Sol-M™ Slip Tip Syringe without Needle, Sol-M™ Eccentric Tip Syringe without Needle
K242099 · Sol-Millennium Medical, Inc. · FMF · Oct 22, 2024 · General Hospital
Device Facts
Record ID
K242099
Device Name
Sol-M™ Slip Tip Syringe without Needle, Sol-M™ Eccentric Tip Syringe without Needle
Applicant
Sol-Millennium Medical, Inc.
Product Code
FMF · General Hospital
Decision Date
Oct 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The syringe without needle is used to inject fluids into, or withdraw fluids from, the body.
Device Story
The Sol-M Slip Tip and Eccentric Tip Syringes are sterile, single-use, 3-piece manual piston syringes. They consist of a polypropylene barrel and plunger with a polyisoprene rubber piston, lubricated with polydimethylsiloxane. The device is operated manually by a clinician to inject or withdraw fluids. It is available in various volumes (1ml to 60ml) and features Luer slip connectors. The device is EO sterilized. By providing a sterile, reliable means for fluid transfer, the syringe facilitates medication delivery or sample collection, benefiting patients by ensuring accurate dosing and maintaining sterility during clinical procedures.
Clinical Evidence
No clinical data. Bench testing only, including compliance with ISO 7886-1 (syringe performance), ISO 80369-7/20 (connector performance), ISO 10993 series (biocompatibility), USP <788> (particulate matter), and ASTM F88/F1929 (packaging integrity).
Technological Characteristics
3-piece manual piston syringe. Materials: Polypropylene (barrel/plunger), Polyisoprene rubber (piston), Polydimethylsiloxane (lubricant). Sterilization: Ethylene Oxide (EO) to 10-6 SAL. Standards: ISO 7886-1, ISO 80369-7, ISO 80369-20, ISO 10993-1/4/5/7/10/11, ISO 11135, USP <788>, ASTM F88/F88M-15, ASTM F1929-15. Connectivity: None (manual).
Indications for Use
Indicated for use in patients requiring injection of fluids into or withdrawal of fluids from the body.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Sterile Hypodermic Syringes for Single Use (K221079)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
October 22, 2024
Sol-Millennium Medical Inc. % Amy Li, Technology Director Shanghai Mind-link Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Rode Jing'an District, Shanghai, Shanghai 201803, China
## Re: K242099
Trade/Device Name: Sol-MIM Slip Tip Syringe without Needle, Sol-MTM Eccentric Tip Syringe without Needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: June 27, 2024 Received: September 27, 2024
## Dear Amy Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Shruti N. Mistry -S
Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242099
#### Device Name
Sol-M™ Slip Tip Syringe without Needle Sol-M™ Eccentric Tip Syringe without Needle
#### Indications for Use (Describe)
The syringe without needle is used to inject fluids into, or withdraw fluids from, the body.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for SolM, which is a company that is building a healthier tomorrow. The logo consists of a green half circle, an orange circle, the word "SolM" in green, and the words "Building a Healthier Tomorrow" in a smaller font. The logo is simple and modern, and it conveys the company's mission of improving health.
# K242099 510(k) Summary
- 1. Date of preparation: October 22, 2024
- 2. Submitter information
Address: 311 S Wacker Drive, Suite 4100, Chicago, Illinois, 60606, United States Contact person: Manu Kalia, Global QARA Director, +1 847-313-9577, mkalia@solm.com Submission correspondent: Amy.li, Technology director, +86 15721449974, amy.li@mind-link.net
- 3. Subject Device
Trade name: Sol-M™ Slip Tip Syringe without Needle, Sol-M™ Eccentric Tip Syringe without Needle Common name: Piston Syringe Regulation: 21 CFR § 880.5860 Product code: FMF Classification: Class II Review Pannel: General Hospital
- 4. Predicate Device
K221079 Trade Name: Sterile Hypodermic Syringes for Single Use Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Review Panel: General hospital
- 5. Indication for use statement
The syringe without needle is used to inject fluids into, or withdraw fluids from, the body
- 6. Device description
The Slip Tip and Eccentric Tip Syringe without Needle is a sterile, single-use, standard 3 piece piston. The device is packed individually and EO sterilized with 10-6 SAL.
- 6.1 The proposed device includes different specifications.
- 1ml Slip Tip Syringe without Needle
- 3ml Slip Tip Syringe without Needle
- 5ml Slip Tip Syringe without Needle
- 10ml Slip Tip Syringe without Needle
- 20ml Slip Tip Syringe without Needle
- 30ml Slip Tip Syringe without Needle
- 60ml Slip Tip Syringe without Needle
- 10ml Eccentric Tip Syringe without Needle
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Sol M. The logo consists of a green half circle, an orange circle, the text "Sol M" in green, and the text "Building a Healthier Tomorrow" in a smaller font size. The text "Sol M" is in a bold, sans-serif font. The logo is simple and modern.
- 20ml Eccentric Tip Syringe without Needle
- 60ml Eccentric Tip Syringe without Needle
- 6.2 Label requirement
- The label shall meet the requirements of 21 CFR Part 801
- 7. Comparison of technological characteristics with the predicate devices
The subject device has the same intended use, technology, and design; and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the subject device and predicate devices do not alter suitability of the proposed device for its intended use.
| Item | Proposed Device Slip Tip Syringe<br>without Needle, Eccentric Tip<br>Syringe without Needle | Predicate Device Sterile Hypodermic<br>Syringes for Single Use<br>K221079 | Remark |
|-------------------------------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Indication for Use | The syringe without needle is<br>used to inject fluids into, or<br>withdraw fluids from, the body. | The Sterile Hypodermic Syringes for Single<br>Use are intended to be used for medical<br>purpose to inject fluid into or withdraw<br>fluid from body. | Same |
| Prescription or OTC<br>(over the counter) | Prescription use | Prescription use | Same |
| Configuration | Barrel | Barrel | Same |
| Configuration | Plunger | Plunger | Same |
| Configuration | Piston | Piston | Same |
| Operation Mode | For manual use only | For manual use only | Same |
| Single Use | Single Use | Single Use | Same |
| Syringe<br>performance | Complied with ISO 7886-1 | Complied with ISO 7886-1 | Same |
| Luer Connector<br>Performance | Complied with ISO 80369-7 | Complied with ISO 80369-7 | Same |
| Volume | Slip tip:1ml, 3ml, 5ml, 10ml,<br>20ml, 30ml, 60ml<br>Eccentric tip: 10ml, 20ml, 60ml | 1ml, 2ml, 3ml, 5ml, 10ml, 20ml, 30ml,<br>50ml, 60ml, 100ml | Similar.<br>Analysis 1 |
| Connector Type | Luer slip(Slip tip, Eccentric tip) | Luer slip/Luer lock | Similar.<br>Analysis 2 |
| Barrel | Polypropylene(PP) | Polypropylene(PP) | Same |
| Plunger | Polypropylene(PP) | Polypropylene(PP) | Same |
| Piston | Polyisoprene rubber(IR) | Polyisoprene | Same |
| Lubricant | Polydimethylsiloxane | Polydimethylsiloxane | Same |
| Sterilization | EO Sterilized | EO Sterilized | Same |
| Shelf life | 5 years | 5 years | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Sol M, a company focused on building a healthier tomorrow. The logo features a green dot, an orange circle, and the text "Sol M" in a dark green, sans-serif font. To the right of "Sol M" are the words "Building a Healthier Tomorrow" in a smaller, lighter font.
#### Analvsis 1:Volume
The subject device has less volume compared with the predicate device. but the predicate device has additional volumes 2ml and 100ml. However, all syringe performance was tested in accordance with ISO 7886-1 and ISO 80369-7. Therefore, this difference in syringe volume will not affect the Substantial Equivalence (SE) between the proposed device and the predicate device.
#### Analysis 2: Connector Type
The subject device has the same connector type compared with the predicate device, but the predicate device has an additional Luer lock type. However, the Luer connector performance complies with ISO 80369-7. Therefore, this difference in connector type will not affect the Substantial Equivalence (SE) between the proposed device and the predicate device.
#### 8. Non-clinical Performance Testing
Non-clinical tests were conducted to verify that the proposed device meets all design specifications. The test results demonstrated that the proposed device complies with the following standards:
- 8.1 ISO 7886-1:2017 Sterile hypodermic syringes for single use- Part 1: Syringes for manual use.
- 8.2 ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare application -- Part 7: Connectors for intravascular or hypodermic application
- 8.3 ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
- 8.4 ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- 8.5 ISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood
- 8.6 ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- 8.7 ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals
- 8.8 ISO 10993-10: 2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
- 8.9 ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity
- 8.10 USP <788> Particulate Matter for injection
- 8.11 ISO 11135:2014 Sterilization of health-care products Ethylene oxide Requirements for the development, validation and routine control of a sterilization process for medical devices
- 8.12ASTM F88/F88M-15, Standard Test Method For Seal Strength Of Flexible Barrier Materials
- 8.13 ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Sol M, a company that is building a healthier tomorrow. The logo features a green circle with a portion cut out, an orange circle, and the text "Sol M" in green. To the right of the "M" is the text "Building a Healthier Tomorrow" in a smaller font.
### 9. Clinical Testing
No clinical study is included in this submission.
- 10.Conclusion
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The subject device is substantially equivalent to the predicate device Sterile Hypodermic Syringes for Single Use (K221079).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.