The Single-use Bronchoscopes have been designed to be used with the Electronic Endoscope Imaging Processor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree. The Bronchoscope System is for use in a hospital environment. The Single-use Bronchoscope is a single-use device designed for use in adults.
Device Story
System comprises sterile single-use flexible bronchoscope and reusable Electronic Endoscope Imaging Processor. Distal tip of bronchoscope contains LED light source and CMOS image sensor. CMOS captures reflected light from body cavity; drive circuit controls acquisition; encoding circuit outputs RGB video signal to processor. Processor receives, processes, and transmits video to external medical monitor; also controls LED brightness. Used in hospital environment by trained surgical physicians for airway/tracheobronchial endoscopy. Provides real-time visualization to assist in diagnosis or therapeutic procedures. Benefits include single-use sterility, eliminating cross-contamination risk.
Clinical Evidence
No clinical test data was used to support the decision of substantial equivalence. Bench testing only.
Technological Characteristics
Flexible bronchoscope with CMOS sensor and LED light source. Materials: Pebax, polyamide, stainless steel, epoxy glue, glass, polycarbonate, PTFE, PET. Connectivity: Wired connection to Electronic Endoscope Imaging Processor. Sterilization: Ethylene oxide (SAL 10^-6). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62366-1, IEC 62471, ISO 10993-1/5/10/11/23, ISO 11607-1, ISO 11135, ISO 10993-7.
Indications for Use
Indicated for use in adults for endoscopic diagnosis or treatment within the airways and tracheobronchial tree in a hospital environment.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Single-use Bronchoscope, Monitor for Single-Use Endoscopy, Bronchoscope System (K231107)
{0}------------------------------------------------
March 14, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Shenzhen Sanping Image Technology Co., Ltd. Vincent Pan General Manager Room 203, No.634, Shajing Road, Buchong Community, Shajing Street, Bao'an District Shenzhen, Guangdong 518104 China
Re: K242010
Trade/Device Name: Bronchoscope System, The Single-use Bronchoscope (Single-use Bronchoscope: SP-Y-ZQGJ2.8, SP-Y-ZQGJ3.8, SP-Y-ZQGJ4.2, SP-Y-ZQGJ4.8, SP-Y-ZQGJ5.2, SP-Y-ZQGJ5.8; Electronic Endoscope Imaging Processor: SP-TXCL012.1) Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: July 10, 2024 Received: July 10, 2024
Dear Pan Vincent:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
### K242010
Device Name
Bronchoscope System、The Single-use Bronchoscope (Single-use Bronchoscope: SP-Y-ZQGJ2.8、SP-Y-ZQGJ3.8、SP-Y-ZQGJ4.2、SP-Y-ZQGJ4.8、SP-Y-ZQGJ5.2、SP-Y-ZQGJ5.8; Electronic Endoscope Imaging Processor: SP-TXCLQ12.1)
#### Indications for Use (Describe)
The Single-use Bronchoscopes have been designed to be used with the Electronic Endoscope Imaging Processor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The Bronchoscope System is for use in a hospital environment. The Single-use is a single-use device designed for use in adults.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
**PRA Statement below**
{4}------------------------------------------------
# 510(k) Summary
### 1. Sponsor Identification:
- Company Name: Shenzhen Sanping Image Technology Co., Ltd. >
- Address: Room 203, No.634, Shajing Road, Buchong Community, Shajing Street, > Bao'an District, Shenzhen, China
- Phone: +86-755-23223975 >
- Fax: +86-755-23223975 >
- > Contact Person (Title): Vincent Pan (General Manager)
- > E-mail: 99020278@qq.com
- > Date of Preparation: December 26, 2024
### 2. Name of the Device:
| Trade Name | Bronchoscope<br>System、The<br>Single-use<br>Bronchoscope | Model<br>Number/Name | Single-use Bronchoscope:<br>SP-Y-ZQGJ2.8、<br>SP-Y-ZQGJ3.8、<br>SP-Y-ZQGJ4.2、<br>SP-Y-ZQGJ4.8、<br>SP-Y-ZQGJ5.2、<br>SP-Y-ZQGJ5.8 |
|------------|----------------------------------------------------------|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| | Model<br>Number/Name | Electronic Endoscope Imaging<br>Processor: SP-TXCLQ12.1 | |
### 3. Common Name and Classification:
- > Device Classification Name: Bronchoscope (Flexible Or Rigid)
- > Classification Product Code: EOQ
{5}------------------------------------------------
- > Regulation Number: 21 CFR 874.4680
- > Class: 2
- > Review Panel: Ear Nose & Throat
- Trade/Proprietary Name: The Single-use Bronchoscopes、Bronchoscope System >
- > Common Name: Bronchoscope (flexible or rigid) and accessories
#### Predicate Device Information1: 4.
- > 510(k) Number: K231107
- Device Classification Name: Bronchoscope (Flexible Or Rigid) >
- > Sponsor: Hunan Endoso Life Technology Co., Ltd.
- Classification Product Code: EOQ >
- > Regulation Number: 21 CFR 874.4680
- > Class: 2
- > Review Panel: Ear Nose & Throat
- Trade/Proprietary Name: Single-use Bronchoscope, Monitor for Single-Use Endoscopy, > Bronchoscope System
- > Common Name: Bronchoscope (flexible or rigid) and accessories
#### Predicate Device Information2: 5.
- > 510(k) Number: K191828
- > Device Classification Name: Bronchoscope (Flexible Or Rigid)
- Sponsor: Hunan Vathin Medical Instrument Co., Ltd. >
- > Classification Product Code: EOQ
- > Regulation Number: 21 CFR 874.4680
- > Class: 2
- > Review Panel: Ear Nose & Throat
- > Trade/Proprietary Name: Vathin® Video Bronchoscope System
- > Common Name: Bronchoscope (flexible or rigid) and accessories
{6}------------------------------------------------
### 6. Device Description
The Single-use Bronchoscope is a sterile single used flexible bronchoscope. The Electronic Endoscope Imaging Processor is a reusable monitor.
The light emitted by the LED cold light source at the distal tip of the Single-use Bronchoscope is irradiated into the body cavity, and the light reflected from the cavity enters the optical system and is captured by the CMOS image sensor. The CMOS acquisition image is controlled by the CMOS drive circuit, and the RGB video signal is output of the Electronic Endoscope Imaging Processor via the encoding circuit. The Electronic Endoscope Imaging Processor receives video signals from the endoscope, processes the video signals, and outputs the processed video signal to the monitor. The Electronic Endoscope Imaging Processor also controls the brightness of the LEDs on the endoscope.
The optical components and their arrangement at the distal tip for all models of the Single-use Bronchoscope are identical.
The Single-use Bronchoscope is single-use medical devices that are sterilized with ethylene oxide and intended for use in adults. This product is accompanied by an electronic endoscopic image processor, which supports the use of medical monitors. The electronic bronchoscope system is used for endoscopic diagnosis or treatment, and the processor collects, processes, stores, and transmits images of the human body cavity to the monitor.
### 7. Indications for Use
The Single-use Bronchoscopes have been designed to be used with the Electronic Endoscope Imaging Processor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The Bronchoscope System is for use in a hospital environment. The Single-use Bronchoscope is a single-use device designed for use in adults.
3 page
{7}------------------------------------------------
## 8. Comparison to the predicate device
Table 1 General Comparison
| Elements of<br>Comparison | Proposed Device | Predicate Device 1<br>K231107 | Predicate Device 2<br>K191828 | Judgment | Review Panel | Ear Nose & Throat | Ear Nose & Throat | Ear Nose & Throat | SE |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|------------------------|--------------------------------------|--------------------------------------|--------------------------------------|----|
| Company<br>Name | Shenzhen Sanping Image<br>Technology Co., Ltd. | Hunan Endoso Life<br>Technology Co., Ltd. | Hunan Endoso Life<br>Technology Co., Ltd. | / | Class | 2 | 2 | 2 | SE |
| Model | Single-use Bronchoscope:<br>SP-Y-ZQGJ2.8、<br>SP-Y-ZQGJ3.8、<br>SP-Y-ZQGJ4.2、<br>SP-Y-ZQGJ4.8、<br>SP-Y-ZQGJ5.2、<br>SP-Y-ZQGJ5.8<br>Electronic Endoscope<br>Imaging Processor:<br>SP-TXCLQ12.1 | Single-use Bronchoscope:<br>BC12-1, BC12-2, BC20-1,<br>BC20-2, BC22-1, BC22-2,<br>BC22-3, BC28-1, BC28-2,<br>BC28-3<br>Monitor for Single-Use<br>Endoscopy: P1-A, P1-B,<br>P1-C, P1-D, P1-E, P2-A,<br>P2-B, P2-C, P2-D, P2-E | Vathin®H-SteriScope™<br>Single-use flexible Video<br>Bronchoscope<br>BCV1-01 BCV1-02<br>BCV1-C1<br>BCV1-C2 BCV1-H1<br>BCV1-H2<br>BCV1-K1 BCV1-K2<br>BCV1-M1<br>BCV1-M2 BCV1-01<br>BCV1-02<br>BCV1-S1 BCV1-S2<br>BCV1-U1<br>BCV1-U2 BCV1-W1<br>BCV1-W2<br>Vathin®VisionCenter™ <br>Digital<br>Video Processor<br>DVP-A1 | / | Prescription<br>or OTC | Prescription Use | Prescription Use | Prescription Use | SE |
| Device Name | Bronchoscope System | Bronchoscope System | Vathin® Video<br>Bronchoscope<br>System | / | Anatomic<br>sites | airways and<br>tracheobronchial tree | airways and<br>tracheobronchial tree | airways and<br>tracheobronchial tree | SE |
| Classification<br>Product Code | EOQ | EOQ | EOQ | SE | Target<br>Population | Adults | Adults | Adults | SE |
| Regulation | 874.4680 | 874.4680 | 874.4680 | SE | Where used | Hospital | Hospital | Hospital | SE |
| Classification<br>Name | Bronchoscope (Flexible Or<br>Rigid) | Bronchoscope (Flexible<br>Or Rigid) | Bronchoscope (Flexible Or<br>Rigid) | SE | | | | | |
{8}------------------------------------------------
{9}------------------------------------------------
| Working<br>place/User | Use in a hospital<br>environment by trained<br>surgical physicians who are<br>familiar with endoscopic<br>procedures. | Use in a hospital<br>environment by trained<br>surgical physicians who are<br>familiar with endoscopic<br>procedures. | Use in a hospital<br>environment by trained<br>surgical physicians who<br>are familiar with<br>endoscopic<br>procedures. | SE |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|
| Technology | The Single-use<br>bronchoscope is inserted<br>through the airways and<br>tracheobronchial tree<br>during Bronchoscopy.<br>Anatomical images are<br>transmitted to the user by<br>the Monitor for Single-Use<br>Endoscopy with a CMOS<br>chip at the distal end of the<br>endoscope and the images<br>showing on the monitor. | The Single-use<br>bronchoscope is inserted<br>through the airways and<br>tracheobronchial tree<br>during Bronchoscopy.<br>Anatomical images are<br>transmitted to the user by<br>the Monitor for Single-Use<br>Endoscopy with a CMOS<br>chip at the distal end of the<br>endoscope and the images<br>showing on the monitor. | The Single-use<br>bronchoscope is inserted<br>through the airways and<br>tracheobronchial tree<br>during Bronchoscopy.<br>Anatomical images are<br>transmitted to the user by<br>the Monitor for Single-Use<br>Endoscopy with a CMOS<br>chip at the distal end of the<br>endoscope and the<br>images showing on the<br>monitor. | SE |
{10}------------------------------------------------
| | | | The<br>Vathin®H-SteriScopeTM I | |
|--------------|---------------------------------------------------|--------------------------------------------------------|-----------------------------------------------------|----|
| | The Single-use<br>Bronchoscopes have been | The Single-use<br>Bronchoscopes have been | Single-use flexible Video<br>Bronchoscope have been | |
| | designed to be used with | designed to be used with<br>the Monitor for Single-Use | designed to be used with | |
| | the Electronic Endoscope | Endoscopy, endotherapy | the | |
| | Imaging Processor, | accessories and other | Vathin®VisionCenterTM I | |
| | endotherapy accessories | ancillary equipment for | Digital Video Processor, | |
| | and other ancillary | endoscopy within the | endotherapy accessories | SE |
| Intended Use | equipment for endoscopy<br>within the airways and | airways and | and other ancillary | |
| | tracheobronchial tree. | tracheobronchial tree. | equipment for endoscopy<br>within the airways and | |
| | The Bronchoscope System | The Bronchoscope System | tracheobronchial tree. | |
| | is for use in a hospital<br>environment. The | is for use in a hospital<br>environment. The | The Vathin® Video | |
| | Single-use Bronchoscope<br>is a single-use device | Single-use Bronchoscope<br>is a disposable device | Bronchoscope System is | |
| | designed for use in adults. | designed for use in adults. | for use in a hospital<br>environment. | |
### Table 2 Safety factor & Performance Comparison
| Safety factor &<br>Performance | Proposed Device | Predicate Device1 | Predicate Device2 | Judgment |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|----------|
| Electrical Safety | Compliance with IEC<br>60601-1 | Compliance with<br>IEC 60601-1 | Compliance with IEC<br>60601-1 | SE |
| EMC | Compliance with IEC<br>60601-1-2 | Compliance with<br>IEC 60601-1-2 | Compliance with IEC<br>60601-1-2 | SE |
| Usability<br>Engineering | Compliance with IEC<br>62366-1 | Compliance with<br>IEC 62366-1 | Compliance with IEC<br>62366-1 | SE |
| Biocompatiblity | Compliance with ISO<br>10993-1 | Compliance with<br>ISO 10993-1 | Compliance with ISO<br>10993-1 | SE |
| Safety factor &<br>Performance | Proposed Device | Predicate Device1 | Predicate Device2 | Judgment |
| Connect to<br>devices | Monitor | Monitor | Monitor | SE |
| Type of Imager | CMOS | CMOS | CMOS | SE |
| Light Source | Internal LED | Internal LED | Internal LED | SE |
| Direction of<br>View (°) | forward | forward | forward | SE |
| Working Length<br>(mm) | 650 | 600mm | 600/700mm | Note 1 |
| Depth of Field (mm) | 3 - 50 | 3 - 50 | 3mm~30mm | SE |
| Type of Scope | Flexible | Flexible | Flexible | SE |
| Distal end<br>outer diameter | <2.9mm (8.7 Fr) | <2.9mm (8.7 Fr) | <3.1mm (9.3 Fr) | SE |
| Deflection Angle | Up & Down 180°, tolerance<br>±10% | Up & Down 180°,<br>tolerance ±10% | Up: 210<br>Down: 210 | SE |
| Distal end<br>diameter(mm) | 3.00/4.05/4.40/4.80/5.20/6.<br>10 | / | 2.2/3.2/4.1/4.7/4.9/5.<br>2/5.8/6.<br>0/6.2 | Note 2 |
| Maximum<br>insertion portion<br>width(mm) | 3.20/4.25/4.55/5.40/5.70/6.<br>35 | / | 2.2/3.2/4.1/4.7/4.9/5<br>2/5.8/6.<br>0/6.2 | Note 2 |
| Minimum<br>insertion channel<br>width(mm) | 1.20/1.60/2.20/2.60/3.00/3.<br>20 | / | 0/1.2/1.7/2.0/2.2/2.4/<br>2.8/3.0<br>/3.2 | Note 2 |
| General material<br>type of main<br>patient-contact | Compliance with ISO10993<br>- 1 | Compliance with<br>ISO10993 - 1 | Compliance with<br>ISO10993 - 1 | SE |
| General<br>material type<br>of main<br>patient-contact<br>part | Pebax, polyamide,<br>Stainless<br>Steel, Epoxy glue, glass,<br>Polycarbonate, PTFE, PET | Pebax, polyamide,<br>Stainless<br>Steel, Epoxy glue,<br>glass,<br>Polycarbonate,<br>PTFE, PET | Pebax, PTFE, LCP,<br>TPU,<br>Fluoro elastomers,<br>Epoxy glue, glass | SE |
| Safety factor &<br>Performance | Proposed Device | Predicate Device1 | Predicate Device2 | Judgment |
| Duration and type<br>of contact | "Surface - Mucosal<br>Membrane" with a contact<br>duration of "Limited (< 24<br>hours)" | "Surface - Mucosal<br>Membrane" with a<br>contact duration of<br>"Limited (< 24<br>hours)" | "Surface - Mucosal<br>Membrane" with a<br>contact duration of<br>"Limited (< 24<br>hours)" | SE |
| Number of Users | Single-Use | Single-Use | Single-Use | SE |
| Sterilization | EO Sterilized,<br>SAL 10-6 | EO Sterilized, SAL<br>10-6 | EO Sterilized, SAL<br>10-6 | SE |
{11}------------------------------------------------
{12}------------------------------------------------
### 8.1 Review of Differences :
#### Note1:
There is slight difference in the working length as compared to the predicate. However, the working length of the device is utilized by many similar devices in clinical practice. Therefore, this variance does not raise different questions of its safety and effectiveness.
### Note 2:
The difference lies in the mechanical specifications, there may be slight size differences within the same specification and all within the tolerance range. All the performance was tested, and the results met the standard requirements, these differences will not raise any issues in safety and effectiveness.
### 8.2 Discussion of Non-clinical Tests Performed for Determination of Substantial Equivalence are as follows:
A series of non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
{13}------------------------------------------------
### 1) Electrical safety
IEC6060-1:2005+A1:2012+A2:2020, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.
### 2) Electromagnetic compatibility (EMC)
IEC 60601-1-2:2014 + A1:2020 / ANSI AAMI IEC 60601-1-2:2014 + A1:2021], Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests.
IEC 60601-2-18:2009, Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
#### 3) Basic Safety and Essential Performance
IEC 60601-1-6:2010 + A1:2020 + A2:2020, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
### 4) Photobiological safety
IEC 62471:2006, Medical electrical equipment, Photobiological safety of lamps and lamp systems.
### 5) Biocompatibility testing
ISO 10993-1:2018 Biological evaluation of medical devices-Parts 1: Evaluation and testing.
ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021 Biological evaluation of medical devices - Parts 10: Tests for skin sensitization
ISO 10993-11:2017 Biological evaluation of medical devices, Part 11: Tests for
{14}------------------------------------------------
systemic toxicity
ISO 10993-23:2021 Biological evaluation of medical devices-Parts 10: Tests for irritation
### 6) Sterilization and Shelf Life & Packaging testing
ISO 11607-1:2019, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.
ISO 11135:2014, Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices.
ISO 10993-7: 2019, Biological evaluation of medical devices: Part 7-Ehylene oxide Sterilization Residuals [Including: Technical Corrigendum 1 (2009), AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019).
The system performance tests were performed including validation of up/down deflection、repeated flexing,water supply system,sealing performance,normal working condition, field of View, direction of view (DOV)、geometric distortion, Signal-To-Noise Ratio, Image Intensity Uniformity (IIU)、and color performance.
All the test results demonstrate Bronchoscope System meets the requirements of its pre-defined acceptance criteria and intended uses, and is substantially equivalent to the predicate device.
### 9. Clinical Tests Performed
No clinical test data was used to support the decision of substantial equivalence.
{15}------------------------------------------------
## 10. Conclusion
The subject devices have all features of the predicate devices. The few differences were evaluated and do not affect the safety and effectiveness of the subject devices compared to the predicate.
Thus, the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.