K242175 · Xod, Inc. · PBX · Feb 27, 2025 · General, Plastic Surgery
Device Facts
Record ID
K242175
Device Name
XOD Diathermia Radiofrequency Device
Applicant
Xod, Inc.
Product Code
PBX · General, Plastic Surgery
Decision Date
Feb 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The XOD diathermia radiofrequency device is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. The XOD device also is intended to provide a temporary reduction in the appearance of cellulite.
Device Story
Device generates high-frequency sinusoidal current (442kHz-454MHz) for monopolar RF therapy; consists of portable power console with touchscreen, resistive threaded electrodes, and silicon return plate. Therapist manipulates handheld electrode while fixed electrode contacts patient to close electrical circuit. Used in clinical settings to elevate tissue temperature (40-45°C). Output provides therapeutic heating; affects clinical decision-making by facilitating pain relief, muscle spasm reduction, and improved circulation. Benefits include non-invasive treatment for specified conditions and temporary cellulite reduction.
Clinical Evidence
Bench testing only. Device performance verified against design specifications. Electrical safety testing per IEC 60601-1 and IEC 60601-1-6; EMC testing per IEC 60601-1-2; biocompatibility assessment per ISO 10993-1.
Technological Characteristics
Portable console with color touchscreen; resistive threaded electrodes; silicon return plate. Energy source: high-frequency sinusoidal current (442kHz-454MHz). Monopolar application. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1/1-6; EMC per IEC 60601-1-2.
Indications for Use
Indicated for patients requiring topical heating for relief of pain, muscle spasms, and increased local circulation, or temporary reduction in the appearance of cellulite.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 27, 2025
XOD Inc. % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172
Re: K242175
Trade/Device Name: XOD Diathermia Radiofrequency Device Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: PBX Dated: January 21, 2025 Received: January 21, 2025
Dear Angela Blackwell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James H. Digitally signed by
James H. Jang -S
Date: 2025.02.27
Jang -S
13:38:15-05'00'
For Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242175
Device Name
XOD Diathermia Radiofrequency Device
### Indications for Use (Describe)
The XOD diathermia radiofrequency device is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. The XOD device also is intended to provide a temporary reduction in the appearance of cellulite.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### K242175
## XOD Diathermia Radiofrequency Device 510K Summary February 25, 2025
#### Name and Address: XOD Inc.
3245 Amber St. Philadelphia, PA 19134 Contact Person: Ron Gardi Email: info@xod.life Telephone: 888-460-9042
Name of device: XOD Diathermia Radiofrequency Device Classification Name: massager, vacuum, radio induced heat CFR: 21 CFR 878.4400 Primary Product Code: PBX
#### Submission Contact:
Angela Blackwell Blackwell Device Consulting P.O. Box 718 Gresham, OR 97030-0172 (704)450-9934 angela@blackwelldevice.com
Device Description: The XOD diathermia radiofrequency device generates high frequency sinusoidal current with a monopolar mode of application using two electrodes. A fixed electrode is placed in contact with the patient and a handheld electrode is manipulated by a therapist. When both electrodes are in contact with a patient the electrical circuit is closed, and RF therapy can be provided. The device can be operated in a resistive monopolar mode.
The product consists of a portable power console with a color touchscreen, and accessories including resistive threaded electrodes, and a silicon return plate. The unit can be adjusted to provide various levels of treatment frequency ranging from 442kHz to 454MHz.
Indications for Use: The XOD diathermia radiofrequency device is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. The XOD device also is intended to provide a temporary reduction in the appearance of cellulite.
Testing Summary: EMC testing shows the device meets IEC 60601-1-2. Electrical safety testing shows the device meets IEC 60601-1 and IEC 60601-1-6. The device components which contact the patient or the user were evaluated according to ISO 10993-1. Performance testing shows the device meets the design specifications.
Predicate Devices: K161458 Indiba Diathermia Radiofrequency Device
{5}------------------------------------------------
## Substantial Equivalence:
XOD's diathermia radiofrequency device is substantially equivalent to Indiba's because they have the same indications, device description, the same principle of operation and similar operating parameters.
| | XOD<br>Diathermia<br>Radiofrequency<br>Device | Indiba Diathermia<br>Radiofrequency<br>Device K161458<br>Predicate Device |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | PBX | PBX |
| Indications for Use | The XOD<br>diathermia<br>radiofrequency<br>device is<br>intended to<br>provide topical<br>heating for the<br>purpose of<br>elevating tissue<br>temperature for<br>the treatment of<br>selected medical<br>conditions such<br>as relief of pain,<br>muscle spasms,<br>and increase in<br>local circulation.<br>The XOD device<br>also is intended<br>to provide a<br>temporary<br>reduction in the<br>appearance of<br>cellulite. | The Indiba<br>Diathermia<br>Radiofrequency<br>Devices are<br>intended to<br>provide topical<br>heating for the<br>purpose of<br>elevating tissue<br>temperature for<br>treatment of<br>selected<br>medical conditions<br>such as: relief of<br>pain, muscle<br>spasms, increase<br>in local<br>circulation.<br>The massage<br>device provided is<br>intended to<br>provide a<br>temporary<br>reduction in the<br>appearance of<br>cellulite. |
| Device Description | The XOD<br>diathermia<br>radiofrequency<br>device generates<br>high frequency<br>sinusoidal<br>current with a<br>monopolar mode | The Indiba<br>Diathermia<br>Radiofrequency<br>Device is a<br>therapeutic device<br>for deep,<br>non-invasive<br>diathermy The |
| of application | device consists of | |
| using two | a console which | |
| electrodes. A | generates a | |
| fixed electrode is | radiofrequency | |
| placed in contact | current which is | |
| with the patient | delivered to the | |
| and a handheld | patient, in | |
| electrode is | monopolar form, | |
| manipulated by a | through two | |
| therapist. When | different types of | |
| both electrodes | electrodes: | |
| are in contact | stainless steel | |
| with a patient the | conductive | |
| electrical circuit | resistive | |
| is closed, and RF | electrodes, and | |
| therapy can be | thin-layer | |
| provided. The | insulated | |
| device can be | capacitive | |
| operated in a | electrodes. The | |
| resistive | electrodes are | |
| monopolar | inserted into a | |
| mode. | handle/handpiece, | |
| The product | one handle for | |
| consists of a | each kind of | |
| portable power | electrode, and the | |
| console with a | handle is | |
| color | connected to the | |
| touchscreen, and | console by means | |
| accessories | of a 2-metre | |
| including | cable. | |
| resistive | In resistive mode | |
| threaded | the system | |
| electrodes, and a | delivers a high- | |
| silicon return | frequency current | |
| plate. The unit | of 448 kHz directly | |
| can be adjusted | to the patient's | |
| to provide | skin surface. In | |
| various levels of | capacitive mode, | |
| treatment | the electrode | |
| frequency | coating creates a | |
| ranging from | layer between the | |
| 442kHz to | electrode and the | |
| 454MHz. | human tissue, | |
| | forming a | |
| | capacitor that | |
| | allows | |
| | a high-frequency | |
| | | |
| | current to pass. | |
| | The high<br>frequency current<br>ranges between<br>400<br>kHz and 449 kHz<br>and is<br>automatically<br>tuned by the<br>equipment<br>according to the<br>patient's<br>impedance.<br>Current returns<br>through the<br>neutral return<br>electrode. The<br>Indiba Diathermia<br>Radiofrequency<br>Device is provided<br>with an<br>electroconductive<br>media which is<br>applied to the<br>patients' skin prior<br>to each treatment<br>session.<br>The RF energy<br>generates a<br>heating profile<br>that produces a<br>moderate<br>temperature<br>rise in the<br>subcutaneous<br>tissue.<br>The temperature<br>on the skin is<br>measured using a<br>separate IR (Infra-<br>red)<br>thermometer and<br>there is an<br>integrated<br>massage device<br>that can be used<br>to | |
| massage the skin during cellulite treatment. |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.