K241951 · BLUECORE COMPANY Co., Ltd. · GEX · Oct 10, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241951
Device Name
Lasya-Trinity
Applicant
BLUECORE COMPANY Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lasya-Trinity diode laser system is intended for use in dermatology procedures requiring coagulation.
Device Story
Diode laser system emitting 755, 808, and 1064 nm wavelengths for dermatological coagulation and hair reduction. System comprises main console, touchscreen GUI, footswitch, and two interchangeable applicators (Trinity and 808). Used in hospitals, physician offices, and medical spas by trained operators. User selects parameters via touchscreen; laser energy delivered to skin surface via handpiece. Device facilitates coagulation of vascular lesions or thermal destruction of hair follicles for permanent hair reduction. Non-sterile, reusable device requiring cleaning between uses.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22), laser safety (IEC 60825-1), software verification/validation (IEC 62304), and biocompatibility.
Indicated for benign vascular and vascular dependent lesions (Trinity applicator) and permanent reduction in hair regrowth (808 applicator) for all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
October 10, 2024
Bluecore Company Co., Ltd. Sang Mi Oh Official Correspondent #1203, 48, Centurm Jungang-Ro, Haeundae-Gu Busan, KR 48059 Korea, South
Re: K241951
Trade/Device Name: Lasya-Trinity Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 13, 2024 Received: July 03, 2024
Dear Sang Mi Oh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@;tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed b TANISHA TANISHA L. HITHE L. HITHE -S Date 23:46:19 -04'00
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical & Infection Control Devices, Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241951
Device Name Lasya-Trinity
Indications for Use (Describe)
The Lasya-Trinity diode laser system is intended for use in dermatology procedures requiring coagulation with the following indications:
The trinity applicator is indicated for benign vascular and vascular dependent lesions.
The 808 applicator is indicated for permanent reduction in hair regrowth.
Use on all skin types (Fitzpatric I-VI), including tanned skin.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) summary (510(k) #: K241951)
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
#### I. Submitter Information [21 CFR 807.92(a) (1)]
| Applicant/Manufacturer Name | BLUECORE COMPANY Co., Ltd. |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Address | #1203, 48, Centum Jungang-Ro, Haeundae-Gu Busan KR 48059<br>Korea, South |
| Contact Person | Sangmi Oh<br>Regulatory Affairs Manager<br>Tel.: +82 070 4870 5146<br>Fax: +82 051 747 4319<br>E-mail: osm7832@bluecorecompany.com |
| Date Prepared | October 8, 2024 |
#### II. Name of device [21 CFR 807.92 (a) (2)]
| Device Trade Name | Lasya-Trinity |
|---------------------|----------------------------------|
| Common Name | Lasya-Trinity Diode Laser System |
| Classification Name | Power Laser Surgical Instrument |
| Regulation Number | 21 CFR 878.4810 |
| Product Code | GEX |
#### III. Predicate Devices [21 CFR 807.92(a) (3)]
| Predicate # | K230371 |
|----------------------|------------------|
| Predicate Trade Name | Soprano Titanium |
| Product Code | GEX |
#### IV. Device Description [21 CFR 807.92(a) (4)]
The Subject device, Trinity-Lasya is a diode laser instrument that use diode as a medium to emit a laser beam of 755 & 808 & 1064nm wavelength.
There are two handpiece, the user can choose from. Moreover, a specific software controls the device functions and allows the user selections through the touchscreen display.
the system is intended to be used in user facilities such as hospitals, physicians' offices and medical spas, and its major components are the main console unit, foot switch, and individual modules.
The device is re-usable and non-sterile; instructions for cleaning its components between uses are provided in the labeling.
Lasya-Trinity Laser System consists of:
- System console (contains the system software, power supply, and various other electronic ● and mechanical parts)
- . Operator control panel with touch-screen technology (GUI)
- Trinity applicator(handpiece) with 1064 nm, 808nm and 755 nm wavelengths applied simultaneously
- . 808 applicator(handpiece) with 808nm wavelength
- . Footswitch and etc.
#### V. Intended use of device and Indications for Use [21 CFR 807.92(a) (5)]
The Lasya-Trinity diode laser system is intended for use in dermatology procedures requiring coagulation.
{5}------------------------------------------------
The indications for use for the Lasya-Trinity include:
Trinity applicator intended use and indications for use: benign vascular and vascular dependent lesions.
808 applicator intended use and indications for use: permanent reduction in hair regrowth. Use on all skin types (Fitzpatric I-VI), including tanned skin.
### VI. Summary of technological characteristics of the device compared to the predicate [21 CFR 807.92(a)(6)]
The subject Lasya-Trinity system shares with its predicate Soprano Titanium system (K230371) the same underlying technology and presents similar technological characteristics. Both systems use laser energy, delivered to the skin surface, via applicators.
The subject Lasya-Trinity is substantially equivalent to the cleared Soprano Titanium with respect to the hardware and software elements of the system, the principles of operation and the product design.
| | Predicate Device | Subject Device |
|---------------------------------|---------------------------------------------------|---------------------------------------------------|
| Product Name | Soprano Titanium<br>(Trio 2 cm² applicator) | Lasya-Trinity<br>(Trinity applicator) |
| 510(k) Number | K230371 | K241951 |
| Manufacturer | Alma Lasers Inc. | BLUECORE COMPANY Co., Ltd. |
| Laser Medium | Diode | Diode |
| Wavelength | 755&810&1064 nm | 755&808&1064 nm |
| Handpiece | Trio 2 cm² | Trinity applicator |
| Operational mode | HR | HR |
| Fluence | 1~120 [J/cm²] | Max.60 [J/cm²] |
| Frequency<br>(Repetition Rate) | up to 10 [Hz] | 1 - 10 [Hz] |
| Pulse Width<br>(Pulse Duration) | Up to 200 [ms] | 5 ~ 200 [ms] |
| Spot Size | 2 [cm²] | 2 [cm²] |
| Indications for<br>Use | Benign vascular and vascular dependent<br>lesions | Benign vascular and vascular dependent<br>lesions |
| | Predicate Device | Subject Device |
| Product Name | Soprano Titanium<br>(808 2 cm² applicator) | Lasya-Trinity<br>(808 applicator) |
| 510(k) Number | K230371 | K241951 |
| Manufacturer | Alma Lasers Inc. | BLUECORE COMPANY Co., Ltd. |
| Laser Medium | Diode | Diode |
| Wavelength | 810 nm | 808nm |
| Handpiece | 810 applicator | 808 applicator |
| Operational mode | HR | HR |
| Fluence | Max. 60 [J/cm²] | Max.60 [J/cm²] |
| Frequency<br>(Repetition Rate) | 0.5 - 3 [Hz] | 1-3 [Hz] |
| Pulse Width<br>(Pulse Duration) | 3.3 ~ 200 [ms] | 5~200 [ms] |
| Spot Size | 20 * 10 [mm²] | 2 [cm²] |
| Indications for use | Permanent reduction in hair regrowth | Permanent reduction in hair regrowth |
#### VII. Performance Testing [21 CFR 807.92(b)(1)]
IEC 60601-1 Medical Electrical Equipment – Part 1: General Requirements for safety and essential performance IEC 60601-1-2 Medical Electrical Equipment 1-2 General Requirements for basic safety and essential performance IEC 60601-2-22 - Medical Electrical Equipment - Part 2-22 - Particular Requirements for basic
{6}------------------------------------------------
safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment IEC 60825-1, Safety of laser products - Part 1: Equipment classification, and requirements
In addition, software verification and validation testing was performed to the requirements of IEC 62304 and biocompatibility conformance to FDA standards was established.
#### VIII. Clinical Data [21 CFR 807.92(b) (2)]
Based on the similarities in the safety and effectiveness profiles of the subject, reference, and primary predicate devices, no animal or clinical studies were deemed needed to support this submission.
#### IX. Conclusions Safety and Effectiveness SE [21 CFR 807.92(b) (3)]
The Lasya-Trinity is as safe and effective as the predicate Soprano Titanium. The proposed Lasya-Trinity has the same intended use and indications, similar technological characteristics, and same principle of operation as its predicate device.
Thus, the Lasya-Trinity is substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.