The Disposable Sphincterotome is indicated for use in transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. This device can also be used to cannulate and inject contrast medium.
Device Story
Sterile, single-use endoscopic device; used with flexible endoscopes for intubation of pancreaticobiliary system. Components include cutting wire, sheath, guide wire connector, injection connector, conductive column, finger ring, handle, and imaging ring. Operates by connecting to an external high-frequency (HF) electrosurgical generator; electrical energy delivered through cutting wire to perform sphincterotomy. Used by physicians in clinical settings. Contrast medium injected via injection connector for visualization. Device facilitates access to Papilla of Vater/Sphincter of Oddi; enables therapeutic cutting and diagnostic imaging. Benefits include minimally invasive access to biliary/pancreatic ducts for therapeutic intervention.
Clinical Evidence
No clinical data included. Bench testing only. Evaluated appearance, dimensions, operational performance, tensile strength, hydraulic leak resistance, connector performance, conduction resistance, contrast injection, cutting function, radiopacity, endoscope compatibility, and thermal effects. Electrical safety and EMC testing performed per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, and IEC 60601-2-2. Biocompatibility verified per ISO 10993.
Technological Characteristics
Materials: PTFE, ABS, SUS304. Energy: High-frequency electrosurgical generator. Dimensions: Working length 1800-2300mm; cutting wire 10-30mm. Connectivity: Compatible with endoscopes (channel ≥2.8mm) and guidewires (≤0.035 inch). Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ISO 10993, ISO 11135, ASTM 1980, IEC 60601 series.
Indications for Use
Indicated for transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi, and for cannulation and injection of contrast medium in patients requiring endoscopic biliary/pancreatic procedures.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 6, 2025
Beijing ZKSK Technology Co.,Ltd. % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China
Re: K242192
Trade/Device Name: Disposable Sphincterotome Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit And Accessories Regulatory Class: Class II Product Code: KNS Dated: January 6, 2025 Received: January 7, 2025
Dear Boyle Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee -S
Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242192
Device Name Disposable Sphincterotome
Indications for Use (Describe)
The Disposable Sphincterotome is indicated for use in transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. This device can also be used to cannulate and inject contrast medium.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K242192
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
# 1.0 Submitter's Information
| Name: | Beijing ZKSK Technology Co., Ltd. |
|----------|------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Building 9, 6 & No.6 Yuan Hengye North 7th Street, Yongle<br>Economic Development Zone, Tongzhou District,Beijing 101105,<br>China |
| Tel: | +86 -13811778090 |
| Contact: | Ma Li |
#### Designated Submission Correspondent
| Contact: | Mr. Boyle Wang |
|----------|-------------------------------------------------------------------------|
| Name: | Shanghai Truthful Information Technology Co., Ltd. |
| Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong, Shanghai,<br>200120 China |
| Tel: | +86-21-50313932 |
| Email: | Info@truthful.com.cn |
Date of Preparation: Jan.14th, 2025
# 2.0 Device Information
| Trade name: | Disposable Sphincterotome |
|------------------|-------------------------------------------------|
| Common name: | Sphincterotome |
| Regulation name: | Endoscopic electrosurgical unit and accessories |
## 3.0 Classification
| Product code: | KNS |
|--------------------|--------------------------|
| Regulation number: | 21 CFR 876.4300 |
| Classification: | Class II |
| Panel: | Gastroenterology/Urology |
#### 4.0 Predicate Device Information
{5}------------------------------------------------
| Manufacturer: | Boston Scientific Corporation (BSC) |
|--------------------|-------------------------------------|
| Trade/Device Name: | Autotome™ RX |
| 510(k) number: | K013153 |
# 5.0 Device Description
The subject device Disposable Sphincterotome is a sterile, single-use endoscopic device, intended to be used with flexible endoscopes for intubation of the pancreaticobiliary system and for sphincterotome.
The disposable sphincterotome consists of cutting wire, sheath, quide wire connector, injection connector, conductive column, finger ring, handle, and imaging ring.
The subject device has 72 specifications. The differences among these models are the Cutting Length, Tip Length, Working Length and Type of guide wire lumen channel.
The subject device is EO sterilized to achieve the Sterility Assurance Level (SAL) of 10-6 and placed in a sterility maintenance package to ensure a shelf life of 3 years.
The main materials used for construction of Disposable Sphincterotome include PTFE, ABS, SUS304.
### 6.0 Indication for Use Statement
Sphincterotome is indicated for use in transendoscopic The Disposable sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. This device can also be used to cannulate and inject contrast medium.
# 7.0 Summary of Non-Clinical Testing
Bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The subject device has been subjected to compliance testing according to, by FDA, recognized consensus standards ISO 10993-1, ISO 10993-7, ISO 11607-1, and Technical Requirements of "Disposable Sphincterotome" provided by Beijing ZKSK Technology Co., Ltd. Results from testing performed confirms that the design requirement specification and user needs have been met. The subject device is confirmed to be safe and effective for the intended use.
7.1 Sterilization and shelf life of Disposable Sphincterotome is delivered sterile and have successfully been tested according to ISO 11135 and ASTM 1980. The label shelf life is 3 vears.
7.2 Biocompatibility testing of Disposable Sphincterotome has successfully been tested for cytotoxicity, sensitization, intracutaneously irritation, acute systemic toxicity and material medicated pyrogenicity. The test results verify that the biocompatibility criteria given in ISO 10993 are fulfilled. Disposable Sphincterotome is non-toxic and biocompatible.
{6}------------------------------------------------
7.3 Performance testing - Bench: The following bench tests were performed on Disposable Sphincterotome:
- -- Appearance,
- -- Dimension,
- -- Operational Performance,
- -- Tensile Performance,
- -- Hydraulic Leak Resistance,
- -- Injection Connector (Luer connectors) performance,
- -- Conduction resistance,
- -- Contrast Agent Injection Function
- -- Cutting Line Function
- -- Radiopacity
- -- Compatible endoscopes tests
- -- Thermal effect test
The results of all testing of Disposable Sphincterotome have been verified.
7.4 Electromagnetic Compatibility and Electrical Safety: Tests on Electromagnetic Compatibility and Electrical Safety were performed in accordance to requirements per IEC 60601-1:2020, IEC 60601-1-2:2020, IEC 60601-2-18:2009 and IEC 60601-2-2:2017.
# 8.0 _Summary of Clinical Testing
No clinical study is included in this submission.
# 9.0 Technological Characteristic Comparison Table
| Item | Subject Device | Predicate Device | Remark |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 510(k) No. | K242192 | K013153 | / |
| Product Code | KNS | KNS | Same |
| Regulation No. | 21 CFR 876.4300 | 21 CFR 876.4300 | Same |
| Class | II | II | Same |
| Intended<br>Use/Indication for<br>Use | The Disposable Sphincterotome<br>is indicated for use in<br>transendoscopic sphincterotomy<br>of the Papilla of Vater and/or the<br>Sphincter of Oddi. This device<br>can also be used to cannulate<br>and inject contrast medium. | The Autotome™ RX is indicated<br>for use in transendoscopic<br>sphincterotomy of the Papilla of<br>Vater and/or the Sphincter of<br>Oddi. This device can also be<br>used to cannulate and inject<br>contrast medium. | Same |
| Injection | Contrast Medium | Contrast Medium | Same |
#### Table 2- Comparison of Technology Characteristics
{7}------------------------------------------------
| Cutting Wire Length | 10/20/25/30mm | 20/30mm | Different |
|-----------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|-----------|
| Compatible<br>endoscopy working<br>channel | ≥2.8mm | ≥2.8mm | Same |
| Compatible guidewire<br>diameter | ≤0.035 inch (Ø0.89mm) | ≤0.035 inch (Ø0.89mm) | Same |
| Energy source | HF generator | HF generator | Same |
| Working Length | 1800/2000/2300mm | 2000mm | Different |
| Tip length | 5/10/15mm | 5mm | Different |
| Rated High-<br>Frequency Voltage | 1600 Vp (3200 Vp-p) | 750Vp (1500Vp-p) | Different |
| Operation Output<br>Power | 30-40W | 30-70W | Different |
| Sterile | Ethylene Oxide, SAL: 10-6 | Ethylene Oxide, SAL: 10-6 | Same |
| Single Use | Single Use | Single Use | Same |
| Biocompatibility | Conform with ISO 10993 -1 | Conform with ISO 10993 -1 | Same |
| Electromagnetic<br>Compatibility and<br>Electrical Safety | Conform with IEC 60601-1;<br>IEC 60601-1-2;<br>IEC 60601-2-2;<br>IEC 60601-2-18 | Conform with IEC 60601-1;<br>IEC 60601-1-2;<br>IEC 60601-2-2;<br>IEC 60601-2-18 | Same |
The technological characteristics of the subject device are identical to those of predicate device. The subject device has the similar basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:
- · Same intended use
- · Same indications for use
- · Similar material types that meet ISO 10993 biocompatibility requirements
- · Same sterilization methods
- · Same fundamental technology/principal of operation/user interface
1)The Cutting Wire Length, Working Length and Tip length range of the subject device are different from those of the predicate device. Performance testing was performed on the representative specifications that represent the worst-case design. The performance comparison testing was conducted with the predicate device and the candidate device to compare their physical performance. The test results showed that the differences don't affect the substantial equivalence. These differences will have no effect on the safety and effectiveness.
2) The Rated High-Frequency Voltage is different from the predicate device, and both devices had passed the safety and EMC testing.
3)Recommendations on Operation Output Power for the subject device is different from the predicate device. Performance comparison testing on Cutting Line Function under 30W, 40W, and 70W was performed and the test results showed that the subject device is as safe and effective as the predicate device.
Therefore, the differences between the proposed device and the predicated device are considered not to affect the Substantially Equivalency between the proposed and predicate devices concerning the safety and effectiveness.
{8}------------------------------------------------
#### Conclusion 10.0
The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device K013153 and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.