Elvie Stride 2 is a powered breast pump to be used by lactating women to express and collect milk from their breasts. Elvie Stride 2 is intended for a single user.
Device Story
Elvie Stride 2 is a portable, wearable powered breast pump for home use by a single lactating user. Device uses a DC motor to drive a membrane vacuum pump, creating cyclic negative pressure to mimic infant suckling. User selects stimulation or expression modes with 10 suction levels and cycle speeds via pump-body buttons or a mobile application. Device features LED indicators for suction level, time, and battery status. Milk is collected into a cup assembly; a physical diaphragm provides backflow protection. Powered by internal rechargeable lithium-ion batteries. Output is expressed breast milk. Device aids in milk collection for breastfeeding mothers.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including vacuum level verification, backflow protection, use-life testing, battery performance, and electrical/EMC safety testing (IEC 60601-1, IEC 62133-2, IEC 60601-1-11, IEC 60601-1-2).
Technological Characteristics
Powered breast pump; DC motor-driven membrane vacuum pump. Materials: food-grade components (21 CFR 174-177). Power: internal rechargeable lithium-ion battery. Connectivity: mobile application. Control: microcontroller-based. Suction: -35 to -300 mmHg. Cycle speeds: 28-103 cycles/min. Backflow protection: physical diaphragm. Form factor: wearable/portable. Non-sterile.
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for single-user, home-use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 26, 2024
Chiaro Technology Ltd Amelia Ortiz Rios Regulatory Affairs Manager 63-66 Hatton Garden London, EC1N 8LE UNITED KINGDOM
Re: K242125
Trade/Device Name: Elvie Stride 2 Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: July 19, 2024 Received: August 29, 2024
Dear Amelia Ortiz Rios:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242125
Device Name Elvie Stride 2
Indications for Use (Describe)
Elvie Stride 2 is a powered breast pump to be used by lactating women to express and collect milk from their breasts. Elvie Stride 2 is intended for a single user.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary - K242125
#### 1. Submitter Information
| Applicant: | Chiaro Technology Ltd |
|------------|-----------------------------------------------|
| Address: | 63-66 Hatton Garden<br>EC1N 8LE, London<br>UK |
#### 2. Correspondent Information
| Contact: | Amelia Ortiz Rios |
|----------|-----------------------------|
| | Regulatory Affairs Manager, |
| Email: | Amelia.ortiz@chiaro.co.uk |
| Phone: | +44 7 470 359 206 |
3. Date prepared: September 26, 2024
## 4. Device Information
| Device Name: | Elvie Stride 2 |
|--------------------|-----------------------------|
| Common Name: | Powered breast pump |
| Regulation Number: | 21 CFR 884.5160 |
| Regulation Name: | Powered Breast Pump |
| Product Code: | HGX (Pump, Breast, Powered) |
| Regulatory Class: | Class II |
#### 5. Predicate Device Information
| Device Name: | Elvie Stride |
|----------------|-----------------------|
| 510(k) Number: | K210936 |
| Manufacturer: | Chiaro Technology Ltd |
The predicate device has not been subject to a design-related recall.
## 6. Device Description
The Elvie Stride 2 is an electrically powered breast pump to be used in a home environment by a single user. The device is provided non-sterile and can be used on one breast (single pumping) or on both breasts at the same time (double pumping).
The subject devices are capable of expression and stimulation modes with ten associated suction levels for each. The subject device features on/off buttons, level up/down buttons, mode selection buttons, suction level/time LED indicators/battery visual indicator displays. The pumps are powered by internal, non-replaceable, rechargeable lithium-ion batteries which are charged using the included cable. The subject device cannot be operated while plugged into AC power.
The breast pumps use cyclic negative pressure (suction) to mimic the suckling patterns of a feeding infant. A DC motor drives a membrane vacuum pump to generate the suction required to stimulate and express breast milk. The timing of this pattern is dependent upon the suction/speed settings selected by the user and is preprogrammed in the devices. Elvie Stride 2 includes the Elvie Stride hub, including the existing firmware and mobile application functionality.
{5}------------------------------------------------
The device is capable of producing peak suction levels between -35 and -260 mmHg when double pumping. -55 and -300 mmHg when single pumping, and operating at speeds between 52 to 103 cycles per minute for stimulation mode and 28 to 76 cycles per minute for expression mode. There are two available pumping modes with 10 distinct levels of vacuum and cycle speed. The subject device ensures backflow protection between the breast shield and the electronic components via a physical barrier (diaphragm) mechanism.
The subject device includes a 28mm breast shield and 5 separate silicone insert sizes (24, 21, 19, 17 and 15 mm). The breast shield may be used with or without the addition of a nipple insert. The cup assembly includes components which are in contact with breastmilk. All food contacting components are composed of food grade material and have been tested to 21 CFR 174-177.
## 7. Indications for Use
Elvie Stride 2 is a powered breast pump to be used by lactating women to express and collect milk from their breasts. Elvie Stride 2 is intended for a single user.
## 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
## Table 1: Comparator Table for Subject and Predicate Devices
| | Elvie Stride 2<br>K242125<br>Subject Device | Elvie Stride<br>K210936<br>Predicate Device | Comparison |
|------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------|-------------------------------------|
| Product Code | HGX | HGX | Same |
| Regulation Number | 21 CFR 884.5160 | 21 CFR 884.5160 | Same |
| Device Class | Class II | Class II | Same |
| Single User | Yes | Yes | Same |
| Wear Configuration | Mobile/portable for use in bra | Mobile/portable for use in bra | Same |
| Single/Double Breast<br>Pump | Yes | Yes | Same |
| Single/double pump | Single or Double | Single or Double | Different |
| Media Separation<br>(Backflow Protection) | Yes | Yes | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Direct User Contact | Yes, breast shield and inserts | Yes, breast shield | Equivalent |
| Cycling Control<br>Mechanism | Microcontroller | Microcontroller | Same |
| Accessories | Nipple Inserts in 5 sizes:<br>24mm, 21mm, 19mm, 17mm,<br>15 mm | None | Different |
| | Power Supply | Rechargeable Lithium-Ion<br>battery | Rechargeable Lithium-Ion<br>battery |
| Suction Strength<br>(stimulation) (Expression) | Single:-55 to -300 mmHg<br>Double:-35 to -260 mmHg | Single:-55 to -300 mmHg<br>Double:-35 to -260 mmHg | Same |
{6}------------------------------------------------
K242125 Page 3 of 4
| Cycle speed (cycles/min):<br>(Stimulation) | 52 to 103 cycles/min | 31 to 83 cycles/min | Equivalent |
|----------------------------------------------|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|------------|
| Cycle speed (cycles/min):<br>(Expression) | 28 to 76 cycles/min | 18 to 51 cycles/min | Equivalent |
| Suction Levels<br>(stimulation)/(expression) | 10 | 10 | Same |
| Visual Indicator | LED display | LED display | Same |
| User Interface | 5-Buttons on Pump Body:<br>on/off, mode select,<br>play/pause, increase vacuum,<br>decrease vacuum | 5-Buttons on Pump Body:<br>on/off, mode select,<br>play/pause, increase vacuum,<br>decrease vacuum | Same |
| Mobile Application | Yes | Yes | Same |
| Overall design | Tabletop | Tabletop | Same |
| Environment of Use | Home | Home | Same |
The indications for use of the subject and predicate device are identical and they have the same intended use - the expression and collection of breast milk.
The subject and predicate devices have similar technological features, including available number of available vacuum levels, single/double pumping operation, and cycle speed specifications. However, as shown in the table above, there are some technological differences between the subject and predicate device. including differences in cup assembly and inclusion of nipple inserts. However, the different technological characteristics of the subject device do not raise different questions of safety and effectiveness.
## 9. Summary of Non-Clinical Performance Testing
## Biocompatibility
Biocompatibility information was provided in accordance with Attachment G, Biocompatibility of Certain Devices in Contact with Intact Skin, of the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process. "
## Electrical Safety
Testing was leveraged from the predicate device and was conducted in accordance with IEC 60601- 1 Edition 3.2 2020-08 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance), IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other nonacid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems, and IEC 60601-1-11 Edition 3.1 2020-07 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
{7}------------------------------------------------
## Electromagnetic Compatibility
Testing was leveraged from the predicate device and was conducted in accordance with IEC 60601-1-2:2014+A1: 2020 Medical Electrical Equipment - Part 1-2: "General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests. "
## Software
Software was leveraged from the predicate device and was evaluated at the equivalent of Basic Documentation level as recommended in the 2023 FDA guidance document "Content of Premarket Submissions for Device Software Functions."
## Performance Testing
Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:
- Vacuum level verification testing at each mode/cycle demonstrated that the devices ● meet mode/cycle specifications.
- . Backflow protection testing was conducted to verify liquid does not backflow into the tubing.
- Use life testing was conducted to demonstrate that the device maintains its specifications ● throughout its proposed use life.
- Battery performance testing was leveraged from the predicate to demonstrate that the battery remains functional during its stated battery use-life.
- . Battery status indicator testing was leveraged from the predicate to demonstrate that the batterv status indicator remains functional during its stated battery life.
## 10. Conclusion
The results of the performance testing described above demonstrate that the Elvie Stride 2 is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.