K242058 · Apple, Inc. · SCW · Oct 21, 2024 · Ophthalmic
Device Facts
Record ID
K242058
Device Name
Digital Prism Correction Feature (DPCF)
Applicant
Apple, Inc.
Product Code
SCW · Ophthalmic
Decision Date
Oct 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1655
Device Class
Class 1
Attributes
Software as a Medical Device, Therapeutic
Indications for Use
The Digital Prism Correction Feature (DPCF) is software that is intended to provide digital image adjustments in Apple Vision Pro in accordance with a user's prism prescription. DPCF is available for users with prism in their eyeglass prescription. When a prescription also includes other parts (e.g., sphere, cylinder, ADD), which can be fulfilled by optical inserts, the DPCF fulfills the prism part of the prescription while using Apple Vision Pro. DPCF supports prism prescriptions up to 7.75 Prism Diopters (PD) in the horizontal and/or vertical dimensions (i.e., base-up, base-down, base-in, base-out), per eye.
Device Story
Software-only device for Apple Vision Pro; provides digital image adjustments to simulate prismatic effect; replaces physical Fresnel prism sheets. User inputs prism prescription; software converts prescription into digital image adjustment parameters; spatial computing system shifts images horizontally/vertically. Used by patients requiring prism correction; operates within Apple Vision Pro environment. Output is visually adjusted digital content; assists users with binocular vision needs; enables use of spatial computing platform with existing prescription requirements. No clinical data; bench testing confirms adherence to ISO 8980-1:2017 prism tolerance standards.
Clinical Evidence
No clinical data submitted. Bench testing only. Validation confirmed DPCF provides prismatic adjustments meeting ISO 8980-1:2017 tolerances for both virtual and digital passthrough content, with and without optical inserts.
Technological Characteristics
Software-only device; operates on Apple Vision Pro spatial computing platform. Implements digital image shifting to achieve prismatic effect. Supports up to 7.75 PD per eye. Verified against ISO 8980-1:2017 prism tolerance standards. Software documentation level: Basic.
Indications for Use
Indicated for users requiring prism correction up to 7.75 Prism Diopters (PD) per eye in horizontal and/or vertical dimensions while using Apple Vision Pro. Compatible with users whose eyeglass prescriptions include sphere, cylinder, or ADD components when used with optical inserts.
Regulatory Classification
Identification
An ophthalmic Fresnel prism is a device that is a thin plastic sheet with embossed rulings which provides the optical effect of a prism. The device is intended to be applied to spectacle lenses to give a prismatic effect.
Predicate Devices
Ophthalmic Fresnel prism (21 CFR 886.1655)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 21, 2024
Ian Marcus Regulatory Affairs Apple Inc. One Apple Park Way Cupertino, CA 95014
Re: K242058
Trade/Device Name: Digital Prism Correction Feature (DPCF) Regulation Number: 21 CFR 886.1655 Regulation Name: Ophthalmic Fresnel Prism Regulatory Class: Class I Product Code: SCW Dated: July 12, 2024 Received: September 10, 2024
Dear Ian Marcus:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# J Angelo Green -S
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242058
Device Name Digital Prism Correction Feature (DPCF)
#### Indications for Use (Describe)
The Digital Prism Correction Feature (DPCF) is software that is intended to provide digital image adjustments in Apple Vision Pro in accordance with a user's prism prescription.
DPCF is available for users with prism in their eyeglass prescription also includes other parts (e.g., sphere, cylinder, ADD), which can be fulfilled by optical inserts, the DPCF fulfills the prism part of the prescription while using Apple Vision Pro.
DPCF supports prism prescriptions up to 7.75 Prism Diopters (PD) in the horizontal and/or vertical dimensions (i.e., baseup, base-down, base-in, base-out), per eye.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
#### 1. Submitter
| Applicant | Apple Inc.<br>One Apple Park Way<br>Cupertino, CA 95014 |
|-----------------------------|--------------------------------------------------------------------------------------------------|
| Submission<br>Correspondent | lan Marcus<br>Regulatory Affairs<br>Phone: 408-761-2934<br>Email: lan_Marcus@apple.com |
| Secondary<br>Correspondent | Sam Surette<br>US Regulatory Affairs Manager<br>Phone: 628-629-4161<br>Email: ssurette@apple.com |
| Date Prepared | July 12, 2024 |
| Premarket Submission | Traditional 510(k), K242058 |
#### 2. Device Names and Classifications
#### Subject Device:
| Name of Device | Digital Prism Correction Feature (DPCF) |
|---------------------|-------------------------------------------|
| Classification | Ophthalmic Fresnel prism, 21 CFR 886.1655 |
| Regulatory Class | Class I |
| Product Code | SCW (Digital prismatic correction) |
| 510(k) Review Panel | Ophthalmic |
#### 3. Predicate Device
Ophthalmic Fresnel prism, 21 CFR 886.1655.
#### 4. Device Description
The Digital Prism Correction Feature (DPCF) is intended to provide a high quality visual experience in Apple Vision Pro spatial computing applications for users with a prism prescription. Specifically, the DPCF is software that is intended to provide digital image adjustments in Apple
{5}------------------------------------------------
Vision Pro in accordance with a user's prism prescription, in the horizontal and/or vertical dimensions. DPCF fulfills the prism part of an eyeglass prescription. When an eyeglass prescription also includes other parts (e.g., sphere, cylinder, ADD), DPCF fulfills the prism part of the prescription, while prescription optical inserts fulfill the other parts of the prescription.
The DPCF achieves its intended use by converting a user's prism prescription into digital image adiustment parameters that are utilized by the spatial computing image system to automatically provide digital image adjustments in horizontal and/or vertical dimensions in accordance with a user's prism prescription. At this time, DPCF supports prism prescriptions up to 7.75 Prism Diopters (PD) in the horizontal and/or vertical dimensions (i.e. base-down, base-in, base-out), per eye.
### 5. Indications for Use
The Digital Prism Correction Feature (DPCF) is software that is intended to provide digital image adjustments in Apple Vision Pro in accordance with a user's prism prescription.
DPCF is available for users with prism in their eyeglass prescription. When a prescription also includes other parts (e.g., sphere, cylinder, ADD), which can be fulfilled by optical inserts, the DPCF fulfills the prism part of the prescription while using Apple Vision Pro.
DPCF supports prism prescriptions up to 7.75 Prism Diopters (PD) in the horizontal and/or vertical dimensions (i.e., base-up, base-down, base-out), per eve.
## 6. Comparison with the Predicate Device
The subject and predicate device, as described by the identification (21 CFR 886.1655), are both intended to provide a prismatic effect. The subject device provides a prismatic effect by providing perceived digital image shift in horizontal and/or vertical dimensions in accordance with a user's prescription; the predicate device also provides perceived prismatic effects. Furthermore, the DPCF can be used in conjunction with ophthalmic lenses, consistent with the intended use of the predicate device, for which the classification's identification describes a device that can be applied to lenses. Since both the subject and predicate device (per 21 CFR 886.1655) are intended to provide prismatic effects in accordance with a prism prescription, and can be used in coniunction with spectacle lenses, thev have the same intended use.
While DPCF prismatically adjusts a digital image using software, and the predicate provides a prismatic effect through a plastic sheet adhered to lenses, these technological differences do not raise different questions of safety or effectiveness, given that devices that provide prismatic effects are well characterized and have a known history of very low risk. Furthermore, the differences in technological characteristics are addressed through acceptable scientific methods, from which the data demonstrate substantial equivalence.
The subject device has been appropriately verified and validated through non-clinical testing to ensure that the device is substantially equivalent to the predicate. A complete comparison of the subject and predicate device can be found in Table 1 below.
{6}------------------------------------------------
| ltem | Subject Device<br>Digital Prism Correction Feature<br>(DPCF) | Predicate Device<br>21 CFR 886.1655 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Device Name | Digital Prism Correction Feature | N/A |
| Manufacturer | Apple, Inc. | N/A |
| Regulation Number | 21 CFR 886.1655 | 21 CFR 886.1655 |
| Product Code | SCW | HKT |
| Regulation Name | Ophthalmic Fresnel prism | Ophthalmic Fresnel prism |
| Device Classification | Class I | Class I |
| OTC/Prescription | Rx Only | N/A |
| Intended Use | Provide prismatic effect, and can be<br>used in conjunction with lenses | Provide prismatic effect, and are used<br>in conjunction with lenses |
| Indications for Use | The Digital Prism Correction Feature<br>(DPCF) is software that is intended to<br>provide digital image adjustments in<br>Apple Vision Pro in accordance with a<br>user's prism prescription.<br>DPCF is available for users with prism<br>in their eyeglass prescription. When a<br>prescription also includes other parts<br>(e.g., sphere, cylinder, ADD), which can<br>be fulfilled by optical inserts, the DPCF<br>fulfills the prism part of the prescription<br>while using Apple Vision Pro.<br>DPCF supports prism prescriptions up<br>to 7.75 Prism Diopters (PD) in the<br>horizontal and/or vertical dimension<br>(i.e., base-up, base-down, base-in,<br>base-out), per eye. | N/A |
| Principle of Operation | Digital image adjustment in<br>accordance with a prism prescription,<br>to provide a prismatic effect. | Optical refraction in accordance with a<br>prism prescription, to provide a<br>prismatic effect. |
| Use Scenario(s) | Used with Apple Vision Pro, which can<br>also include use in conjunction with<br>optical inserts. | Used in conjunction with spectacle<br>lenses |
| Intended Users | Users in need of prism correction. | Users in need of prism correction. |
| Performance | Meets standardized prism tolerance<br>requirements | Meets standardized prism tolerance<br>requirements |
# Table 1: DPCF Comparison with the Predicate
{7}------------------------------------------------
The subject device shares the same intended use and represents the same generic device type as the predicate device, as established by 21 CFR 886.1655. While there are differences of technological characteristics, these differences do not raise different questions of safety and effectiveness. Data from acceptable methods have been provided in this 510(k) submission. including results from bench validation testing, to substantiate the intended use of the subject device and demonstrate the subject device is substantially equivalent to the predicate device. The differences between the subject and predicate device do not present any new issues of safety or effectiveness as they do not increase the risk of the device, and can be appropriately verified and validated through existing methods. Thus, the DPCF is substantially equivalent to the predicate device, ophthalmic Fresnel prism (21 CFR 886,1655).
# 7. Summary of Non-Clinical Testing
#### Non-clinical Testing Summary
Apple conducted non-clinical testing on DPCF with passing results, supporting a determination of substantial equivalence. Non-clinical testing conducted included the following:
#### Software Verification and Validation
Software verification and validation was conducted in accordance with Apple's robust quality system and documented to address the recommendations in FDA's "Content of Premarket Submissions for Device Software Functions" quidance document. DPCF was determined to be a Basic Documentation Level. Apple's good software engineering practices, as demonstrated by the 510(k) submission's documentation, supports a conclusion that DPCF was appropriately designed, verified, and validated.
#### Bench Validation Testing
The bench validation testing demonstrated that the DPCF provides prismatic adjustments in accordance with a prism prescription. Specifically, the results validate that the digital image adiustments provided by DPCF meet the prism tolerance requirements specified in ISO 8980-1:2017, and demonstrates that the DPCF provides reliable and acceptable prism adjustments for the available prism adjustment range, including for both virtual and digital passthrough content. Furthermore, the results demonstrate that the DPCF performs as intended with and without optical inserts, and that the general purpose spatial computing platform (i.e., Apple Vision Pro) and it's "other functions" do not adversely affect the DPCF's ability to meet standardized prism tolerances.
#### General Purpose Spatial Computing Platform Assessment
DPCF is a software-only device, available on compatible general purpose spatial computing platforms (i.e., Apple Vision Pro); therefore, medical device hardware testing is not applicable. However, as a multiple function device product, the impact of the general purpose spatial computing platform on DPCF was assessed as part of the feature's risk management, verification, and validation activities, and determined to be acceptable.
#### 8. Summary of Clinical Testing
No clinical testing data has been submitted.
{8}------------------------------------------------
## 9. Conclusion
The Digital Prism Correction Feature is substantially equivalent to the predicate device as they share the same intended use and where there are differences in technological characteristics, these differences do not raise different questions of safety and effectiveness. Furthermore, the scientific methods used for evaluating the different technological characteristics of the subject device are acceptable and the results of that testing supports a decision of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.