This product is intended to direct a catheter to the desired anatomical location in the peripheral vasculature (excluding coronary and cerebral vessels).
Device Story
CROSSLOOP is a steerable guidewire (0.018" diameter; 200-300 cm lengths) used to navigate catheters through peripheral vasculature. Device consists of stainless-steel core wire, Pt-Ni alloy coil, and Au-Sn solder. Distal tip features loop structure; surface includes hydrophilic/silicone coatings distally and hydrophobic PTFE coating proximally to enhance lubricity. Operated by physicians during interventional procedures; wire is manually manipulated to traverse tortuous vessels. Radiopaque coil allows visualization under fluoroscopy. Device facilitates catheter placement, aiding in diagnostic or therapeutic interventions. Benefits include improved navigation through complex peripheral anatomy.
Clinical Evidence
Bench testing only. Performance testing included tensile strength, torque strength, torqueability, tip flexibility, coating integrity, catheter compatibility, visual inspection, corrosion resistance, kink resistance, radio detectability, and particulate characterization. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous irritation, systemic toxicity, USP rabbit pyrogen, hemolysis, partial thromboplastin time, in vivo thrombogenicity, and Sc5b-9 complement activation. All tests met acceptance criteria.
Indicated for directing catheters to desired anatomical locations in peripheral vasculature, excluding coronary and cerebral vessels.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Asahi Intecc Co., Ltd. Cynthia Valenzuela Director, Quality Systems, Regulatory Affairs/Compliance Asahi Intecc USA, Inc. 3002 Dow Avenue Suite 212 Tustin, California 92780
Re: K241962
Trade/Device Name: Crossloop Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: February 20, 2025 Received: February 20, 2025
Dear Cynthia Valenzuela:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jenny R.
Katsnelson -S
for Lydia Glaw
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Digitally signed by Jenny R.
Katsnelson -S
Date: 2025.03.27 12:02:30 -04'00'
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241962
Device Name CROSSLOOP
Indications for Use (Describe)
This product is intended to direct a catheter to the desired anatomical vasculature (excluding coronary and cerebral vessels).
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary (as required by 21 CFR § 807.92)
# SAHI INTECC CO.,LTD.
Global Headquarters and R&D Center 3-100 Akatsuki-cho, Seto-shi, Aichi 489-0071 Japan TEL : +81-561-48-5551 FAX : +81-561-48-5552 http://www.asahi-intecc.co.jp/
#### CROSSLOOP
#### 510(k) K241962
| Date Prepared: | 25 JUNE 2024 | | | | | |
|------------------------|------------------------------------------|---------------------------|--|--|--|--|
| Applicant: | ASAHI INTECC CO., LTD. | | | | | |
| | 3-100 Akatsuki-cho, | | | | | |
| | Seto, Aichi 489-0071 Japan | | | | | |
| Contact: | Mrs. Cynthia Valenzuela | | | | | |
| | Director, Quality and Regulatory Affairs | | | | | |
| | ASAHI INTECC USA, Inc. | | | | | |
| | 3002 Dow Avenue, Suite 212 | | | | | |
| | Tustin, CA 92780 | | | | | |
| | Tel: (949) 745-1617 | | | | | |
| | e-mail: cynthiav@asahi-intecc-us.com | | | | | |
| Trade Name: | CROSSLOOP | | | | | |
| Common Name: | Guidewire | | | | | |
| Device Classification: | Class 2 per 21 CFR § 870.1330 | | | | | |
| Classification Name: | Catheter, Guide, Wire | | | | | |
| Product Code: | DQX - Catheter Guide Wire | | | | | |
| Predicate Device: | Astato 30 K071721, K163426 | | | | | |
| Reference Devices: | CROSSLEAD Penetration | K230377 | | | | |
| | SION blue | K122468, K163426, K191464 | | | | |
| | Astato XS 40 | K153443, K163426 | | | | |
| | V-18 Control Wire | K033742 | | | | |
#### INTENDED USE/INDICATIONS FOR USE:
This product is intended to direct a catheter to the desired anatomical location in the peripheral vasculature (excludinq coronary and cerebral vessels).
#### DEVICE DESCRIPTION:
The CROSSLOOP is steerable guide wire with a maximum diameter of 0.018 inches (0.46 mm) and available in various lengths of 200 cm, 235 cm and 300 cm. The CROSSLOOP consists of a stainless-steel (SUS) core wire with a Pt-Ni alloy coil, soldered with Au-Sn solder. The distal tip is loop-structured. The coil is radiopaque to achieve visibility and can be made to bend easily to accommodate vessel tortuosity. A silicone and hydrophilic coating are applied to the distal portion of the guide wire. A hydrophobic coating (PTFE) is applied to proximal portion. The purpose of these
ASAHI INTECC CO., LTD.
{5}------------------------------------------------
surface coatings is to provide lubricity when the guide wire is passed through percutaneous catheters.
The basic structure, construction and materials of the CROSSLOOP are similar to those of the predicate Astato 30 (K071721, K163426) and reference CROSSLEAD Penetration (K230377) and SION blue (K122468, K163426, K191464).
## COMPARISON WITH PREDICATE AND REFERENCE DEVICES:
Comparisons of the CROSSLOOP and predicate / reference devices show that the technological characteristics of the subject device such as the components, design, materials, sterilization method and operating principle are similar to the currently marketed predicate and reference devices. The intended use of the subject device and the predicate is similar.
{6}------------------------------------------------
| Device name | Subject | Predicate | Reference | | | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CROSSLOOP | Astato 30 | CROSSLEAD Penetration | SION blue | Astato XS 40 | V-18 Control Wire |
| 510(k) | TBD | K071721,K163426 | K230377 | K122468, K163426, K191464 | K153443,K163426 | K033742 |
| Manufacturer | ASAHI INTECC CO., LTD. | | | Boston Scientific | | |
| Regulation<br>Number | 21 CFR 870.1330 | | | | | |
| Regulation<br>Name | Catheter Guide Wire | | | | | |
| Regulatory<br>Class | II | | | | | |
| Product code | | | | DQX | | |
| Intended and/or<br>Indication for<br>Use | This product is<br>intended to<br>direct a catheter<br>to the desired<br>anatomical<br>location in the<br>peripheral<br>vasculature<br>(excluding<br>coronary and<br>cerebral<br>vessels). | This product is<br>intended to<br>facilitate the<br>placement and<br>exchange of<br>diagnostic and<br>therapeutic<br>devices during<br>intravascular<br>procedures. This<br>device is intended<br>for peripheral<br>vascular use only. | This product is<br>designed to<br>direct a catheter<br>to the desired<br>anatomical<br>location in the<br>peripheral<br>vasculature<br>during<br>diagnostic or<br>interventional<br>procedures. | ASAHI PTCA<br>Guide Wires are<br>intended to<br>facilitate the<br>placement of<br>balloon dilatation<br>catheters during<br>percutaneous<br>transluminal<br>coronary<br>angioplasty<br>(PTCA) and<br>percutaneous<br>transluminal<br>angioplasty<br>(PTA). The<br>ASAHI PTCA<br>Guide Wires are<br>not to be used in<br>the<br>neurovasculature | This product is<br>intended to<br>facilitate the<br>placement and<br>exchange of<br>diagnostic and<br>therapeutic<br>devices during<br>intravascular<br>procedures. This<br>device is intended<br>for peripheral<br>vascular use only. | The 110cm V-18<br>Control Wire is<br>intended for<br>general<br>intravascular use<br>including the<br>placement of PTA<br>balloon catheters<br>requiring an 0.018"<br>guide wire in<br>hemodialysis AV<br>access<br>procedures. The<br>wire can be<br>torqued to facilitate<br>placement of<br>diagnostic or<br>therapeutic<br>catheters. This<br>device is intended<br>for peripheral<br>vascular use only.<br>A torque device is<br>included with each<br>wire to facilitate<br>directional<br>manipulation of the<br>guide wire. |
| Device name | Subject | Predicate | Reference | | | |
| | CROSSLOOP | Astato 30 | CROSSLEAD<br>Penetration | SION blue | Astato XS 40 | V-18 Control<br>Wire |
| Nominal OD | 0.46 mm<br>(0.018 inch) | 0.45 mm<br>(0.018 inch)<br>with tapered end | 0.36 mm<br>(0.014 inch)<br>0.46 mm<br>(0.018 inch) | 0.36 mm<br>(0.014 inch) | 0.36 mm<br>(0.014 inch)<br>with tapered end | 0.46 mm<br>(0.018 inch) |
| Overall Length | 200 cm, 235 cm,<br>300 cm | 180 cm, 300 cm | 200 cm, 235 cm,<br>300 cm | 180 cm,190 cm,<br>300 cm | 180 cm, 300 cm | 110 cm |
| Coil | Pt-Ni alloy | Pt-Ni alloy | Pt-Ni alloy | Austenitic<br>Stainless Steel<br>Pt-Ni alloy | Pt-Ni alloy | unknown |
| Tapered Core<br>Wire | Austenitic<br>Stainless Steel | Austenitic<br>Stainless Steel | Austenitic<br>Stainless Steel | Austenitic<br>Stainless Steel | Austenitic<br>Stainless Steel | Stainless Steel |
| Tip Shape | Straight, Pre-<br>shape | Straight | Straight, Pre-<br>shape | Straight, J-shape,<br>Pre-shape | Straight | Straight |
| Distal Tip<br>Structure | Round | Round | Micro-corn | Round | Round | Round |
| Coating | Hydrophilic<br>PTFE<br>Silicone | Hydrophilic<br>PTFE | Hydrophilic<br>PTFE | Hydrophilic<br>PTFE<br>Silicone | Hydrophilic<br>PTFE | Hydrophilic<br>PTFE |
| Solder | Ag-Sn<br>Au-Sn | Ag-Sn<br>Au-Sn | Au-Sn | Ag-Sn<br>Au-Sn | Ag-Sn<br>Au-Sn | unknown |
| Sterilization | Ethylene oxide<br>SAL 10-6 | Ethylene oxide<br>SAL 10-6 | Ethylene oxide<br>SAL 10-6 | Ethylene oxide<br>SAL 10-6 | Ethylene oxide<br>SAL 10-6 | unknown |
{7}------------------------------------------------
The following technological differences exist between the predicate or the reference devices:
- Coil structure
- Core shaft
- Distal tip structure
{8}------------------------------------------------
## NON-CLINICAL TESTING/PERFORMANCE DATA:
Non-clinical laboratory testing was performed on the CROSSLOOP to determine substantial equivalence. The following testing/assessments were performed:
- Tensile Strenath ●
- Torque Strength ●
- Torqueability ●
- Tip Flexibility ●
- Coating Integrity
- Catheter Compatibility
- Visual Inspection
- Corrosion Resistance
- Kink Resistance
- Radio Detectability ●
- . Dimensional Verification
- Coating Integrity / Acute Particulate Characterization ●
The in vitro bench tests demonstrated the CROSSLOOP met all acceptance criteria and performed similarly to the predicate and reference devices. Performance data demonstrates the device functions as intended and has a safety and effectiveness profile that is similar to the predicate device.
#### BIOCOMPATIBILITY:
Testing was performed to assess biocompatibility of the device. The following tests were performed:
- Cytotoxicity ●
- Sensitization
- Intracutaneous Irritation
- Systemic Toxicity ●
- USP Rabbit Pyrogen, Material Mediated ●
- . Hemolysis
- Partial Thromboplastin Time ●
- In Vivo Thrombogenicity
- Sc5b-9 Complement Activation ●
#### CONCLUSION:
The CROSSLOOP has the same intended use and the same or similar technological characteristics such as components, design, materials, sterilization method and operating principles as the predicate and reference devices. The conclusions drawn from the nonclinical tests demonstrate that the device has a similar safety and effectiveness profile as the predicate device. Therefore, the CROSSLOOP is substantially equivalent to the predicate device.
ASAHI INTECC CO., LTD.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.