The HydroCision SpineSite System is intended to be used as a single-use endoscopic video camera in a variety of endoscopic diagnostic and surgical procedures for spine applications. The device is also intended to be used as an accessory for microscopic surgery.
Device Story
The HydroCision SpineSite System is an endoscopic visualization platform consisting of a sterile, single-use disposable endoscope and a reusable Video Processing Unit (VPU). The endoscope features distal LED illumination and a high-resolution video sensor; it includes a working channel for micro-instrumentation. The endoscope connects to the VPU via a proprietary edge card connector, which provides power and supports video processing. The VPU processes the video signal for display on a user-provided monitor and digital documentation. The system is used in clinical settings for spine-related diagnostic and surgical procedures. Healthcare providers use the real-time video output to visualize the surgical site, facilitating minimally invasive interventions. The device is not MR-safe.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), mechanical and optical design verification/validation, electrical/thermal safety (IEC 60601-1, IEC 60601-2-18), human factors/usability (IEC 60601-1-6), and electromagnetic compatibility (IEC 60601-1-2).
Technological Characteristics
System comprises a sterile, single-use endoscope and a reusable VPU. Endoscope uses LED illumination and a high-resolution video sensor. Materials meet ISO 10993-1 for limited duration contact (<24 hrs). Connectivity via proprietary edge card connector; VPU outputs via HDMI/USB. Sterilization of endoscope via ethylene oxide. Electrical safety per IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2.
Indications for Use
Indicated for use as a single-use endoscopic video camera in endoscopic diagnostic and surgical procedures for spine applications and as an accessory for microscopic surgery.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 30, 2025
HydroCision, Inc.
Mark Lewis
Vice President Operations and Regulatory Affairs
267 Boston Road
Suite 28
North Billerica, Massachusetts 01862
Re: K241990
Trade/Device Name: HydroCision SpineSite System
Regulation Number: 21 CFR 888.1100
Regulation Name: Arthroscope
Regulatory Class: Class II
Product Code: HRX
Dated: July 2, 2025
Received: July 2, 2025
Dear Mark Lewis:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K241990 - Mark Lewis
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K241990 - Mark Lewis
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSE MUIR
-S
Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K241990 | |
| Device Name | |
| HydroCision SpineSite System | |
| Indications for Use (Describe) | |
| The HydroCision SpineSite System is intended to be used as a single-use endoscopic video camera in a variety of endoscopic diagnostic and surgical procedures for spine applications. The device is also intended to be used as an accessory for microscopic surgery. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K241990
# 510(k) Summary
# HydroCision SpineSite System
Date Prepared: July 29, 2025
## Submitter Information
Submitter: HydroCision™, Inc.
Address: 267 Boston Rd, North Billerica, MA 01862
Contact Person: Mr. Richard Lytle
Chief Executive Officer
Phone: +1 (630) 779-1723
Email: rlytle@hydrocision.com
## Device Information
Device Proprietary Name: HydroCision SpineSite System
Common Name: Arthroscope
Device Classification: Class II
Regulation: 21 CFR 880.1100
Classification Name: Arthroscope and accessories
Classification Panel: Orthopedic
Product Code: HRX
Predicate Device: Arthrex Nanoscope System, K201134
Product Codes: GCJ, HRX
Classifications: 876.1500 and 888.1100
## Purpose of Submission:
The purpose of this Traditional 510(k) submission is to gain regulatory clearance per section 510(k) of the Federal Food, Drug and Cosmetic Act for the HydroCision SpineSite System, consisting of a sterile, single-use, disposable endoscope and a reusable, non-sterile capital equipment Video Processing Unit (VPU).
## Device Description:
The HydroCision SpineSite System is comprised of (i) sterile disposable endoscope and (ii) reusable Video Processing Unit (VPU). The HydroCision SpineSite System provides illumination, image processing and digital documentation for endoscopic procedures. The HydroCision SpineSite System is not suitable for use in the MR environment.
Page 1 of 4
{5}
K241990
The SpineSite Endoscope provides distal LED illumination via LEDs surrounding a high-resolution video sensor. The SpineSite Endoscope contains a working channel for the passage of micro instrumentation to the surgical site. The SpineSite Endoscope is provided sterile, via ethylene oxide sterilization.
The SpineSite Endoscope is designed to be connected to the SpineSite VPU via a proprietary edge card connector which provides power to the endoscope and supports video processing capability. The SpineSite VPU is powered via connection to an external wall outlet via a 12V power adapter.
## Indications for Use:
The HydroCision SpineSite System is intended to be used as a single-use endoscopic video camera in a variety of endoscopic diagnostic and surgical procedures for spine applications. The device is also intended to be used as an accessory for microscopic surgery.
## Performance Testing:
The following non-clinical testing was performed on the HydroCision SpineSite System to confirm the device meets product specification and performance requirements:
- Biocompatibility per ISO 10993-1
- Design verification/validation to mechanical and optical specifications
- Electrical, Mechanical and Thermal (EMT) safety testing per IEC 60601-1, IEC 60601-2-18
- Human Factors/Usability per IEC 60601-1-6
- Electromagnetic compatibility testing per IEC 60601-1-2
- Software validation
## Substantial Equivalence Summary:
The HydroCision SpineSite System is substantially equivalent in intended use, technological design, performance and principles of operation to the predicate device; Arthrex Nanoscope System (K201134). Both systems include the same principal components- (i) a single use, disposable endoscopic camera system and (ii) video processing technology. The disposable endoscope component of both devices consists of a plastic handle, LED light transmission, image processing and digital image documentation capability. The HydroCision SpineSite System has the same intended use as the predicate device as a video camera system providing illumination and visualization at the surgical site. The indications for use for the HydroCision SpineSite Endoscope System represent a subset of the cleared indications for the predicate Arthrex Nanoscope System. As a subset of the indications for the cleared device, this difference does not raise any new risks or issues related to safety or efficacy.
The HydroCision SpineSite System and the predicate device consist of the same technology and are sterilized with same acceptable method. Both the devices are fabricated of biocompatible materials and comply with applicable electrical safety standards.
Below is a comparison table that provides a top-level overview of the substantial equivalent comparison between proposed and predicate devices.
| Comparative Characteristics | Proposed Device: HydroCision SpineSite System | Predicate Device: Arthrex Nanoscope System (K201134) |
| --- | --- | --- |
| Single Use | SpineSite Endoscope is single-use disposable | Nanoscope is single-use disposable |
{6}
K241990
| Comparative Characteristics | Proposed Device: HydroCision SpineSite System | Predicate Device: Arthrex Nanoscope System (K201134) |
| --- | --- | --- |
| | SpineSite Video Processing Unit – reusable, does not enter sterile field | Nanoscope Console – reusable, does not enter sterile field |
| Principle System Components (Sterile, Disposable, Single Use) | SpineSite Endoscope | Nanoscope |
| Principle System Components (Reusable) | SpineSite Video Processing Unit | Nanoscope Control Unit |
| Sterilization | SpineSite Endoscope Supplied Sterile, EO Sterilization | Nanoscope Supplied Sterile, EO Sterilization |
| Electrical Safety | | |
| Compliance to applicable Standards | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18 | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18 |
| Materials | | |
| Biocompatibility of Materials | Meets ISO 10993-1 requirements for tissue/bone/dentin contact of limited duration (<24 hrs) | Meets ISO 10993-1 requirements for tissue/bone/dentin contact of limited duration (<24 hrs) |
| Technical Features/Design - Endoscope | | |
| Optical Resolution | 400 × 400 pixels | 400 × 400 pixels |
| Field of View | 120° | 120° |
| Angle of View | 0° | 0° |
| Depth of field | 5 mm-50 mm | 3 mm × 100 mm |
| Working Length | 200 mm | 125 mm, 180 mm, and 250 mm lengths |
| Overall Diameter | 5.0 mm | Scope diameter 1.9mm; utilizes accessory scope sheath of up to 3.4 mm diameter |
| Packaging Description | Blister tray, Tyvek lid seal, Tyvek pouch, unit box. | Sealed mylar/Tyvek pouch or a PETG tray with a sealed Tyvek lid. |
| Technical Features/Design – Video Processing Unit | | |
| Control Unit | Reusable video processing unit | Reusable 13 in HD monitor/console |
| Endoscope connection | Via edge card connector from endoscope | Via edge card connector from endoscope |
| Output connectors | HDMI, USB ports | Ethernet, HDMI, USB ports |
| Power Source | 110 V via 12 V power adapter | Universal input 110/240V medical grade - 19 V /110 W |
| Audio capability | n/a | microphone |
| Internet Connectivity | n/a | Ethernet and Wi-Fi |
Page 3 of 4
{7}
K241990
| Comparative Characteristics | Proposed Device: HydroCision SpineSite System | Predicate Device: Arthrex Nanoscope System (K201134) |
| --- | --- | --- |
| Image Display | User provided monitor | 11.5 inches x 6.5 inches |
Conclusion:
The proposed HydroCision SpineSite System is substantially equivalent to the predicate Arthrex Nanoscope System (K201134). The safety and effectiveness of the HydroCision SpineSite System are adequately supported by the non-clinical performance data, substantial equivalence information, and comparison of design characteristics provided within this premarket notification. Any differences between the proposed device and the predicate device are considered minor and do not raise new questions concerning safety or effectiveness.
Based on intended use, technological characteristics and the summary of data submitted in this premarket notification, the HydroCision SpineSite System is determined to be substantially equivalent to the predicate Arthrex Nanoscope device.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.