K241737 · Sway Medical, Inc. · POM · Feb 15, 2025 · Neurology
Device Facts
Record ID
K241737
Device Name
Sway System Sports Plus
Applicant
Sway Medical, Inc.
Product Code
POM · Neurology
Decision Date
Feb 15, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1471
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Sway System Sports Plus is intended for use as a computerized cognitive assessment test to aid in the assessment and management of concussion in individuals ages 18-24. The Sway System Sports Plus is indicated for use when there is a suspected concussion or head injury.
Device Story
Sway Sports+ is a prescription software-as-a-medical-device (SaMD) for concussion management. It runs on iOS/Android mobile devices, utilizing internal accelerometers, gyroscopes, and touch-screen inputs to record patient responses to cognitive and balance tasks. The device measures reaction time, impulse control, inspection time, and memory; it also includes a balance test. Used in clinical (offices, clinics) and non-clinical (sideline) settings by healthcare providers, the device displays summary statistics (e.g., mean reaction time) via a mobile app, web portal, or PDF. It does not provide diagnostic or treatment recommendations, nor does it identify the presence/absence of concussion; it serves as an assessment aid to be used alongside other diagnostic tools. By providing objective cognitive performance data, it assists clinicians in monitoring recovery and making informed management decisions.
Clinical Evidence
Evidence includes a normative database (n=126,000) and multiple studies. Validity studies (n=88, n=85) showed significant correlations (r=0.22-0.42) with traditional neuropsychological tests and ImPACT. Reliability studies (n=55, n=97) demonstrated strong test-retest reliability (ICC 0.68-0.92). Sensitivity/specificity study (n=46) showed 69.6% sensitivity for concussion group membership, comparable to ImPACT (58.7%, p=0.383). Data supports the device as a reliable tool for monitoring cognitive performance in concussed populations.
Technological Characteristics
Software-only device; runs on Apple iOS (v13+) and Android (v9.0+). Utilizes mobile device internal accelerometer, gyroscope, and touch screen for input. Standalone application; no patient contact. Complies with ISO 14971, IEC 62304, and IEC 62366. Connectivity via mobile OS; data exportable via web portal (CSV/PDF).
Indications for Use
Indicated for individuals ages 18-24 with a suspected concussion or head injury to aid in cognitive assessment and management.
Regulatory Classification
Identification
The computerized cognitive assessment aid for concussion is a prescription device that uses an individual's score(s) on a battery of cognitive tasks to provide an indication of the current level of cognitive function in response to concussion. The computerized cognitive assessment aid for concussion is used only as an assessment aid in the management of concussion to determine cognitive function for patients after a potential concussive event where other diagnostic tools are available and does not identify the presence or absence of concussion. It is not intended as a stand-alone diagnostic device.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Software, including any proprietary algorithm(s) used by the device to arrive at its interpretation of the patient's cognitive function, must be described in detail in the software requirements specification (SRS) and software design specification (SDS). Software verification, validation, and hazard analysis must be performed.
(2) Clinical performance data must be provided that demonstrates how the device functions as an interpretation of the current level of cognitive function in an individual that has recently received an injury that causes concern about a possible concussion. The testing must:
(i) Evaluate device output and clinical interpretation.
(ii) Evaluate device test-retest reliability of the device output.
(iii) Evaluate construct validity of the device cognitive assessments.
(iv) Describe the construction of the normative database, which includes the following:
(A) How the clinical workup was completed to establish a “normal” population, including the establishment of inclusion and exclusion criteria.
(B) Statistical methods and model assumptions used.
(3) The labeling must include:
(i) A summary of any clinical testing conducted to demonstrate how the device functions as an interpretation of the current level of cognitive function in a patient that has recently received an injury that causes concern about a possible concussion. The summary of testing must include the following:
(A) Device output and clinical interpretation.
(B) Device test-retest reliability of the device output.
(C) Construct validity of the device cognitive assessments.
(D) A description of the normative database, which includes the following:
(
*1* ) How the clinical workup was completed to establish a “normal” population, including the establishment of inclusion and exclusion criteria.(
*2* ) How normal values will be reported to the user.(
*3* ) Representative screen shots and reports that will be generated to provide the user results and normative data.(
*4* ) Statistical methods and model assumptions used.(
*5* ) Whether or not the normative database was adjusted due to differences in age and gender.(ii) A warning that the device should only be used by health care professionals who are trained in concussion management.
(iii) A warning that the device does not identify the presence or absence of concussion or other clinical diagnoses.
(iv) A warning that the device is not a stand-alone diagnostic.
(v) Any instructions technicians must convey to patients regarding the administration of the test and collection of cognitive test data.
In combination with the general controls of the FD&C Act, the Computerized Cognitive Assessment Aid for Concussion is subject to the following special controls:
{0}------------------------------------------------
February 15, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Sway Medical, Inc. % Robert Steurer Consultant Steurer Consulting Group, LLC 800 Blue Quail Rd Keller, Texas 76248
Re: K241737
Trade/Device Name: Sway System Sports Plus Regulation Number: 21 CFR 882.1471 Regulation Name: Computerized Cognitive Assessment Aid For Concussion Regulatory Class: Class II Product Code: POM Dated: January 15, 2025 Received: January 15, 2025
Dear Robert Steurer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay R. Gupta -S
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Sway System Sports Plus
#### Indications for Use (Describe)
The Sway System Sports Plus is intended for use as a computerized cognitive assessment and management of concussion in individuals ages 18-24.
The Sway System Sports Plus is indicated for use when there is a suspected concussion or head injury.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Date Prepared: | February 12, 2025 |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Sway Medical, Inc.<br>32 S. Lewis Ave.<br>Tulsa, OK 74101 |
| Submitter Contact: | Sway Medical, Inc.<br>32 S. Lewis Ave.<br>Tulsa, OK 74101<br>Name: Alex Pettigrew<br>Sway Medical Inc<br>Phone: (281) 726-3820<br>Email: alex.pettigrew@swaymedical.com |
| Application<br>Correspondent: | Robert Steurer<br>Principal Consultant<br>Steurer Consulting Group<br>800 Blue Quail Rd.<br>Keller, TX 76248<br>steurerbob@gmail.com<br>+1 (425) 358-1072 |
| Manufacturing<br>Site: | Sway Medical, Inc.<br>32 S. Lewis Ave.<br>Tulsa, OK 74101 |
| Trade Name: | Sway Sports+ |
| Common Name: | Neurological Diagnostic Device |
| Classification<br>Name:<br>Device Class: | Computerized Cognitive Assessment Aid for Concussion<br>Class II |
| Primary<br>Classification<br>Regulation: | 21 CFR §882.1471 |
{5}------------------------------------------------
Primary Product Code:
POM
Predicate Device: Impact Applications, Inc. ImPACT Quick Test, K170551
Reference
Predicate Device: Impact Version 4, K202485
| Substantially<br>Equivalent<br>Devices: | New Device | Predicate<br>Device | Reference Predicate<br>Device |
|-----------------------------------------|--------------------------------------------------|-----------------------------------------------------------|----------------------------------------------------------|
| | Sway Sports Plus (Sports+)<br>Sway Medical, Inc. | ImPACT Quick Test<br>K170551<br>Impact Applications, Inc. | ImPACT Version 4<br>K202485<br>ImPACT Applications, Inc. |
{6}------------------------------------------------
Sway Sports Plus (referred to as Sway Sports+) is a licensed variant of the Device Sway System and is a software solution (SaMD) that provides a computerized Description: cognitive assessment aid for concussion. This prescription device uses an individual's score(s) on a batterv of cognitive tasks to provide an indication of the current level of cognitive function in response to a suspected concussion. The computerized cognitive assessment aid for concussion is used only as an assessment aid in the management of concussion to determine cognitive function for patients after a potential concussive event where other diagnostic tools are available and does not identify the presence or absence of concussion. The examiner may have the patient complete the Sway Balance test to provide an additional indicator in their assessment of concussion. The Sway System is not intended as a stand-alone diagnostic device. After the patient completes the Sway Sports+ test(s), the examiner can view the summary statistics for each test such as mean reaction time (speed).
> Sway Sports+ does not create new information through analysis or interpretation, nor does it offer any diagnostic or treatment recommendations. The Sway Sports+ is not intended to be used as a standalone diagnostic device.
Sway Sports+ operates in the Apple iOS or Android operating systems, enabling it to be used on the hardware of Apple iPhone. iPad, and iPod Touch. or Android mobile devices. The internal accelerometer is accessed to analyze motion of the device in response to stimuli during testing. The mobile device hardware is not provided by Sway Medical.
Environments where Sway Sports+ may be used may include both clinical and non-clinical settings. Clinical settings may include physician offices, hospitals, physical or occupational therapy clinics, skilled nursing facilities, assisted living centers and other healthcare facilities where cognitive tests may be evaluated. In addition, non-clinical environments may include sports team's locations where the application may be used on the sideline to perform a cognitive assessment, e.g., fitness centers, gymnasiums, or stadiums. It is recommended that the tests be taken in a supervised environment for baseline and post-injury testing.
Sway Sports+ consists of the Sway Balance (K121590) and the Sway Computerized Cognitive Assessment Aid. The Sway Computerized Cognitive Assessment Aid (PTY) is exempt from 510(k).
{7}------------------------------------------------
| Intended Use: | The Sway System Sports Plus is intended for use as a computerized cognitive assessment test to aid in the assessment and management of concussion in individuals ages 18-24. The Sway System Sports Plus is indicated for use when there is a suspected concussion or head injury |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Substantial Equivalence: | The Intended Use of Sway Sports+ is the same as the predicate device, ImPACT QT, and the reference predicate device, ImPACT. The patient age range for Sway Sports+ is contained within that of the predicate and reference devices. Sway Sports+ and the predicate device function on mobile devices incorporating Apple iOS; Sway Sports+ also functions on Android OS devices.<br>Sway Sports+, the predicate device, and the reference device all enable the administration of neuropsychological tests to evaluate a test taker's cognitive state. All devices are similar in terms of technological characteristics, as they administer neurocognitive test modules using standard off-the-shelf computing devices and record objective cognitive performance measurements as the test taker responds to stimuli presented on the screen. Sway Sports+ also includes balance assessments; balance has been documented in the scientific literature as a potential indicator of the effects of concussion.<br>Sway Sports+ differs from the predicate and reference devices in the content and design of the cognitive tasks. However, both Sway Sports+ and the predicate device provide the same reporting on the mobile application, on a web portal, and in .pdf format. The reference device also has reporting available on a web portal and in .pdf format. Both summary and raw data are available as an export for all devices.<br>Sway Sports+ uses internal accelerometer and gyroscope data from mobile devices as well as touch input to record patient responses. The predicate device relies solely on touch input, and the reference device uses input from computer peripherals. This is not a significant difference, as all devices include a response signal to the underlying operating system as part of their algorithms. |
| Technology Comparison: | The Sway Sports+ employs the same technological characteristics as the predicate device and the reference device. |
{8}------------------------------------------------
| Table 1 - Substantial Equivalence Comparison | | | |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Sway Sports Plus<br>(Sports+) | Predicate Device<br>ImPACT Quick Test<br>(K170551) | Reference Device<br>ImPACT Version 4<br>(K202485) |
| Operating System | Apple iOS and Android<br>mobile devices | Apple iOS mobile<br>devices | Desktop/laptop<br>computers |
| Use Cases | Measures change over<br>time | Same | Same |
| Patient Population | 18-24 years old | 12-70 years old | 12-80 years old |
| How Provided | Software only | Same | Same |
| Reporting Features | Report viewable on the<br>Sway Mobile Application<br>and on the Sway Web<br>Portal and in .pdf<br>format. Summary and<br>Raw Data is available as<br>csv export from the web<br>portal data. | Report viewable in PDF<br>format. Summary and<br>Raw Data available in<br>csv, xml, and html file<br>formats. | Same |
| Technology | Stand-alone software<br>application | Same | Same |
| Stimulus Presentation | Mobile device screen | Same | Desktop or laptop screen |
| Stimulus Capture<br>(test taker response) | User input (motion<br>response) via mobile<br>device motion sensors<br>and via touch screen | User input (finger touch)<br>via mobile device touch<br>screen | User input via computer<br>peripherals. |
| Cognitive Tasks<br>Administered | Test battery consisting of<br>multiple test modules:<br>-Module 1: Reaction<br>Time<br>-Module 2: Impulse<br>Control<br>-Module 3: Inspection<br>Time<br>-Module 4: Memory Test,<br>including three letter<br>delayed recall and<br>working memory | Test battery consisting of<br>multiple test modules:<br>-Module 1: Symbol<br>Match<br>-Module 2: Three letter<br>memory and reverse<br>number counting<br>-Module 3: Attention<br>Tracker | Test battery consisting<br>of multiple test modules:<br>-Module 1: Word<br>Memory and Delayed<br>Memory Recognition<br>-Module 2: Design<br>Memory and Delayed<br>Design Recognition<br>-Module 3: X's and O's<br>-Module 4: Symbol<br>Matching<br>-Module 5: Color Match<br>Module 6: Three Letter<br>Memory |
| Additional Tests/Input | Balance Test; Symptom<br>Tracker; Patient<br>Demographics | Symptom Tracker;<br>Patient Demographics | Same as predicate<br>device |
| Results Interpretation | The Sway System does<br>not provide a<br>recommendation that<br>the patient is impaired<br>vs. unimpaired. Clinical<br>interpretation of the<br>results includes a<br>normative database for<br>Clinical comparison. | Same | Same |
| Psychometric<br>Properties | Demonstrates construct<br>validity with traditional<br>neuropsychological test<br>and test-retest reliability | Same | Same |
| Standards Used | ISO 14971<br>IEC 62304<br>IEC 62366 | ISO 14971<br>IEC 62304 | Same |
| Key Clinical Outcomes | Validity: r values range<br>from 0.22 to 0.42 when<br>compared to traditional<br>neuropsychological<br>testing; similar small to<br>moderate correlations<br>when compared to the<br>predicate device.<br><br>Reliability: ICC values<br>range from 0.68 to 0.92<br>for one-week test-<br>retests intervals | Validity: r values range<br>from 0.28 to 0.61 when<br>compared to traditional<br>neuropsychological<br>testing.<br><br>Reliability: r values<br>range from 0.18 to 0.83 | Validity: r values range<br>from 0.37 to 0.70 when<br>compared to traditional<br>neuropsychological<br>testing.<br><br>Reliability: ICC values<br>range from 0.50 to 0.88<br>for test-retest intervals<br>less than 30 days |
{9}------------------------------------------------
{10}------------------------------------------------
| Sterilization and<br>Shelf-life | Sway Sports+ is a software product. Sway Sports+ does not have a shelf-<br>life. Therefore, this section is not applicable. |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | Sway Sports+ is a software product and does not directly nor indirectly contact<br>the patient. Therefore, this section is not applicable. |
| Software Testing | Sway Sports+ was designed and developed according to Sway Medical, Inc.'s,<br>internal software development process in accordance with IEC 62304. Sway<br>Sports+ was tested using verification and validation methods, the results of<br>which indicate the Sway Sports+ complies with its specifications. |
| Electrical Safety | Sway Sports+ is a software product. Therefore, this section is not applicable. |
| Electromagnetic<br>Compatibility Testing | Sway Sports+ is a software product. Therefore, this section is not applicable. |
{11}------------------------------------------------
Sway Sports+ was developed and validated in accordance with IEC 62304, Medical Performance Devices – Software Life Cycle and the FDA Guidance documents "General Principles Testing — Bench of Software Validation" and "Applying Human Factors and Usability Engineering to Optimize Medical Device Design" that are appropriate for software development. Software verification and validation activities, including design phase reviews, risk management, traceability and software testing for performance, were all executed to demonstrate compliance with predetermined specifications.
Non-Clinical A summary of the Non-Clinical Performance Bench Testing is included in Table 2.
#### Table 2 - Summary of Non-Clinical Performance Testing Documents: Software Testing as part of Verification and Validation; Documents: Test PRJ-2303-02 Verification and Validation Report . Performed PRJ-2211-04 Usability Validation Protocol ● . PRJ-2303-03 Usability Validation Results Device The Sway Sports+ is a software product capable of running on Android v9.0 Description or later and Apple iOS v13 or later. The software was tested on both platforms. The Sway Sports+ was tested per Verification and Validation protocol and Test Method/ criteria to ensure product meets product requirements. The results of the Applicable testing are included in PRJ-22303-02 Verification and Validation Report Standards along with Usability Testing Acceptance The actual (observed) results conform to the specified expected results. Criteria Unexpected Results/ None Significant Deviations Results All tests pass with no exceptions. System meets requirements specifications.
Testing
{12}------------------------------------------------
## Performance Testing
– Clinical
This 510(k) submission includes a literature review of studies that demonstrate the validity, reliability, normative data, and clinical use of Sway Sports+. Data was collected to examine test-reliability in healthy controls, test-retest reliability in concussed patients, and the sensitivity & specificity of the product in the concussed population.
While Sway Sports+ is not intended to be used to classify a patient as having a concussion, it may be useful to create criteria based on the device's output to classify a patient as concussion positive or negative, to allow comparison to predicate devices to support a substantial equivalence determination. This information, along with additional data that demonstrate the device's test-reliability, construct validity, and reliability of outputs in patients with concussion, was used to support a determination of substantial equivalence.
## Summary of Clinical Testing:
## Normative Database
De-identified baseline test data was collected from over 126,000 users aged 18 to 24 years old. Subjects were selected based on the following inclusion criteria: Completed a baseline test in English, completed a baseline test in the United States of America, primary language was English, did not report a concussion within the last 6 months, did not report neurological conditions, did not report learning disability, had valid scores on all tests. Data were stratified across age and gender. Mann-Whitney U statistical tests were used to assess inter- and intra-sex and age differences in Sway balance and cognitive test module scores. Descriptive statistics (mean, standard deviation, median, and interquartile range [10th, 25th, 75th, and 90th]) were calculated for all Sway tests. Statistical significance was set a priori at a p-value of 0.05. Age/gender group samples were required to have a minimal size of 50 profiles to be considered for normative analysis. Results indicate an increase in performance in Sway Sports+ modules with age, with scores peaking in the early to mid-20's. Specific tests also show gender-based normative value differences. Thus, Sway Sports+ is designed to use both age and gender when relating individual scores to normative values.
## Validity
- VanRavenhorst-Bell et al., 20214: This work demonstrated the construct validity of Sway Sports+ Reaction Time, Impulse Control, and Inspection Time modules compared to ImPACT QT. The sample included 88 healthy adults aged 18 to 48 years (mean 22.09 ± 4.47 years). Participants completed the ImPACT QT battery and Sway Sports+ tests during the same session, and scores from Sway Sports+ were compared to ImPACT QT composite and sub-test scores. The Sway Reaction Time module was statistically significantly correlated (r = -0.46 to 0.22, p ≤ 0.05) with several ImPACT QT reaction time measures. The Sway Impulse Control and Inspection Time measures were also statistically significantly correlated (r = -0.25 to -0.46, p ≤ 0.05) with five ImPACT QT reaction time measures. Overall, these findings suggest that Sway Sports+ cognitive test battery is clinically comparable to the FDA-cleared concussion management system.
- Clark & VanRavenhorst-Bell, 2022-: This study investigated the construct validity ●
{13}------------------------------------------------
of Sway Sports+ in comparison to ImPACT and traditional neurocognitive assessments, including the Wechsler Adult Memory Scale-IV (WMS-IV), Wechsler Adult Intelligence Scale-IV (WAIS-IV), Delis-Kaplan Executive Function System (D-KEFS), and California Verbal Learning Test – 3rd Edition (CVLT-III). A total of 85 adults (mean age 23.1 years; 68% female) participated, with all tests administered by trained professionals in one session. Results indicated significant correlations between Sway Sports+ and both ImPACT and traditional tests, with significance at both p < 0.001 and p < 0.05. The correlation strength between traditional tests served as a benchmark for the Sway Sports+ assessments, showing that correlations varied from weak (0.20) to strong (0.56), particularly between the WAIS-IV Cancellation and WMS-IV Symbol Search tests. Notably, Sway Sports+ modules exhibited correlation values ranging from 0.22 to 0.42 with traditional tests, while also aligning with ImPACT measures. The Reaction Time module, characterized by rapid cognitive processing, did not correlate significantly with traditional tests, consistent with their differing formats. Moreover, Sway Sports+ demonstrated stronger correlation strengths with traditional neurocognitive tests compared to the correlations observed between ImPACT and these tests. Findings support the utility of Sway Sports+ for cognitive assessments, highlighting its robust associations with both traditional neurocognitive tests and the ImPACT system.
| Table 3 - Validity between Sway Sports+ and ImPACT QT & ImPACT | | | | |
|----------------------------------------------------------------|--------------------------------|--------------------------------|----------------------------|--------------------------------|
| | Sway<br>Reaction<br>Time | Sway<br>Impulse<br>Control | Sway<br>Inspection<br>Time | Sway<br>Memory |
| ImPACT QT<br>Three Letter Count Correct | 0.22 | 0.17 | 0.06 | |
| ImPACT QT<br>Three Letter Time First Click | -0.08 | -0.01 | -0.17 | |
| ImPACT QT<br>Rectangular Average Time | -0.44 | -0.46 | -0.19 | |
| ImPACT QT<br>Figure Eight Average Time | -0.46 | -0.36 | -0.29 | --- |
| ImPACT QT<br>Complex Average Time | -0.32 | -0.31 | -0.20 | |
| ImPACT QT<br>Symbol Match Correct Visible | -0.27 | -0.19 | 0.04 | |
| ImPACT QT<br>Symbol Match Correct Hidden | -0.35 | -0.06 | -0.05 | |
| ImPACT QT<br>Symbol Match Incorrect Hidden | -0.32 | -0.10 | -0.25 | |
| ImPACT<br>Visual Motor Composite Scale | $r_s = 0.26$<br>$p = 0.015 *$ | $r_s = 0.43$<br>$p < 0.001 *$ | | |
| IMPACT<br>Reaction Time Composite Scale | $r_s = -0.31$<br>$p = 0.005 *$ | $r_s = -0.41$<br>$p < 0.001 *$ | | |
| IMPACT<br>Memory Composite Verbal | | | --- | $r_s = 0.13$<br>$p = 0.251$ |
| IMPACT<br>Memory Composite Visual | | | | $r_s = 0.26$<br>$p = 0.015 *$ |
| IMPACT<br>Design Memory | | | | $r_s = 0.24$<br>$p = 0.026 *$ |
| IMPACT<br>Symbol Match | | | | $r_s = -0.22$<br>$p = 0.045 *$ |
| IMPACT<br>XO Memory | | | | $r_s = -0.22$<br>$p = 0.046 *$ |
{14}------------------------------------------------
| ImPACT<br>Three Letters | | $r_s = 0.22$<br>$p = 0.041$ | |
|----------------------------------------------------------------------------|-----------------------------|-------------------------------|------------------------------|
| Table 4 – Validity between Sway and Traditional Neuropsychological Testing | | | |
| | Sway<br>Reaction Time | Sway<br>Impulse<br>Control | Sway<br>Memory |
| WAIS-IV Coding | $r_s = 0.13$<br>$p = 0.241$ | $r_s = 0.02$<br>$p = 0.840$ | |
| WAIS-IV Symbol Search | $r_s = 0.07$<br>$p = 0.504$ | $r_s = 0.22$<br>$p = 0.043 *$ | |
| WAIS-IV Cancellation | $r_s = 0.10$<br>$p = 0.381$ | $r_s = 0.08$<br>$p = 0.501$ | --- |
| D-KEFS, Trail Making Condition 1 | $r_s = 0.09$<br>$p = 0.437$ | $r_s = 0.42$<br>$p < 0.001 *$ | |
| D-KEFS, Trail Making Condition 2 | $r_s = 0.06$<br>$p = 0.437$ | $r_s = 0.20$<br>$p = 0.064$ | |
| D-KEFS, Trail Making Condition 3 | $r_s = 0.15$<br>$p = 0.170$ | $r_s = 0.33$<br>$p = 0.003 *$ | |
| WMS-IV Symbol Span | | | $r_s = 0.00$<br>$p = 0.995$ |
| WAIS-IV Digit Span | | | $r_s = -0.07$<br>$p = 0.542$ |
| WAIS-IV Letter-Number Sequencing | --- | --- | $r_s = 0.00$<br>$p = 0.958$ |
| CVLT-III Short Delayed Free Recall | | | $r_s = 0.07$<br>$p = 0.551$ |
| CVLT-III Short Delayed Cued Recall | | | $r_s = 0.20$<br>$p = 0.070$ |
## Reliability
- Van Patten et al., 20213: This work demonstrated both the inter- and intra-. session reliability of Sway Sports+ Reaction Time, Impulse Control, Inspection Time, and Memory modules. Participants were assessed remotely over 3 subsequent weeks. The mean age of the sample was 26.69 years (± 9.89; range = 18-58). Of the 55 adults in the sample, 69% were women and 89% used an iPhone. ANOVA results indicated non-significant differences in scores across all three weeks for all measures. ICC values for weeks 1-2 ranged from 0.60 to 0.8, and ICC values for weeks 1-3 ranged from 0.68 to 0.88, indicating adequate or better test-retest reliability for all tested measures. The authors also established reliable change estimates for the included Sway Sports+ modules.
- Caccese et al., 20224: This work demonstrated inter- and intra-session reliability . for Sway Sports+ Balance module using the same sample and test procedure as Van Patten et al., 2021. Participants were assessed remotely over 3 subsequent weeks. The sample included 55 healthy adults (69% women) with a mean age of 26.69 years (SD = 9.89; range = 18-58). The ICC for Weeks 1 & 2 was 0.82, and the ICC for Weeks 1 & 3 was 0.88. ANOVA demonstrated no significant differences between test dates, demonstrating strong reliability between test and retest sessions. The authors also established reliable change estimates for the Sway Sports+ Balance module.
- An additional study was undertaken that evaluated the test-retest reliability of ● the Sway Sports+ battery in a sample of healthy adults aged 18-35. This
{15}------------------------------------------------
research included detailed method agreement analyses and established reliable change estimates relevant to the clinical interpretation of test score fluctuations. A total of 97 participants were initially recruited, with 56 completing all test and retest sessions, with a one-week interval. Converging evidence from paired sample comparisons, intraclass correlation coefficients (ICCs), Bland-Altman plots, and Deming regression analyses indicate no statistically significant differences in scores between test and retest sessions for any Sway Sports+ module. ICC values ranged between 0.82 and 0.92. This research demonstrates the reliability of the Sway Sports+ test battery and offers critical insights into the interpretation of score changes associated with external events, thus reinforcing the clinical utility of Sway Sports+ in assessing cognitive and functional performance.
- An additional study was undertaken that investigated the test-retest reliability of the Sway Sports+ test battery among concussed individuals, building on prior research that established reliability in healthy adults. Given the challenges in conducting an exact replication of baseline assessments due to variability in postconcussion recovery trajectories, a within-session design was employed. Participants completed multiple trials across different test modules during both baseline and post-concussion assessment sessions. The findings revealed that the Sway Sports+ battery maintains a robust level of reliability during postconcussion testing, comparable to that observed in baseline evaluations. This study utilized a dataset derived from NCAA member schools, ensuring that clinical standards for concussion diagnosis were strictly followed according to NCAA Concussion Safety Protocol guidelines. The results support the application of the Sway Sports+ system as a reliable tool for concussion assessment, contributing to enhanced understanding and management of concussion outcomes in sports.
| Table 5 - Test-Retest Reliability of the Sway Sports+ Test Battery | | |
|--------------------------------------------------------------------|----------------------------------------------------------------|-----------------------------|
| | ICC – 1 week interval | ICC – 1 week interval |
| | (published, Van Patten et al., 2021 &<br>Caccese et al., 2022) | (510(k) submission studies) |
| Sway Reaction Time | 0.83 | 0.88 |
| Sway Impulse Control | 0.68 | 0.88 |
| Sway Inspection Time | 0.75 | 0.82 |
| Sway Memory | 0.88 | 0.83 |
| Sway Balance | 0.82 | 0.92 |
## Sensitivity & Specificity
- Chikar & Curtiss, 20235: This study evaluated the effectiveness of Sway Sports+ . in classifying individuals with sports-related concussions, demonstrating comparable performance to the ImPACT tool. Presented at the Big Sky Sports Medicine Conference in Big Sky, MT (Jan 29 - Feb 2, 2023), the research utilized data from athletic organizations that conducted both ImPACT and Sway assessments on athletes with confirmed concussions between November 2020 and October 2022. Participants included 46 athletes (mean age 19.6, SD=1.4; 59% male) who completed pre-season baseline and post-injury testing with both tools. The mean interval between injury and post-injury Sway testing was 1.1
{16}------------------------------------------------
days (SD=1.4). While not all schools provided exact timings for ImPACT postinjury tests, they were administered within a week post-injury. Results indicated that Sway testing achieved a sensitivity of 69.6% (32/46) for predicting concussion group membership, outperforming ImPACT, which showed a sensitivity of 58.7% (27/46) at an 80% confidence interval. A McNemar's test revealed no significant difference between these sensitivities (p=0.383). Notably, the sensitivity and specificity of Sway Sports+ tests are strong overall, although they diminish when analyzing individual modules. This research emphasizes the necessity for multi-modal assessments in concussion evaluation and treatment, consistent with the Sway Sports+ test battery.
- An additional study was undertaken that evaluated the effectiveness of the Sway Sports+ test battery in distinguishing between concussed and nonconcussed individuals. The sensitivity and specificity of the Sway Sports+ system were assessed based on score changes from initial baseline to subsequent evaluations. Results indicate high sensitivity but moderate specificity, with high negative predictive values (NPVs) and low positive predictive values (PPVs). While the Sway Sports+ system is effective in detecting concussion effects, the moderate specificity suggests a tendency to over-identify concussed individuals, leading to potential false positives. Nevertheless, in a clinical context where Sway is utilized as part of a comprehensive evaluation by healthcare professionals, such false positives can be eliminated. Importantly, the system demonstrates a very low rate of false negatives, thereby mitigating the risk of untreated concussion cases. The findings underscore the utility of Sway Sports+ in concussion management protocols while highlighting areas for further refinement in specificity.
- . An additional pilot study was conducted to compare the effectiveness of ImPACT and Sway in the evaluation and management of sports-related concussions in student-athletes. The research involved 24 high school and college-aged participants who underwent baseline and post-injury testing with both tools. This study documented 47 synchronous testing sessions postconcussion (i.e., ImPACT and Sway testing on the same day) and reported concordance rates of 59.6% between the two tools, with discordance rates of 40.4%. Concordance was seen in both flagged and non-flagged scores. ImPACT identified more abnormal results, suggesting greater sensitivity, while SWAY provided critical measures that were missing in ImPACT. On average, recovery to baseline scores occurred in 12.7 days with ImPACT and 9.9 days with SWAY. While the findings support the complementary use of both tools, the limitations include a small sample size and differences in the functional domains assessed. This study underscores the importance of combining objective digital tools to enhance concussion evaluation and management in clinical and athletic settings.
Conclusion: Sway Sports+ falls within the type of device regulated under 21 CFR §882.1471, Computerized Cognitive Assessment Aid for Concussion (product code POM). The differences between Sway Sports+, the predicate device ImPACT QT, and the reference device ImPACT, described above, do not affect the safety or effectiveness of Sway Sports+ for its intended use. The safety and effectiveness of Sway Sports+ have been demonstrated through software verification and validation, non-clinical assessments, and clinical performance testing. Therefore, Sway Sports+ is substantially equivalent to the predicate device.
{17}------------------------------------------------
References:
- 1. Vanravenhorst-Bell, H. A., Muzeau, M. A., Luinstra, L., Goering, J. & Amick, R. Z. Accuracy of the SWAY Mobile Cognitive Assessment Application. Int J Sports Phys Ther 16, 991 (2021).
- 2. Clark, C. B. & Vanravenhorst-Bell, H. A. Construct Validity of Sway Mobile Application Cognitive Testing. (2022).
- 3. Van Patten, R., Iverson, G. L., Muzeau, M. A. & VanRavenhorst-Bell, H. A. Test-Retisbility and Reliable Change Estimates for Four Mobile Cognitive Tests Administered Virtually in Community-Dwelling Adults. Front Psychol 12, (2021).
- 4. Caccese, J. B. et al. Test-Retest reliability and preliminary reliable change estimates for Sway Balance tests administered remotely in community-dwelling adults. Front Digit Health 4, 217 (2022).
- ഗ് Chikar, J. & Curtiss, C. The Use of the Sway Medical Mobile Application in Concussion Management. in Big Sky Athletic Training & Sports Medicine Conference (2023).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.