K242029 · H.S Hospital Service S.P.A · KNW · Oct 9, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K242029
Device Name
TRAP EASY
Applicant
H.S Hospital Service S.P.A
Product Code
KNW · Gastroenterology, Urology
Decision Date
Oct 9, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The device is used for drawing osteomedullary substance and/or for bone marrow explantation.
Device Story
Trap Easy is a manual bone marrow biopsy needle used by physicians to harvest bone marrow specimens. The device consists of an AISI 304 stainless steel cannula with a cylindrical tip, an ergonomic ABS handle, a removable pyramidal stylet, an integrated sample trapping mechanism, and an extraction wire. During the procedure, the physician inserts the needle into the bone; the integrated trapping mechanism secures the sample within the cannula. A safety button and side lever are used to activate the trap and retrieve the specimen. The device is supplied sterile, is for single-use only, and is disposed of as infectious waste after the procedure. By facilitating the collection of intact bone marrow samples, the device assists clinicians in diagnostic evaluations of bone marrow tissue.
Clinical Evidence
Bench testing only. Performance evaluated via 50 samples to assess sample integrity, length, diameter, and removal ease. Results showed all samples maintained integrity without fragmentation; internal mechanisms remained undamaged post-biopsy. Average lengths and diameters met acceptance criteria. Biocompatibility assessed per ISO 10993-1 for externally communicating devices with limited contact (<24 hours). Sterilization validated per ISO 11135:2014/A1:2018; packaging validated per ISO 11607-1:2019.
Technological Characteristics
Manual biopsy needle. Materials: AISI 304 stainless steel (cannula/stylet), ABS (handle). Sizes: 8G and 11G (100mm/150mm lengths). Principle: Mechanical sample trapping via internal cannula mechanism. Sterilization: Ethylene Oxide (EtO). Packaging: Tyvek/OPA/PE pouch or Tyvek/PET/PE blister. No software or electronic components.
Indications for Use
Indicated for bone marrow tissue biopsies in patients requiring osteomedullary substance collection or bone marrow explantation. Intended for use by experienced, trained physicians.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
October 9, 2024
HS Hospital Service SPA % Guido Bonapace Consultant Isemed Srl Via P. Togliatti, 19/X Imola (BO), 40026 Italy
Re: K242029
Trade/Device Name: Trap Easy Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: July 3, 2024 Received: July 11, 2024
Dear Guido Bonapace:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Colin K. Chen -S Digitally signed by Colin K. Chen -S Date: 2024.10.09 14:10:58 -04'00'
Jessica Carr Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
for
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242029
Device Name
TRAP EASY
Indications for Use (Describe)
The device is used for drawing osteomedullary substance and/or for bone marrow explantation.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### General Information
#### Applicant:
HS HOSPITAL SERVICE SPA Via Zosimo 13, Rome IT 00178 - Italy Phone: +39 06 9201961 Establishment Registration Number: 8010312 Contact: Mr. Massimiliano Improta
### Correspondent/Consultant:
Guido Bonapace Isemed Srl Via P. Togliatti. 19/X, Imola, BO 40026 ITALY Mob. +39 335 537 8686 Tel. +39 0542 683803 Fax +39 0542 698456 Email: regulatory@isemed.eu
### Date prepared: 9th October, 2024
#### Device Name
Device Trade Name: TRAP EASY Common Name: Gastroenterology-urology biopsy instrument Classification name: Instrument, Biopsy Regulation Number: 876.1075 Product Code: KNW
#### Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name | Product code |
|-------------|----------------------|--------------|
| K033616 | TRAPMATIC SYSTEM SET | KNW |
#### Device Description Summary
Trap Easy is a bone marrow biopsy needle, consisting of a cannula with cylindrical tip and ergonomic handle, a removable pyramidal stylet, a sample trapping device integrated into the cannula and an extraction wire.
The device is used for drawing osteomedullary substance and/or for bone marrow explantation. This device is intended to be used only by experienced and properly trained physicians and is intended to be used exclusively for bone marrow tissue biopsies.
The device is available in different models:
- 8G x 100 mm
- 8G x 150 mm
- 11G x 100 mm
- 11G x 150 mm
{5}------------------------------------------------
# HS HOSPITAL SERVICE SPA K242029
Trap Easy is supplied in sterile packaging and is designed for single use, non-reusable. At the end of its use, the device should be disposed of as infectious hospital waste.
# Intended use / Indications for Use
The device is used for drawing osteomedullary substance and/or for bone marrow explantation.
## Indications for Use Comparison
Indications for use of subject device are the same of primary predicate devices are intended to be used in the procedure for harvesting bone marrow specimens.
Both subject device and predicate devices are intended to be used by a physician.
# Technological Comparison
The Trap Easy medical device has the same technological characteristics, design, material and principle of operation as predicate device Trapmatic System set (K033616). As a matter of facts, both the Trap Easy medical device and the Trapmatic System set consist of an outer cannula with handle and trapping element inside, and an extraction wire. As accessories, both devices possess a safety and luer lock cap.
Trap Easy medical device is characterized by the same principle of operation of primary predicate: both devices trap the sample inside and both devices are characterized by a safety and an activation mechanism of the trap. For the Trap Easy medical device, the safety is represented by the white button, while in the Trapmatic System, this safety feature is the white drawer. Then, to activate the mechanism, subject device is equipped with a lever on the side, while in the Trapmatic device, you need to push down the luer connector. Therefore, the only difference with the primary predicate consists in the safety feature that is present for the activation of the mechanism of action: in the Trap Easy this safety feature is represented by a white button, while in the Trapmatic System, it consists in a white drawer. The devices can therefore be considered substantially equivalent.
Subject device is available in sizes 8G and 11G, which fall within those available for the primary predicate device. In addition, the devices are made of the same materials: both consists of AISI 304 stainless steel cannula and ABS handle.
Both devices are manual instrument, provided for single-use and sterilized with Eto, therefore being substantially equivalent.
The devices can be considered substantially equivalent as the differences do not raise concerns about safety or effectiveness.
## Non- Clinical and/or Clinical Tests Summary & Conclusions
Performance of subject device have been evaluated by testing sample integrity after biopsy, sample length of single biopsy, sample diameter, sample removal and by testing mechanical functioning of the needle.
The tests have been conducted in a manner as similar as possible to how the biopsy device is used in a medical procedure.
Performed tests show that all the 50 samples have a good integrity and there were no samples ruined or divided in small parts. Moreover, the sample removal was easy for all samples and the needles didn't show any damages to the internal mechanism after all biopsies. The average lengths and diameters were compliant with acceptance criteria.
{6}------------------------------------------------
# HS HOSPITAL SERVICE SPA K242029
Obtained results demonstrated that the device's performances are maintained for the intended use of the device.
In addition, in order to guarantee performance and compliance, devices are subjected to inspections after the manufacturing process.
Nonclinical testing data demonstrate the technological characteristics of subject device Trap Easy and that the medical device perform as intended for its use.
The medical device Trap Easy is sterilized using Ethylene Oxide (EtO), and the sterilization process is validated in accordance with the requirements of ISO 11135:2014/A1:2018. The manufacturer, HS Hospital Service Spa, produces several sterile medical devices, all of which fall under the same process validation. These devices are categorized into four different families based on product type. The sterilization process has been validated and is periodically requalified to ensure the continued ability to effectively sterilize the products.
Additionally, the limits for Ethylene Oxide (EtO) residues and its by-products, Ethylene Glycol (EG) and Ethylene Chlorohydrin (ECH), are defined in accordance with ISO 10993-7, taking into account the limited exposure of the medical device.
Moreover, the Trap Easy medical device is available in two different packaging configurations that serve as the sterile barrier system for the product:
- A pouch made of Tyvek® 2FS and an OPA/PE film (16x44 cm) ●
- A blister made of Coated Tyvek® CR27 2FS and PET+PE600 (12x22.5 cm) .
Both packaging configurations have been validated in compliance with the requirements of ISO 11607-1:2019.
Regarding the biocompatibility of the medical device, based on its characteristics, Trap Easy is classified according to Annex A of ISO 10993-1 as an externally communicating medical device with limited (< 24 hours) contact with bone/tissue. The material in direct contact is AISI 304 stainless steel, which is used for both the needle (cannula) and the mandrel (stylet). AISI 304 stainless steel is a well-established material, historically used in the manufacturing of this type of medical device due to its proven safety and performance.
Additionally, Trap Easy in its final finished form is identical to the Trapmatic System set in terms of formulation, processing, and sterilization, and no other chemical have been added.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.