K241789 · Hony Medical Co., Ltd. · MUI · Jul 24, 2024 · Radiology
Device Facts
Record ID
K241789
Device Name
Non-Sterile Ultrasound Transmission Gels
Applicant
Hony Medical Co., Ltd.
Product Code
MUI · Radiology
Decision Date
Jul 24, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Non-sterile Ultrasound Transmission Gels are ultrasound couplants intended to be used on intact skin during non-invasive medical ultrasound procedures to couple sound waves between a patient and the medical imaging electronics. The gels are intended for use in all diagnostic ultrasound procedures which require coupling gel or fluid. The gels can be used during the procedures that involved adults and pediatrics in professional healthcare facility.
Device Story
Non-Sterile Ultrasound Transmission Gels function as acoustic couplants; applied to intact skin during diagnostic ultrasound procedures. Gel enables tight bond between patient skin and ultrasound transducer; facilitates transmission of sound waves to underlying tissues for imaging. Used in professional healthcare facilities by clinicians. Device is a water-based gel; formulated to improve acoustic transmission. Benefits patient by ensuring high-quality ultrasound image acquisition. Available in various volumes (20g to 5000g/ml).
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Water-based gel composed of purified water, glycerin, polyethylene glycol, carbomer, phenoxyethanol, triethanolamine, and methylparaben. Acoustic properties: velocity 1520-1620m/s, impedance 1.5-1.7x10^6 Pa·s/m, attenuation ≤0.1dB/(cm·MHz). Physical properties: viscosity ≥15Pa·s, density 987-1049kg/m³, pH 5.5-8.0. Biocompatibility per ISO 10993-1 (sensitization, irritation, acute systemic toxicity). Shelf life 2 years (ASTM F1980).
Indications for Use
Indicated for use as an ultrasound couplant on intact skin during non-invasive medical ultrasound procedures in adults and pediatrics within professional healthcare facilities.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
EcoVue® Sterile and Non-Sterile Ultrasound Gels (K181363)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Hony Medical Co., Ltd. % Wang Boyle General Manager Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai. 200120 CHINA
Re: K241789
Trade/Device Name: Non-Sterile Ultrasound Transmission Gels Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: MUI Dated: June 21, 2024 Received: June 21, 2024
Dear Wang Boyle:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
July 24, 2024
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241789
Device Name
Non-Sterile Ultrasound Transmission Gels
ndications for Use (Describe)
The Non-sterile Ultrasound Transmission Gels are ultrasound couplants intended to be used on intact skin during non-invasive medical ultrasound procedures to couple sound waves between a patient and the medical imaging electronics.
The gels are intended for use in all diagnostic ultrasound procedures which require coupling gel or fluid.
The gels can be used during the procedures that involved adults and pediatrics in professional healthcare facility.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K241789 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
| Name: | Hony Medical Co., Ltd. |
|----------|---------------------------------------------------------------------|
| Address: | No.12,9 Road,Changlong Ind.Zone,Taishan City,Guangdong China 529222 |
| Tel: | +86 15916327827 |
| Contact: | Zhu Huina |
#### Designated Submission Correspondent
| Contact: | Mr. Boyle Wang |
|----------|------------------------------------------------------------------------|
| Name: | Shanghai Truthful Information Technology Co., Ltd. |
| Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120<br>China |
| Tel: | +86-21-50313932 |
| Email: | Info@truthful.com.cn |
Date submitted: Mar.28,2024
#### 2.0 Device Information
| Trade name: | Non-Sterile Ultrasound Transmission Gels |
|----------------------|------------------------------------------|
| Common name: | Diagnostic ultrasonic transducer |
| Classification name: | Media, Coupling, Ultrasound |
| Production code: | MUI |
| Regulation number: | 21 CFR 892.1570 |
| Classification: | Class II |
| Panel: | Radiology |
#### 3.0 Predicate Device Information
| Manufacturer: | HR Pharmaceuticals, Inc. |
|--------------------|-------------------------------------------------|
| Trade/Device Name: | EcoVue® Sterile and Non-Sterile Ultrasound Gels |
| 510(k) number: | K181363 |
#### 4.0 Device Description
The proposed device, Non-Sterile Ultrasound Transmission Gels, are non- irritating,
{4}------------------------------------------------
non sensitizing acoustic couplants intended for use as a scanning gel in medical diagnostic ultrasound procedures. The gel is an accessory used to couple sound waves between the patient and the medical imaging electronic transducers during diagnostic ultrasound procedures. Ultrasound gel is a type of conductive medium that enables a tight bond between the skin and the probe or transducer, allowing the waves to transmit directly to the tissues beneath and to the parts that need to be imaged. As such, the gel is formulated to improve the acoustic transmission of sound waves to create the ultrasound image.
The subject device is available in the below sizes: 20g/bag,250ml/bottle, 1000ml/bottle,2000ml/bottle,5000ml/bottle,250g/bottle,1000g/bottle,2000 g/bottle,5000g/bottle.
#### 5.0 Indication for Use Statement
The Non-sterile Ultrasound Transmission Gels are ultrasound couplants intended to be used on intact skin during non-invasive medical ultrasound procedures to couple sound waves between a patient and the medical imaging electronics.
The gels are intended for use in all diagnostic ultrasound procedures which require coupling gel or fluid.
The gels can be used during the procedures that involved adults and pediatrics in professional healthcare facility.
## 6.0 Summary of Non-Clinical Testing
Summary of non-clinical and performance testing Bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. Results from testing performed confirms that the design requirement specification and user needs have been met. The subject device is confirmed to be safe and effective for the intended use.
6.1 Shelf life –The label shelf life of the Non-sterile Ultrasound Transmission Gels is 2 years. An accelerated aging test was conducted per ASTM F1980. The service life is 2 years is reasonable and effective.
6.2 Biocompatibility testing - Non-Sterile Ultrasound Transmission Gels have successfully been tested for sensitization, intracutaneously irritation, and Acute Systemic Toxicity Test Testing. The test results verify that the biocompatibility criteria given in ISO 10993-1 are fulfilled. Non-Sterile Ultrasound Transmission Gels are non-sensitizing, non-irritating and non-acute systemic toxicity.
6.3 Performance testing - Bench The performance of Ultrasound Transmission Gels has been verified. Tests as described in table 1 have been completed.
{5}------------------------------------------------
| Test Item | Description | Test Results |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Sound Velocity<br>(Acoustic<br>Velocity) | Measured at 35°C ,The Sound Velocity<br>(Acoustic Velocity) shall be 1520-1620m/s | 1538.1m/s ~1540.3m/s<br>Pass |
| Acoustic<br>Impedance | Measured at 35°C ,Acoustic Impedance shall be $1.5×10^6$ ~ $1.7×10^6$ Pa·s/m | $1.59×10^6$ Pa·s/m~ $1.60×10^6$ Pa·s/m<br>Pass |
| Sound<br>Attenuation | Measured at 35°C ,Sound Attenuation shall be ≤0.1dB/(cm • MHz) | 0.06~0.07dB/(cm • MHz)<br>Pass |
| Viscosity | Measured at 25°C, the viscosity of the product should not be less than 15Pa-s. | 101Pa·s~103Pa·s<br>Pass |
| Density | The density of the product shall be 987-1049kg/m³ | 1033.42kg/m³~1038.46kg/m³<br>Pass |
| pH | The pH value of the product should be 5.5~8.0 | 6.74~6.79<br>Pass |
| Antimicrobial<br>Effectiveness | Bacteria: Not less than a 2.0 log reduction from the initial count at 14 days, and no increase from the 14 days count at 28 days<br>Yeast and Molds: No increase from the initial calculated count at 14 and 28 days | Pass |
| Appearance | The product is colorless or light-colored transparent gel without insoluble foreign matter. Under normal storage conditions of the product, there will be no delamination, mildew and odor | Pass |
| Table 1: Performance testing summary - Bench | | |
|----------------------------------------------|--|--|
| | | |
## 7.0 _Summary of Clinical Testing
No clinical study is included in this submission.
#### 8.0 Technological Characteristic Comparison Table
## Table 2- Comparison of Technology Characteristics
| Item | Subject Device | Predicate Device |
|------------|----------------|------------------|
| 510(k) No. | K241789 | K181363 |
{6}------------------------------------------------
| Product Code | MUI | MUI |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation No. | 21 CFR 892.1570 | 21 CFR 892.1570 |
| Class | II | II |
| Intended<br>Use/Indication for<br>Use | The Non-sterile Ultrasound<br>Transmission Gels are ultrasound<br>couplants intended to be used on<br>intact skin during non-invasive<br>medical ultrasound procedures to<br>couple sound waves between a<br>patient and the medical imaging<br>electronics.<br>The gels are intended for use in all<br>diagnostic ultrasound procedures<br>which require coupling gel or fluid.<br>The gels can be used during the<br>procedures that involved adults<br>and pediatrics in professional<br>healthcare facility. | EcoVue® Sterile and Non-Sterile<br>Ultrasound Gels are ultrasound<br>couplants intended to be used on<br>intact skin during non-invasive<br>medical ultrasound procedures to<br>couple sound waves between a<br>patient and the medical imaging<br>electronics.<br>The gels are intended for use in<br>all diagnostic ultrasound<br>procedures which require<br>coupling gel or fluid. |
| Patient-contacting<br>material | Purified water, glycerin,<br>polyethylene glycol, carbomer,<br>phenoxyethanol, triethanolamine<br>and methylparaben. | Water-based gel<br>(sterile and non-sterile) |
| Target Population | Pediatric and adult | Pediatric and adult |
| Environment of Use | Healthcare Facility | Healthcare Facility |
| Model | Single Use:20g/bag,<br>Multiple Uses:<br>250ml/bottle,500ml/bottle,1000ml/<br>bottle,2000ml/bottle,<br>5000ml/bottle,250g/bottle,<br>500g/bottle,1000g/bottle,<br>2000g/bottle,5000g/bottle. | 20g individual packets (both<br>sterile and non-sterile)<br>Multiple Uses:<br>250g pouch (non-sterile) |
| Sterile | Non-sterile | Both in Gamma sterilization<br>and non-sterile |
| Shelf Life | 2 years | 1 years |
| Appearance | The product is colorless or<br>light-colored transparent gel<br>without insoluble foreign matter. | Clear to Hazy Color; free from<br>foreign matter |
| Sound Velocity<br>(Acoustic Velocity) | 1520-1620m/s | 1398-1750 m/s |
| Acoustic<br>Impedance | $1.5×10^6~1.7×10^6 Pa·s/m$ | $1.40×10^6-1.80 ×10^6 Pa·s/m$ |
| Sound<br>Attenuation | ≤0.1dB/(cm-MHz) | 0.32-0.95 dB/cm at 5 MHz |
| | | 0.65-1.10 dB/cm at 7.5 MHz |
| | | 0.85-1.55 dB/cm at 10 MHz |
| Viscosity | $\ge$ 15Pa·s (15,000 cP) | >35,000 cP |
| Density | 987-1049kg/m³ | 850-1150 kg/cm³ |
| pH | 5.5~8.0 | 5.5 – 7.8 |
| Biocompatibility | Conform with ISO10993-1<br>(ISO10993-10, ISO10993-11) | Conform with ISO 10993<br>standards |
{7}------------------------------------------------
The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:
- Same intended use
- · Same indications for use
- Same Technological Characteristics
- · Same Similar Physical Characteristics
- · Same Similar Performance Specifications
- Biocompatible
- · Water-based gels
Physical and Chemical Properties of the subject device are a little different with those of the predicate device, but all the required values are within those of predicate device.
The predicate's Ultrasound Gels are including sterile and non-sterile Ultrasound Gels in its 510(k) clearance under K181363. Only Non-sterile Ultrasound Transmission Gels are included in this current 510(k). The subject device's indications are compared to the Ultrasound Gels. The proposed device's Indications for Use are, therefore, more narrow than the Indications for Use stated in K181363, which applied to both the sterile and non-sterile Ultrasound Gels, as the subject device's indications are only applicable to the Non-sterile Ultrasound Transmission Gels, there is no significant risk raised by the difference.
## 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device in K181363 and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.