Single-Use Video Hysteroscope (RP-G-C24, RP-G-C0101)
K241987 · Guangzhou Red Pine Medical Instrument Co., Ltd. · HIH · Aug 2, 2024 · Obstetrics/Gynecology
Device Facts
Record ID
K241987
Device Name
Single-Use Video Hysteroscope (RP-G-C24, RP-G-C0101)
Applicant
Guangzhou Red Pine Medical Instrument Co., Ltd.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Aug 2, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1690
Device Class
Class 2
Indications for Use
Single-Use Video Hysteroscope is intended to be used for viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures. The device is suitable for professional healthcare facility environments such as hospitals and clinics.
Device Story
Single-Use Video Hysteroscope consists of an endoscope with an insertion portion and handle; includes two-way valves and silicone cap accessories. Device provides visualization of adult cervical canal and uterine cavity for diagnostic and operative procedures. Used in hospitals and clinics by professional healthcare providers. Operates as a video hysteroscope to capture and display internal images. Healthcare providers view output on a connected display to guide procedures and inform clinical decision-making. Single-use design eliminates reprocessing requirements, reducing cross-contamination risk.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including mechanical performance, optical performance (DOF), biocompatibility (ISO 10993), sterilization validation (ISO 11135), and packaging/shelf-life integrity (ASTM/ISO standards).
Technological Characteristics
Single-use video endoscope. Materials evaluated per ISO 10993-1. Sterilization via ethylene oxide (ISO 11135). Connectivity via Luer taper (ISO 80369-7/20). Depth of field: 3mm-50mm (RP-G-C0101). Accessories: two-way valves, silicone cap. Packaging validated per ISO 11607 and ASTM standards.
Indications for Use
Indicated for viewing the adult cervical canal and uterine cavity for diagnostic and operative procedures in professional healthcare facilities.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
Single-Use Video Hysteroscope: RP-G-C24, RP-G-C0101 (K232003)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
August 2, 2024
Guangzhou Red Pine Medical Instrument Co., Ltd. Ping Yang Regulatory Affairs Manager 12 F, No.87 Luoxuan Avenue, Guangzhou International Bioisland, Huangpu District Guangzhou, Guangdong 510000 CHINA
Re: K241987
Trade/Device Name: Single-Use Video Hysteroscope (RP-G-C24, RP-G-C0101) Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: July 8, 2024 Received: July 8, 2024
Dear Ping Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Reginald K. Avery -S
for Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241987
Device Name
Single-Use Video Hysteroscope (RP-G-C24, RP-G-C0101)
Indications for Use (Describe)
Single-Use Video Hysteroscope is intended to be used for viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures.
The device is suitable for professional healthcare facility environments such as hospitals and clinics.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <div> <object data="data:image/svg+xml;base64,PHN2ZyB4bWxucz0iaHR0cDovL3d3dy53My5vcmcvMjAwMC9zdmciIHdpZHRoPSIyMCIgaGVpZ2h0PSIyMCI+PHJlY3Qgd2lkdGg9IjIwIiBoZWlnaHQ9IjIwIiBmaWxsPSJ3aGl0ZSIgc3Ryb2tlPSIjMDAwIiAvPjxwYXRoIGQ9Ik0gMywzIEwgMTcsMTcgTSAzLDE3IEwgMTcsMyIgZmlsbD0ibm9uZSIgc3Ryb2tlPSIjMDAwIiAvPjwvc3ZnPg==" type="image/svg+xml"></object> Prescription Use (Part 21 CFR 801 Subpart D) </div> </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <div> <object data="data:image/svg+xml;base64,PHN2ZyB4bWxucz0iaHR0cDovL3d3dy53My5vcmcvMjAwMC9zdmciIHdpZHRoPSIyMCIgaGVpZ2h0PSIyMCI+PHJlY3Qgd2lkdGg9IjIwIiBoZWlnaHQ9IjIwIiBmaWxsPSJ3aGl0ZSIgc3Ryb2tlPSIjMDAwIiAvPjwvc3ZnPg==" type="image/svg+xml"></object> Over-The-Counter Use (21 CFR 801 Subpart C) </div> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K241987 Page 1 / 4
# 510(k) SUMMARY
### 510(k) Number: K241987
#### I. Contact Details
Submitter Name: Guangzhou Red Pine Medical Instrument Co., Ltd.
Submitter Address: 12 F, No.87 Luoxuan Avenue, Guangzhou International Bioisland, Huangpu District Guangzhou Guangdong 510000 China
Submitter Contact Telephone: +86 13902971205
Submitter Contact: Mr. Weihua Yang
Submitter Contact Email: regulation(@gzredpine.com
Correspondent Name: Guangzhou Red Pine Medical Instrument Co., Ltd.
Correspondent Address: 12 F, No.87 Luoxuan Avenue, Guangzhou International Bioisland, Huangpu District Guangzhou Guangdong 510000 China
Correspondent Contact Telephone: +86 18565651640
Correspondent Contact: Mrs. Ping Yang
Correspondent Contact Email: regulation(@gzredpine.com
Date prepared: August 2, 2024
#### II. Subject Device
Device Trade Name: Single-Use Video Hysteroscope (RP-G-C24, RP-G-C0101)
Common Name: Hysteroscope and accessories
Classification Number: 21 CFR 884.1690
Classification Name: Hysteroscope and accessories
Product Code: HIH
#### III. Legally Marketed Predicate Device
Predicate 510(k) Number: K232003 Device name: Single-Use Video Hysteroscope: RP-G-C24, RP-G-C0101 Product Code: HIH Note: This predicate device has not been subject to a design-related recall.
{5}------------------------------------------------
### IV. Device Description Summary
The subject device, Single-Use Video Hysteroscope is intended to be used for viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures.
The device is suitable for professional healthcare facility environments such as hospitals and clinics.
The subject device consists of an endoscope (a hysteroscope) which includes an insertion portion and handle, and the following accessories: 2 pieces of two-way valves and 1 piece of silicone cap.
The subject Single-Use Video Hysteroscope is the same as the predicate Single-Use Video Hysteroscope (cleared under K232003) with respect to its indications for use, design, operating principle, and fundamental technology.
The subject device differs from the predicate device showed as below,
- 1. All features of the new device (subject device) are identical to the predicate, with 2 accessories added to its primary package, consisting of 2 pieces of two-way valves and a silicone cap;
- 2. The depth of field (DOF) of RP-G-C0101 is changed from 5 mm~50 mm to 3 mm~50 mm.
### V. Indications for use
Single-Use Video Hysteroscope is intended to be used for viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures.
The device is suitable for professional healthcare facility environments such as hospitals and clinics.
### VI. Indications for Use Comparison
The subject device, Single-Use Video Hysteroscope, has the same indications for use in comparison to the predicate device, Single-Use Video Hysteroscope.
#### VII. Technological Comparison
The subject Single-Use Video Hysteroscope is the same as the predicate Single-Use Video Hysteroscope (cleared under K232003) with respect to its indications for use, design, operating principle, and fundamental technology.
The subject device differs from the predicate device showed as below,
- 1. All features of the new device (subject device) are identical to the predicate, with 2 accessories added to its primary package, consisting of 2pcs of two-way valves and a silicone cap;
- The depth of field (DOF) of RP-G-C0101 is changed from 5 mm~50 mm into 3 mm~50 mm. 2.
These differences between the subject and predicate device do not raise different questions of safety or effectiveness. As part of demonstrating substantial equivalence to the predicate, a risk analysis was completed to identify the risks associated with the above changes. Verification testing was conducted to evaluate the modifications. The subject device passed all testing in accordance with internal requirements and applicable standards to support substantial equivalence of the subject device.
### VIII. Non-Clinical Tests Summary
As part of demonstrating substantial equivalence to the predicate, a risk analysis was completed to
{6}------------------------------------------------
K241987 Page 3 / 4
identify the risks associated with the design changes. Verification testing including performance testing using previously submitted methods and acceptance criteria cleared under K232003, additional biocompatibility testing per ISO 10993 standards, shelf life testing, and simulated transportation testing were conducted to evaluate the modifications. The subject device passed all the testing in accordance with internal requirements and applicable standards.
#### A Biocompatibility Summary
The biocompatibility evaluation for the patient contacting components was conducted in accordance with the 2023 FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The following testing was conducted based on contact category of "Surface – Breached or Compromised Surface" with a contact duration of "Limited (< 24 hours)":
- Cytotoxicity test per ISO 10993-5:2009
- Sensitization test per ISO 10993-10:2011
- Intracutaneous Reactivity test per ISO 10993-23:2021
- Acute Systemic Toxicity test per ISO 10993-11:2017
- Material-Mediated Pyrogenicity test per USP <151>
#### A Sterilization Validation
The subject device is sterilized via ethylene oxide (EO). The sterilization method has been validated to ISO 11135:2014 half-cycle method, which has thereby determined the routine control and monitoring parameters.
EO/ethylene chlorohydrin (ECH) residual testing was performed according to ISO 10993-7:2008 and was below the stated residual limits.
#### A Shelf Life and simulated transportation distribution followed by sterile packaging integrity test
Shelf Life and simulated transportation distribution followed by packaging integrity testing have been validated according to the following applicable standards:
- ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- ISO 11607-1:2019 Packaging for terminally sterilized Medical Device Part 1: Requirement for materials, sterile barrier systems and packaging systems
- ISO 11607-2:2019 Packaging for terminally sterilized Medical Device Part 2: Validation Requirement for forming, sealing and assembly process
- ASTM F 1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- | ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- ASTM D4169-23 Standard Practice for Performance Testing of Shipping Containers and Systems
{7}------------------------------------------------
#### > Performance
The following performance testing was conducted using previously submitted methods and acceptance criteria cleared under K232003 in support of the substantial equivalence, as shown in the table below.
| Test category | Test Item |
|---------------------------|---------------------------------------------------------------------------------------|
| Mechanical<br>performance | Surface and Edges Visual Assessment |
| | Water Delivery System (Flow Rate and Leakage Testing) |
| | Basic Size (Dimensional Analysis and Compatibility Testing of the<br>Two-way Valve) |
| | Comprehensive Performance of Luer Taper per ISO 80369-7:2021<br>and ISO 80369-20:2015 |
| | Pullout Force of Silicone Cap |
| | Compatibility with Accessories When Installing Silicone Cap |
| | Sealing Test after Installing the Two-way Valve (Leakage Testing<br>under Pressure) |
| Optical<br>performance | Depth of field |
# IX. Clinical Evidence
N/A.
# X. Conclusion
The results of nonclinical performance testing demonstrate that the subject device is as safe and effective as the predicate device to support a substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.