AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K242123 · Jan 6, 2025
Brainomix 360 e-CTA
Brainomix Limited
Retrospective clinical CTA brain scans from Boston Medical Centre and Mayo Clinic Rochester
A retrospective study of 308 clinical CTA scans was conducted to assess the standalone performance of the device's vessel delineation algorithm compared to the predicate device, ensuring generalizability across diverse patient demographics, scanner manufacturers, and clinical sites.
Retrospective study of 308 CTA brain scans from Boston Medical Centre and Mayo Clinic Rochester
—
Indications for Use
Brainomix 360 e-CTA is an image processing software package to be used by trained professionals, including, but not limited to physicians and medical technicians. The software runs on standard "off the-shelf" hardware (physical or virtualized) and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM compliant imaging devices. Brainomix 360 e-CTA provides viewing and analysis capabilities for imaging datasets acquired with CTA (CT Angiography). Brainomix 360 e-CTA is not intended for mobile diagnostic use. Brainomix 360 e-CTA vessel density asymmetry ratio applies only to the MCA region.
Device Story
Brainomix 360 e-CTA is a DICOM-compliant medical image processing software package running on off-the-shelf server hardware. It ingests CTA images from hospital PACS; utilizes AI/ML algorithms to distinguish bone from vessels and reduce image grain; performs image registration and vessel delineation. The device outputs include 3D-colored overlays visualizing MCA vessel density asymmetry, contrast intensity measurements, and estimated phase. Used in clinical/hospital environments by radiologists and clinicians to aid in clinical decision-making. Clinicians may receive email notifications with compressed preview images; however, diagnostic decisions require viewing non-compressed images on a diagnostic workstation. The device provides quantitative analysis of hemispheric vessel density differences, assisting physicians in evaluating cerebral vascular status.
Clinical Evidence
Retrospective standalone performance study (N=308) using clinical CTA scans from multiple scanner manufacturers (Siemens, GE, Philips, Canon/Toshiba). Primary endpoints: vessel delineation (Vessels DSC) and parenchyma mask (Parenchyma DSC). Results: Vessels DSC 0.955 (95% CI: 0.953, 0.957); Parenchyma DSC 0.999 (95% CI: 0.999, 1.000). Subgroup analyses performed by age, gender, race, clinical site, scanner manufacturer, and stenosis presence. Digital phantom validation confirmed algorithm performance with reduced Mean Absolute Error (MAE) compared to the predicate.
Technological Characteristics
Software-only medical image management and processing system. Runs on standard off-the-shelf physical or virtual servers. Connectivity via DICOM. Features include image registration, vessel delineation, and 3D-colored opacity-graded overlays. Algorithm utilizes a CNN-based AI/ML model for bone/vessel distinction and grain reduction. Compliant with ISO 14971:2019 (risk management), FDA Cybersecurity Guidance, and IEC 81001-5-1.
Indications for Use
Indicated for use by trained professionals (physicians, medical technicians) to perform viewing, processing, and analysis of DICOM-compliant CT Angiography (CTA) images. Vessel density asymmetry ratio analysis is limited to the MCA region. Not intended for mobile diagnostic use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that.
Brainomix Limited Zsolt Szrnka Regulatory Affairs Manager First Floor, Seacourt Tower West Way Oxford, OX2 0JJ United Kingdom
January 6, 2025
Re: K242123
Trade/Device Name: Brainomix 360 e-CTA Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ, OIH Dated: November 22, 2024 Received: November 27, 2024
Dear Zsolt Szrnka:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K242123
Device Name
Brainomix 360 e-CTA
Indications for Use (Describe)
Brainomix 360 e-CTA is an image processing software package to be used by trained professionals, including, but not limited to physicians and medical technicians. The software runs on standard "off the-shelf" hardware (physical or virtualized) and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM compliant imaging devices.
Brainomix 360 e-CTA provides viewing and analysis capabilities for imaging datasets acquired with CTA (CT Angiography).
Brainomix 360 e-CTA is not intended for mobile diagnostic use.
Brainomix 360 e-CTA vessel density asymmetry ratio applies only to the MCA region.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K242123
Image /page/4/Picture/1 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is a gradient of blue and green, while the letters "AI" in "BRAINOMIX" are also in the same gradient of blue and green, while the rest of the letters are in dark gray.
Head office First Floor, Seacourt Tower, West Way Oxford OX2 0JJ, United Kingdom
# 510(K) Summary Brainomix Limited – Brainomix 360 e-CTA
| Date Prepared: | 26Nov2024 |
|--------------------------|----------------------------------------------------------------------------|
| Applicant's Name: | Brainomix Limited |
| Applicant's Address: | First Floor, Seacourt Tower, West Way<br>Oxford, OX2 0JJ<br>United Kingdom |
| Official Contact: | Zsolt Szrnka<br>+44 (0) 1865 582730<br>regulatory@brainomix.com |
| Device Proprietary Name: | Brainomix 360 e-CTA |
| Device Common Name: | Medical Image Management And Processing System |
| Regulatory Class: | Class II |
| Product Code: | Primary: LLZ, Secondary: QIH |
| Regulation: | 21 C.F.R. §892.2050 |
## 1. Predicate Devices
Brainomix 360 e-CTA is Substantially Equivalent to the following Legally Marketed devices:
Trade Name: Brainomix 360 e-CTA (Primary Predicate) Manufacturer: Brainomix Ltd. Regulation Number: 21 C.F.R. §892.2050 Regulatory Class: Class II Regulation Name: Medical Image Management And Processing System Product Code: LLZ Submission Number: K192692
Trade Name: Rapid (Secondary Predicate) Manufacturer: iSchemaView Regulation Number: 21 C.F.R. §892.2050 Regulatory Class: Class II Regulation Name: Medical Image Management And Processing System Product Code: LLZ, QIH Submission Number: K233582
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, with the word "BRAINOMIX" in capital letters to the right of the icon. The brain icon is a gradient of blue and teal, while the letters "BRA, I" are in a similar gradient, and the letters "NOMIX" are in a dark gray color.
## 2. Device Description
Brainomix 360 e-CTA is a medical image visualization and processing software package compliant with the DICOM standard and running on an off-the-shelf physical or virtual server.
Brainomix 360 e-CTA allows for the visualization, analysis and post-processing of DICOM compliant CTA images which, when interpreted by a trained physician or medical technician, may yield information useful in clinical decision making.
Brainomix 360 e-CTA provides a wide range of basic image viewing, processing and manipulation functions, through multiple output formats. Functionality includes image registration and visualization of large cerebral vessels to provide an analysis of hemispheric difference via contralateral comparison (displayed as a relative percentage).
Brainomix 360 e-CTA processes the images using Al/ML algorithms where the input channels will help the software distinguish bone from vessels and reduce image grain.
Brainomix 360 e-CTA automatically provides a colored overlay to provide a visual reference of the MCA hemisphere of the brain with lower vessel density, and corresponding contrast intensity measurements and estimated phase.
Brainomix 360 e-CTA can connect with other DICOM-compliant devices, for example to transfer CTA scans from a Picture Archiving and Communication System (PACS) to Brainomix 360 e-CTA software for processing. Results and images can be sent to a PACS via DICOM transfer and can be viewed on a PACS workstation or via a web user interface on any machine and accessed within a hospital network and firewall and with a connection to the Brainomix 360 e-CTA software (e.g. a LAN connection).
Brainomix 360 e-CTA notification capabilities enable clinicians to preview images via e-mail notification with result image attachments. Images that are previewed via e-mail are compressed, are for informational purposes only, and not intended for diagnostic use beyond notification.
Brainomix 360 e-CTA is not intended for mobile diagnostic use. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests.
# 3. Intended Use / Indications for Use
Brainomix 360 e-CTA is an image processing software package to be used by trained professionals, including, but not limited to physicians and medical technicians. The software runs on standard "off the-shelf" hardware (physical or virtualized) and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM compliant imaging devices.
Brainomix 360 e-CTA provides viewing and analysis capabilities for imaging datasets acquired with CTA (CT Angiography).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The "AI" in the word "BRAINOMIX" is in a gradient of blue and green, while the rest of the letters are in a dark gray color.
Brainomix 360 e-CTA is not intended for mobile diagnostic use.
Brainomix 360 e-CTA vessel density asymmetry ratio applies only to the MCA region.
## 4. Performance Data
#### Summary of Digital Phantom Validation 4.1
Digital phantom evaluation was performed to validate the "vessel density ratio value" output as calculated by Brainomix 360 e-CTA.
The tests have passed and demonstrate that Brainomix 360 e-CTA performs following its requirements and that the algorithm as implemented in Brainomix 360 e-CTA performs by the required performance criteria, which is summarized below in Table 1.
| Metric Name | Value | Criteria | Pass/Fail |
|---------------|-------|----------|-----------|
| Left-MAE | 6.444 | <10 | Pass |
| Left-MAE-STD | 9.269 | <15 | Pass |
| Right-MAE | 5.611 | <10 | Pass |
| Right-MAE-STD | 8.610 | <15 | Pass |
Table 1. Vessel density ratio validation results
#### AI/ML Comparison Digital Phantom Validation 4.2
The algorithm version of the primary predicate device (K192692) was compared against that of the proposed device having the AI/ML model add-on to increase the quality of the vessel mask. The metrics are summarized in Table 2.
| Device | Left MAE (%) | Right MAE (%) |
|-------------------------------------|--------------|---------------|
| Predicate device with NO-CNN (n=18) | 7.333 | 6.889 |
| Proposed device with CNN (n=18) | 3.000 | 6.278 |
| Difference between devices | 4.333 > 0 | 0.611 > 0 |
Table 2. Comparison of performance metrics between predicate device and proposed device
The proposed device shows a reduction in the Mean Absolute Error (MAE) for both hemispheres (4.333% left; 0.611% on the right).
These results corroborate that the algorithm approach of the proposed device is safe, effective and does not raise different questions of safety and effectiveness to Brainomix 360 e-CTA while slightly improving its performance.
#### 4.3 Summary of Standalone Performance Study
A retrospective study has been carried out to assess the standalone performance of the processing step of vessel delineation in Brainomix 360 e-CTA in comparison to the predicate (Brainomix 360 e-CTA, K192692).
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in capital letters. The brain icon is a gradient of light blue to dark blue. The letters "AI" in "BRAINOMIX" are also in a similar blue gradient, while the rest of the letters are in a dark gray color.
A sample size of 308 Computed Tomography (CTA) brain scans (studies) were obtained from clinical sites in the U.S. The majority of studies came from Boston Medical Centre (BMC), or from referring hospitals in the Massachusetts area (N=179). The remaining studies came from Mayo Clinic Rochester (MCR; N=129). The patient cohort was enriched to ensure the distribution of clinical and demographic variables (e.g., age, gender, race) allows generalizability to the patient population for whom use is intended.
The imaging was collected on a variety of different scanner models and manufacturers (Table 8). The dataset included 133 cases from SIEMENS scanners, 96 from GE scanners, 76 from Philips scanners and 4 from Canon/Toshiba scanners.
The slice thickness distribution was median 0.75mm [IQR: 0.625 – 0.8] with a minimum of 0.5mm (n=2) and a maximum of 1mm (n=42). In-plane (axial) resolution showed a range of 0.381mm to 0.72mm with a median of 0.49mm. Four different tube voltage (KvP, peak kilovoltage) values were encountered in the dataset (90KvP n=6, 100KvP, n=64; 120KvP n=137 and 140KvP, n=99). Tube current was median of 526mA [IQR: 445mA - 645mA]. With respect to contrast phase during acquisition, 62 studies were Early Arterial (EA), 130 Peak Arterial (PA), 110 in Equilibrium phase (EQ), 3 in Peak Venous (PV), and 3 in Late Venous (LV). All the images used a standard or a soft (low-pass filter) convolution kernel for reconstruction (e.g., STANDARD in GE MEDICAL SYSTEMS).
The cases (n=308) were all successfully processed with the algorithm. The global results are shown below in Table 3, the Dice Similarity Score is 0.955 reaching for the vessel delineation and 0.999 for the parenchyma mask as the desired requirement.
Table 3. Summary performance metrics from full sample of 308 patients, with 95% confidence intervals where approgriate.
| Metrics | All Cases (N=308) |
|----------------|----------------------|
| Vessels DSC | 0.955 (0.953, 0.957) |
| Parenchyma DSC | 0.999 (0.999, 1.000) |
In order to assess the generalization of the performance in different patient subgroups, further stratifications were derived from the complete dataset by age (Table 4), by gender (Table 5), by race (Table 6), by clinical site (Table 7), by scanner manufacturer (Table 8), by the presence of stenosis (Table 9).
Table 4. Summary performance metrics from patients sub-categorized by age, with 95% confidence intervals where appropriate.
| Metrics | Age 22-50 | Age 50-70 | Age 70+ |
|----------------|----------------------|----------------------|----------------------|
| Total N | 57 | 132 | 119 |
| Vessels DSC | 0.958 (0.953, 0.962) | 0.959 (0.957, 0.962) | 0.949 (0.946, 0.953) |
| Parenchyma DSC | 0.999 (0.999, 1.000) | 0.999 (0.999, 1.000) | 0.999 (0.999, 1.000) |
Table 5. Summary performance metrics from patients sub-categorized by gender, with 95% confidence intervals where appropriate.
| Metrics | Male | Female |
|-------------|----------------------|----------------------|
| Total N | 140 | 168 |
| Vessels DSC | 0.952 (0.949, 0.955) | 0.958 (0.955, 0.960) |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain image on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain image is a gradient of light blue to white. The letters "B", "R", "O", "M", "X" are dark gray, while the letters "A", "I", "N" are a gradient of dark gray to light blue.
#### Parenchyma DSC 0.999 (0.999, 1.000) 0.999 (0.999, 1.000)
Table 6. Summary performance metrics from patients sub-categorized by race/ethnicity, with 95% confidence intervals where appropriate. * CI could not be calculated due to lack of evidence (low N).
| Metrics | White | Black/African<br>American | Hispanic/<br>Latino | Asian/Asian<br>American | Unknown/<br>Refused |
|-------------------|-------------------------|---------------------------|-------------------------|-------------------------|-------------------------|
| Total N | 148 | 85 | 27 | 7 | 40 |
| Vessels DSC | 0.954 (0.951,<br>0.956) | 0.957 (0.953,<br>0.961) | 0.957 (0.949,<br>0.964) | 0.958 (0.948,<br>0.967) | 0.955 (0.951,<br>0.959) |
| Parenchyma<br>DSC | 0.999 (0.999,<br>1.000) | 0.999 (0.999,<br>1.000) | 0.999 (0.999,<br>1.000) | 0.999 (0.999,<br>1.000) | 0.999 (0.999,<br>1.000) |
Table 7. Summary performance metrics from patients sub-categorized by clinical site, with 95% confidence intervals where appropriate. MCR = Mayo Clinic Rochester, BMC = Boston Medical Center.
| Metrics | MCR | BMC |
|----------------|----------------------|----------------------|
| Total N | 129 | 179 |
| Vessels DSC | 0.952 (0.949, 0.956) | 0.957 (0.955, 0.959) |
| Parenchyma DSC | 0.999 (0.999, 1.000) | 0.999 (0.999, 1.000) |
Table 8. Summary performance metrics from patients sub-categorized by scanner manufacturer, with 95% confidence intervals where appropriate.
| Metrics | SIEMENS | GE Medical | Philips | Canon/ Toshiba |
|-------------------|----------------------|----------------------|-------------------------|----------------------|
| | | Systems | | |
| Total N | 133 | 96 | 75 | 4 |
| Vessels DSC | 0.953 (0.95, 0.956) | 0.954 (0.95, 0.957) | 0.961 (0.958,<br>0.964) | 0.964 (0.947, 0.981) |
| Parenchyma<br>DSC | 0.999 (0.999, 1.000) | 0.999 (0.999, 1.000) | 0.999 (0.999,<br>1.000) | 0.999 (0.999, 1.000) |
Table 9. Summary performance metrics from patients with or without stenosis as noted by the truthers, with 95% confidence intervals where appropriate.
| Metrics | With Stenosis | Without Stenosis |
|----------------|----------------------|----------------------|
| Total N | 30 | 278 |
| Vessels DSC | 0.957 (0.952, 0.962) | 0.955 (0.953, 0.957) |
| Parenchyma DSC | 0.999 (0.999, 1.000) | 0.999 (0.999, 1.000) |
The standalone performance testing demonstrated that the proposed device provides accurate vessel and parenchyma delineation masks against the predicate device (head-to-head comparison) under a range of clinically relevant variables associated with the intended use of the software. These results corroborate that the algorithm approach of the proposed device adds no risks to Brainomix 360 e-CTA while slightly improving its performance.
### 5. Prescriptive Statement
Caution: Federal law restricts this device to sale by or on the order of a physician.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is a gradient of blue and teal, while the letters "A" and "I" in "BRAINOMIX" are also in the same gradient, providing a visual connection between the icon and the text.
## 6. Safety and Effectiveness
Brainomix 360 e-CTA has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with ISO 14971:2019 (risk management).
# 7. Cybersecurity
Brainomix 360 e-CTA has been designed and validated to the FDA Cybersecurity Guidance and IEC 81001-5-1.
## 8. Summary of Technological Characteristics
Brainomix 360 e-CTA includes similar CTA processing features and technological characteristics as compared to the predicate devices. Where the proposed device differs from the predicate devices, is that it provides three new claims listed below:
- 1. Algorithm: The proposed device introduces an Al/ML algorithm which was not present in the primary predicate (Brainomix e-CTA). The input channels are intended to help the software distinguish bone from vessels which leads to improved image quality (less grain) as well as improved vessel detection, which has been extensively validated through digital phantom testing. In summary, the phantom validation results show a reduction in the Mean Absolute Error (MAE) for both hemispheres (4.333% left; 0.611% on the right), which corroborates that the proposed device is safe and effective and does not raise different questions of safety and effectiveness. Furthermore, the use of Al/ML algorithms is similar to the technological characteristics of the secondary predicate (Rapid).
- 2. Accessible MIP views: The proposed device introduces posterior views to be accessed through Brainomix 360 e-CTA which was not present in the primary predicate (Brainomix 360 e-CTA). The current MIP views of the primary predicate device (Brainomix 360 e-CTA) already display posterior vessels where this new added view just enables focused visualization of posterior large vessels. Furthermore, there are no separate density analysis performed on the posterior MIPs and therefore this change does not raise different questions of safety and effectiveness.
- 3. 3D-colored overlay: The proposed device introduces a 3-dimensional opacity graded overlay to the primary predicate device (Brainomix 360 e-CTA) that helps to visualize the asymmetry in blood vessel density between MCA hemispheres which is calculated via contralateral comparison. This colored overlay is a similar technological characteristic when compared to the secondary predicate device's (Rapid) 2D-colored overlay. The proposed 3D-colored overlay compares to the secondary predicate in that:
- Both the outputs 3D-colored overlay in the subject device and the 2D-colored overlay of a. the secondary predicate device are similar as they are a mathematical quantification of image outputs displayed in percentage that provides a reference for the clinician. Furthermore, the intended user is responsible to follow the standard of care and to engage in appropriate patient evaluation based on relevant clinical background before making care-related decisions when using both the proposed device and secondary predicate
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "BRAINOMIX" in a sans-serif font. The first letter, "B", is in dark gray, while the "A" is in a gradient of blue and green. To the left of the word is a stylized image of a brain, also in a gradient of blue and green.
device. Where the features differ is that the proposed overlay (3D-colored overlay) uses a transparency scale instead of a multi-colored scale (secondary predicate device).
- b. The proposed overlay (3D-colored overlay) is defined as such given the user's ability to view it through different slices of the DICOM image. However, per slice it is presented as a 2D overlay. The presentation of the outputs of the regions of asymmetry in a transparency scale as compared to a multi-colored scale would not change how the user interacts with the device and the intended use of the overlay and therefore, does not raise new questions of safety and effectiveness.
The proposed device (Brainomix 360 e-CTA) and the secondary predicate (Rapid) include similar CTA processing features and technological characteristics as compared to the secondary predicate device. Where technological differences exist, these are predominantly where the predicate device offers more features, functionality and a broader set of imaging indications as compared to the proposed device. The differences in technological characteristics for the proposed device do not raise any new or unanswered questions of safety or effectiveness.
## 9. Substantial Equivalence
| Characteristics/<br>Parameter | Proposed Device<br>Brainomix 360 e-CTA<br>(K242123) | Primary Predicate Device<br>Brainomix 360 e-CTA<br>(K192692) | Secondary Predicate Device<br>iSchemaView Rapid<br>(K233582) |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K242123 | K192692 | K233582 |
| Product Code | LLZ, QIH | LLZ | LLZ, QIH |
| Regulation<br>Number | 21 CFR §892.2050 | 21 CFR §892.2050 | 21 CFR §892.2050 |
| Intended<br>Use/Indications<br>for Use | Brainomix 360 e-CTA is an<br>image processing software<br>package to be used by trained<br>professionals, including, but<br>not limited to physicians and<br>medical technicians. The<br>software runs on standard "off<br>the-shelf" hardware (physical<br>or virtualized) and can be used<br>to perform image viewing,<br>processing, and analysis of<br>images. Data and images are<br>acquired through DICOM<br>compliant imaging devices.<br>Brainomix 360 e-CTA provides<br>viewing and analysis<br>capabilities for imaging<br>datasets acquired with CTA (CT<br>Angiography). | Brainomix 360 e-CTA is an<br>image processing software<br>package to be used by trained<br>professionals, including, but<br>not limited to physicians and<br>medical technicians. The<br>software runs on standard "off<br>the-shelf" hardware (physical<br>or virtualized) and can be used<br>to perform image viewing,<br>processing, and analysis of<br>images. Data and images are<br>acquired through DICOM<br>compliant imaging devices.<br>Brainomix 360 e-CTA provides<br>viewing and analysis<br>capabilities for imaging<br>datasets acquired with CTA (CT<br>Angiography). | Rapid is an image processing<br>software package to be used by<br>trained professionals, including<br>but not limited to physicians<br>(medical analysis and decision<br>making) and medical<br>technicians (administrative case<br>processing). The software runs<br>on a standard off-the-shelf<br>computer or a virtual platform,<br>such as VMware, and can be<br>used to perform image viewing,<br>processing, and analysis of<br>images. Data and images are<br>acquired through DICOM<br>compliant imaging devices.<br>Rapid is indicated for use in<br>Adults only. Rapid provides both<br>viewing and analysis capabilities<br>for functional and dynamic<br>imaging datasets acquired with<br>CT, CT Perfusion (CTP), CT<br>Angiography (CTA), C-arm CT |
| Characteristics/<br>Parameter | Proposed Device<br>Brainomix 360 e-CTA<br>(K242123) | Primary Predicate Device<br>Brainomix 360 e-CTA<br>(K192692) | Secondary Predicate Device<br>iSchemaView Rapid<br>(K233582) |
| | Brainomix 360 e-CTA is not<br>intended for mobile diagnostic<br>use. | Brainomix 360 e-CTA is not<br>intended for mobile diagnostic<br>use. | Perfusion and MRI including a<br>Diffusion Weighted MRI (DWI)<br>Module and a Dynamic Analysis<br>Module (dynamic contrast-<br>enhanced imaging data for MRI,<br>CT, and C-arm CT). Rapid C-arm<br>CT Perfusion can be used to<br>qualitatively assess cerebral<br>hemodynamics in the<br>angiography suite. The CT<br>analysis includes NCCT maps<br>showing areas of hypodense<br>and hyperdense tissue. The DWI<br>Module is used to visualize local<br>water diffusion properties from<br>the analysis of diffusion -<br>weighted MRI data. The<br>Dynamic Analysis Module is<br>used for visualization and<br>analysis of dynamic imaging<br>data, showing properties of<br>changes in contrast over time.<br>This functionality includes<br>calculation of parameters<br>related to tissue flow<br>(perfusion) and tissue blood<br>volume. Rapid CT Perfusion and<br>Rapid MR Perfusion can be used<br>by physicians to aid in the<br>selection of acute stroke<br>patients (with known occlusion<br>of the intracranial internal<br>carotid artery or proximal<br>middle cerebral artery).<br>Instructions for the use of<br>contrast agents for this<br>indication can be found in<br>Appendix A of the User's<br>Manual. Additional information<br>for safe and effective drug use is<br>available in the product-specific<br>iodinated CT and gadolinium-<br>based MR contrast drug<br>labeling. In addition to the<br>Rapid imaging criteria, patients<br>must meet the clinical<br>requirements for<br>thrombectomy, as assessed by |
| | Brainomix 360 e-CTA vessel<br>density asymmetry ratio<br>applies only to the MCA region. | | |
| Characteristics/<br>Parameter | Proposed Device<br>Brainomix 360 e-CTA<br>(K242123) | Primary Predicate Device<br>Brainomix 360 e-CTA<br>(K192692) | Secondary Predicate Device<br>iSchemaView Rapid<br>(K233582) |
| | | | the following contraindications<br>or exclusions:<br>Bolus Quality: absent or<br>inadequate bolus. Patient Motion: excessive<br> |
| Environment of<br>use | Clinical/Hospital environment | Same | Same |
| Energy used<br>and/or delivered | None - software only<br>application. The software<br>application does not deliver or<br>depend on energy delivered to<br>or from patients | Same | Same |
| Primary Users | Radiologist/Clinician | Same | Same |
| Design:<br>Supported<br>Modalities for<br>image processing<br>and visualization | CTA only | CTA only | CT [NCCT, CT, CTA, C-arm<br>CT(CBCT)] or MR (MR, MRA) |
| MIP Views<br>accessible | Maximum intensity projections<br>(MIPs) of the vascular system<br>in axial, coronal, left and right<br>hemisphere sagittal, and<br>posterior views | Maximum intensity projections<br>(MIPs) of the vascular system<br>in axial, coronal and left and<br>right hemisphere sagittal views | Maximum intensity projections<br>(MIPs) of the vascular system in<br>axial, coronal, left and right<br>hemisphere sagittal. |
| Technical<br>Implementation | Mixed Traditional and AI/ML | Traditional | Mixed Traditional and AI/ML |
| Image Overlay | Vessel Density with visual<br>representation (3D-colored | No | Vessel Density with visual<br>representation (2D-colored<br>overlay) |
| Characteristics/<br>Parameter | Proposed Device<br>Brainomix 360 e-CTA<br>(K242123) | Primary Predicate Device<br>Brainomix 360 e-CTA<br>(K192692) | Secondary Predicate Device<br>iSchemaView Rapid<br>(K233582)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.