BioSieve™ Fentanyl FIA Home Test Kit; BioSieve™ Fentanyl FIA Pro Test Kit; BioSieve™ Toxismart Reader
Applicant
Vivachek Biotech (Hangzhou) Co., Ltd.
Product Code
NGL · Clinical Toxicology
Decision Date
Oct 4, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3650
Device Class
Class 2
Indications for Use
BioSieve™ Fentanyl FIA Home Test Kit is a fluorescence immunoassay (FIA) for the qualitative determination of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with BioSieve™ Toxismart Reader. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. BioSieve™ Fentanyl FIA Pro Test Kit is a fluorescence immunoassay (FIA) for the qualitative determination of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with BioSieve™ Toxismart Reader. It is for in vitro diagnostic use only. The tests provide only preliminary results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive. BioSieve™ Toxismart Reader is a portable fluorescence instrument for in vitro diagnostic use only. The Reader is designed to perform in vitro diagnostic tests on urine specimens. This Reader is intended for OTC use.
Device Story
System comprises test strip in plastic cassette, dropper, and BioSieve™ Toxismart Reader. Operates via competitive fluorescence immunochromatography; urine sample applied to test device; fluorescently labeled antibodies compete with fentanyl-BSA conjugate on nitrocellulose membrane. Reader measures fluorescence signal at test (T) and control (C) lines. Used in home or clinical settings; operated by lay users or professionals. Output is qualitative (positive/negative) result displayed on reader. Assists in preliminary drug screening; clinical judgment required for positive results. Reader is portable, factory-calibrated, and supports cleaning/disinfection protocols.
Clinical Evidence
No clinical studies performed. Bench testing included a lay-user study (n=140) across three sites with participants aged 18 to >50. Samples spiked with fentanyl at -100%, +/-75%, +/-50%, and +/-25% of the 1.0 ng/mL cutoff were tested. Results showed 100% agreement with LC/MS-confirmed concentrations. Flesch-Kincaid analysis confirmed package insert readability at <8th-grade level.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding principle. Materials: nitrocellulose membrane, fentanyl-BSA conjugate, fluorescent microsphere-labeled antibodies. Reader: portable fluorescence instrument, LED light source, 1.54-inch LCD, 3.7V Li-ion battery, Type-C/USB charging. Dimensions: 12.45cm x 7.25cm x 4cm; weight ~0.36 lbs. Connectivity: Bluetooth disabled. Sterilization: not specified.
Indications for Use
Indicated for qualitative determination of fentanyl in human urine at 1.0 ng/mL cutoff. Intended for OTC use by lay users (Home Test Kit) and professional use (Pro Test Kit). Preliminary results only; requires confirmatory testing via GC/MS or LC/MS.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K241869
B Applicant
VivaChek Biotech (Hangzhou) Co., Ltd
C Proprietary and Established Names
BioSieve™ Fentanyl FIA Home Test Kit; BioSieve™ Fentanyl FIA Pro Test Kit; BioSieve™ Toxismart Reader
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NGL | Class II | 21 CFR 862.3650 - Opiate Test System | TX - Clinical Toxicology |
| KHO | Class I | 21 CFR 862.2560 - Fluorometer for clinical use | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
Addition of new claims (OTC) to a cleared device.
B Measurand:
Fentanyl
C Type of Test:
Qualitative, fluorescence immunoassay (FIA)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
BioSieve™ Fentanyl FIA Home Test Kit is a fluorescence immunoassay (FIA) for the qualitative determination of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with BioSieve™ Toxismart Reader.
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.
BioSieve™ Fentanyl FIA Pro Test Kit is a fluorescence immunoassay (FIA) for the qualitative determination of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with BioSieve™ Toxismart Reader.
It is for in vitro diagnostic use only.
The tests provide only preliminary results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the preferred confirmatory method.
Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.
BioSieve™ Toxismart Reader is a portable fluorescence instrument for in vitro diagnostic use only. The Reader is designed to perform in vitro diagnostic tests on urine specimens. This Reader is intended for OTC use.
C Special Conditions for Use Statement(s):
OTC - Over the Counter
D Special Instrument Requirements:
BioSieve™ Toxismart Reader
IV Device/System Characteristics:
A Device Description:
The test system consists of a test strip enclosed in a plastic cassette (the test device), a dropper, the BioSieve™ Toxismart Reader, and associated labeling. External quality control materials are recommended but not included. The test device is the same physical device as was cleared in k240124.
K241869 - Page 2 of 10
{2}
K241869 - Page 3 of 10
## B Principle of Operation:
The BioSieve™ Fentanyl FIA Test Kit uses the principle of competitive and fluorescence immunochromatography. The nitrocellulose membrane test area (T) of the test strip is coated with fentanyl-bovine serum albumin conjugate, and the quality control area (C) is coated with goat anti-rabbit IgG polyclonal antibody. Both Fentanyl monoclonal antibody and rabbit IgG polyclonal antibody labeled with fluorescent microspheres are embedded on the conjugate pad. The labeled antibody will flow forward with the sample when the urine sample is applied to the sample well of the test device. When the concentration of fentanyl in the sample is higher than or equal to the cut-off of the product, it will compete with the corresponding conjugate coated on the test area (T) to bind to the fluorescently labeled monoclonal antibody. Thus, fluorescence signal rendering of the test line is inhibited, and the result is positive. If the sample does not contain fentanyl or its concentration is lower than the cut-off of the product, the corresponding conjugate on the test line reacts with sufficient fluorescently-labeled monoclonal antibodies. Thus, the test line will develop a fluorescence signal and the result is negative. The quality control area (C) will also develop a fluorescence signal, indicating that the testing has been performed correctly.
## C Instrument Description Information:
1. Instrument Name:
BioSieve™ Toxismart Reader
2. Specimen Identification:
Each sample is assigned a numerical identifier.
3. Specimen Sampling and Handling:
Specimen collection, preparation, and handling conditions are described in the package insert of the assay.
4. Calibration:
The analyzer is factory calibrated and does not need to be recalibrated by the operator.
5. Quality Control:
Quality control materials are recommended but not provided.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
BioSieve™ Fentanyl FIA Test Kit; BioSieve™ ToxiSmart FIA Reader
{3}
B Predicate 510(k) Number(s):
K240124
C Comparison with Predicate(s):
Comparison of BioSieve™ Fentanyl FIA Home/Pro Test Kit to the Predicate Device
| Device & Predicate Device(s): | K241869 | K240124 |
| --- | --- | --- |
| Device Trade Name | BioSieve™ Fentanyl FIA Home Test Kit; BioSieve™ Fentanyl FIA Pro Test Kit | BioSieve™ Fentanyl FIA Test Kit |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For the qualitative determination of fentanyl in human urine. | Same |
| Calibrator and Cut-Off Values | Fentanyl (FTY) 1 ng/mL | Same |
| Methodology | Competitive binding, lateral flow immunochromatographi c assays based on the principle of antigen antibody immunochemistry | Same |
| Specimen Type | Urine | Same |
| Configurations | Cassette | Same |
| Analyzer | Immunofluorescence Analyzer | Same |
| Storage | 2-30 °C | Same |
| General Device Characteristic Differences | | |
| Use Settings | Over-the-counter use | Prescription use only |
| | | |
Comparison of BioSieve™ Toxismart Reader to the Predicate Device
| Device & Predicate Device(s): | K241869 | K240124 |
| --- | --- | --- |
| Device Trade Name | BioSieve™ Toxismart Reader | BioSieve™ ToxiSmart FIA Reader |
| General Device | | |
K241869 - Page 4 of 10
{4}
| Device & Predicate Device(s): | K241869 | K240124 |
| --- | --- | --- |
| Characteristic Similarities | | |
| Intended Use/Indications For Use | Immunofluorescence analyzer designed to perform in vitro diagnostic tests on clinical specimens including drug urine test. | Same |
| Principles of Assay Operation | Competitive immunofluorescence immunoassay | Same |
| Calibration Check | A Quality control test device is supplied with the Reader and used to check the Reader optics and calculation systems. | Same |
| General Device Characteristic Differences | | |
| Test Modes | Standard test | Standard test
Quick test |
VI Standards/Guidance Documents Referenced:
None referenced.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Provided in k240124.
2. Linearity:
Not applicable, this device is intended for qualitative use only.
3. Analytical Specificity/Interference:
Provided in k240124.
4. Assay Reportable Range:
Not applicable, this device is intended for qualitative use only.
K241869 - Page 5 of 10
{5}
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Provided in k240124.
6. Detection Limit:
Provided in k240124.
7. Assay Cut-Off:
Provided in k240124.
8. Accuracy (Instrument):
See accuracy study in Section VII.B.1.
9. Carry-Over:
Provided in k240124.
## B Comparison Studies:
1. Method Comparison with Predicate Device:
Provided in k240124.
2. Matrix Comparison:
Not applicable. The assay is intended to be used with urine samples only.
## C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
## D Clinical Cut-Off:
Not applicable.
K241869 - Page 6 of 10
{6}
E Expected Values/Reference Range:
Not applicable.
F Other Supportive Instrument Performance Characteristics Data:
1) Lay User Study:
A lay user study was performed at three intended user sites with 140 lay persons. They had diverse educational and professional backgrounds and ranged in age from 21 to >50 years. Urine samples were prepared at the following concentrations: -100%, +/-75%, +/-50%, +/-25% of the 1 ng/mL cut-off, by spiking fentanyl into drug free-pooled urine specimens. Each sample was aliquoted into individual containers, blind-labeled and randomized. Testing was conducted using one lot of BioSieve™ Fentanyl FIA Test Kit and a total of 8 BioSieve™ Toxismart Readers. Each participant was provided with the package insert, one blind labeled sample and a device. The concentrations of the samples were confirmed by LC/MS. The results are summarized below:
| % of Cutoff | Number of samples | Fentanyl Concentration by GC/MS (ng/mL) | Lay Person Results | | Percentage of Correct Results (%) |
| --- | --- | --- | --- | --- | --- |
| | | | No. of Positive | No. of Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 0.248 | 0 | 20 | 100 |
| -50% Cutoff | 20 | 0.504 | 0 | 20 | 100 |
| -25% Cutoff | 20 | 0.745 | 0 | 20 | 100 |
| +25% Cutoff | 20 | 1.267 | 20 | 0 | 100 |
| +50% Cutoff | 20 | 1.508 | 20 | 0 | 100 |
| +75% Cutoff | 20 | 1.768 | 20 | 0 | 100 |
Lay-users were also given a questionnaire on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on the package insert and the score revealed a reading grade level of less than 8.
2) Effect of Temperature and Humidity:
A study was conducted to investigate the effect of using a test device outside of the temperature and humidity operating conditions. The product labeling instructs the operator to allow the test device to reach operating temperature (64.4-77°F, 18-25°C) prior to use. The claimed humidity operating range for the system is 10 to 90% RH. Temperatures ranging from 45-110°F (7.2-43.3°C) and relative humidity from 10-90% were tested. Each drug concentration was confirmed by LC-MS/MS. The following conditions were tested:
- Condition 1: ambient condition (20°C and 50% RH)
- Condition 2: excessive temperature (40°C) and humidity's of 5% RH
- Condition 3: excessive temperature (40°C) and humidity's of 95% RH
K241869 - Page 7 of 10
{7}
- Condition 4: stored in the refrigerator, 2-8°C for 1 hour. The test devices were run immediately after taking out of the refrigerator.
- Condition 5: low temperature (7.2 °C) and 5% RH
- Condition 6: low temperature (7.2 °C) and 95% RH
Samples with concentrations of -100% cutoff, -50% cutoff and +50% cutoff for fentanyl cutoff concentration of 1 ng/mL were tested. 3 lots of test devices and 3 readers were used for testing. All samples generated expected results, indicating that the test device is tolerant to inappropriate temperature and humidity.
3) Effect of Vibration:
A study was conducted to evaluate the effect of vibrations produced by nearby laboratory equipment on the performance of the fentanyl test device when run on the analyzer. Test devices were prepared with samples at with concentrations of -100% cutoff, -50% cutoff and +50% cutoff. 3 lots of cartridges and 3 analyzers were used for testing. Samples were tested at ambient temperature, placed in close proximity (within 1 foot) to a centrifuge running at 9,500 rpm or to a centrifuge running at 15,000 rpm. All samples generated expected results, indicating that the device is tolerant to vibrations coming from a centrifuge at the rpm tested.
4) Effect of Time Lapse:
A study was conducted to evaluate the effect of time lapse between applying the sample on the test device and inserting the test device into the reader. The product labeling instructs that during the test procedure, insert the sample-added test device into the slot of BioSieve™ Toxismart Reader immediately. The testing was performed by evaluating time lapse from 0 to 3 minutes (i.e., 0, 1, 2, and 3 minutes). Samples with concentrations of -100% cutoff, -50% cutoff and +50% cutoff were tested. 3 lots of test devices and 3 readers were used for testing. All samples generated expected results, indicating that that the test device is tolerant to inappropriate time lapse between applying the sample on the test device and inserting the test device into the reader.
5) Effect of Drop
A was conducted to evaluate the potential of the invisible damage to the test device when inadvertently dropped or knocked over during the procedure both before and after sample addition to the device.
- Condition 1: reader plus test device were dropped from the test height (6.5 foot, 2 meters) onto the floor prior to sample addition. After dropping, samples were added, and each test device was tested.
- Condition 2: reader plus test device were dropped from the test height (6.5 foot, 2 meters) onto the floor after sample addition. After dropping, each device was tested.
Samples with concentrations of -100% cutoff, -50% cutoff and +50% cutoff were tested. 3 lots of test devices and 3 readers were used for testing. All samples generated expected results, indicating that the test device is tolerant to drop of 6.5 feet (higher than the normal expected height when an operator drops a test device).
6) Effect of Sample Volume:
The study was conducted to investigate the effect of adding insufficient or excess number of drops from the specimen collection container into the sample well. The product labeling instructs that during the test procedure, user should add three drops (~75 μL) of the sample to the sample well of the test device using the Dropper. The testing was
K241869 - Page 8 of 10
{8}
performed by evaluating number of drops from 0.67x to 1.67x the nominal volume (i.e., 2, 3, 4, and 5 drops of sample). Samples with concentrations of -100% cutoff, -50% cutoff and +50% cutoff were tested. 3 lots of test devices and 3 readers were used for testing. All samples generated expected results, indicating that the test device is tolerant to incorrect sample addition to the sample well.
7) Effect of Using Expired Test Devices:
A study was conducted to evaluate the effect of using expired test devices. Three lots of expired cassettes (different expiration dates) were used in this study. Results support that the test device yields an error code when expired cassettes are used.
8) Effect of Analyzer Assembly Held at an Angle While Reading Results:
A study was conducted to evaluate the effect of placing the assembled fentanyl test device in holder with the analyzer at different angle facing the operator reading of the results. Test devices were prepared with sample (with concentrations of -100% cutoff, -50% cutoff and +50% cutoff) and tested at different angle of from 15-90 degrees (i.e., 15, 30, 45, 60, 75 and 90 degree). 3 lots of test devices and 3 readers were used for testing. All samples generated expected results, indicating that holding the analyzer assembly at an angle while reading results has no impact on the performance.
9) Effect of Incorrect Orientation of Test Device in Holder:
A study was conducted to evaluate the effect of incorrect orientation of the test device in the holder on reading the results by the Analyzer. Three incorrect orientations were tested with respect to the correct orientation of the test device in the holder:
- Condition 1: correct orientation
- Condition 2: upside down from the correct orientation
- Condition 3: 180° rotation from correct orientation
- Condition 4: 180° rotation and upside down from correct orientation
Samples with concentrations of -100% cutoff, -50% cutoff and +50% cutoff were tested. 3 lots of test devices and 3 readers were used for testing. Results support that when the orientation is incorrect, the test device cannot be scanned.
10) Effect of Reader Button Pressed Throughout the Run:
A study was to evaluate the effect of touching the left and/or right reader button by mistake throughout the run (from start of the run to result display). Samples with concentrations of -100% cutoff, -50% cutoff and +50% cutoff were tested. 3 lots of test devices and 3 readers were used for testing. All samples generated expected results, indicating that pressing of the left and/or right reader button by mistake during the run does not affect the final test results.
11) Effect of Environmental Light:
A she study was conducted to evaluate the effect of environmental light (with and without direct sunlight) on reading of test results by BioSieve™ Toxismart Reader. The testing was performed across the following three conditions: (1) indoors with only light and no sunlight; (2) outdoors in direct sunlight and (3) indoors with indirect sunlight. Samples with concentrations of -100% cutoff, -50% cutoff and +50% cutoff were tested. 3 lots of test devices and 3 readers were used for testing. All samples generated expected results, indicating that the test device is tolerant to different lighting conditions.
K241869 - Page 9 of 10
{9}
12) Cleaning and Disinfection Robustness Evaluation:
The sponsor provided information to support the potency of the chosen disinfectant, Clorox Healthcare Bleach Germicidal Wipes. Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the reader after 365 cycles of cleaning and disinfection using the chosen disinfectant, simulating cleaning and disinfection over the 3-year use life of the reader. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K241869 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.