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology may be present that is near or smaller than the scanning resolution.
The analysis results produced by the software and provided to the Healthcare Professional are not intended to replace the skill and judgment of a qualified medical practitioner. The analysis results should be reviewed with other clinical information which may include but is not limited to: The patient's original CT images, clinical history, symptoms, clinical risk factors, results of other diagnostic tests, and the clinical judgement of appropriately qualified Healthcare Professionals.
## Comparison to the Predicate Device
The overall intended use for CaRi-Plaque is the same as the predicate device: both devices are intended to process radiological images and provide measurements of anatomical components in those images. Furthermore, the subject device's indications for use statement is similar to the predicate device. Both devices are indicated to provide an optimized non-invasive application to analyze coronary anatomy and pathology from CT angiographic images, to provide a post-processing tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenosis, and to provide analysis results to aid in the determination of treatment paths by the healthcare professional, in conjunction with other patient data. Additionally, both devices are indicated for use by trained operators.
The differences in indications for use between the subject device and predicate device do not cause the subject device to have a different overall intended use. In particular, the location of use for the software operator differs, as CaRi-Plaque is indicated as a web-based application that can be operated off-site with imaging data transmitted via web-based software, while the predicate device is indicationbased desktop device for use at clinical sites. However, for both devices the final analysis results are indicated to be provided to a healthcare professional for use at a clinical location. Therefore, the indicated recipients and locations of application of the final device output are the same for both devices. Of note, the location of use and workflow of the CaRi-Plaque software is the same as the
{5}------------------------------------------------
CaRiCloud reference device – i.e., trained operators located in regional data centers use the software device to perform an analysis and generate a report that is subsequently provided to a healthcare professional end user.
## Comparison of Technological Characteristics
The technology of CaRi-Plaque is similar to the predicate device (see Table 1). Both devices analyze CT angiographic images for the presence and extent of coronary plaque and luminal stenosis. In addition, they both require trained operators to perform the assessment. Both devices provide equivalent quantitative measurements in a report to the healthcare professional after completing the analysis. These reports are intended to be reviewed by the healthcare professional and used in conjunction with other patient information, such as the original CT images, clinical history, symptoms, clinical risk factors, results of other diagnostic tests, and the clinical judgement of the ordering prescriber. Both devices have demonstrated agreement between clinical expert readers and software measurements in clinical performance testing and have demonstrated appropriate inter-operator and intra-operator agreement.
There are three key technological differences between the subject device and the predicate. First, the location of the DICOM image data is different: The CaRi-Plaque is web-based and receives DICOM images via transmittal by a secure picture archiving and communication system (PACS), while the AutoPlaque predicate software resides on a commercial computer platform at the and receives DICOM image data via loading from the local computer hard drive. Second, the location of the trained software operator may be different: The subject device operator can be at a remote site, while the predicate device operator is at the clinical site on a desktop computer. For both devices, the software operators are trained to review anatomy, define and correct the vessels of interest and generate a report. In both cases, the healthcare professional end user can repeat (or request of) the analysis if needed. The third difference is that the methods of developing segmentation and plaque calculation algorithms differ for the subject versus the predicate device, as the subject device uses threshold-based segmentation while the predicate device uses machine learning to train the algorithms. This technological methodology difference does not raise questions of differences in safety or effectiveness because the algorithms of both devices are intended to accomplish the same indicated uses and were verified and validated using the same clinical testing approach in which device outputs are compared to expertdetermined ground truth. Furthermore, as discussed below in the Performance Testing section, the CaRi-Plaque software met all acceptance criteria in the clinical validation study, showing an excellent correlation between CaRi-Plaque measurements and ground truth by expert readers for the same endpoints that the predicate device used in its clinical performance testing.
Therefore, the minor technological differences between the subject and predicate device do not raise different questions of safety or effectiveness.
{6}------------------------------------------------
| Parameter | CaRi-Plaque<br>Subject Device | Autoplaque 3.0 (K212758)<br>Predicate Device | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----|------|---------------------|-----|------|-------------------------|-----|------|-------------------------------------------|-----|------|----------------------------------------------------------|-----|------|---------------|----------------------------------------------------------------|-----|--------------------------------------------------------------------|----|-----|--------------------------------------------------------------------------------------|----|-----|--------------------------------------------------------------|----|-----|--------------------------------------------------------------------|-----|------|--------------------------------------------------------------------------------------------------|----|-----|---------------------------------------------------------------------------------------------|----|-----|---------------------------------------------------------------|----|-----|
| Computer Operating System | Client-side Google Chrome<br>application | Windows / Mac OS | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Stand-alone software | Yes | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| DICOM Compliance | DICOM 3 | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| 2D Imaging | Review of coronary vessels in 2D<br>MPR, curved MPR, and<br>straightened view. | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| 2D Measurement | 2D measurement tools of vessel<br>diameter and contour | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| 3D Imaging | Review of structures in 3D | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Maximum intensity projection (MIP) | N/A; this visualization method is<br>not necessary analyzing the<br>presence and extent of coronary<br>plaque and luminal stenosis | MIP with interactive control | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Multiplanar reformatting (MPR) | MPR with oblique slicing.<br>Variable slab thickness is not<br>necessary analyzing the<br>presence and extent of coronary<br>plaque and luminal stenosis | MPR with oblique slicing and<br>variable slab thickness | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Quantitative Measurements | | | Diameter stenosis: Maximum diameter<br>stenosis, with respect to proximal and<br>distal references | Yes | Same | Total Plaque Volume | Yes | Same | Calcified Plaque Volume | Yes | Same | NCP volume: noncalcified plaque<br>volume | Yes | Same | LD-NCP volume: low-density<br>noncalcified plaque volume | Yes | Same | Vessel Volume | No. This is calculated but not<br>reported as a device output. | Yes | NCP burden: noncalcified plaque<br>volume / analyzed vessel volume | No | Yes | LD-NCP burden: low-density<br>noncalcified plaque volume / analyzed<br>vessel volume | No | Yes | CP burden: calcified plaque<br>volume/analyzed vessel volume | No | Yes | Total plaque burden: total plaque<br>volume/analyzed vessel volume | Yes | Same | Plaque composition NCP: noncalcified<br>plaque composition (NCP volume /<br>total plaque volume) | No | Yes | Plaque composition CP: calcified<br>plaque composition (CP volume / total<br>plaque volume) | No | Yes | Plaque composition LDNCP: low-<br>density noncalcified plaque | No | Yes |
| Quantitative Measurements | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Diameter stenosis: Maximum diameter<br>stenosis, with respect to proximal and<br>distal references | Yes | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Total Plaque Volume | Yes | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Calcified Plaque Volume | Yes | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| NCP volume: noncalcified plaque<br>volume | Yes | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| LD-NCP volume: low-density<br>noncalcified plaque volume | Yes | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Vessel Volume | No. This is calculated but not<br>reported as a device output. | Yes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| NCP burden: noncalcified plaque<br>volume / analyzed vessel volume | No | Yes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| LD-NCP burden: low-density<br>noncalcified plaque volume / analyzed<br>vessel volume | No | Yes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| CP burden: calcified plaque<br>volume/analyzed vessel volume | No | Yes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Total plaque burden: total plaque<br>volume/analyzed vessel volume | Yes | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Plaque composition NCP: noncalcified<br>plaque composition (NCP volume /<br>total plaque volume) | No | Yes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Plaque composition CP: calcified<br>plaque composition (CP volume / total<br>plaque volume) | No | Yes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Plaque composition LDNCP: low-<br>density noncalcified plaque | No | Yes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
# Table 1. Technology Comparison to Predicate
{7}------------------------------------------------
| Parameter | CaRi-Plaque<br>Subject Device | Autoplaque 3.0 (K212758)<br>Predicate Device |
|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| composition (LD-NCP volume / NCP<br>volume) | | |
| QCAD: Maximum diameter stenosis | Yes, refers to a single reference<br>site | Yes, refers to proximal and distal<br>references |
| Remodeling index: ratio of maximum<br>vessel area / proximal and distal<br>references | Yes | Same |
| Area stenosis: maximum area stenosis,<br>with respect to proximal and distal<br>references | Yes | Same |
| Plaque length: diseased vessel length | No. This is calculated but not<br>reported as a device output. | Yes |
| Contrast density difference: maximum<br>difference in contrast density over<br>lesion with respect to proximal | No; not used as a common<br>reference standard | Yes |
| MLD: minimal luminal dimeter over<br>lesion | No. This is calculated but not<br>reported as a device output. | Yes |
| MLA: minimum luminal area over<br>lesion | No | Yes |
| Vessel profile: area, maximum<br>diameter, minimum diameter<br>measured from selected vessel cross<br>section | Yes | Same |
| Lumen profile: area, maximum<br>diameter, minimum diameter<br>measured from selected lumen cross<br>section | Yes | Same |
| Vessel, plaque, and lumen<br>segmentation | Yes- semi-automatic (threshold-<br>based segmentation with full<br>option to edit); output<br>demonstrated to meet pre-<br>determined acceptance criteria<br>in clinical performance study | Yes- semi-automatic (deep<br>learning based with full option to<br>edit) |
## Performance Testing
The 510(k) submission provided performance data to establish the substantial equivalence of CaRi-Plaque (the subject device) to the predicate device. A summary of these performance tests is provided below.
## Software Verification and Validation:
Software verification and validation testing were conducted to demonstrate that the subject device meets specifications and works as intended.
## Clinical Validation Study:
A clinical performance test was conducted to compare plaque characteristics reported with CaRi-Plaque to ground truth determined by qualified independent medical experts. The study determined the degree of consistency between plaque characteristics reported by CaRi-Plaque and independent medical expert readers using a multi-center, international patient population in 117 subjects (85 men, 32 women) aged 27 to 85 years of age from four (4) sites (2 US and 2 OUS). Individual subject-level ethnicity was collected at two of the four study sites. This gave the following breakdown for 57 subjects: White 76%, Asian 4%,
{8}------------------------------------------------
Middle Eastern 1%, other 14%, and unknown 5%. For the two remaining sites, ethnicity was estimated from local population census data (www.census.gov). This gave the following breakdown for the first site (48 subjects): White (not Hispanic or Latino) 44%, Black 22%, Hispanic or Latino 27%, and Asian 5%; and for the second site (12 subjects): White (not Hispanic or Latino) 46%, Black 19%, Hispanic or Latino 26%, and Asian 13%. The CT scanners to collect images included commercial CT scanners from Toshiba, GE, Phillips, and Siemens. Qualified independent medical experts performed ground truthing.
The study established agreement between the expert readers and the CaRi-Plaque device measurements for Maximum Stenosis, Total Plaque Volume, Calcified Plaque Volume and Noncalcified Plaque Volume using the acceptance criteria of Pearson Correlation Coefficients for each metric. Agreement was also reported for Plaque Burden, Low Density Noncalcified Plaque Volume and Remodeling Index. The Pearson Correlation Coefficients for Maximum Stenosis, Total Plaque Volume, Calcified Plaque Volume, Noncalcified Plaque Volume, Plaque Burden and Low Density Noncalcified Plaque Volume were 0.930, 0.985, 0.999, 0.977, 0.885, 0.817, respectively. These results met the predetermined acceptance criteria. Remodeling Index was assessed using Cohen's Kappa with an agreement of 63.3% (K = 0.42). Subgroup analysis by vessel, sex and age did not reveal any bias in performance.
Agreement between the expert readers and the CaRi-Plaque device measurements at the branch level were reported using Pearson Correlation Coefficients (or Cohen's Kappa for remodeling index). For RCA, R-PDA and R-PLB: Maximum Stenosis, Total Plaque Volume, Calcified Plaque Volume, Noncalcified Plaque Volume, Plaque Burden, Low Density Noncalcified Plaque Volume and Remodeling Index were 0.863, 0.958, 0.964, 0.953, 0.924, 0.633, 90.3% (K=0.31), respectively. For LAD, D1, D2 and Ramus: Maximum Stenosis, Total Plaque Volume, Calcified Plaque Volume, Noncalcified Plaque Volume, Plaque Burden, Low Density Noncalcified Plaque Volume and Remodeling Index were 0.929, 0.959, 0.997, 0.942, 0.913, 0.802, 87.2% (K=0.44), respectively. For LCX, OM1 and OM2: Maximum Stenosis, Total Plaque Volume, Calcified Plaque Volume, Noncalcified Plaque Volume, Plaque Burden and Remodeling Index were 0.911, 0.948, 0.991, 0.934, 0.888, 0.633, 90.3% (K=0.31), respectively. There was insufficient low density non-calcified plaque in the LCX, OM1 and OM2 to generate a meaningful correlation.
Additionally, there was good inter-operator agreement and repeatability at the branch level within the CaRi-Plaque device arm and between readers in the Ground Truth arm.
## Conclusion
The information provided above supports that CaRi-Plaque is as safe and effective as the predicate device. The device has the same intended use as the predicate device, and minor differences in indications for use do not alter the intended use of the device and do not affect its safety and effectiveness. In addition, the minor technological differences between the devices do not raise different questions of safety or effectiveness, and they are addressed through software verification and validation testing and a clinical validation study. Therefore, CaRi-Plaque is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.