← Product Code [LFL](/productcode/LFL) · K241700

# Tenex 2nd Generation System (K241700)

_Trice Medical, Inc. · LFL · Nov 18, 2024 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K241700

## Device Facts

- **Applicant:** Trice Medical, Inc.
- **Product Code:** [LFL](/productcode/LFL.md)
- **Decision Date:** Nov 18, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

The Tenex® 2nd Generation System is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of both soft and hard (e.g.: bone) tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery, and Plastic and Reconstructive Surgery. The Tenex® 2nd Generation System is also indicated for use in the debridement of wounds, such as, but not limited to diabetic ulcers, in application, in which, in the physician's judgement would require the use of an ultrasonic aspirator with sharp debridement.

## Device Story

Ultrasonic surgical instrument for fragmentation, emulsification, and aspiration of soft/hard tissue and wound debridement. System comprises console with touchscreen, ultrasonic MicroTips, and foot switch. Console generates ultrasonic energy (26.5 kHz) via piezoelectric transducers in MicroTips; manages irrigation fluid delivery and vacuum aspiration. Used in office, clinic, or hospital settings by physicians. User controls cutting levels, irrigation flow, and aspiration vacuum via console interface and foot switch. Emulsified tissue/fluid aspirated into integrated collection bag in single-use MicroTip. Provides precise tissue removal; benefits include improved occlusion prevention and enhanced surgical control compared to previous generation. Output visualized on console display; assists surgeon in real-time tissue management.

## Clinical Evidence

Bench testing only. No clinical data. Performance validated via electro-mechanical bench testing (aspiration, irrigation, console/MicroTip function), EMC/electrical safety (IEC 60601-1, IEC 60601-1-2), software V&V (IEC 62304), biocompatibility (ISO 10993-1), sterilization (ISO 11137), and packaging/shelf-life (ISO 11607). Simulated use testing in bovine tissue and cadaveric models validated design, durability, and usability (IEC 60601-1-6, IEC 62366-1).

## Technological Characteristics

Ultrasonic surgical instrument; 26.5 kHz piezoelectric vibration. Console-driven; 100-240V power. MicroTips: single-use, gamma-sterilized, stainless steel horn/needle, biomedical grade PVC/silicone tubing. Irrigation: forced delivery. Aspiration: vacuum pump. Connectivity: USB for data/display. Software: Windows-based application on tablet, DSP microcontroller firmware. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 11137, ISO 11607.

## Predicate Devices

- Tenex Health TX System with the TXP MicroTip ([K181367](/device/K181367.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 18, 2024

Trice Medical, Inc. David Vancelette VP of Quality Assurance & Regulatory Affairs 40 General Warren Blvd. Suite 100 Malvern, Pennsylvania 19355

Re: K241700

Trade/Device Name: Tenex 2nd Generation System Regulatory Class: Unclassified Product Code: LFL Dated: June 11, 2024 Received: June 13, 2024

Dear David Vancelette:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Long H. Chen-S Digitally signed by Long H.Chen-S
Date: 2024.11.18 15:01:04 -05'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K241700

Device Name Tenex® 2nd Generation System

#### Indications for Use (Describe)

The Tenex® 2nd Generation System is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of both soft and hard (e.g.: bone) tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery, and Plastic and Reconstructive Surgery.

The Tenex® 2nd Generation System is also indicated for use in the debridement of wounds, such as, but not limited to diabetic ulcers, in application, in which, in the physician's judgement would require the use of an ultrasonic aspirator with sharp debridement.

| Type of Use (Select one or both, as applicable)                                                |                                                                                               |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K241700

# 510(k) Summary

| Date:                      | November 7, 2024                                                                                                                      |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Owner:                     | Trice Medical, Inc.<br>40 General Warren Blvd.<br>Suite 100<br>Malvern, PA 19355 USA<br>Phone: (610) 989-8080<br>www.tricemedical.com |
| Contact Person:            | David Vancelette<br>Vice President of Quality Assurance & Regulatory Affairs<br>P: 949-238-8220<br>E: dvancelette@tricemedical.com    |
| Type of 510(k) Submission: | Traditional                                                                                                                           |
| Trade Name:                | Tenex® 2nd Generation System                                                                                                          |
| Common Name:               | Ultrasonic Surgical Instrument                                                                                                        |
| Classification Name:       | Instrument, Ultrasonic Surgical                                                                                                       |
| Device Class:              | Unclassified                                                                                                                          |
| Panel:                     | General & Plastic Surgery                                                                                                             |
| Product Code:              | LFL                                                                                                                                   |
| Predicate Devices:         | K181367, Tenex Health TX System with the TXP MicroTip, Product<br>Code: LFL                                                           |

### Device Description:

The Tenex 2nd generation system is an ultrasonic surgical instrument that fragments, emulsifies and aspirates soft and hard tissue. It also provides sharp debridement of soft and hard tissue in wounds such as neuropathic diabetic ulcers. The Tenex 2nd generation system is intended for use in an office-based setting, clinical or hospital environment. The system consists of a Console, ultrasonic MicroTips, a Foot Switch and a number of accessories.

The Console includes a touch screen tablet with a graphical user interface, motors to drive diaphragm pumps in the MicroTip, ultrasonic driver electronics and Software. The Console generates the ultrasonic energy required to operate the MicroTips and manages irrigation fluids. The Console is non-sterile when used and does not require processing.

The MicroTips connect to the Console for electrical power and to facilitate the transfer of irrigation fluid from the saline bag to the surgical site, and then aspirate the emulsified tissue and fluid to a collection bag incorporated into single use, disposable the MicroTip assembly. Motors in the Console couple to the diaphragm pumps in the MicroTips to control the fluid flows. The MicroTips are constructed from

{5}------------------------------------------------

various biocompatible polymers and metals. The MicroTip tubing is made of biomedical grade PVC and silicone.

The MicroTips are single use and sterile packaged. They are sterilized via gamma irradiation. The MicroTips are the ultrasonic surgical hand pieces that house the transducers, horns and needles with beveled tips.

The Foot Switch is used by the user to remotely control the Console. It connects to the back of the Console via an electrical cable. It provides on/off functionality for cutting, irrigation, and aspiration. The Foot Switch is IPX8 rated for protection against liquids.

The System includes additional components. a Packaging Hard Case is provided for shipping, handling. and storage of the Console, and a power cord is provided for connecting the Console to facility electrical power to operate the device. These items are non-sterile when used and do not require processing.

Associated accessories include: a USB Flash Drive for saving procedure information, USBC-HDMI cable for connecting to an external display, and a USB-Smartphone adaptor for connecting USB output to a smartphone.

## Intended Use / Indications for Use:

The Tenex® 2nd Generation System is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of both soft and hard (e.g.: bone) tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery, and Plastic and Reconstructive Surgery.

The Tenex® 2nd Generation System is also indicated for use in the debridement of wounds, such as, but not limited to diabetic ulcers, in application, in which, in the physician's judgement would require the use of an ultrasonic aspirator with sharp debridement.

### Technical Characteristics (Compared to Predicate):

The subject and predicate devices are ultrasonic surgical instruments that use electrical energy in the ultrasonic frequency range to drive an applicator needle for the fragmentation and emulsification of diseased tissue. The systems use irrigation with the needle at the treatment site to emulsify the fragmented tissue. The systems also use aspiration to remove the fragmented and emulsified material from the treatment site.

| Characteristic                   | Subject Device<br>Tenex® 2nd Generation<br>System | Predicate Device<br>Tenex Health TX System<br>with TXP MicroTip<br>(K181367) |
|----------------------------------|---------------------------------------------------|------------------------------------------------------------------------------|
| Energy source                    | 100-240V 50/60Hz                                  | Same                                                                         |
| Fragmentation/<br>Emulsification | Ultrasonic energy                                 | Same                                                                         |
| Vibration system                 | Piezoelectric                                     | Same                                                                         |
| Frequency                        | 26.5 ±1.5kHz                                      | Same                                                                         |
| Aspiration                       | Vacuum                                            | Same                                                                         |

The following technological characteristics are identical for the subject and predicate device:

{6}------------------------------------------------

K241700

| Characteristic                | Subject Device<br>Tenex® 2nd Generation<br>System | Predicate Device<br>Tenex Health TX System<br>with TXP MicroTip<br>(K181367) |
|-------------------------------|---------------------------------------------------|------------------------------------------------------------------------------|
| Irrigation                    | Forced                                            | Same                                                                         |
| System Control                | Foot switch                                       | Same                                                                         |
| MicroTip Horn/Needle          | One-piece, stainless steel                        | Same                                                                         |
| Tip sheaths                   | Stainless steel sheath                            | Same                                                                         |
| Reusable System<br>Components | Console and Foot Switch                           | Same                                                                         |

The following technological characteristics of the subject device and predicate were compared and found to have differences which do not affect the mode of operation:

| Characteristic                            | Subject Device<br>Tenex® 2nd Generation<br>System                              | Predicate Device<br>Tenex Health TX<br>System (K181367)                | Comparison                                                                  |
|-------------------------------------------|--------------------------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Cutting levels                            | User selected: low,<br>medium and high                                         | Same                                                                   | Updated ultrasonic driver                                                   |
| Irrigation flow<br>control                | User controllable, forced<br>delivery of saline from<br>an external saline bag | Same                                                                   | Using a fluid pump instead<br>of a pressure cuff                            |
| Irrigation flow rates                     | In excess of 60 cc/min                                                         | In excess of 30 cc/min<br>with a full 1000cc saline<br>irrigation bag. | Enhanced to stabilize flow<br>rate                                          |
| Aspiration flow<br>control                | User controllable, suction<br>provided by a vacuum<br>pump                     | Same                                                                   | Enhanced the pump                                                           |
| Aspiration flow rates                     | User selectable, 3 levels<br>between 100-200 cc/min                            | User selectable, 3 levels<br>between 10-30 cc/min                      | Enhanced to improve<br>occlusion prevention and<br>removal                  |
| Aspiration potential<br>(vacuum pressure) | User selectable, 3 levels<br>between 0 - 520mmHg                               | User selectable, 3 levels<br>between 100 - 500<br>mmHg                 | Enhanced to improve time<br>to vacuum pressure and<br>occlusion prevention. |
| Aspiration Waste<br>Container             | 1000cc sterile collection<br>bag integrated into the<br>MicroTip               | 1000cc sterile collection<br>bag integrated into the<br>MicroTip       | Subject device incorporates<br>additional filtration.                       |
| MicroTip Case Nose                        | ABS over-molded<br>stainless-steel sheath                                      | Polycarbonate over-<br>molded stainless-steel<br>sheath                | Both have equivalent<br>biocompatible polymer<br>over-molding               |
| Tip diameters                             | Multiple MicroTips,<br>between 1.5 - 1.9mm<br>(.059 - .074 in)                 | Multiple MicroTips,<br>between 1.1 - 1.9mm<br>(.042 - .074 in)         | Improved tip strength and<br>durability.                                    |
| Tip lengths                               | Multiple MicroTips,<br>between 1.0 - 3.0 in                                    | Multiple MicroTips,<br>between 1.0 - 2.0 in                            | Improved depth of reach.                                                    |

{7}------------------------------------------------

K241700

| Characteristic                    | Subject Device<br>Tenex® 2nd Generation<br>System                                                                                            | Predicate Device<br>Tenex Health TX<br>System (K181367)                            | Comparison                                                                                    |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Single-Use, Sterile<br>Components | MicroTips                                                                                                                                    | MicroTips<br>Supply Kit                                                            | Convenience kit removed                                                                       |
| User Interface                    | Color touchscreen,<br>component of an internal<br>MS Tablet                                                                                  | Color touchscreen,<br>connected to the internal<br>computer.                       | Higher resolution, greater<br>processing power for more<br>display and interface<br>features. |
| Software/Firmware                 | Windows Operating<br>System running<br>Application Software on<br>Touch Screen Tablet,<br>interfaces with Firmware<br>on DSP Microcontroller | Linux Operating System<br>running Application<br>Software on ARM<br>Microprocessor | Improved programming,<br>display and interface<br>options with current<br>technology          |
| Ultrasonic driver                 | Digital Voltage and<br>Frequency Controlled                                                                                                  | Analog Frequency<br>Controlled                                                     | Enhanced control of power<br>and acoustics                                                    |

### Non-Clinical Performance Data:

The following non-clinical performance tests were conducted to evaluate differences noted above and demonstrate substantial equivalence to the predicate device:

- 1) Electro-mechanical bench testing verified that the system met requirements for:
	- Aspiration
	- Irrigation ●
	- Cassette attachment and detection ●
	- Console functions ●
	- Console life cycle
	- Environmental operating conditions
	- Hardware verification
	- MicroTip functions ●
	- . MicroTip strength
	- Transit conditions

2) EMC and electrical safety testing verified the system met the standards:

- IEC 60601-1:2005+A1:2012+A2:2020 / Edition 3.2 2020-08 Consolidated Version, Medical . electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014+A1:2021, Medical electrical equipment Part 1-2: General requirements . for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests.
- 3) Software Verification & Validation ...

{8}------------------------------------------------

- Software was developed, verified and validated per IEC 62304:2006/A1:2016, Medical device ● software - Software life cycle processes
- 4) Biocompatibility evaluation verified the system met the standard: ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- 5) Sterilization validation of the sterile packaged MicroTip showed the device met the standards:
	- . ISO 11137-1 Third edition 2018-01 [Including AMD1:2021], Sterilization of health care products - Radiation - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
	- ISO 11137-2 Third edition 2013-06 [Including AMD1:2022], Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose
- 6) Packaging & Shelf Life Validation of the sterile packaged MicroTip showed the device met the standards:
	- . ISO 11607-1 Second edition 2019-02 [Including AMD1:2023], Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
	- ISO 11607-2 Second edition 2019-02, Packaging for terminally sterilized medical devices Part ● 2: Validation requirements for forming, sealing and assembly processes
- 7) Simulated use testing in bovine tissue
	- Validated the system's performance and durability in tissue. ●
- 8) Simulated use testing in cadaver validated the system's design and usability.
	- . Usability Validation per IEC 60601-1-6 Edition 3.2 2020-07 Consolidated Version, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance -Collateral standard: Usability, and:
	- ANSI AAMI IEC 62366-1:2015+AMD1:2020 Medical devices Part 1: Application of usability . engineering to medical devices, including Amendment 1
	- . Cadaveric design validation of the device in a simulated surgical setting for its Indications for Use.

# Conclusions:

The Tenex® 2ªd Generation System met all specified performance requirements. The non-clinical data, including the simulated use testing results, demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device. Evaluation did not raise any new questions regarding safety and effectiveness. The Tenex® 2nd Generation System is determined to be substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K241700](https://fda-staging.innolitics.com/device/K241700)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
