K242188 · Riverain Technologies, Inc. · JAK · Dec 3, 2024 · Radiology
Device Facts
Record ID
K242188
Device Name
ClearRead CT CAC
Applicant
Riverain Technologies, Inc.
Product Code
JAK · Radiology
Decision Date
Dec 3, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Agatston score category
—
weighted kappa > 0.85
kappa of 0.959 (95% CI: 0.943-0.975)
Approximately 400 cases held out from development.
—
Clinical assessment of 491 cases (143 CAC negative, 348 CAC positive) from four CT manufacturers across five clinical sites.
3 (radiologists)
Agatston score
—
—
Bland-Altman limits of agreement of -12.20 ± 437.84
Approximately 400 cases held out from development.
—
Clinical assessment of 491 cases (143 CAC negative, 348 CAC positive) from four CT manufacturers across five clinical sites.
3 (radiologists)
Coronary artery calcification segmentation
—
—
Dice score of 0.892 (95% CI: 0.871, 0.911)
Approximately 400 cases held out from development.
—
Clinical assessment of 491 cases (143 CAC negative, 348 CAC positive) from four CT manufacturers across five clinical sites.
3 (radiologists)
Indications for Use
ClearRead CT CAC is image processing software designed to aid physicians in assessing coronary artery calcification on non-gated, non-contrast, standard, or low-dose chest CT scans of adult patients 30 years of age or older. The primary output of ClearRead CT CAC is the Agatston score. ClearRead CT CAC locates calcified coronary lesions and assigns them to one of several coronary arteries: the right coronary artery (RCA), the left main artery (LM), the left anterior descending artery (LAD), and the left circumflex artery (LCX). The software calculates the Agatston score as the sum for each artery while also providing the total across all coronary arteries. The total Agatston score is calibrated to 3.0mm slice thickness and is used to derive a CAC category and the arterial age. The provided segmentations are for illustrative purposes only and are not intended for diagnostic use. ClearRead CT CAC provides adjunctive information and is not intended to be used without clinical expert review.
Device Story
ClearRead CT CAC is image processing software for radiologists; inputs are DICOM-formatted non-gated, non-contrast, standard, or low-dose chest CT scans (up to 3mm slice thickness). The device uses machine learning to automatically detect and segment calcified coronary lesions; assigns lesions to specific coronary arteries (RCA, LM, LAD, LCX); calculates Agatston scores per artery and total; derives CAC category and arterial age. Output is provided to the physician for review within the clinical workflow; segmentations are illustrative. The device provides adjunctive information to aid clinical decision-making; it does not replace the original scan or report. Benefits include automated quantification of coronary calcium burden to assist in patient assessment.
Clinical Evidence
Standalone performance assessment using 491 cases (143 negative, 348 positive for CAC) from 5 sites and 4 CT manufacturers. Primary endpoint: quadratic weighted Kappa for Agatston score category agreement was 0.959 (95% CI: 0.943-0.975). Secondary endpoints: Bland-Altman analysis showed LoA of -12.20 ± 437.84; Dice score comparison for segmentation yielded 0.892 (95% CI: 0.871-0.911) vs. 0.910 (95% CI: 0.893-0.927) for ground truth consensus. Results demonstrate high accuracy in automated CAC quantification.
Technological Characteristics
Software-based image processing; DICOM input/output; automatic detection and segmentation of coronary calcifications; Agatston scoring algorithm; calibrated to 3.0mm slice thickness; 130 HU default threshold. Compliant with IEC 62304, IEC 62366-1, ISO 14971, and NEMA PS 3.1-3-20. Moderate level of concern software.
Indications for Use
Indicated for adult patients 30 years of age or older undergoing non-gated, non-contrast, standard, or low-dose chest CT scans for assessment of coronary artery calcification.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
December 3, 2024
Riverain Technologies, Inc. % Jonathan Jackson Director of RAOA 3130 South Tech Boulevard MIAMISBURG, OH 45342
Re: K242188
Trade/Device Name: ClearRead CT CAC Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: November 13, 2024 Received: November 14, 2024
Dear Jonathan Jackson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K242188
Device Name ClearRead CT CAC
#### Indications for Use (Describe)
ClearRead CT CAC is image processing software designed to aid physicians in assessing coronary artery calcification on non-gated, non-contrast, standard, or low-dose chest CT scans of adult patients 30 years of age or older. The primary output of ClearRead CT CAC is the Agatston score. ClearRead CT CAC locates calcified coronary lesions and assigns them to one of several coronary arteries: the right coronary artery (RCA), the left main artery (LM), the left anterior descending artery (LAD), and the left circumflex artery (LCX). The software calculates the Agatston score as the sum for each artery while also providing the total across all coronary arteries. The total Agatston score is calibrated to 3.0mm slice thickness and is used to derive a CAC category and the arterial age. The provided segmentations are for illustrative purposes only and are not intended for diagnostic use.
ClearRead CT CAC provides adjunctive information and is not intended to be used without clinical expert review.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="unicode-bidi:isolate; direction:ltr;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="unicode-bidi:isolate; direction:ltr;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### K242188
# 510(k) Summary
### Submitter Information:
| Company Name: | Riverain Technologies, Inc. |
|------------------|----------------------------------------------------|
| Company Address: | 3130 South Tech Blvd.<br>Miamisburg, OH 45342-4860 |
| Contact Person: | Jonathan Jackson |
| | Director of RAQA |
| | Riverain Technologies, Inc.<br>937.531.5092 |
jjackson@riveraintech.com
### Device Information:
| Trade Name: | ClearRead CT CAC |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number: | 21 CFR 892.1750 |
| Regulation Name: | Computed tomography x-ray system |
| Regulatory Class: | Class II |
| Product Code: | JAK |
| Device Description: | ClearRead CT CAC is a computer aided quantification system<br>intended to aid in the assessment of coronary artery calcification.<br>The system receives chest Computed Tomography (CT) studies as<br>input, in DICOM® format, and generates output in DICOM (or<br>other) format. |
| Intended Purpose: | ClearRead CT CAC is intended to aid physicians in assessing<br>Coronary Artery Calcification while reviewing non-gated, non-<br>contrast, standard, and low-dose chest CT exams. |
| Indications for Use: | ClearRead CT CAC is image processing software designed to aid<br>physicians in assessing coronary artery calcification on non-gated,<br>non-contrast, standard, or low-dose chest CT scans of adult<br>patients 30 years of age or older.<br>The primary output of ClearRead CT CAC is the Agatston score.<br>ClearRead CT CAC locates calcified coronary lesions and<br>assigns them to one of several coronary arteries: the right coronary<br>artery (RCA), the left main artery (LM), the left anterior<br>descending artery (LAD), and the left circumflex artery (LCX).<br>The software calculates the Agatston score as the sum for each |
{5}------------------------------------------------
artery while also providing the total across all coronary arteries. The total Agatston score is calibrated to 3.0mm slice thickness and is used to derive a CAC category and the arterial age. The provided segmentations are for illustrative purposes only and are not intended for diagnostic use.
ClearRead CT CAC provides adjunctive information and is not intended to be used without clinical expert review.
#### Bunkerhill, Inc. Predicate Device: (K230223) iCAC Device
Class II
# Comparison to Predicate Device Technical Characteristics
Riverain Technologies, Inc. is of the opinion that ClearRead CT CAC is substantially equivalent, both in intended use as well as to the technical characteristics of the listed predicate device. Differences in the design and performance of the cited predicate device do not affect either the safety or the effectiveness of ClearRead CT CAC for its intended use. Table 1 shows the predicate devices listed along with the subject device, including the Product Code as well as the Indications for Use for each device.
| | Predicate: | Subject Device: |
|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | iCAC Device<br>(Bunkerhill, Inc.) | ClearRead CT CAC<br>(Riverain Technologies, Inc.) |
| | K230223 | |
| Product<br>Code | JAK | JAK |
| Indications for<br>Use | iCAC is a software device<br>intended for use in estimating<br>presence and quantity of<br>coronary artery calcium for<br>patients aged 30 years and above<br>during routine care. The device<br>automatically analyzes non-<br>gated, non-contrast chest<br>computed tomography (CT)<br>images collected during routine<br>care and outputs a visual<br>representation of estimated<br>coronary artery calcium | ClearRead CT CAC is image<br>processing software designed to<br>aid physicians in assessing<br>coronary artery calcification on<br>non-gated, non-contrast, standard,<br>or low-dose chest CT scans of<br>adult patients 30 years of age or<br>older.<br>The primary output of ClearRead<br>CT CAC is the Agatston score.<br>ClearRead CT CAC locates<br>calcified coronary lesions and<br>assigns them to one of several |
| | informational purposes only) and<br>both exact and four-category<br>quantitative estimates of the<br>patient's coronary artery calcium<br>burden in Agatston units.<br>The output of the subject device<br>is made available to the<br>physician on-demand as part of<br>his or her standard workflow.<br>The device generated calcium<br>score or score group can be<br>viewed in the patient report at<br>the discretion of the physician,<br>and the physician also has the<br>option of viewing the device-<br>generated calcium segmentation<br>in a diagnostic image viewer.<br>The subject device output in no<br>way replaces the original patient<br>report or the original chest CT<br>scan; both are still available to be<br>viewed and used at the discretion<br>of the physician.<br>The device is intended to provide<br>information to the physician to<br>provide assistance during review<br>of the patient's case. Results of<br>the subject device are not<br>intended to be used on a stand-<br>alone basis and are solely<br>intended to aid and provide<br>information to the physician. In<br>all cases, further action taken on<br>a patient should only come at the<br>recommendation of the physician<br>after further reviewing the<br>patient's results. | coronary arteries: the right<br>coronary artery (RCA), the left<br>main artery (LM), the left anterior<br>descending artery (LAD), and the<br>left circumflex artery (LCX). The<br>software calculates the Agatston<br>score as the sum for each artery<br>while also providing the total<br>across all coronary arteries. The<br>total Agatston score is calibrated<br>to 3.0mm slice thickness and is<br>used to derive a CAC category and<br>the arterial age. The provided<br>segmentations are for illustrative<br>purposes only and are not intended<br>for diagnostic use.<br>ClearRead CT CAC provides<br>adjunctive information and is not<br>intended to be used without<br>clinical expert review. |
| | | |
| Intended User | Radiologists | Radiologists |
| Modality | Routine, non-contrast, non-gated<br>chest CT Series | Non-contrast, non-gated, standard,<br>or low-dose chest CT scans |
| Anatomical<br>Region | Chest | Chest |
| Clinical<br>Condition | Coronary Artery Calcification | Coronary Artery Calcification |
| Quantification<br>of Calcium<br>Burden | Yes | Yes |
| Segmentation<br>of Calcium | Yes | Yes |
| Number of<br>Detection<br>Categories | 4 | 4 |
| Calculation of<br>Exact Calcium<br>Score | Yes | Yes |
| Type of<br>Information | Adjunctive | Adjunctive |
| Intended<br>location | Medical Facility | Medical Facility |
| Prescriptive<br>Use | Yes | Yes |
| Measurement<br>Scale | Agatston Units | Agatston Units |
| Image Format | DICOM | DICOM |
| Slice<br>Thickness | Up to 5mm | Up to 3mm |
| Calcification<br>Detection | Automatic | Automatic |
| Default<br>Threshold of<br>Calcium | 130 HU (Hounsfield Units) | 130 HU (Hounsfield Units) |
| Coronary<br>Artery<br>Calcification<br>Quantification<br>Method | CAC detection category (based<br>on Agatston score), exact<br>Agatston score. | CAC detection category and<br>arterial age based on the measured<br>Agatston score. |
| Annotation of<br>Detected<br>Calcium | Yes | Yes |
| Generate<br>Patient Report | Optional to copy result to<br>clipboard, insert in report,<br>DICOM Secondary Capture | Optional to copy result to<br>clipboard, insert in report, DICOM<br>Secondary Capture |
| Report of the<br>Calcium Score | Yes, Coronary Calcium<br>Detection Category and exact<br>Agatston score 4 categories (for<br>detection category):<br>• 0<br>• 1-99<br>• 100-399<br>• 400 + | Yes, Coronary Calcium Detection<br>Category and measured Agatston<br>score 4 categories (for detection<br>category):<br>• 0<br>• 1-99<br>• 100-399<br>• 400 + |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Traditional 510(k) Premarket Notification ClearRead™ CT CAC
Table 1: Predicate Devices vs. Subject Device
Key features identified that differ from the predicate are discussed below.
## Slice Thickness
ClearRead CT CAC operates slice thickness up to 3mm, whereas the predicate device indicates use up to 5mm slice thickness. A limit of 3mm is in keeping with the current standards used for gated protocols, to which the Agatston score is calibrated to. For values exceeding 3mm, it is unclear how the performance of the measurement or detection of CAC scores degrades and, therefore, is excluded from consideration.
# Testing Summary
## Non-clinical Testing
Non-clinical tests were conducted during the development process in accordance with the Riverain Technologies Design Control Process, which is compliant with the FDA Quality System Regulations, ISO 13485:2016 with MDSAP, and the following standards.
- IEC 62304:2006/AMD1:2016, Medical devices - Software life cycle processes
- . IEC62366-1:2015, Medical device - Part1: Application of usability engineering to medical devices
- ISO14971:2019, Medical devices Application of Risk Assessment to Medical Devices ●
- NEMA PS 3.1-3-20, Digital Imaging and Communications in Medicine (DICOM) Set ● 2016
- Guidance for Industry and Food and Drug Administration Staff - Cybersecurity in Medical Devices: Ouality System Considerations and Content of Premarket Submissions. Issues September 27, 2023
Testing verified the requirements according to the ClearRead CT CAC device specifications.
The ClearRead CT CAC Risk Assessment was created for this device, with risk control measures implemented to mitigate identified hazards. Documentation required for software with
{9}------------------------------------------------
### Traditional 510(k) Premarket Notification ClearRead™ CT CAC
a Moderate Level of Concern is included as part of this submission. Device labeling, together with the results from verification and validation testing, demonstrate that the device is safe and effective.
Approximately 400 cases were collected and held out from development of any kind to estimate the test performance of the CAC scoring system. Regions associated with coronary artery calcification were manually outlined, including the particular artery, from which the ground truth Agatston score was obtained.
To assess the measurement of the Agatston score, we used weighted kappa scores (weighted, owing to the ordinal nature of the categories) and Bland-Altman analysis. Pre-test acceptance criteria were defined based on existing literature and internal investigation from visually inspected validation performance. If kappa scores were greater than 0.85, these would be considered excellent. Actual CAC scores are those derived from the use of the ground truth mask, while predicted scores come from the ClearRead CT CAC system. Figure 1 below shows a plot of the predicted versus true CAC categories. A weighted kappa score of 0.958 was achieved, demonstrating good agreement between the automatic method and manual processing.
| Predicted | 65 | 5 | 0 | 0 |
|-----------|----|-----|----|----|
| | 0 | 115 | 2 | 0 |
| | 0 | 18 | 91 | 1 |
| | 0 | 0 | 10 | 95 |
Ground Truth
Figure 1: Confusion table of predicted versus ground truth CAC categories
A Bland-Altman plot, as shown in Figure 2, demonstrates strong agreement with the true score compared to the estimated score.
{10}------------------------------------------------
Image /page/10/Figure/2 description: This image is a scatter plot with the x-axis ranging from 0 to 4000 and the y-axis ranging from -1000 to 200. The plot includes a horizontal line indicating the mean at -19.64, with a confidence interval (CI) of -31.54 to -7.734. Additionally, there are two dashed lines representing the limits of agreement (LoA) at 163.1 (CI: 142.4-183.7) and -202.3 (CI: -223--181.7).
Figure 2: Bland-Altman analysis plot for the Agatston scores within the internal test set
## Clinical Performance Testing
A clinical study was performed to validate the clinical efficacy of ClearRead CT CAC.
This study was a standalone model performance assessment of ClearRead CT CAC. This software, as a medical device product, was developed using machine learning to measure the amount of arterial calcification in non-contrast chest CT cases. This assessment involved comparing the algorithm outputs to ground-truth measurements on non-contrast, non-gated chest CT cases to determine the algorithm accuracy in automated measurements of the Agatston score (scaled to 3mm slice thickness). The Agatston score was measured overall and for the four component arteries of the right coronary artery, left anterior descending artery, left circumflex artery, and left main coronary artery.
In total, 491 cases were used in the clinical assessment of the device: 143 cases with no sign of coronary artery calcification and 348 with evidence of CAC, as determined by a consensus ground truth review by three radiologists. These nearly 500 cases included samples from four CT manufacturers across five clinical sites. In the CAC positive cases, selection was performed to ensure the entire range of scores was spanned.
For clinical assessment, the following predefined endpoints were used:
- . Primary endpoint: the accuracy of ClearRead CT CAC in calculating the Agatston score category (0, 1-99, 100-399, ≥400) as defined by the quadratic weighted Kappa coefficient when compared with ground truth radiologist Agatston score category.
- . Secondary endpoint: the accuracy of ClearRead CT CAC in calculating the overall continuous Agatston score as assessed by Bland-Altman analysis.
- Secondary endpoint: the accuracy of ClearRead CT CAC in segmenting the overall coronary artery calcification as assessed by Dice score comparison.
The results of the primary and secondary endpoints are as follows:
- . Primary endpoint: kappa of 0.959 (95% CI: 0.943-0.975). In addition to kappa scoring, percentage agreement scores for CAC positive (PPA) and negative (NPA) cases were
{11}------------------------------------------------
assessed. These calculations assessed the proportion of cases correctly identified in the relevant category (for PPA) or in all other categories (for NPA). Figure 3 summarizes all percentage agreement scores as stratified by CAC category.
- Secondary endpoint: Bland-Altman analysis (see Figure 4 for a data plot) yielding limits ● of agreement (LoA) of -12.20 ± 437.84.
- . Secondary endpoint: Dice score comparison (see Table 2) between the annotator ground truth and automatic outp…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.