HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available tinted and used to enhance or alter the apparent color of the eye. Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients. Disposable Wear: Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When prescribed for Daily Disposable Wear, the lenses are not to be used with disinfecting systems as they are to be discarded after a single use. Frequent / Planned Replacement Wear: Eyecare practitioners may prescribe any of the above lenses for frequent / planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
Device Story
HPM38 is a spherical, soft (hydrophilic) contact lens made of polymacon (62% polymer, 38% water). It acts as a refractive medium to focus light on the retina. Available in clear or cosmetic tinted versions; tinted lenses use a 'sandwich' structure to encapsulate color additives between two polymacon layers. Prescribed by eyecare practitioners for daily wear, either as daily disposable or frequent/planned replacement. For frequent replacement, patients use chemical disinfection systems. The lens conforms to the corneal curvature when hydrated. It provides vision correction for myopia and low astigmatism. Benefits include visual acuity correction and cosmetic eye color alteration.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including physical, optical, and physicochemical performance (ISO 18369-2/3/4, ASTM D882-18), biocompatibility (ISO 10993 series), and sterilization validation (ISO 17665-1).
Technological Characteristics
Material: Polymacon (HEMA cross-linked with EGDMA). Water content: 38%. Refractive index: 1.437. Oxygen permeability: 8.99x10^-11 (cm²/s) [ml O2/(ml·mmHg)] at 35°C. Manufacturing: Lathe-cut (semi-molded). Sterilization: Moist heat. Packaging: Foil blister pack with borate buffer solution. Standards: ISO 18369, ISO 10993, ISO 17665, ISO 11987, ASTM D882, ASTM D4169.
Indications for Use
Indicated for correction of visual acuity in aphakic and non-aphakic persons with non-diseased eyes with myopia and refractive astigmatism of 0.50 diopters or less.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on the first line and "ADMINISTRATION" on the second line.
March 20, 2025
Hunan Haipuming Technology Co., Ltd. % Grace Liu Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square Nanshan District Shenzhen, Guangdong 518000 China
Re: K241884
Trade/Device Name: HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens (HPM38) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: February 13, 2025 Received: February 13, 2025
Dear Grace Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketingsafety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-deviceadvice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
J Angelo Green -S
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241884
Device Name
HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens (HPM38)
Indications for Use (Describe)
HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available tinted and used to enhance or alter the apparent color of the eye.
Daily wear replacement schedules may vary from patient and should be decided by evecare practitioners in consultation with their patients.
Disposable Wear:
Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When prescribed for Daily Disposable Wear, the lenses are not to be used with disinfecting systems as they are to be discarded after a single use.
Frequent / Planned Replacement Wear:
Eyecare practitioners may prescribe any of the above lenses for frequent / planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary - K241884
# 1. Contact Details
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
### 1.1 Applicant information
| Applicant Name | Hunan Haipuming Technology Co., Ltd. |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | The Intersection of Chuangye Avenue and Lvting Avenue, Dahetang<br>Sub-district Office, Shaodong City, Shaoyang City, Hunan Province,<br>China |
| Contact person | Zhang Mingxuan |
| Phone No. | +86-739-2712889<br>+86-18107399879 |
| E-mail | 164059332@qq.com |
| Date Prepared | 2025.03.14 |
### 1.2 Submission Correspondent
| Image: Logo | Shenzhen Joyantech Consulting Co., Ltd. |
|-------------------------|-----------------------------------------------------------------------|
| | 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District, |
| | Shenzhen, Guangdong Province, China |
| Phone No. | +86-755-86069197 |
| Contact person | Grace Liu |
| Contact person's e-mail | grace@cefda.com |
| Website | http://www.cefda.com |
# 2. Device Information
| Trade name | HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens (HPM38 |
|---------------------|---------------------------------------------------------------------|
| Common name | Soft (Hydrophilic) Contact Lens |
| Classification | II |
| Classification name | Lenses, Soft Contact, Daily Wear |
| Product code | LPL, MVN |
| Regulation No. | 21 CFR 886.5925 |
# 3. Legally Marketed Predicate Device
| Trade Name | Neo Cosmo (polymacon) Soft (hydrophilic) Contact Lens |
|---------------|-------------------------------------------------------|
| 510(k) Number | K142275 |
| Product Code | LPL |
| Manufacturer | Neo Vision Co., Ltd. |
The predicate device has not been subject to a design-related recall.
# 4. Device Description
HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is a spherical lens. It is fabricated from
{5}------------------------------------------------
polymacon which has been adopted by the United Stated Adopted Names Council (USAN). The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.
The lens material, polymacon, is a hydrophilic polymer of 2-hydroxyethyl methacrylate (HEMA) cross-linked with ethylene glycol dimethacrylate (EGDMA), initiated by 2, 2'-azobisisobutyronitrile (AIBN). The lens consists of 62% polymacon and 38% water by weight when immersed in standard saline solution.
HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is available cosmetic tinted to enhance or alter the apparent color of the eye. It is tinted in an annular pattern, providing a clear optic zone, using a combination of one or more of the following 'listed' color additives:
| Color additive | Listing No. |
|----------------------|-----------------|
| Reactive black 5 | 21 CFR 73.3121 |
| Titanium dioxide | 21 CFR 73.3126 |
| Iron oxides | 21 CFR 73.3125 |
| Solvent Yellow 18 | 21 CFR 73.3122 |
| Carbazole violet | 21 CFR 73.3107 |
| Pigment Blue 36 | 21 CFR 73.3110a |
| Phthalocyanine green | 21 CFR 73.3124 |
HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens contains only the amount of color additives to accomplish the intended coloring effect. The proposed color lenses are manufactured by sandwiching the pattern layer containing color additives between two layers of lens materials (polymacon). As part of the manufacturing process, the color lenses are thoroughly washed to remove the unbound reactive color additives. Because of the 'sandwich' structure, the color additives will not be removed by lens handling and cleaning procedures.
In the hydrated state, HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus, and it acts as a refracting media to focus light rays on the retina. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution compatible with the eye.
HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is available in the spherical configuration with the following features and properties:
| Back vertex power | -10.00D to 0.00D, step: 0.25D |
|-------------------|------------------------------------|
| Base curve | 8.40 mm to 8.70 mm, step: 0.10 mm |
| Total diameter | 13.80 mm to 14.50 mm, step:0.10 mm |
The physical properties of the proposed lens are:
| Specific gravity | 1.124 |
|------------------|-------|
| Refractive index | 1.437 |
{6}------------------------------------------------
| Light Transmission | 93% ± 5% |
|---------------------|--------------------------------------------------------|
| Surface character | Hydrophilic |
| Water content | 38% |
| Oxygen permeability | $8.99\times10^{-11}$ (cm²/s) [ml O₂/(ml·mmHg)] at 35°C |
HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is supplied sterile in the foil blister pack containing the borate buffer solution.
# 5. Intended Use/Indication for Use
HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available tinted and used to enhance or alter the apparent color of the eye.
Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.
### Disposable Wear:
Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When prescribed for Daily Disposable Wear, the lenses are not to be used with disinfecting systems as they are to be discarded after a single use.
Frequent / Planned Replacement Wear:
Eyecare practitioners may prescribe any of the above lenses for frequent / planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
| Comparison<br>item | Proposed Device<br>(K241884) | Predicate Device<br>(K142275) | Comment |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Hunan Haipuming Technology Co.,<br>Ltd. | Neo Vision Co., Ltd. | / |
| Product Name | HPM38 (polymacon) Daily Wear Soft<br>(hydrophilic) Contact Lens | Neo Cosmo (polymacon) Soft<br>(hydrophilic) Contact Lens | / |
| Regulation<br>Number | 21 CFR 886.5925 | 21 CFR 886.5925 | Same |
| Classification | Class II | Class II | Same |
| Prescription<br>Use | Yes | Yes | Same |
| Intended Use /<br>Indications for<br>Use | HPM38 (polymacon) Daily Wear Soft<br>(Hydrophilic) Contact Lens is<br>indicated for the correction of visual | The Neo Cosmo (Polymacon) Soft<br>(hydrophilic) Contact Lens is<br>indicated for daily wear for the | Equivalent |
| | | | |
| | acuity in aphakic and not-aphakic<br>persons with non-diseased eyes with | correction of visual acuity in aphakic<br>and non-aphakic persons with | |
| | myopia. The lens may be worn by | non-diseased eyes with myopia or | |
| | persons who exhibit refractive | hyperopia. The lens may be worn by | |
| | astigmatism of 0.50 diopters or less | persons who exhibit refractive | |
| | where the astigmatism does not | stigmatism of 0.50 diopters or less | |
| | interfere with visual acuity. The lens is | where the astigmatism does not | |
| | available tinted and used to enhance | interfere with visual acuity. The lens is | |
| | or alter the apparent color of the eye. | available clear or colored and may be | |
| | Daily wear replacement schedules<br>may vary from patient to patient and | used to enhance or alter the apparent<br>color of the eye. | |
| | should be decided by eyecare | Eye care practitioners may prescribe | |
| | practitioners in consultation with their | the lens frequent replacement wear | |
| | patients. | with cleaning, disinfecting and | |
| | Disposable Wear:<br>Eyecare practitioners may prescribe | schedule replacement. The lens may<br>be disinfected using a chemical (not<br>heat) lens care system only. | |
| | any of the above lenses for Daily | | |
| | Disposable Wear. When prescribed | | |
| | for Daily Disposable Wear, the lenses | | |
| | are not to be used with disinfecting | | |
| | systems as they are to be discarded | | |
| | after a single use. | | |
| | Frequent/Planned<br>Replacement<br>Wear: | | |
| | Eyecare practitioners may prescribe | | |
| | any of the above lenses for frequent / | | |
| | planned replacement wear, with | | |
| | cleaning, disinfection and scheduled | | |
| | replacement. When prescribed for | | |
| | frequent/planned replacement wear, | | |
| | the lens may be disinfected using a | | |
| | chemical disinfecting system. | | |
| Functionality | The contact lenses act as a refractive<br>medium that focus light rays from<br>near and distant objects on the retina. | The contact lenses act as a refractive<br>medium that focus light rays from<br>near and distant objects on the retina. | Same |
| Lens Design | Spherical | Spherical, Toric | Equivalent |
| Manufacturing<br>Process | Lathe-cut<br>(semi-molded) | Lathe-cut<br>(semi-molded) | Same |
| Lens Material | Polymacon | Polymacon | Same |
| Material<br>Classification | Group I<br>Low water content (<50%), Nonionic | Group I<br>Low water content (<50%), Nonionic | Same |
| Color Additives | Reactive Black 5 | Reactive Black 5 | Different |
| | | | |
| | | | |
| | (21CFR 73.3121)<br>Titanium Dioxide<br>(21 CFR 73.3126)<br>Iron Oxides<br>(21CFR 73.3125)<br>Solvent Yellow 18<br>(21CFR 73.3122)<br>Carbazole Violet<br>(21CFR 73.3107)<br>Pigment Blue 36<br>(21 CFR 73.3110a)<br>Phthalocyanine Green<br>(21 CFR 73.3124) | (21CFR 73.3121)<br>Titanium Dioxide (White)<br>(21 CFR 73.3126)<br>Iron Oxides(Red)<br>(21CFR 73.3125)<br>C.I Pigment Green 7<br>(21CFR 73.3124)<br>(Phthalocyaninato(2-)) Copper<br>(21CFR 74.3045) | |
| Storage Solution | Borate buffer solution | NaCl 0.85~0.95%(w/v) solution | Different |
| Primary Packaging | Blister pack | Blister pack and Vial | Equivalent |
| Total Diameter | 13.80 mm to 14.50 mm | 13.5 mm to 14.5 mm | Equivalent |
| Base Curve | 8.40 mm to 8.70 mm | 8.3 mm to 9.0 mm | Equivalent |
| Center Thickness<br>(varies with power) | 0.050 mm to 0.090 mm | 0.03 mm to 0.50 mm | Equivalent |
| Power Range | -10.00D to 0.00D | -25.00D to +25.00D | Equivalent |
| Water Content | 38% ±2 % | 38% ±2 % | Same |
| Spectral Transmission | 93% ± 5% | >90% | Same |
| Refraction Index | 1.437 | 1.428 | Same |
| Oxygen Permeability | 8.99×10-11 (cm²/s) [ml O₂/(ml·mmHg)]<br>at 35°C | 9.77×10-11 (cm²/s) [ml O₂/(ml·mmHg)]<br>at 35°C | Different |
| Modulus | 0.61 MPa | 0.350 MPa | Different |
| Tensile Strength | 0.62 MPa | 0.425 MPa | Different |
| Elongation at break | 279% | 127% | Different |
| Sterilization | Moist Heat Sterilization | Moist Heat Sterilization | Same |
| Biocompatibility | Biocompatible | Biocompatible | Same |
| | | | |
# 6. Substantial Equivalence Comparison
{7}------------------------------------------------
{8}------------------------------------------------
HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens has the similar indication for use as the predicate device as well as similar technical characteristics, and the differences don't raise any additional questions for safety and effectiveness, therefore, the proposed device is substantially equivalent to the predicate device.
# 7. Summary of Non-clinical Testing
{9}------------------------------------------------
Non-clinical tests were conducted in accordance with the FDA guidance - Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses" issued on June 27, 1994. The non-clinical performance tests had demonstrated the safety and effectiveness of HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens and the substantial equivalence to the predicate device.
#### > Performance Testing
The physical, optical, physicochemical performance tests were performed in accordance with ISO 18369-2:2017, ISO 18369-3:2017, ISO 18369-4:2017 and ASTM D882-18, and the test results showed that the proposed device meets the requirements.
- Appearance
- । Physical parameters (Total diameter, Base curve, Center thickness)
- Optical parameter (Back vertex power) ।
- Spectral transmission ।
- । Refractive index
- Water content ।
- -Specific gravity
- i Extractables
- i Oxygen permeability
- Mechanical properties i
- pH and Osmolality of the storage solution ।
In addition, the compatibility test of the lens with the lens care regimen has been conducted as per ISO 11981:2017, and the results demonstrate the proposed device is compatible with the marketed lens care regimens. The leachability of the color additives has also been conducted, and the results also demonstrate that the color additives contained in the proposed device are stable.
#### > Biocompatibility Testing
The following biocompatibility tests were performed respectively on the contact lens and primary packaging according to FDA guidance - Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses" issued on June 27, 1994, and the test results demonstrated that the contact lens and primary packaging have no cytotoxicity, no ocular irritation, no skin sensitization and no acute systemic toxicity.
- Cytotoxicity (ISO 10993-5:2009) ।
- i Ocular Irritation (ISO 10993-23:2021)
- i Skin Sensitization (ISO 10993-10:2021)
- Acute System Toxicity (ISO 10993-11:2017) ।
In addition, the following biocompatibility tests were performed on the storage solution, and the test
{10}------------------------------------------------
results demonstrated that the storage solution had no cytotoxicity, no ocular irritation and no skin sensitization.
- Cytotoxicity (ISO 10993-5:2009) ।
- Ocular Irritation (ISO 10993-23:2021) i
- Skin Sensitization (ISO 10993-10:2021)
In conclusion, the results of the above biocompatibility tests showed that the proposed device has no biocompatibility issues.
#### A Sterilization
A Sterility Assurance Level (SAL) of 10° has been validated in accordance with the requirements of ISO 17665-1:2006.
#### A Shelf Life
Accelerated aging testing has been performed to assure the shelf-life of 5 years in accordance with ISO 11987:2012.
#### A Simulated Transportation Testing
The simulated transportation testing has been conducted according to ASTM D4169-22.
# 8. Clinical Testing
This 510(k) submission does not utilize clinical study for establishing substantial equivalence therefore this section does not apply.
### 9. Conclusions
The results of comparing the design specifications and non-clinical testing between the proposed device and the legally marketed predicate device (K142275) show that they are Substantially Equivalent (SE).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.