TheRay Collared and Collarless Total Hip System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to: • A severely painful and/or disabled joint from osteoarthritis, theumatoid arthritis, avascular necrosis, or congenital hip dysplasia; · Acute traumatic fracture of the femoral head or neck; · Failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement. TheRay Collared and Collarless Total Hip System femoral stems are intended for cementless fixation. The acetabular cup is intended for cementless fixation. The porous structured surfaces on the acetabular cup provide biological fixation in a cementless application. The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System Femoral Stem or TheRay Collared or Collarless Femoral Stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate: · Femoral neck and trochanteric fractures of the proximal femur; · Osteonecrosis of the femoral head; · Revision procedures where other devices or treatments for these indications have failed.
Device Story
TheRay Collared and Collarless Femoral Stem is a hip prosthesis component for cementless fixation. It consists of a forged Ti-6Al-4V ELI substrate with a dual coating of plasma-sprayed commercially pure titanium (CPTi) and hydroxyapatite (HA). The device is available in collared (to seat on native resected calcar) and collarless versions. It is used by orthopedic surgeons in a clinical/OR setting during total hip replacement or hemiarthroplasty procedures. The stem is designed to be used with previously cleared femoral heads, acetabular cups, liners, and bone screws. The device provides structural support and biological fixation for the hip joint, aiming to restore joint function and alleviate pain in patients with degenerative or traumatic hip conditions. It may be used in combination with the iNSitu Bipolar Hip System for hemiarthroplasty.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included Range of Motion (ISO 21535), distal stem fatigue (ISO 7206-4), neck fatigue (ISO 7206-6), femoral head disassembly performance, and biocompatibility (ISO 10993, ANSI-AAMI ST72, ANSI-AMMI ST98, USP <643>).
Technological Characteristics
Substrate: Forged Ti-6Al-4V ELI (ASTM F136). Coating: Plasma-sprayed CPTi and hydroxyapatite. Bipolar system components: UHMWPE (ASTM F648), Cobalt Chrome (ASTM F75). Cementless fixation. Collared and collarless configurations. Sterilization method not specified.
Indications for Use
Indicated for skeletally mature individuals requiring total hip replacement due to osteoarthritis, rheumatoid arthritis, avascular necrosis, congenital hip dysplasia, acute traumatic femoral head/neck fracture, or failed previous hip surgery. Also indicated for hemiarthroplasty in cases of femoral neck/trochanteric fractures, osteonecrosis, or revision procedures.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
September 13, 2024
NextStep Arthropedix Garrett Spurgeon VP - Compliance, Regulatory, & Strategy 1800 Triplett Blvd Akron, Ohio 44306
Re: K241875
Trade/Device Name: TheRay Collared and Collarless Femoral Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: MEH, KWY, OQG, LZO, OQI Dated: September 5, 2024 Received: September 5, 2024
Dear Garrett Spurgeon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Limin Sun -S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241875
Device Name
TheRay Collared and Collarless Femoral Stem
#### Indications for Use (Describe)
TheRay Collared and Collarless Total Hip System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to:
• A severely painful and/or disabled joint from osteoarthritis, theumatoid arthritis, avascular necrosis, or congenital hip dysplasia;
· Acute traumatic fracture of the femoral head or neck;
· Failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement.
TheRay Collared and Collarless Total Hip System femoral stems are intended for cementless fixation. The acetabular cup is intended for cementless fixation. The porous structured surfaces on the acetabular cup provide biological fixation in a cementless application.
The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System Femoral Stem or TheRay Collared or Collarless Femoral Stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
· Femoral neck and trochanteric fractures of the proximal femur;
· Osteonecrosis of the femoral head;
· Revision procedures where other devices or treatments for these indications have failed.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
The following 510(k) Summary is provided in accordance with 21 CFR 807.92.
## 510(k) Owner and Registration
| Owner's Name: | Theken Companies, LLC |
|------------------------------------|--------------------------------------|
| | Subsidiary: NextStep Arthropedix |
| Address: | 1800 Triplett Blvd., Akron, OH 44306 |
| Phone Number: | (330) 733-7600 |
| Fax Number: | (330) 733-7602 |
| Date Summary Prepared: | 9/12/2024 |
| Establishment Registration Number: | 3002498892 |
## 510(k) Contact
| Contact: | Theken Companies |
|-----------------|--------------------------------------|
| Address: | 1800 Triplett Blvd., Akron, OH 44306 |
| Phone Number: | 330-733-7600 |
| Fax Number: | 330-733-7600 |
| Contact Person: | Garrett Spurgeon |
#### Device Name and Classification
| Device Trade Name: | TheRay Collared and Collarless Femoral<br>Stem |
|------------------------------------|------------------------------------------------|
| Device Common Name: | Femoral Stem |
| Regulation Number and Description: | 21 CFR 888.3353 |
| Device Class: | Class II |
| Product Codes: | MEH<br>KWY<br>OQG<br>LZO<br>OQI |
| Advisory Panel: | 87 (Orthopedic) |
#### Legally Marketed Predicate
The DePuy-Synthes Actis Duofix Hip Prosthesis is considered the primary predicate (similar indications for use), while the NextStep Arthropedix iNSitu Total Hip System and the NextStep Arthropedix iNSitu Bipolar Hip System are considered reference devices and included for indications for use and performance comparisons.
{5}------------------------------------------------
| Predicate | Company | Device Name | 510(k) Number(s) | Clearance Date |
|-------------------|----------------------|---------------------------|------------------|----------------|
| Primary Predicate | DePuy Synthes | Actis | K150862 | 9/25/2015 |
| Reference Devices | NextStep Arthropedix | iNSitu Total Hip System | K161184 | 10/14/2016 |
| | | iNSitu Bipolar Hip System | K191927 | 9/17/2019 |
| | | iNSitu Total Hip System | K191936 | 8/20/2019 |
#### Device Description
TheRay Collared and Collarless Hip System includes the new subject femoral stems, along with previously cleared femoral heads (K161184, K220336), acetabular cups (additively manufactured – K161184 and K191936), acetabular liners (vitamin E polyethylene – K161184), acetabular bone screws (K161184), screw hole covers for the screw holes in the acetabular cups (K161184), and apical hole covers for the acetabular cups (K161184). The subject femoral stem implant features a forged triplaner Ti-6Al-4V ELI substrate and an applied dual coating comprised of plasma-sprayed commercially pure titanium (CPTi) and hydroxyapatite (HA). The collared version of the subject TheRay Femoral Stem features a collar on the medial aspect of the stem above the coating that is designed to seat on the native resected calcar. The collarless version of the subject TheRay Femoral Stem is identical to the collared version, less the medial collar.
The iNSitu Bipolar Hip System has been previously cleared for use in K191297 and consists of a factory assembled UHMWPE (ASTM F648) liner in a cobalt chrome (ASTM F75) outer shell, and UHMWPE (ASTM F648) retention ring with a Ti-6Al-4V ELI (ASTM F136) spring. These Bipolar Heads include outer diameters ranging from 38 to 60 mm, in 1 mm increments, to properly fit the patient anatomy. The smaller Bipolar Heads (38 to 43 mm) have an inner diameter that mates with a 22mm diameter Femoral Head; the larger Bipolar Heads (44 to 60 mm) have an inner diameter that mates with a 28 mm diameter Femoral Head. The iNSitu Bipolar Hip System may be used in conjunction with the subject TheRay Collared and Collarless Hip System Femoral Stems.
#### Indications for Use:
TheRay Collared and Collarless Total Hip System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to:
- A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, ● avascular necrosis, or congenital hip dysplasia;
- . Acute traumatic fracture of the femoral head or neck;
- Failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement.
TheRay Collared and Collarless Total Hip System femoral stems are intended for cementless fixation. The acetabular cup is intended for cementless fixation. The porous structured surfaces on the acetabular cup provide biological fixation in a cementless application.
The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System Femoral Stem or the TheRay Collared or Collarless Femoral Stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
- Femoral neck and trochanteric fractures of the proximal femur;
- Osteonecrosis of the femoral head; .
- . Revision procedures where other devices or treatments for these indications have failed.
## Summary of Technological Characteristics
TheRay Collared and Collarless Femoral Stem is manufactured from the same materials as the predicate stems (substrate - Ti-6Al-4V ELI, coating - CPTi + HA) and feature similar stem lengths, offsets, taper specifications, and head centers as the primary predicate and reference devices. TheRay Collared and Collarless Femoral Stems have identical indications and packaging configurations as the reference devices. The subject stem is therefore substantially equivalent to the predicates based on comparisons of intended use, design features, and technological characteristics.
{6}------------------------------------------------
## Performance Testing
Extensive preclinical performance testing was conducted on the subject TheRay Collared and Collarless Femoral Stem to evaluate the device and to demonstrate substantial equivalence. The results confirm that the TheRay Collared and Collarless Femoral Stem exhibits substantially equivalent performance compared to the predicate devices.
- . Range of Motion (ROM) evaluation (ISO 21535)
- Distal stem fatigue testing (ISO 7206-4)
- . Neck fatigue testing (ISO 7206-6)
- . Femoral head disassembly performance engineering rationale
- . Biocompatibility testing (ISO 10993, ANSI-AAMI ST72, ANSI-AMMI ST98, USP <643>) and rationale
#### Conclusions
TheRay Collared and Collarless Femoral Stem has the same indications for use as predicate hip systems. A comparison of technological characteristics and performance testing demonstrates that TheRay Collared and Collarless Hip Stem is substantially equivalent to the predicate stems.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.