K242114 · Icu Medical, Inc. · FRN · Apr 2, 2025 · General Hospital
Device Facts
Record ID
K242114
Device Name
Plum Solo™ Precision IV Pump
Applicant
Icu Medical, Inc.
Product Code
FRN · General Hospital
Decision Date
Apr 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Plum Solo™ Precision IV Pump is intended for administration of parenteral fluids, medications, and whole blood and blood products through the following routes of administration: intravenous, intra-arterial, and subcutaneous. The Plum Solo™ Precision IV Pump is intended for use in clinical environments in the hospital and other outpatient healthcare facilities by licensed healthcare professionals. These healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use. The Plum Solo™ Precision IV Pump is intended for adults, pediatric (including infants and children), and neonatal patient populations.
Device Story
Plum Solo™ Precision IV Pump is a single-channel large volume infusion pump (LVP) for hospital and outpatient use. It accepts dedicated PlumSet™ administration sets with cassettes to deliver fluids from primary/secondary containers. Operates via volumetric piston mechanism; delivers fluids serially (piggyback) or concurrently. Features include continuous, multistep, loading dose, and bolus delivery modes. Clinicians program concentration, flow rate, and volume-to-be-infused (VTBI) via a 7-inch color touchscreen. Device integrates with ICU Medical LifeShield™ Infusion Safety Software Suite for drug library management. Enhanced motor mechanism improves flow rate accuracy to +/- 3% (TIR 101 standard). Sensors monitor air, occlusion, temperature, and cassette presence. Output includes visual/audible alarms for safety. Benefits include precise medication delivery, automated piggyback flushing, and configurable intermittent infusion settings to reduce manual errors. Device is pole-mounted; powered by universal AC supply or internal Lithium Iron Phosphate battery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including design verification, human factors validation, reliability testing, software validation, electrical safety, and electromagnetic compatibility testing.
Indicated for administration of parenteral fluids, medications, whole blood, and blood products via IV, intra-arterial, or subcutaneous routes. Intended for use by trained licensed healthcare professionals in hospital and outpatient clinical environments. Applicable to adult, pediatric (including infants and children), and neonatal patient populations.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 2, 2025
ICU Medical Inc
Yuliya Matlin
Sr. Director of Regulatory Affairs
600 N. Field Drive
Lake Forest, Illinois 60045
Re: K242114
Trade/Device Name: Plum Solo™ Precision IV Pump
Regulation Number: 21 CFR 880.5725
Regulation Name: Infusion Pump
Regulatory Class: Class II
Product Code: FRN
Dated: March 4, 2025
Received: March 5, 2025
Dear Yuliya Matlin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242114 - Yuliya Matlin
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K242114 - Yuliya Matlin
Page 3
Sincerely,
Rebecca Dorsey -S
For
Jake Lindstrom, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K242114
Device Name
Plum Solo™ Precision IV Pump
Indications for Use (Describe)
The Plum Solo™ Precision IV Pump is intended for administration of parenteral fluids, medications, and whole blood and blood products through the following routes of administration: intravenous, intra-arterial, and subcutaneous.
The Plum Solo™ Precision IV Pump is intended for use in clinical environments in the hospital and other outpatient healthcare facilities by licensed healthcare professionals. These healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use.
The Plum Solo™ Precision IV Pump is intended for adults, pediatric (including infants and children), and neonatal patient populations.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
K242114
icumedical
human connections
# 510(k) Summary – K242114
This 510(k) Summary is prepared in accordance with the requirements of 21 CFR Part 807.92.
| Submitter Information | |
| --- | --- |
| Name | ICU Medical, Inc |
| Address | 600 North Field Drive, Lake Forest, IL 60045 |
| Phone Number | 224- 706-2419 |
| Name Establishment Registration Number | 3013319212 |
| Name of contact person | Yuliya Matlin, MS, MBA |
| Date prepared | March 26, 2025 |
| Name of device | |
| Trade or proprietary name | Plum Solo™ Precision IV Pump |
| Common or usual name | Infusion Pump |
| Classification | II |
| Classification Reason | 21 CFR 880.5725 |
| Panel | 80 |
| Product Code(s) | FRN |
| Legally marketed device(s) to which equivalence is claimed | Plum Duo™ Infusion System (K223607) |
| Reason for 510(k) submission | The submission is the traditional pre-market notification for next generation of Plum family devices Plum Solo™ Precision IV Pump |
| Device description | The Plum Solo™ Precision IV Pump is a large volume pump (LVP) with one pump channel that can deliver fluid to a patient from 1 to 2 lines independently. The Plum Solo™ Precision IV Pump can only be used with dedicated PlumSet™ administration sets (not subject of this filing). The pump channel accepts a cassette that is part of a PlumSet™ administration set and can connect to a primary and/or secondary container. The fluid is delivered from the upstream lines either serially (piggyback) or concurrently through the cassette to the downstream line. The flow rate accuracy precision has been improved (lower allowed variance) by implementing the new motor mechanism, as well as increased precision for programming concentration, flow rate and VTBI entries. The overall delivery accuracy of the system has improved to +/- 3% per TIR 101 standard condition testing and +/-5% for non- standard conditions. |
| Intended Use of Device/Indication for Use | The Plum Solo™ Precision IV Pump is intended for administration of parenteral fluids, medications, and whole blood and blood products through the following routes of administration: intravenous, intra-arterial, and subcutaneous. The Plum Solo™ Precision IV Pump is intended for use in clinical environments in the hospital and other outpatient healthcare facilities by licensed healthcare professionals. These healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use. The Plum Solo™ Precision IV Pump is intended for adults, pediatric (including infants and children), and neonatal patient populations. |
{5}
K242114
Page 2 of 7
icumedical
human connections
| Summary of the technological characteristics of the device compared to the predicate device | | | |
| --- | --- | --- | --- |
| Characteristic | Subject Device | Predicate (K223607) | Comparison |
| Device Name | Plum Solo™ Precision IV Pump | Plum Duo™ Infusion System | N/A |
| Type of Pump | Large Volume Infusion Pump | | Same |
| Intended Use/Indications for Use | The Plum Solo™ Precision IV Pump is intended for administration of parenteral fluids, medications, and whole blood and blood products through the following routes of administration: intravenous, intra-arterial, and subcutaneous.
The Plum Solo™ Precision IV Pump is intended for use in clinical environments in the hospital and other outpatient healthcare facilities by licensed healthcare professionals. These healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use.
The Plum Solo™ Precision IV Pump is intended for adults, pediatric (including infants and children), and neonatal patient populations. | The Plum Duo Infusion System is intended for parenteral fluids and medications through clinically acceptable routes (limited to intravenous, intra-arterial, and subcutaneous therapies). The Plum Duo Infusion System is intended for adult, pediatric (including infants and children), and neonatal patient populations. | Substantially equivalent |
| Patient Population | The Plum Solo Precision IV Pump is intended for adult, pediatric (including infants and children), and neonatal patient populations. | The Plum Duo Infusion System is intended for adult, pediatric (including infants and children), and neonatal patient populations. | No change to the intended population. |
| Environment of Use | Hospital environments and other outpatient healthcare facilities that exclude hyperbaric or oxygen-rich environments. Not for MRI environment. The Plum Solo Precision IV Pump is intended for use in clinical environments in the hospital and other outpatient healthcare facilities by licensed healthcare professionals. These healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use. | | Same |
| Route of Administration | Intravenous, intra-arterial, and subcutaneous. | Parenteral, (limited to intravenous, subcutaneous, and intra-arterial therapies). | Same |
| Set Compatibility | Compatible with currently marketed Plum™ Administration Sets | | Same |
{6}
K242114
Page 3 of 7
icumedical
human connections
| Summary of the technological characteristics of the device compared to the predicate device | | | |
| --- | --- | --- | --- |
| Characteristic | Subject Device | Predicate (K223607) | Comparison |
| ICU Medical Safety Software/Drug Library | Compatible with ICU Medical LifeShield™ Infusion Safety Software Suite | | Same |
| Drug Library | Custom Drug Library (CDL) (automatically opt-in upon start of programming) No Default Drug Library (DDL) | | Same |
| Delivery Methods/Therapies | 1. Continuous
2. Multistep
3. Loading Dose (standalone or with underlying continuous therapy)
4. Bolus (standalone or with underlying continuous therapy) | | Same |
| Therapy Modes | 1. Piggyback
• optional flush feature
• optional Infuse to Empty feature for medications configured as intermittent in the drug library
2. Concurrent
• optional disablement of VTBI Complete alarm for medications configured as intermittent in the drug library
3. Deliver alone (primary delivery only) | 1. Piggyback (optional flush feature)
2. Concurrent
3. Deliver alone (primary delivery only) | Substantially Equivalent
The addition of the piggyback optional Infuse to Empty feature is an automation of the existing manual workflow and ensures the patient receives the entire piggyback medication at the programmed rate, including the volume of the upstream tubing.
The addition of the concurrent optional disablement of VTBI Complete alarm for medications configured as intermittent in the drug library allows the clinician to disable the VTBI Complete Alarm when no further secondary infusion in concurrent mode is required.
No impact on Safety and Effectiveness. |
| Rate Ranges | • 0.1 to 9.99 mL/hr (in 0.01 mL/hr increments)
• 10. to 99.9 mL/hr (in 0.1 mL/hr increments)
• 100 to 999 mL/hr (in 1 mL/hr increments) | • 0.1 to 99.9 mL/hr (in 0.1 mL/hr increments)
• 100 to 999 mL/hr (in 1 mL/hr increments) | Substantially equivalent
The higher precision rate entry at the low-rate range gives the clinician greater precision and flexibility to tailor the therapy to meet the patient's specific medical needs. |
| Bolus and Loading Dose Rate Ranges | • 0.1 to 9.99 mL/hr (in 0.01 mL/hr increments)
• 10 to 99.9 mL/hr (in 0.1 mL/hr increments)
• 100 to 999 mL/hr (in 1 mL/hr increments) | • 0.1 to 99.9 mL/hr (in 0.1 mL/hr increments)
• 100 to 999 mL/hr (in 1 mL/hr increments) | Substantially equivalent
The higher precision rate entry at the low-rate range gives the clinician greater precision and flexibility to tailor the therapy to meet the patient's specific medical needs. |
{7}
K242114
icumedical
human connections
| Summary of the technological characteristics of the device compared to the predicate device | | | |
| --- | --- | --- | --- |
| Characteristic | Subject Device | Predicate (K223607) | Comparison |
| Flow Rate Delivery Accuracy | • 0.1 - 999 mL/hr: +/- 3%
• 0 to 96 hours of cassette service
Bolus delivery accuracy is +/-5% for delivery volumes as low as 0.1 ml at 999 ml/hr. | • 0.1 - 999 mL/hr: +/- 5%
• 0 to 96 hours of cassette service
Bolus delivery accuracy is +/-5% for delivery volumes as low as 0.1 ml at 999 ml/hr. | Substantially equivalent
Flow rate accuracy improvements provide greater precision of delivery. |
| WiFi | Yes | | Same |
| Pump Operation | Volumetric Piston Type | | Same |
| Number of Channels | 1 Channel: Right channel has primary (Line 1) and secondary (Line 2) | 2 Channels: Left and Right Channel. Each channel has primary (Line 1) and secondary (Line 2) | Substantially equivalent |
| Sensors | Air detection, Occlusion, Temperature, Cassette detection, Ambient light sensors | | Same |
| Alarm Summary | Check cassette alarms, proximal (upstream) and distal (downstream) occlusion, door open, air detection, callback, lockout, low battery, and internal system monitor alarms. | | Substantially equivalent. Subject device has enhanced air detection alarms. |
| Clinical Advisories | Optional clinical advisory field/message on the pump interface.
Cannot be configured or disabled by pump user. | | Same |
| Operating Environment and Storage Specifications | Operating Temperature: 41°F to 104°F (5°C to 40°C)
Storage Temperature: -4°F to 140°F (-20°C to 60°C)
Atmospheric Pressure: 0 to 10,000 feet (0 to 3000 meters) or equivalent atmospheric pressure
Relative Humidity: 10% to 90% (maximum dew point of 30°C) | | Same
Details are located in the product labeling |
{8}
K242114
Page 5 of 7
icumedical
human connections
| Summary of the technological characteristics of the device compared to the predicate device | | | |
| --- | --- | --- | --- |
| Characteristic | Subject Device | Predicate (K223607) | Comparison |
| Physical Specifications | Single Channel Device: Dimensions: Approximately 8.5 H x 6.5 W x 6.5 D inches 21.6 cm H x 16.5 cm W x 16.5 cm D) (excluding pole clamp extrusion and power cord storage) Mass: Approximately 7 lbs (3.2 kg) with battery Casing: High-impact plastic. | Two Channel Device: Duo Dimensions: Approximately 9 H x 11.75 W x 6.5 D inches (23 cm H x 30 cm W x 17 cm D) (excluding pole clamp extrusion and power cord storage) Mass: Approximately 10.6 lbs (4.8 kilograms) with battery Casing: High-impact plastic. | Substantially equivalent |
| Power Requirements | Off the Shelf (OTS) universal power supply acceptable for 100 to 240 VAC, 50-60 Hz mains supply. | | Same |
| Battery | Lithium Iron Phosphate Battery (2500mAhr) with battery management electronics. Battery life greater than 2 years. | Lithium Iron Phosphate Battery (3000 mAhr) with battery management electronics. Battery life greater than 2 years. | Substantially equivalent |
| Display | 7" color display with WXGA resolution and Gorilla glass to protect LCD. | 10.1" color display with WXGA resolution and Gorilla glass to protect LCD. | Substantially equivalent |
{9}
K242114
Page 6 of 7
icumedical
human connections
# Summary of Non-Clinical Testing
To demonstrate substantial equivalence between the subject and predicate device the following non-clinical tests were performed:
- Verification testing of product requirements
- Human factors validation testing of product requirements associated with critical tasks
- Testing for the reliability goals of the device
The safety assurance case was provided for the Plum Solo™ Precision IV Pump as recommended in the FDA Guidance document, Infusion Pumps Total Product Life Cycle issued December 2, 2014.
The safety assurance case is used to build a robust argument that the Plum Solo™ Precision IV Pump is safe for its intended use in its intended environment. The argument is made by mitigating the following three arguments wherein risk may be produced within the context of the device's intended use.
The assurance case defines the device system, including the operational description, system definition, indications for use, patient population, intended users and use environments. The supporting assurance arguments covered the following attributes:
- Plum Solo™ Precision IV Pump hazards are adequately identified and addressed.
- Plum Solo™ Precision IV Pump design is adequately reliable.
- Plum Solo™ Precision IV Pump design requirements are adequate and are adequately verified and validated.
The following evidence was included in the safety assurance case:
- Design verification and validation testing confirmed the Plum Solo™ Precision IV Pump met user needs, risk controls, and design inputs. Testing results conformed with acceptance criteria.
- Flow rate and bolus accuracy testing were conducted by following AAMI TIR101.
- Device reliability activities, testing and statistical analysis confirmed the Plum Solo™ Precision IV Pump met its reliability goal at the system, device subsystem, and subsystem component level.
- Software verification and validation were performed per FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices issued May 11, 2005, and Content of Premarket Submissions for Device Software Functions Draft Guidance for Industry and Food and Drug Administration Staff, issued November 4, 2021.
- Human factors evaluations were conducted to validate the effectiveness of safety critical user-related features/functionality and use error-related mitigations in the associated use environments. The 2016 FDA guidance document, Applying Human Factors and Usability Engineering to Medical Devices, and IEC 62366-1 Medical Devices Part 1: Application of usability engineering to medical devices were followed.
Electrical and Electromagnetic Compatibility testing were conducted. The Plum Solo™ Precision IV Pump complies with the following standards:
- Electrical Safety per IEC 60601-1 and Electromagnetic Compatibility per IEC 60601-1-2.
{10}
K242114
Page 7 of 7
icumedical human connections
Cybersecurity testing performed confirmed the system is effective in addressing cybersecurity threats. FDA Cybersecurity Guidance followed during development include Content of Premarket Submissions for Management of Cybersecurity, issued October 2, 2014 and Postmarket Management of Cybersecurity in Medical Devices, issued December 28, 2016.
Risk management activities have been incorporated into the design in accordance with ISO 14971:2019 +A11 2021 and have been tested for correct implementation and effectiveness as part of design verification and validation
## Clinical Testing
Clinical evaluation is not required for this submission to support substantial equivalence.
## Conclusion
The differences between the predicate and subject device do not raise different questions of safety or effectiveness. The Plum Solo™ Precision IV Pump is substantially equivalent to the predicate device Plum Duo™ Infusion System (K223607).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.