The AGILON® XO Shoulder Replacement System is indicated for use for total or hemi shoulder replacement in cases of: - Non- inflammatory degenerative joint disease including osteoarthritis and avascular necrosis, - Post-traumatic osteoarthritis. - Fractures, - Rheumatoid arthritis. The main indications for the implantation of an AGILON® hemi shoulder prosthesis are: - Multi-fragmental comminuted fractures of the humeral head, - 3- and 4-Fragment-fractures of the proximal humerus, - Head-splitting fractures, - Dislocated head-splitting fractures, - Humeral head depression with more than 40% of joint surface depressed, - Interlocking chronic dislocation with deep HILL-SACHS lesion, - Fracture instability following internal fixation attempt in 3-fragment fractures (secondary dislocation, material loosening), - Posttraumatic humeral head necrosis, - Omarthrosis. AGILON® CTA heads are destined for treatment of stable types of rotator cuff tear arthropathy. In order to achieve satisfactory results with the CTA heads the fornix humeri and the subscapularis tendon must be intact. A CTA cap is intended for the use as a hemi-arthroplasty, to treat a patient after an inverse shoulder has failed. It is not combined with a glenoid implant. It can be used in primary and revision cases. The main indications for the implantation of an AGILON® inverse (reverse) shoulder prosthesis are: - Rotator cuff tear arthropathy, - Chronic trauma shoulder, - Decentering of the humeral head after implantation of a humeral head prosthesis. Please note, that the patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary. In case of revision surgery the available bone stock has to be evaluated to allow for implantation of well-fixed stems. Conversion of the system can be performed in revision cases as follows: - From Hemi Shoulder Arthroplasty to Anatomic Total Shoulder Arthroplasty - From Hemi Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty - From Hemi Shoulder Arthroplasty to Inverse (Reverse) Total Shoulder Arthroplasty - From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi Shoulder Arthroplasty as salvage procedure - From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty as salvage procedure - From Anatomic Total Shoulder Arthroplasty to Inverse (Reverse) Total Shoulder Arthroplasty - From Anatomic Total Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty The device is intended for adults. The AGILON® XO Shoulder Replacement System Titanium alloy stems are intended for cementless fixation. The PE-glenoids of the AGILON® XO Shoulder Replacement System are intended for cemented fixation. The qlenoid cementless is intended for cementless use with the addition of screws for fixation.
Device Story
Modular shoulder replacement system; components include humeral heads, glenoids, glenospheres, humeral stems, metaphyseal components, and fixation/cancellous screws. Subject submission adds new cancellous screws for total reverse shoulder replacement. Used in orthopedic surgery; implanted by surgeons. System allows conversion between hemi, anatomic total, and inverse (reverse) configurations. Titanium alloy stems for cementless fixation; PE-glenoids for cemented fixation; cementless glenoids use screws. Provides structural joint replacement to restore function in patients with degenerative disease, fractures, or failed prior arthroplasty.
Clinical Evidence
Bench testing only. Mechanical characterization of new cancellous screws performed per ASTM F543 (torsional properties, driving torque, axial pull-out strength). Biocompatibility evaluation per ISO 10993-1. Cleaning, sterilization, packaging, and shelf-life data leveraged from predicate K231657.
Technological Characteristics
Modular shoulder prosthesis. Materials: Titanium alloy (stems), PE (glenoids). Fixation: Cementless (stems, screws) and cemented (PE-glenoids). Mechanical testing per ASTM F543. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for adults requiring total or hemi shoulder replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), post-traumatic osteoarthritis, fractures, or rheumatoid arthritis. Specific indications include rotator cuff tear arthropathy, chronic trauma, and revision cases. Contraindicated if joint anatomy/structure is unsuitable or deltoid muscle is non-functional.
Regulatory Classification
Identification
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
March 10, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
implantcast, GmbH % Sarah Pleaugh Director of Regulatory Affairs MCRA, LLC 803 7th Street NW. Floor 3 Washington, District of Columbia 20001
Re: K241944
Trade/Device Name: AGILON® XO Shoulder Replacement System Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWT, HSD, PHX Dated: February 11, 2025 Received: February 11, 2025
Dear Juliane Hoppner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Farzana Sharmin -S
Digitally signed by Farzana Sharmin -S Date: 2025.03.10 15:46:21 -04'00'
Farzana Sharmin, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241944
Device Name
AGILON® XO Shoulder Replacement System
Indications for Use (Describe)
The AGILON® XO Shoulder Replacement System is indicated for use for total or hemi shoulder replacement in cases of:
- · Non- inflammatory deqenerative joint disease including osteoarthritis and avascular necrosis,
- · Post-traumatic osteoarthritis.
- · Fractures,
- · Rheumatoid arthritis.
The main indications for the implantation of an AGILON® hemi shoulder prosthesis are:
- · Multi-fragmental comminuted fractures of the humeral head,
- · 3- and 4-Fragment-fractures of the proximal humerus,
- · Head-splitting fractures,
- · Dislocated head-splitting fractures,
- · Humeral head depression with more than 40% of joint surface depressed,
- · Interlocking chronic dislocation with deep HILL-SACHS lesion,
- · Fracture instability following internal fixation attempt in 3-fragment fractures (secondary dislocation, material loosening),
- · Posttraumatic humeral head necrosis,
· Omarthrosis.
AGILON® CTA heads are destined for treatment of stable types of rotator cuff tear arthropathy. In order to achieve satisfactory results with the CTA heads the fornix humeri and the subscapularis tendon must be intact. A CTA cap is intended for the use as a hemi-arthroplasty, to treat a patient after an inverse shoulder has failed. It is not combined with a glenoid implant. It can be used in primary and revision cases.
The main indications for the implantation of an AGILON® inverse (reverse) shoulder prosthesis are:
- · Rotator cuff tear arthropathy,
- · Chronic trauma shoulder,
· Decentering of the humeral head after implantation of a humeral head prosthesis.
Please note, that the patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary.
In case of revision surgery the available bone stock has to be evaluated to allow for implantation of well-fixed stems. Conversion of the system can be performed in revision cases as follows:
- · From Hemi Shoulder Arthroplasty to Anatomic Total Shoulder Arthroplasty
- · From Hemi Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty
- · From Hemi Shoulder Arthroplasty to Inverse (Reverse) Total Shoulder Arthroplasty
· From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi Shoulder Arthroplasty as salvage procedure
· From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty as salvage procedure
- From Anatomic Total Shoulder Arthroplasty to Inverse (Reverse) Total Shoulder Arthroplasty
- · From Anatomic Total Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty The device is intended for adults.
The AGILON® XO Shoulder Replacement System Titanium alloy stems are intended for cementless
{4}------------------------------------------------
fixation.
The PE-glenoids of the AGILON® XO Shoulder Replacement System are intended for cemented fixation. The qlenoid cementless is intended for cementless use with the addition of screws for fixation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
| Device Trade Name: | AGILON® XO Shoulder Replacement System |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | implantcast. GmbH<br>Lüneburger Schanze 26<br>21614 Buxtehude, Germany |
| Contact: | Juliane Höppner, Director of Regulatory & Clinical & Biological Affairs<br>implantcast. GmbH<br>Office: 49-4161-744-135 |
| Prepared by: | MCRA, LLC<br>803 7th Street, NW, 3rd Floor<br>Washington, DC 20001<br>Office: 202.552.5800 |
| Date Prepared: | July 2, 2024 |
| Classifications: | 21 CFR 888.3650, Shoulder joint metal/polymer non-constrained cemented<br>prosthesis<br>21 CFR 888.3660, Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis<br>21CFR 888.3690, Shoulder joint humeral (hemi-shoulder) metallic<br>uncemented prosthesis |
| Common Name: | Shoulder Prosthesis |
| Class: | II |
| Product Codes: | KWT, HSD, PHX |
| Primary Predicate: | AGILON® XO Shoulder Replacement System (K231657) |
| Additional Predicate: | AGILON® XO Shoulder Replacement System (K222482)<br>AGILON® XO Shoulder Replacement System (K191433) |
## Indications For Use:
The AGILON® XO Shoulder Replacement System is indicated for use for total or hemi shoulder replacement in cases of:
- · Non- inflammatory degenerative joint disease including osteoarthritis and avascular necrosis,
- Post-traumatic osteoarthritis, .
- Fractures, .
{6}------------------------------------------------
- . Rheumatoid arthritis.
The main indications for the implantation of an AGILON® hemi shoulder prosthesis are:
- Multi-fragmental comminuted fractures of the humeral head, .
- 3- and 4-Fragment-fractures of the proximal humerus, ●
- Head-splitting fractures, ●
- Dislocated head-splitting fractures, ●
- Humeral head depression with more than 40% of joint surface depressed, ●
- Interlocking chronic dislocation with deep HILL-SACHS lesion, ●
- Fracture instability following internal fixation attempt in 3-fragment and 4-fragment fractures (secondary dislocation, material loosening),
- Posttraumatic humeral head necrosis, ●
- . Omarthrosis.
AGILON® CTA heads are destined for treatment of stable types of rotator cuff tear arthropathy. In order to achieve satisfactory results with the CTA heads the fornix humeri and the subscapularis tendon must be intact. A CTA cap is intended for the use as a hemi-arthroplasty, to treat a patient after an inverse shoulder has failed. It is not combined with a glenoid implant. It can be used in primary and revision cases.
The main indications for the implantation of an AGILON® inverse (reverse) shoulder prosthesis are:
- Rotator cuff tear arthropathy, ●
- Chronic trauma shoulder,
- Decentering of the humeral head after implantation of a humeral head prosthesis. ●
Please note, that the patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary.
In case of revision surgery the available bone stock has to allow for implantation of well-fixed stems. Conversion of the system can be performed in revision cases as follows:
- From Hemi Shoulder Arthroplasty to Anatomic Total Shoulder Arthroplasty .
- From Hemi Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty ●
- From Hemi Shoulder Arthroplasty to Inverse (Reverse) Total Shoulder Arthroplasty ●
- From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi Shoulder Arthroplasty as ● salvage procedure
- From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty as salvage procedure
- From Anatomic Total Shoulder Arthroplasty to Inverse (Reverse) Total Shoulder ● Arthroplasty
- . From Anatomic Total Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty
The device is intended for adults.
{7}------------------------------------------------
The AGILON® XO Shoulder Replacement System Titanium alloy stems are intended for cementless fixation.
The PE-glenoids of the AGILON® XO Shoulder Replacement System are intended for cemented fixation. The glenoid cementless is intended for cementless use with the addition of screws for fixation.
### Device Description:
The AGILON® XO Shoulder Replacement System is a modular shoulder replacement system offering various components that can be combined to replace the shoulder joint with options depending upon the size and anatomical needs of each patient. The subject submission is for new cancellous screws for use in a total reverse shoulder replacement with the previously cleared K231657 components:
- Humeral Head Components ●
- Glenoid and Glenosphere Components ●
- Humeral Stems and Stem Extension Pieces
- Metaphyseal Components ●
- Fixation and Cancellous Screws ●
The subject line extension components are intended for use with previously cleared AGILON XO Shoulder System instruments. There are no changes to packaging, sterility, shelf life, or materials subject to this submission.
#### Predicate Device:
implantcast submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, AGILON® XO Shoulder Replacement System is substantially equivalent in indications, design principles, and performance to the predicate AGILON® XO Shoulder Replacement System (K231657).
### Performance Testing Summary:
Engineering analysis and a review of the previously completed testing of the predicate AGILON XO Shoulder Replacement System demonstrated no new worst-case was created for the system of devices requiring new construct testing. The new cancellous screws were fully characterized per ASTM F543 testing:
- Torsional Properties ●
- Driving Torque ●
- Axial Pull-out Strength ●
The testing demonstrated sufficient performance for the intended use, met the predetermined acceptance criteria, and supports substantial equivalence.
The biocompatibility evaluation was completed per ISO 10993-1.
The cleaning, sterilization, packaging, shelf-life, and endotoxin testing were leveraged from the predicate AGILON® XO Shoulder Replacement System (K231657).
{8}------------------------------------------------
## Substantial Equivalence Conclusion:
The data included in this submission demonstrate substantial equivalence to the predicate device. AGILON® XO Shoulder Replacement System is as safe, as effective, and performs as well as, or better, than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.