Unicorn+ RF System (Unicorn+); Unicorn+ RF System (Unicorn+ I); Unicorn+ RF System (Unicorn+ II); Unicorn+ RF System (Unicorn+ III)
K241832 · Boston Aesthetics, Inc. · GEI · Feb 13, 2025 · General, Plastic Surgery
Device Facts
Record ID
K241832
Device Name
Unicorn+ RF System (Unicorn+); Unicorn+ RF System (Unicorn+ I); Unicorn+ RF System (Unicorn+ II); Unicorn+ RF System (Unicorn+ III)
Applicant
Boston Aesthetics, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Unicorn+ RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The RF System is used for the relief of minor muscle aches and pain, muscle spasm, and temporary improvement of local blood circulation.
Device Story
Unicorn+ RF System consists of host, trolley, footswitch, power cord, and various handpieces/electrodes. Device delivers RF energy to tissue via bipolar or monopolar modes to achieve electrocoagulation, hemostasis, or therapeutic heating. Operated by clinicians in clinical settings; user controls parameters via touchscreen. System utilizes three electrode types (MicroRF, Artist, Pure+) acting at varying skin depths. Energy output is a function of power and duration. Device provides thermal effects for pain relief and circulation improvement. Clinical benefit derived from controlled thermal tissue interaction; safety verified through bench testing and thermal damage comparisons against predicates.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including biocompatibility (ISO 10993 series), electrical safety (IEC 60601-1, 60601-2-2), EMC (IEC 60601-1-2), software validation, and ex vivo thermal testing on tissue samples to confirm therapeutic temperature maintenance (40-45°C).
Technological Characteristics
Radiofrequency electrosurgical system; 1MHz frequency; 50W max output power. Bipolar and monopolar modes. Components: host, trolley, footswitch, handpieces, electrodes. Sterilization: Ethylene Oxide (ISO 11135). Biocompatibility: ISO 10993. Electrical safety: IEC 60601-1, IEC 60601-2-2. Shelf-life: ASTM F1980-16.
Indications for Use
Indicated for patients requiring dermatologic and general surgical procedures for electrocoagulation and hemostasis, or relief of minor muscle aches, pain, muscle spasms, and temporary improvement of local blood circulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
February 13, 2025
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that.
Boston Aesthetics Inc Hongmei Cao General Manager 1521 Concord Pike Suite 201 Wilmington, Delaware 19803
Re: K241832
Trade/Device Name: Unicorn+ RF System (Unicorn+); Unicorn+ RF System (Unicorn+ I); Unicorn+ RF System (Unicorn+ II); Unicorn+ RF System (Unicorn+ III) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI, PBX Dated: November 20, 2024 Received: January 17, 2025
Dear Hongmei Cao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James H. Digitally signed by
James H. Jang -S
Date: 2025.02.13
Jang -S
17:23:19-05'00'
For Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Submission Number (if known)
K241832
Device Name
Unicorn+ RF System (Unicorn+); Unicorn+ RF System (Unicorn+ I); Unicorn+ RF System (Unicorn+ II);
Unicorn+ RF System (Unicorn+ III)
### Indications for Use (Describe)
The Unicorn+ RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The RF System is used for the relief of minor muscle aches and pain, muscle spasm, and temporary improvement of local blood circulation.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
### General Information 1.
| Submitter: | Boston Aesthetics INC<br>1521 Concord Pike Suite 201 Wilmington DE 19803.US |
|---------------------------|------------------------------------------------------------------------------------------|
| Contact Person: | Hongmei Cao<br>General Manager<br>Tel: 001 949-792-816<br>Email: bsnaesthetics@gmail.com |
| Summary Preparation Date: | June 14,2024 |
### 2. Device Name and Code
Trade/Common Name: Unicorn+ RF System Classification Name: Electrosurgical, Cutting & Coagulation & Accessories Classification: Class II Product Code: GEI, PBX Regulation Number: 21 CFR 878.4400 Review Panel: General & Plastic Surgery
### Predicate Device 3.
Predicate Device 1:Potenza, cleared under Traditional 510(k) K192545 Predicate Device 2: AGNES, cleared under Traditional 510(k) K160469 Predicate Device 2: Capenergy C Equipment RF System, cleared under Traditional 510(k) K222260
### 4. Device Description
Unicorn+ RF System is consist of a host, a trolley, a footswitch, a power cord, three kinds of handpieces and matched electrodes.
#### Indications for Use 5.
The Unicorn+ RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The RF System is used for the relief of minor muscle aches and pain, muscle spasm, and temporary improvement of local blood circulation.
### Technological Characteristics in Comparison to the Predicate Device 6.
Although this subject device has three Predicate devices, they have same fundamental technology theory. Operators control host through touch screen, outputs different amounts of RF energy to handpieces and matched electrodes. These three kinds of electrodes(including invasive or non-invasive electrodes) act on different skin depth to achieve different treatment modes.
The subject device has the same intended use and indications for use and the same fundamental scientific technology as the Predicate devices. The subject device design, technology, and the
{5}------------------------------------------------
principles of operation are the same as the predicate devices. Although subject device output has minor differences with some predicate devices, and RF Energy= RF Output power × On Time, subject device conducts treatment duration to achieve same treatment with predicates devices. Therefore, the minor differences do not raise any new safety and effectiveness results because the parameters of the Subject device are similar to those of the predicate device.
#### Substantial Equivalence Table 7.
A comparison of the intended use and technological characteristics of the subject device and predicate device is provided in the table below:
| Items | Subject device | Predicate device<br>(K192545) | Predicate device<br>(K160469) | Predicate device<br>(K222260) |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | Unicorn+ RF<br>System | POTENZA | AGNES | Capenergy C<br>Equipment RF System<br>- C25, C50, C100,<br>C200, C300, C400,<br>C500 |
| Manufacturer | Boston Aesthetics<br>INC | Jeisys Medical<br>Incorporated | Gowoonsesang<br>Cosmetics Co.,<br>Ltd | Capenergy Medical<br>S.L. |
| Classification<br>Product Code | GEI,PBX | GEI | GEI,KCW | PBX |
| Indication for<br>Use | The Unicorn+<br>RF System is<br>intended for use<br>in dermatologic<br>and general<br>surgical<br>procedures for<br>electrocoagulatio<br>n and hemostasis.<br>The RF System is<br>used for the relief<br>of minor muscle<br>aches and pain,<br>muscle spasm,<br>and temporary<br>improvement of<br>local blood<br>circulation | The POTENZA is<br>intended for use in<br>dermatologic and<br>general surgical<br>procedures for<br>electrocoagulation<br>and hemostasis. | AGNES is<br>indicated for use<br>in dermatological<br>and general<br>surgical<br>procedures for<br>electrocoagulation<br>and hemostasis. | Capenergy C<br>equipments RF System<br>- C25, C50, C100,<br>C200, C300, C400,<br>C500 are intended to<br>provide topical heating<br>for the purpose of<br>elevating tissue<br>temperature for<br>treatment of selected<br>medical conditions<br>such as: relief of pain,<br>muscle<br>spasms, increase in<br>local circulation, and<br>dermatological<br>procedures.<br>The massage device<br>provided is intended to<br>provide a temporary<br>reduction in the<br>appearance of cellulite. |
{6}------------------------------------------------
| | | | | Capenergy C<br>equipments RF System<br>- C25, C50, C100,<br>C200, C300, C400,<br>C500 are indicated for<br>use in dermatological<br>and general surgical<br>procedures for<br>electrocoagulation and<br>hemostasis. |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|--------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription/Ov<br>er-the Counter<br>Use | Prescription | Prescription | Prescription | Prescription |
| System Type | Bipolar RF<br>(Radiofrequency)<br>and Monopolar<br>RF<br>(Radiofrequency) | Bipolar RF<br>(Radiofrequency)<br>and Monopolar<br>RF(Radiofrequency) | Monopolar RF<br>(Radiofrequency) | Monopolar RF<br>(Radiofrequency) |
| Frequency | 1MHz | 1MHz, 2Mhz | 1MHz | 1MHz |
| Max Output<br>Power | 50W | 50W | 46W | 45W |
| Treatment<br>Duration(Time) | MicroRF:10-600<br>ms<br>Artisit:<br>50ms-5000ms<br>Pure+B1:1-30min | 5-500ms | Min 50ms / Max<br>1,000ms | 15-660 sec |
| Tips | single electrode:<br>ArtistD2.0,<br>2.0mm<br>ArtistD3.5,<br>3.5mm<br>ArtistD4.5,<br>4.5mm<br>ArtistD6.0,<br>6.0mm<br>Pure+B1, 3cm^2<br>dual electrodes:<br>MicroRF 49,<br>0.5-3.5mm<br>MicroRF 25,<br>0.5-3.5mm | Depth: Maximum of<br>4.0mm<br>Thickness : 0.25mm | Needle length :<br>0.8/1.25/1.5/ 2.0<br>mm<br>Thickness :<br>0.2mm | N.A. |
| | MicroRF 9N,<br>0.5-3.5mm<br>MicroRF 25N,<br>0.5-3.5mm | | | |
| Single Use<br>or Reusable | Single<br>Use/Resusable | Single<br>Use/Resusable | Single Use | Reusable |
| Sterilization | EO gas | EO gas | EO gas | N.A. |
{7}------------------------------------------------
Unicorn+ RF System belongs Product Code GEI and PBX in Classification,and does not own Needle-type epilator,so that not belong to KCW. Therefore, Unicorn+ RF System Classification code are all same as three predicate devices code. The Unicorn+ RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. This description is same as POTENZA and AGNES. The RF System is used for the relief of minor muscle aches and pain, muscle spasm, and temporary improvement of local blood circulation. This description is same as Capenergy C Equipment RF System statement, which Capenergy C equipments RF System -C25, C50, C100, C200, C300, C500 are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation, and dermatological procedures. The Unicorn+RF system owns 3 modes matched 3 different types of tips. MicroRF, is Bipolar RF (Radiofrequency) .which is same as POTENZA Bipolar tips in 1Mhz mode, Moreover, Artist&Pure+ tips are Monopolar RF (Radiofrequency) same as others. Treatment depends on device supplied energy , the energy is affected by treatment duration and power supply. Micro RF tips treatment duration is minimal different from POTENZA ones. Artist tips treatment is 50ms-5000ms, much longer than AGNES treatment duration, however, Artist tips max output power is designed 20W, and AGNES is 46W, thus the influence on energy with tissue is also minimal. Finally on Pure+ tips is not a point of use, this tip is effect on skin area, thus the treatment duration depends on treatment area size. MicroRF tips length is no more than POTENZA tips maximum length; Some of Artist tips lengths are longer than AGNES, but tips lengths is effect on treatment target tissue area under the skin, because all tips divided into non-insulated and insulated part, only non-insulated release energy to target tissue.
Conclusion, Unicorn+RF system is designed similar with predicate devices, they are same indications for use, and the operating principle and technological characteristics. Although there are some differences as output power, treatment duration and tip physical dimensions. the thermal damage tests which is compare with the predicate devices and performance test reports are supported to the safety and effectiveness of the subject device.
#### 8. Performance Date
The following performance data were provided in support of the substantial equivalence determination:
## Biocompatibility
The biocompatibility evaluation for electrode was conducted in accordance with the International Standard ISO 10993-1 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing
{8}------------------------------------------------
within a Risk Management Process, as recognized by FDA. The subject device is contacted with patient during the treatment for duration of less than 24 hours. The biocompatibility testing includes the following:
- · In Vitro Cytotoxicity Test: ISO 10993-5: 2009
- · Skin Sensitization Test: ISO 10993-10:2021
- · Intracutaneous Reactivity Test: ISO 10993-23:2021
- · Acute Systemic Toxicity Test: ISO 10993-11:2017
- · Pyrogen Test: ISO 10993-11:2017
Sterilization Validation Sterilization validation of the electrode has been conducted per the standard ISO 10993-7:2008 for Ethylene Oxide(standard ISO 11135:2014).
# Shelf-life Validation
The shelf-life validation of the device has been conducted using the accelerated aging method in accordance to ASTM F1980-16.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety testing and EMC testing were conducted on the Radio frequency System complying with the IEC 60601-1 standard, the IEC 60601-2-2 standard ,the IEC 60601-1-6 standard for safety and the IEC 60601-1-2 standard for EMC.
# Software Verification and Validation Testing
Software Verification and Validation Testing were conducted and documentation was provided as recommended by FDA relevant Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device was considered as a 'moderate' level of concern, since a failure or latent flaw in the software could indirectly result in minor injury to the patient or operator.
# Ex Vivo testing
Ex Vivo testing conducted on four types of tissue-Live, Kidney, Skin and Muscle under GLP Thermal testing in accordance with FDA's "Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", August 15, 2014
## Summary discussion Non-clinical tests data
Based on bench test, the device demonstrated ability to reach and maintain therapeutic temperature 40-45°C on the surface of skin, and the actual skin temperature test was conducted. No preclinical and clinical data was provided in this submission
#### 9. Conclusion
In comparing between the subject device and the predicate devices, there are the same indications for use, the operating principle and technological characteristics .Although there are some differences as output power, treatment duration and tip physical dimensions. the thermal damage test which is compare with the predicate devices and performance test reports are supported to the safety and effectiveness of the subject device.
{9}------------------------------------------------
In this regard, we conclude that the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.