STERiJECT Ophthalmic Needle; STERiJECT Ophthalmic Needle Low Dead Space (LDS)
K242073 · Tsk Laboratory International Japan KK · QYM · Oct 23, 2024 · General Hospital
Device Facts
Record ID
K242073
Device Name
STERiJECT Ophthalmic Needle; STERiJECT Ophthalmic Needle Low Dead Space (LDS)
Applicant
Tsk Laboratory International Japan KK
Product Code
QYM · General Hospital
Decision Date
Oct 23, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended to inject into or aspirate from, the body. The device is intended for intravitreal use.
Device Story
STERiJECT Ophthalmic Needle is a sterile, single-use hypodermic needle consisting of a stainless steel tube sharpened at one end and attached to a polycarbonate hub. Designed for manual operation by healthcare professionals, the device attaches to luer lock or luer slip piston syringes. Available in standard and low dead space (LDS) configurations, the LDS hub is designed to minimize medication waste. The device is used for subcutaneous, intramuscular, intradermal, and specifically intravitreal injections. By facilitating precise delivery of ophthalmic medications, the device supports clinical treatment of ocular conditions. It is provided sterile via gamma irradiation and is compatible with standard syringe connections.
Clinical Evidence
Bench testing only. Performance testing included conformance to ISO 9626 (tubing), ISO 7864 (hypodermic needles), and ISO 80369-7 (connectors). Biocompatibility was evaluated per ISO 10993-1, -5, -10, -11, -23, and ASTM F756. Sterility, packaging integrity, and shelf-life (5 years) were validated per ISO 11137, ISO 11607, ASTM F3039, ASTM D4169, and ASTM F1980. No clinical data was required or provided.
Technological Characteristics
Stainless steel needle tube; polycarbonate hub; epoxy adhesive; silicone oil lubricant. Gauge range 27G-34G; lengths 4mm-13mm. Luer lock/slip connectivity compliant with ISO 80369-7 (standard hubs). Sterilization via gamma irradiation (SAL 10^-6). Biocompatibility per ISO 10993 series. Packaging validated per ISO 11607-1 and ASTM F3039.
Indications for Use
Indicated for intravitreal injection or aspiration in patients requiring such procedures.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
STERiJECT Low Dead Space, STERiJECT The INVISIBLE Needle (K231734)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 23, 2024
TSK Laboratory International Japan KK Seika Shiba Regulatory Affairs Specialist 2-1-5 Hirayanagi-Cho Tochigi-Shi, Tochigi 328-0012 Japan
#### Re: K242073
Trade/Device Name: STERiJECT Ophthalmic Needle; STERiJECT Ophthalmic Needle Low Dead Space (LDS) Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: QYM Dated: September 23, 2024 Received: September 24, 2024
#### Dear Seika Shiba:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
## Sincerely,
# Shruti N. Mistry -S
Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K242073
Device Name
STERiJECT Ophthalmic Needle; STERiJECT Ophthalmic Needle Low Dead Space (LDS)
Indications for Use (Describe) The device is intended to inject into or aspirate from, the body. The device is intended for intravitreal use.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| <div> <span> </span> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image features the letters "TSK" in a bold, sans-serif font. The letters are white and are set against a blue background. The blue background has a slight gradient effect, with the top portion appearing lighter than the bottom.
### 510 (k) Summary
#### Preparation Date: October 22nd, 2024
#### 1. Contact Details
Applicant Name: TSK Laboratory International Japan KK Applicant Address: 2-1-5 Hirayanagi-Cho Tochigi-Ken 328-0012 Japan Applicant Contact Telephone: 81 282 25 5200 Applicant Contact: Seika Shiba Applicant Contact Email: shiba@tsklab.com
#### 2. Device Name
Device Trade Name: STERiJECT Ophthalmic Needle; STERiJECT Ophthalmic Needle Low Dead Space (LDS) Device: Ophthalmic Needle Classification Name: Ophthalmic Needle Regulation Name: Hypodermic single lumen needle Medical Specialty: General Hospital Regulation Number: 21 CFR 880.5570 Product Code(s): QYM (Ophthalmic Needle) Device Class: Class II
#### 3. Legally Marketed Predicate Device
| Predicate # | Predicate Trade Name | Product Code |
|-------------|--------------------------------|--------------|
| K230951 | Terumo Injection Filter Needle | QYM |
| Reference # | Reference Device Trade Name | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K231734 | STERiJECT Low Dead Space, STERiJECT The INVISIBLE Needle | QNS |
#### 4. Device Description Summary
The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a luer lock or luer slip piston syringe. The device can be used by health care professionals. The hub of the needle can be a standard size or a reduced size (low dead space) that is designed to reduce medication waste. The device is operated manually by attaching it to a piston syringe. Target injection sites include subcutaneous, intramuscular, intradermal, and intravitreal.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the letters "TSK" in white font on a blue background. The letters are bold and sans-serif. The background is a solid blue color, and the letters are centered on the background. The image is simple and clean, with a focus on the letters.
#### Models:
| Model | Connector Type | Thread Type | Dead Space |
|-------|----------------------|-------------|------------|
| LDS | Luer lock, Luer slip | Thread | Low |
| HPC | Luer lock, Luer slip | Thread | Standard |
| PRC | Luer lock, Luer slip | Lug | Standard |
#### Product sizes and color codes of the label:
| Nominal OD | Nominal Gauge Size | Label Color | Length |
|------------|--------------------|-------------|-------------|
| (mm) | Ref Std | | (mm) |
| Φ 0.18 | ISO 34G | Orange | 4, 6, 9 |
| Φ 0.20 | ISO 33G | Black | 4, 6, 9, 13 |
| Φ 0.23 | ISO 32G | Deep Green | 4, 6, 9, 13 |
| Φ 0.24 | ENE 33G | Black | 4, 6, 9, 13 |
| Φ 0.25 | ISO 31G | White | 4, 6, 9, 13 |
| Φ 0.26 | ENE 32G | Deep Green | 4, 6, 9, 13 |
| Φ 0.28 | ENE 31G | White | 4, 6, 9, 13 |
| Φ 0.30 | ISO 30G | Yellow | 4, 6, 9, 13 |
| Φ 0.33 | ISO 29G | Red | 4, 6, 9, 13 |
| Φ 0.36 | ISO 28G | Blue Green | 4, 6, 9, 13 |
| Φ 0.40 | ISO 27G | Medium Grey | 4, 6, 9, 13 |
#### 5. Intended Use/Indication for Use
The device is intended to inject into or aspirate from, the body. The device is indicated for intravitreal use.
#### 6. Indications for Use Comparison
The subject device has completed testing to show that the device meets its intended use/indication for use and demonstrates substantial equivalence to the predicate device K230951.
#### 7. Technological Comparison
The table below includes a comparison of the technological characteristics between the subject device and those of the predicate device and reference device:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image features the letters 'TSK' in a bold, sans-serif font. The letters are white and stand out against a blue background. The overall design is simple and clean, with a focus on readability and contrast.
| Technological<br>Characteristic | Subject Device | Predicate Device | Reference Device | Discussions |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|---------------------------------|
| Trade name | STERIJECT<br>ophthalmic<br>Needle; STERIJECT<br>Ophthalmic<br>Needle Low Dead<br>Space (LDS), | Terumo Injection Filter<br>Needle (NF-<br>3013RBKE05M)<br>(K230951) | STERIJECT Low<br>Dead Space (LDS);<br>STERIJECT The<br>INVISIBLE Needle | |
| Indications<br>for Use | The device is<br>intended to inject<br>into or aspirate<br>from, the body.<br>The device is<br>indicated for<br>intravitreal use. | The Terumo Injection<br>Filter Needle is<br>indicated for general<br>application - for<br>treatment - injection<br>of fluids into parts of<br>the body below the<br>surface of the skin and<br>into the vitreous. | (K231734)<br>To inject fluids<br>into or aspirate<br>from, the body. | Same |
| Regulation | 880.5570 | 880.5570 | 880.5570 | Same |
| Product Code | QYM | QYM | QNS | Same |
| Class | Class II | Class II | Class II | Same |
| Gauge | 27G - 34G | 30G | 21G - 34G | Different.<br>See Discussion #2 |
| Length | 4mm - 13mm | 12mm ( ½" ) | 4mm - 120mm | Different.<br>See Discussion #2 |
| Dead Volume<br>OPHTHALMIC<br>NEEDLE<br>OPHTHALMIC<br>NEEDLE LDS | Standard | Unspecified | 0.0028ml | Different.<br>See Discussion #4 |
| 0.0028ml | | | | |
| Safety<br>Feature | No safety feature | unspecified | No safety feature | See Discussion #1 |
| Syringe<br>compatibility | Luer Lock or<br>Luer Slip | Unspecified | Luer Lock or<br>Luer Slip | Same |
| Hub | Polycarbonate | PMMA/Masterbatch | Polycarbonate | Different.<br>See Discussion #5 |
| Needle tube | Stainless Steel | Stainless Steel | Stainless Steel | Same |
| Adhesive | Epoxy | Acrylic Glue | Epoxy | Different.<br>See Discussion #5 |
| Lubricant | Silicon oil | Silicone<br>(Polydimethylsiloxane) | Silicon oil | Same |
| Sterilization | Gamma | EO gas | Gamma | Different |
| method | irradiation | | irradiation | See Discussion #3 |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image features a blue rectangular button with the letters 'TSK' in white. The letters are bold and slightly stylized, giving them a modern appearance. The blue background is a solid color, providing a strong contrast to the white letters, making them easily readable.
#### Discussions of differences in technological characteristics
#### Discussion #1
The subject device does not have a safety feature. In general, the safety feature is optional for hypodermic or ophthalmic needles and not having the safety feature does not raise new or different questions of safety and effectiveness when compared to the predicate device.
#### Discussion #2
The difference in specifications does not raise new questions of safety and effectiveness. The subject device was tested to the same standards as the predicate device K230951. The substantial equivalence of the device is confirmed through performance testing.
#### Discussion #3
The sterility of the subject device has been validated to achieve SAL of 10-6 in accordance with ISO 11137-1 and ISO 11137-2. Packaging integrity tests were conducted according to ISO 11607-1 and ASTM F3039 to ensure that the hard pack packaging serving as the sterile barrier system is maintained over the life of the device. K230951 is sterilized using EtO gas sterilization which uses standards ISO 11135 whereas the subject device uses gamma irradiation. K231734, which does use gamma irradiation, has been validated to achieve SAL of 10-6 in accordance with ISO 11137-1 and ISO 11137-2, therefore the difference between the subject and predicate device does not raise new or different questions of safety and effectiveness.
#### Discussion #4
The subject device has standard hubs or low dead space hubs. Standard hubs meet all the requirements for ISO 80369-7 which defines the connector requirements for standard needle hub connections. Low dead space hub meets the requirement for ISO 80369-7 with the exception of the needle hub depth. There are currently no recognized standards for low dead space needles, however the physical characteristic of the low dead space hub is identical to K231734. The difference between the subject and predicate device does not raise new or different questions of safety and effectiveness.
#### Discussion #5
Although the materials of the subject device are different the from predicate device K230951, the biocompatibility of the materials was tested and meets the biocompatibility requirements of the applicable standards in the ISO 10993 series. Shelf-life study was conducted, and the result demonstrates the subject device continues to meet its specifications for 5 years after gamma irradiation.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the letters "TSK" in white font on a blue background. The letters are bold and sans-serif. The background is a solid, light blue color. The image is simple and clean, with a focus on the text.
The substantial equivalence of the subject device has been demonstrated by non-clinical testing data, hence the differences between the subject device and the predicate do not raise any new or different questions of safety or effectiveness. Therefore, it is concluded that the subject device is substantially equivalent to the legally marketed predicate device K230951.
#### 8. Non-clinical and/or Clinical Tests Summary & Conclusions
The subject device was tested and demonstrated to be in conformance with the following FDA recognized standards.
- ISO 9626: 2016 Stainless steel needle tubing for the manufacture of medical devices Requirements and test methods.
- ISO 7864: 2016 Sterile hypodermic needles for single-use Requirements and test methods
- ISO 80369-7: 2021 Small-bore connectors for liquid and gases in healthcare applications Part 7: Connectors for intravascular or hypodermic applications.
The standard hub needle models meet the requirements for ISO 80369-7. The subject device meets the requirements for ISO 80369-7 with the exception of the needle hub length which creates the dead space for the low dead space needle models. There are currently no FDA recognized standards for low dead space needles, however, it is equivalent to K231734 which also has a low dead volume hub. The subject device's ability to meet the connection requirements of ISO 80369-7 confirms the device's ability to secure connection for the low dead space needles.
#### Biocompatibility
In accordance with ISO 10993-1 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", the needle is classified as: Externally Communicating Device, Blood Path Indirect, Limited Contact (<24 hours). For Ophthalmic biological effects the device is also evaluated for Ocular Irritation, Intravitreal Injection Irritation, and as per USP788/789 "Particulate Matter in Ophthalmic Solutions", and as per FDA Guidance "Endotoxin Testing Recommendations for Single-Use Intraocular Ophthalmic Devices" each lot is tested for 0.2 EU/Device.
The following testing was conducted:
- ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image is a blue rectangle with rounded corners. Inside the rectangle, there is the text "TSK" in white, sans-serif font. The letters are bold and slightly stylized, giving the logo a modern and clean appearance.
- ISO 10993-23:2021 Biological evaluation of medical devices – Part 23: Tests for irritation
- ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
- . USP Chapter <151> , ISO 10993-11 Pyrogen Test
#### Sterility, Shipping and Shelf-Life
Sterilization validation was performed in accordance with standards listed below.
- . ISO 11137-1:2006/AMD 1:2013 Sterilization of health care products-Radiation -Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 11137-2:2006 Sterilization of health care products-Radiation - Part 2: Establishing the sterilization dose
Packaging validation was performed in accordance with the standards listed below.
- . ISO 11607-1:2019 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- . ASTM F3039-15 Standard Test Method for Detecting Leaks in Nonporous Packaging or Flexible Barrier Materials by Dye Penetration
- ASTM D4169 – 22 Standard Practice for Performance Testing of Shipping Containers and Systems
Shelf life of 5 years validated using the standard listed below.
- . ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
The result of non-clinical testing demonstrates the substantial equivalence of the subject device.
#### 9. Conclusion
The differences between the predicate devices and the subject device do not raise any new or different questions of safety or effectiveness. The subject device is substantially equivalent to K230951.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.