The device is intended to provide electrical stimulation and neuromuscular reeducation for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.
Device Story
Levina Pelvic Floor Muscle Stimulator (RS-48) is a non-implantable, home-use device for treating urinary incontinence. System consists of a remote control with 1.77" TFT LCD screen, a stimulator pod, and a reusable, single-patient use vaginal dual-electrode probe. Powered by 3.7V lithium-ion polymer batteries. Remote control sends parameter settings to the stimulator pod, which delivers biphasic, symmetrical electrical stimulation to pelvic floor muscles via the probe to induce Kegel-like contractions. Device strengthens pelvic floor muscles to maintain urinary continence. Operated by the patient in a home environment. Clinical benefit includes rehabilitation of weak muscles and improved continence control.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), home healthcare environment performance (IEC 60601-1-11), nerve/muscle stimulator safety (IEC 60601-2-10), software verification/validation, and electrical performance verification.
Technological Characteristics
Non-implantable electrical stimulator; biphasic, symmetrical rectangular waveform. Powered by 3.7V Li-ion polymer batteries. Dual-electrode vaginal probe (6cm² electrode area). Frequency 12-50 Hz; pulse width 200-300 µsec. Connectivity: remote control to stimulator pod. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ISO 10993-1.
Indications for Use
Indicated for women with stress, urge, or mixed urinary incontinence or weak pelvic floor muscles requiring neuromuscular reeducation and rehabilitation.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 27, 2025
ZMI Electronics Ltd. Lawrence Liu Section Manager 6F-1, 286-4, Shin Ya Road, Chien Chen District Kaohsiung, 806 TAIWAN
Re: K241862
> Trade/Device Name: Levina Pelvic Floor Muscle Stimulator (RS-48) Regulation Number: 21 CFR 876.5320 Regulation Name: Nonimplanted Electrical Continence Device Regulatory Class: II Product Code: KPI Dated: February 25, 2025 Received: February 25, 2025
Dear Lawrence Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Angel A. Soler-garcia -S
for
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241862
Device Name
Levina Pelvic Floor Muscle Stimulator Model: RS-48
Indications for Use (Describe)
The device is intended to provide electrical stimulation and neuromuscular reeducation for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510 (k) Summary
#### Submitter Information 1.
| 510 (k) submitter: | ZMI Electronics Ltd. |
|---------------------------|--------------------------------------------------|
| Address: | 6F-1, 286-4, Shin-Ya Road, Kaohsiung, Taiwan 806 |
| | Phone: +886-7-8150053 |
| | Fax: +886-7-8150057 |
| | Email: mail@zmi-electronics.com |
| Applicant Contact Person: | Lawrence Liu |
| | Regulatory Affairs Manager |
| | Phone: +886-7-8150053 Ext 351 |
| | Fax: +886-7-8150057 |
| | E-mail: lawrence@zmi-electronics.com |
| Preparation date: | March 26, 2025 |
#### 2. Device Name
| Trade Name of the Device: | Levina Pelvic Floor Muscle Stimulator Model: RS-48 |
|----------------------------|-----------------------------------------------------------|
| Common Name: | Nonimplanted electrical continence device |
| Classification Name: | Stimulator, electrical, non-implantable, for incontinence |
| Classification Regulation: | 21 CFR 876.5320 |
| Device Class: | II |
| Panel: | Gastroenterology/Urology |
| Product Code: | KPI |
#### 3. Predicate and Reference Devices
| | 510(k)# | Trade Name |
|-----------|---------|-------------------------|
| Predicate | K160773 | Yarlap II |
| Reference | K081480 | Kegel8 |
| Reference | K230983 | Unicare (K-UNICARE-USA) |
The predicate and reference devices have not been subject to a design related recall.
#### Device Description 4.
Levina Pelvic Floor Muscle Stimulator is a non-implantable, home use device for the treatments of stress, urge, and mixed urine incontinence through the intravaginal probe's electrical stimuli to the muscles of the pelvic floor to induce Kegel-like contractions, which help users to achieve pelvic floor muscle strengthening for maintaining urinary continence in women.
{5}------------------------------------------------
The subject device includes a remote control, a stimulator pod and a vagina stimulation probe. Both the remote control and the stimulator pod are powered by an off-the-shelf 3.7V lithium-ion polymer battery. The remote control has a 1.77" TFT LCD screen and 4 buttons to control the stimulator pod. The stimulator pod has two output channels and will perform electrical stimulation according to the parameter settings sent by the remote control.
The device is supplied with a reusable (single-patient use) vaginal, dual-electrode, stimulation probe.
#### ਹੈ, Indications For Use
The device is intended to provide electrical stimulation and neuromuscular reeducation for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.
#### Comparison of the Technological Characteristics with Predicate Device 6.
| Technological/ Performance<br>Characteristics | | Subject Device | Predicate Device |
|-----------------------------------------------|-------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Device Name | | Levina Pelvic Floor Muscle<br>Stimulator | Yarlap II |
| 510(k) Number | | K241862 | K160773 |
| Indications for Use | | The device is intended to<br>provide electrical stimulation<br>and<br>neuromuscular<br>reeducation for the purpose<br>of rehabilitation of weak<br>pelvic floor muscles for the<br>treatment of stress, urge and<br>mixed urinary incontinence in<br>women | The device is intended to provide<br>electrical stimulation and<br>neuromuscular re-education for<br>the purpose of rehabilitation of<br>weak pelvic floor muscles for the<br>treatment of stress, urge and<br>mixed urinary incontinence in<br>women |
| Environments of Use | | Home environment | Home environment |
| Target Population | | women with incontinence or<br>weak pelvic floor muscles | women with incontinence or weak<br>pelvic floor muscles |
| Prescription or Over-the-Counter<br>(OTC) | | Prescription Use & OTC | Prescription Use & OTC |
| Electrode Types | | Incontinence probe | Incontinence probe |
| Waveform | | Biphasic, Symmetrical | Biphasic, Symmetrical |
| Shape | | Rectangular | Rectangular |
| Maximum<br>Output<br>Voltage<br>(volts) | @500Ω | 40 | 40 |
| | @2kΩ | 50 | 50 |
| | @10kΩ | 50 | 95 |
| Maximum<br>Output<br>Current (mA) | @500Ω | 80 | 80 |
| | @2kΩ | 25 | 50 |
| | @10kΩ | 5 | 19 |
| Pulse Width (usec) | | 200~300 | 200~250 |
| Frequency (Hz) | | 12~50 | 2~100 |
{6}------------------------------------------------
K241862 Page 3 of 4
| For multiphasic waveforms only: | Symmetrical phases | Yes | Yes |
|--------------------------------------------------------|--------------------|--------|---------|
| Phase Duration (µsec) | | 20-150 | 100~125 |
| Maximum Phase Charge (@500Ω) (µC) | | 12 | 10 |
| Electrode Area | | 6cm² | 6.4cm² |
| Maximum (peak) Current Density (mA/cm²) | | 13.33 | 12.5 |
| Maximum Current Density (@500Ω) (mA/cm²) rms | | 1.63 | 1.98 |
| Maximum Average Current (average absolute value), (mA) | | 1.2 | 2 |
| Maximum Average Power Density (@500Ω) (mW/cm2) | | 8 | 12.5 |
As evidenced by the above table, both the subject and the predicate devices have the same intended use, but the subject and predicate devices have different technological characteristics. The reference devices were used to support the stimulation parameters of the subject device. Additionally, performance testing was conducted on the subject device to evaluate its safety and effectiveness. Overall, it was established that the differences in technological characteristics between the subject and the predicate does not raise different questions of safety or effectiveness.
# 7. Non-Clinical Testing
Below is a list of the tests that were performed and successfully completed for the subject device per the specified guidance and standards:
- Biocompatibility testing according to ISO 10993-1:2018 Biological evaluation of medical O devices - Part 1: Evaluation and testing within a risk management process and FDA Guidance "Use of International Standard ISO 10993-1" (2016).
- Electrical Safety testing according to IEC 60601-1: 2020 Medical electrical equipment Basic O safety and essential performance
- Electromagnetic Compatibility testing according to IEC 60601-1-2: 2020 General requirements O for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests
- Software Verification and Validation Testing according to FDA's Guidance for the O Content of Premarket Submissions for Software Contained in Medical Devices"
- o Cybersecurity risk management activities according to FDA Guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, issued September 27, 2023
- Performance testing according to IEC 60601-1-11, Medical electrical equipment -- Part 1-11: O General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical systems used in the home healthcare environment
- Performance test according to IEC 60601-2-10, Medical electrical equipment -- Part 2-10: O Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
{7}------------------------------------------------
- Electrical performance testing to verify the stimulation parameters O
All pre-determined acceptance criteria were met.
#### 8. Conclusions
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.