The Morph DNA Steerable Introducer Sheath is intended to provide a pathway through which medical instruments, such as balloon dilatation catheters, guidewires, or other therapeutic devices, may be introduced into the peripheral vasculature or chambers and coronary vasculature of the heart.
Device Story
Sterile, single-use bidirectional deflectable catheter; ergonomic handle with integrated swiveling hemostasis valve, brake, and 3-way stopcock. Distal tip includes fluoroscopic marker and vent holes. Used in clinical settings by physicians to provide access to peripheral and coronary vasculature. Handle actuation via rocker mechanism tensions pull wires to deflect distal tip in two directions. Brake holds deflection angle. Device facilitates introduction of therapeutic instruments (e.g., guidewires, balloon catheters). Benefits include improved navigation and access to target anatomy within heart chambers or peripheral vessels.
Clinical Evidence
Bench testing only. Testing included distal tip deflection, dimensional verification, catheter/handle functionality (leakage, torsional rigidity, kink resistance, pull-wire tensile strength), fatigue resistance, and dilator verification. Simulated use testing confirmed navigation, deflection/brake mechanism performance, and device compatibility. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, thromboresistance, complement activation) leveraged from predicate/reference devices.
Technological Characteristics
Materials: PTFE inner lining, Pebax external extrusion jacket. Bidirectional deflection via rocker mechanism and pull wires. Dimensions: 7.5F inner diameter, variable working lengths. Connectivity: None. Sterilization: Ethylene Oxide (ISO 11135). Standards: ISO 10993 (biocompatibility), ISO 10555, ISO 11070, ASTM D4332-14, ASTM D7386-16, ASTM F2096-11.
Indications for Use
Indicated for patients requiring a pathway for the introduction of medical instruments, such as balloon dilatation catheters, guidewires, or other therapeutic devices, into the peripheral vasculature or chambers and coronary vasculature of the heart.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
8F Morph® DNA Deflectable Guide Catheter (K192774)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 23, 2024
BioCardia Peter Altman President and CEO 320 Soquel Way Sunnyvale, California 94085
Re: K242169
Trade/Device Name: Morph DNA Steerable Introducer Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: July 23, 2024 Received: July 24, 2024
Dear Peter Altman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Misti L. Malone -S
Misti Malone, PhD Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242169
Device Name
Morph DNA Steerable Introducer Sheath
Indications for Use (Describe)
The Morph DNA Steerable Introducer Sheath is intended to provide a pathway through which medical instruments, such as balloon dilatation catheters, guidewires, or other therapeutic devices, may be introduced into the peripheral vasculature or chambers and coronary vasculature of the heart.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This summary of 510(k) is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Date Prepared: July 15, 2024
510(k) number: K242169
#### Applicant Information:
BioCardia, Inc. 320 Soquel Way Sunnyvale, CA 94085
#### Contact Person
Edward Gillis Sr. VP of Devices 408 205-0475 egillis@biocardia.com
#### Device Information:
| Trade Name: |
|----------------------|
| Classification: |
| Classification Name: |
| Product Code: |
Morph DNA Steerable Introducer Sheath 870.1340. - Class II Catheter Introducer DYB
| Predicate Device: 8F Morph® DNA Deflectable Guide Catheter | K192774 |
|------------------------------------------------------------|---------|
| Reference Device: Avance 8.5 F Steerable Introducer | K190941 |
| Reference Device: BioCardia Morph® Sheath Guide | K090999 |
# Device Description:
The Morph DNA Steerable Introducer Sheath is sterile, single-use bidirectional deflectable catheter with an ergonomic handle that has an integrated swiveling hemostasis valve. When the handle is fully actuated, the distal tip deflects to a nominal angle in either of two directions based on the diameter and the working length of the catheter. The catheter tip includes a fluoroscopic marker to help visualize the tip location and vent holes to facilitate aspiration and flushing of the lumen. The Morph DNA handle includes a brake to hold the deflection angle, an integrated hemostasis valve with a swiveling sideport, and 3-way stopcock.
Devices are packaged with a compatible dilator on a plastic backing card within a sealed Tyvek pouch, all of which are contained in a product box.
#### Indications for Use:
The Morph DNA Steerable Introducer Sheath is intended to provide a pathway through which medical instruments, such as balloon dilatation catheters, guidewires, or other therapeutic devices, may be introduced into the peripheral vasculature or chambers and coronary vasculature of the heart.
{4}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device:
The Morph DNA Steerable Introducer Sheath is substantially equivalent to the predicate with respect to the following attributes:
- . Intended use
- Classification (870.1250. - Class II)
- Basic Design
- Fundamental Technology
- Method of operation ●
- Inner PTFE lining with external Pebax extrusion jacket
- Rocker mechanism in handle for deflecting catheter tip via pull wire tensioning.
- . Catheter mounted onto inner backer card, sealed by Tyvek pouch, and enclosed in product box
- Sterilization via Ethylene Oxide
- . Compliance with ISO 10993
- Compliance with ISO 10555
- Compliance with ISO 11070 ●
- Compliance with ISO 11135
The specific design differences between Morph DNA Steerable Introducer Sheath and the predicate and reference devices are the following:
- Addition of 7.5F (inner diameter) sizes vs. 5.5F (Morph DNA Deflectable Guide), 8.5F . (Avance) and 6.5F (Morph Sheath Guide)
- Additional working length of 30cm vs. 115cm (Morph DNA Deflectable Guide).71cm ● (Avance), and 45 cm and 90 cm Morph Sheath Guide
- . Minor difference in the dilator, where subject product utilizes straight dilator vs, no dilator in Morph DNA Deflectable Guide and pre-curved tip dilator in Avance. Morph Sheath Guide similarly uses a straight dilator.
- . The maximum tip deflection is greater than predicate and reference devices in shorter lenaths.
- The range for distal curvature is greater than predicate and reference devices.
# Supporting Data/Information:
# Performance
The following tests were conducted to confirm that the modifications made between the Morph DNA Steerable Introducer Sheath and predicate/reference devices did not affect the ability to meet all product specification requirements.
| In-Vitro Bench Top Testing |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Distal Tip – right/left deflection angle, curve distance, residual curvature in neutral<br>position, angular deviation from plane |
| Catheter Dimensional Verification – inner diameter, outer diameter, working length |
| Catheter and Handle Functionality – Freedom from leakage, dilator and HMV<br>connections, torsional rigidity, bend kink resistance, column support, max<br>actuation force, curve size, deflection angle, deflection mechanism robustness,<br>pullwire tensile strength, brake curve retention, push kink resistance |
| Fatigue Resistance – Deflection fatigue resistance, torque fatigue resistance |
| Dilator Verification - Inner diameter, tube joint tensile strength, dilator tube to hub<br>tensile, dilator extension length |
{5}------------------------------------------------
Additional testing results applicable for Morph DNA Steerable Introducer Sheath
# Packaging Integrity
Atmospheric Conditioning ASTM D4332-14 and Shipping Simulation ASTM D7386-16 Test Schedule 3 - Manual Handling, Random Vibration, Low Pressure Hazard, Tip Over, Rotational Edge Drop, Bridge Impact, Concentrated Impact, Gross Leak (Bubble Test) ASTM F2096-11 Leveraged from Morph DNA predicate and Avance reference devices.
| In-Vitro Simulated Use Testing |
|-----------------------------------------------------|
| Catheter Navigation to all intended anatomy |
| Deflection Mechanism |
| Brake Mechanism |
| Device Compatibility to guide therapeutic catheters |
# Biocompatibility
The materials used in the BioCardia Morph DNA Steerable Introducer Sheath meet the requirements of ISO 10993-1 and FDA guidance, "Use of International Standard ISO 10993-1 -Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Results for the following biocompatibility tests were leveraged from the predicate and reference devices:
- Cytotoxicity ●
- Sensitization ●
- . Irritation
- Acute Systemic Toxicity ●
- Material Mediated Pyrogenicity
- Hemolysis
- Thromboresistance
- Complement Activation
# Sterilization
The Morph DNA Steerable Introducer Sheath was adopted into an existing ethylene oxide sterilization validation for BioCardia products per TIR 28:2009.
# Summary:
Based on the intended use, product performance, biocompatibility and sterilization information provided in this notification, the BioCardia Morph DNA Steerable Introducer Sheath has been shown to be substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.