Retrace Ureteral Access Sheath (ASXL10, ASXL12, ACXL10, ACXL12, AXXL10, AXXL12, ALXL10, ALXL12)

K242173 · Coloplast Corp. · FED · Nov 12, 2024 · Gastroenterology, Urology

Device Facts

Record IDK242173
Device NameRetrace Ureteral Access Sheath (ASXL10, ASXL12, ACXL10, ACXL12, AXXL10, AXXL12, ALXL10, ALXL12)
ApplicantColoplast Corp.
Product CodeFED · Gastroenterology, Urology
Decision DateNov 12, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

To establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.

Device Story

ReTrace Ureteral Access Sheath is a sterile, single-use device used in hospital settings during ureteroscopic procedures. It consists of a reinforced polymeric sheath, introducer/dilator, connector, and clip. The device is inserted over a guidewire into the urinary tract; the dilator facilitates gradual passage enlargement and trackability. Once positioned, the dilator is removed, leaving a hollow conduit for the passage of endoscopes and instruments. The sheath features a hydrophilic coating for lubricity, kink-resistant reinforcement, and a radiopaque distal tip. It is operated by physicians to provide access to the upper and lower urinary tract, potentially improving procedural efficiency and patient outcomes by protecting the ureter during repeated instrument exchanges.

Clinical Evidence

No clinical data or animal studies were provided. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), dimensional inspection, simulated use, tensile strength, kink resistance, friction testing, and shelf-life validation.

Technological Characteristics

Materials: PEBA, PTFE, stainless steel, PVC, polycarbonate, Nitinol. Features: Hydrophilic coating, radiopaque distal ring, kink-resistant reinforced tube. Dimensions: 10/12 Fr and 12/14 Fr diameters; 28-55 cm lengths. Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: None (mechanical device).

Indications for Use

Indicated for patients requiring ureteroscopic procedures in a hospital environment to establish a conduit for endoscopes and instruments in the upper and lower urinary tract.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 12, 2024 Coloplast Corp. Jennifer Tvrdik Senior Regulatory Affairs Specialist 1601 West River Road North Minneapolis, Minnesota 55411 Re: K242173 Trade/Device Name: Retrace Ureteral Access Sheath (ASXL10, ASXL12, ACXL10, ACXL12, AXXL10, AXXL12, ALXL10, ALXL12) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: July 24, 2024 Received: October 7, 2024 Dear Jennifer Tvrdik: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2}------------------------------------------------ Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark R. Kreitz -S for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K242173 Device Name Retrace Ureteral Access Sheath (ASXL10, ASXL12, ACXL10, ACXL12, AXXL10, AXXL12, ALXL10, ALXL12) Indications for Use (Describe) To establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 2.03 510(k) SUMMARY ### I. SUBMITTER | 510(K) Owner's Name: | Coloplast A/S | |-----------------------------|------------------------------------------------------------------------------------------------| | Legal Manufacturer Address: | Holtedam 1<br>3050 Humlebaek, Denmark | | Phone/Fax/Email: | Phone: (612) 998-4517<br>Email: usjtvr@coloplast.com | | Name of Contact Person: | Jennifer Tvrdik<br>Sr. Regulatory Specialist | | Address/Contact: | 1601 West River Road North<br>Minneapolis, MN 55411 | | Date Prepared: | July 12, 2024 | | II. DEVICE | | | Trade or Proprietary Name: | ReTrace® Ureteral Access Sheath | | Common or Usual Name: | Endoscopic Access Overtube, Gastroenterology-Urology | | Classification Name: | Endoscope Access Overtube<br>(21 CFR section 876.1500)<br>Product Code: FED<br>Device Class: 2 | | Classification Panel: | Gastroenterology-Urology | ### III. PREDICATE DEVICE | Primary Predicate: | ReTrace Ureteral Access Sheath | |--------------------|--------------------------------| | 510(k) Holder: | Coloplast A/S | | 510(k) Number: | K181811 | This predicate has not been subject to a design-related recall. No reference devices were used in this submission. {5}------------------------------------------------ ### IV. DEVICE DESCRIPTION The ReTrace Ureteral Access Sheath is a sterile, single use device that is inserted over a guidewire into the upper or lower urinary tract of patients requiring ureteroscopic procedure sin a hospital environment. The ReTrace Ureteral Access Sheath consists of: - Reinforced tube/sheath ● - Introducer/dilator ● - Connector ● - Clip ● The reinforced polymeric introducer sheath includes a hydrophilic coating on the exterior and a lubricious inner surface. The sheath is reinforced to provide kink resistance. The distal tip of the sheath is fitted with a radiopaque ring. A white connector is fitted on the proximal end of the sheath and an orange connector is on the proximal end of the introducer to allow the two components to be attached. The dilator is fitted with a Luer connector on the proximal end and has a hydrophilic coating on the distal end. A guidewire entry and exit eye and exit holes for fluid delivery are located at the distal end of the introducer. The dilator protrudes out the distal end and allows for a gradual enlargement of the passage if needed and improves trackability along the guidewire. Once in the desired location, the dilator is removed to establish a hollow conduit for insertion of endoscopes and other instruments into and out of the urinary tract. The ReTrace Ureteral Access Sheaths are available in two diameters 10/12 Fr and 12/14 Fr. Lengths range between 28 and 55 cm. This submission is for modifications to device materials, labeling, packaging, and connector design. #### V. INDICATIONS FOR USE To establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract. ### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE The modified ReTrace Ureteral Access Sheath is substantially equivalent in performance, indication, design, and materials to the previously cleared, ReTrace Ureteral Access Sheath (K181811) as shown in Table 1. {6}------------------------------------------------ | Device Feature/<br>Component | ReTrace Ureteral Access Sheath<br>(Subject Device) | ReTrace Ureteral Access Sheath<br>K181811<br>(Primary Predicate) | |----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Coloplast A/S | Coloplast A/S | | 510 (k) number | 510(k) Pending | K181811 | | Regulation name | Endoscopic Access Overtube | Endoscopic Access Overtube | | Regulation number | 21 CFR 876.1500 | 21 CFR 876.1500 | | Classification<br>product code | FED | FED | | Classification | II | II | | Prescription device | Yes | Yes | | Intended Use | To establish a continuous conduit<br>during urological endoscopic<br>procedures facilitating the in and out<br>passage of endoscopes and other<br>instruments into the urinary tract. | Used to establish a continuous conduit<br>during urological endoscopic procedures<br>facilitating the in and out passage of<br>endoscopes and other instruments into the<br>urinary tract. | | Duration of Use | During procedure <24 hrs | During procedure <24 hrs | | Target Patient<br>Population | Patients requiring ureteroscopic<br>procedures. | Patients requiring ureteroscopic<br>procedures. | | Anatomical sites | Upper and lower urinary tract | Upper and lower urinary tract | | Where used<br>(hospital, home,<br>ambulance, etc.) | Hospital | Hospital | | Components | The ReTrace UAS is comprised of<br>four components:<br>● Reinforced tube/sheath<br>● Introducer/dilator<br>● Connector<br>● Clip | The ReTrace UAS is comprised of four<br>components:<br>● Reinforced tube/sheath<br>● Introducer/dilator<br>● Connector<br>● Clip | | Sizes | 10 FR ID / 12 FR OD<br>12 FR ID / 14 FR OD | 10 FR ID / 12 FR OD<br>12 FR ID / 14 FR OD | | Length | 28 cm through 55 cm | 28 cm through 55 cm | | Performance | Testing of dimensions, device<br>integrity, device functionality,<br>simulated use, guidewire pullout<br>force, kink resistance, injection<br>testing, and shelf-life testing to<br>support 5 years | Testing of dimensions, device integrity,<br>device functionality, simulated use,<br>guidewire pullout force, kink resistance,<br>injection testing, and shelf-life testing to<br>support 5 years | | Insertion<br>Technique | Insertion over a guidewire | Insertion over a guidewire | | Device Materials | PEBA, PTFE, stainless steel, PVC,<br>polycarbonate, Nitinol | PEBA, PTFE, stainless steel, PVC,<br>polycarbonate, Nitinol | | Coating | Hydrophilic coating | Hydrophilic coating | | Single use | Yes | Yes | | Biocompatibility<br>per ISO 10993 | Meets Standard | Meets Standard | | Sterilization<br>method | Ethylene Oxide | Ethylene Oxide | | Device Feature/<br>Component | ReTrace Ureteral Access Sheath<br>(Subject Device) | ReTrace Ureteral Access Sheath<br>K181811<br>(Primary Predicate) | | Sterility Assurance<br>Level | $10^{-6}$ | $10^{-6}$ | | Shelf Life | 5 years | 5 years | # Table 1: Comparison of Technological Characteristics {7}------------------------------------------------ ### VII. PERFORMANCE DATA The following testing data was provided in support of the substantial equivalence determination and changes to the product line. ## Biocompatibility Testing Biocompatibility testing was conducted based upon ISO 10993-1 (2018): Biological evaluation of medical devices – Part 1: "Evaluation and testing within a risk management process" and FDA Guidance for Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" - Guidance for Industry and Food and Drug Administration Staff - September 2020. ## Mechanical Testing Mechanical testing was completed in support of the substantial equivalence determination and changes to the product line. - Visual Inspection ● - Dimensional Inspection ● - Viscous fluid test ● - Simulated use test - Compatibility with accessories ● - Compatibility with Solutions ● - Peelability test - Tensile strength on introducer - Tensile strength between introducer and luer - Tensile strength between connector and sheath ● - Folding resistance on sheath - Radiopacity test on introducer - Friction test on sheath - Friction test on introducer tip ● - No perforation test - Tip flexibility ● - Nitinol tube compression / position ● - Shelf life testing to support 5 years ● ## Sterilization {8}------------------------------------------------ The ReTrace Ureteral Sheaths are sterilized using ethylene oxide in a validated cycle demonstrating a microbial assurance level of 10-6. ## Packaging and Distribution Packaging testing included dye penetration testing, peel strength, peelability and visual inspection including through five year simulated accelerated aging. No animal studies or clinical testing were provided to support substantial equivalence between the subject and predicate device. ## VIII. CONCLUSIONS The modifications to the ReTrace Ureteral Access Sheath have been demonstrated to be substantially equivalent to the primary predicate, ReTrace Ureteral Access Sheath, based on the non-clinical data provided. The test results demonstrate that the design and labeling changes do not raise new questions of safety or effectiveness and are substantially similar to the predicate.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...