The DUO sensors are intended for any dental practice that uses X-ray equipment for intraoral diagnostic purposes. DUO sensors can be used by trained dental professionals for patients receiving intraoral X-ray examinations and procedures for capturing digital X-ray images. Captured digital X-ray images can be used for examinations and diagnostic purposes with the help of optional image analysis software. The optional image analysis software is not part of this submission. DUO sensors can be used with dental positioning devices and holders to assist with aligning an X-ray source beam with the sensor and anatomy.
Device Story
USB-driven intraoral X-ray sensor; captures dental radiographic images. Scintillating material converts X-rays to visible light; CMOS imager converts light to digital data. Used in dental practices by trained professionals. Device requires connection to PC via USB 2.0 and separate, non-included image analysis software. Sensor assembly supports automatic X-ray detection once armed via software command. Output is digital image displayed on PC screen for diagnostic aid. Benefits include immediate visualization of anatomical structures and pathologies to assist clinical decision-making.
Clinical Evidence
Clinical performance testing involved evaluation of images captured by both the DUO sensor and the predicate device (Brasseler GEM). Images were reviewed by US dentists to confirm the system functions as intended and performs similarly or better than the predicate.
Technological Characteristics
CMOS-based indirect conversion X-ray detector; Cesium Iodide scintillator. USB 2.0 interface. IPx8-equivalent housing (SABIC resin). Biocompatible per ISO 10993. Not sterile. Standards: ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 62304.
Indications for Use
Indicated for acquisition of intraoral radiographic images in dental patients. Operated by trained healthcare professionals. May be used with positioning devices or hand-positioned with patient assistance.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
July 5, 2024
Zinnovi Inc. % Abdel Halim President & MRP Global Quality and Regulatory Services 10 Scenic Way MONROE, NJ 08831
#### Re: K241649
Trade/Device Name: DUO1 and DUO2 Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: June 7, 2024 Received: June 7, 2024
### Dear Abdel Halim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241649
Device Name DU01 and DUO2
#### Indications for Use (Describe)
The DUO Sensors are USB driven intraoral sensors which are intended to acquire intraoral radiographic images. The DUO Sensors shall be operated by trained healthcare professionals, who are educated and competent to perform the acquisition of intraoral radiographs.
The DUO Sensors can be used either in combination with special positioning devices to facilitate positioning and alignment with the X-ray beam or they may also be positioned by hand with the assistance of the patient.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K241649
# 510(k) Summary
Date Prepared: June 14th, 2024
# Manufacturer and 510(k) Owner:
Name : Zinnovi, Inc. Address : 135 W. Central Rd, Suite 204, Schaumburg, IL 60195 Telephone: +1 (224) 801-5902 Official Contact: Swaroop Patel (CEO and Co-founder) Email : spatel@zinnovi.com
#### Representative/Consultant:
Name : Erica Livingston, RAC Company : Qualomics LLC Address : 5938 Priestly Dr, Suite 103, Carlsbad, CA 92008 Telephone: +1 (858) 247-6804 Email: erica@qualomics.com
# Identification of New Device
| Owner/Operator: | Zinnovi Inc. |
|--------------------------------|-------------------------------|
| Premarket Notification Number: | K241649 |
| Trade Name: | DUO1 and DUO2 |
| Common Name: | Dental digital X-ray sensor |
| Classification Name: | Extraoral Source X-ray System |
| Product Code: | MUH |
| Submission Type: | 510(k) |
| Classification: | Class II |
| Panel/Medical Specialty: | Radiology/Dental |
| Regulation Number: | 21 CFR 872.1800 |
| Manufacturer: | Zinnovi Inc. |
# Identification of Predicate Device
| Owner/Operator: | RealCloud Imaging Inc. doing business as RealCloud Imaging |
|-----------------------------------------------|------------------------------------------------------------|
| 510(k) Premarket Notification Number: K221955 | |
| Trade Name: | Brasseler GEM |
{4}------------------------------------------------
Common Name: Dental digital X-ray sensor Classification Name: Extraoral Source X-ray system Product Code: MUH Classification: Class II Panel/Medical Specialty: Radiology/Dental Regulation Number: 21 CFR 872.1800 Manufacturer: RealCloud Imaging Inc. doing business as RealCloud Imaging
# Device Description
The DUO sensors are USB-driven digital X-ray sensors designed for health care professionals who are already acquainted with the standard procedures for acquiring dental intraoral radiographs. Digital X-ray imaging is an aide for diagnosis and should always be confirmed by the doctor using appropriate additional diagnostic aides, professional judgment, and experience.
The DUO Sensors are indirect converting X-ray detectors. A scintillating material converts the incident Xrays into visible light, this light is coupled optically to a CMOS technology light detection imager, and then converted to digital data.
The design of the sensor assembly supports the automatic detection of incident X-rays to generate digital images for intraoral applications, once armed via a software command. The digital image created is immediately visible on the screen of a personal computer connected to the DUO sensor through the standard USB port. For DUO sensors to be used in a dental practice, an optional image analysis software will be necessary. Image analysis software is not part of the submission. DUO captured X-ray images are suitable for recognition of normal anatomical structures, dental pathologies, and abnormal conditions.
The DUO sensors support USB2.0 connectivity to computers using a dedicated electronic assembly and a sensor software driver. Functions of the DUO sensors are controlled by software drivers and utilities support sensor activation and settings.
The DUO sensors are manufactured with the same device firmware as the predicate device, Brasseler GEM.
# Indications for Use
The DUO Sensors are USB driven intraoral sensors which are intended to acquire intraoral radiographic images. The DUO Sensors shall be operated by trained healthcare professionals, who are educated and competent to perform the acquisition of intraoral radiographs.
The DUO Sensors can be used either in combination with special positioning devices to facilitate positioning and alignment with the X-ray beam or they may also be positioned by hand with the assistance of the patient.
{5}------------------------------------------------
#### Intended Use
The DUO sensors are intended for any dental practice that uses X-ray equipment for intraoral diagnostic purposes. DUO sensors can be used by trained dental professionals for patients receiving intraoral X-ray examinations and procedures for capturing digital X-ray images. Captured digital X-ray images can be used for examinations and diagnostic purposes with the help of optional image analysis software. The optional image analysis software is not part of this submission. DUO sensors can be used with dental positioning devices and holders to assist with aligning an X-ray source beam with the sensor and anatomy.
# Comparison Table
The following comparison table compares Zinnovi DUO to the predicate device, Brasseler GEM, with respect to intended use, indications of use, environment of use, limitations of use, technical performance, and technological characteristics, and provides more detailed information regarding the basis for the determination of substantial equivalence.
| Descriptive<br>Information | Zinnovi DUO | Brasseler GEM | VERDICT |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) Number | K241649 | K221955 | NA |
| Manufacturer | Zinnovi Inc. | RealCloud Imaging Inc. doing<br>business as RealCloud<br>Imaging | NA |
| Trade Name | DUO1<br>DUO2 | Brasseler GEM10<br>Brasseler GEM15<br>Brasseler GEM20 | NA |
| Classification Name | Digital X-ray Sensor | Digital X-ray Sensor | Same |
| Classification | Class II | Class II | Same |
| Classification Panel/<br>Medical Specialty | Radiology/Dental | Radiology/Dental | Same |
| Regulation Number | 21 C.F.R. §872.1800 | 21 C.F.R. §872.1800 | Same |
| Product Code | MUH | MUH | Same |
| Number of Sensors | 2 | 3 | Different |
| Sensor Exterior Sizes | 36.36 mm x 24.53.mm | 36.36 mm x 24.53.mm | Same |
| | | 38.83 mm x 29.63 mm | NA |
| | 41.80 mm x 30.48 mm | 41.80 mm x 30.48 mm | Same |
| Sensor Imaging Sizes | 30.26 mm x 20.32 mm | 30.26 mm x 20.32 mm | Same |
| | | 33.15 mm x 26.25 mm | NA |
| | 36.08 mm x 26.25 mm | 36.08 mm x 26.25 mm | Same |
| | All with four clipped corners | All with four clipped corners | Same |
| Overall Imaging Areas | 615 mm2 | 615 mm2 | Same |
| | | 870 mm2 | NA |
| | 947.1mm2 | 947.1mm2 | Same |
| | | | |
| Pixel Size | 19.5 μm | 19.5 μm | Same |
| Image Resolution | 1539 x 1026 pixels<br>(1.70 M pixels) | 1539 x 1026 pixels<br>(1.70 M pixels) | Same |
| | | 1692 x 1324 pixels<br>(2.2 M Pixels) | NA |
| | 1842 x 1324 pixels<br>(2.40 M pixels) | 1842 x 1324 pixels<br>(2.40 M pixels) | Same |
| | | | |
| X-Ray Resolution | 20 visible lp/mm | 20+ visible lp/mm | Different |
| Dynamic Range | 16,384:1 | 16,384:1 | Same |
| Technology | CMOS | CMOS | Same |
| Scintillator Technology | Cesium lodide | Cesium lodide | Same |
| Operating System | Microsoft Windows 10 and<br>Windows 11 | Microsoft Windows 7 and 10 | Different |
| Extraoral source | X-ray system | X-ray system | Same |
| Interface to PC | USB 2.0, Type A | USB 2.0, Type A | Same |
| Cable Length | 0.6m and 1.9m | 1.9 m and 2.9 m | Different |
| Power Consumption | 0.8 Watts Max | 0.8 Watts Max | Same |
| Electrical Rating | DC 5V, 350 mA max | DC 5V, 350 mA max | Same |
| Sterilization | Not suitable for sterilization | Not suitable for sterilization | Same |
| Housing | IPx8 Equivalent ISO 10993-1<br>Biocompatible | IPx8 Equivalent ISO 10993-1<br>Biocompatible | Same |
| Indications for use | The DUO Sensors are USB<br>driven digital intraoral sensors<br>which are intended to acquire<br>intraoral radiographic images.<br>The DUO Sensors shall be<br>operated by trained<br>healthcare professionals, who<br>are educated and competent<br>to perform the acquisition of<br>intraoral radiographs.<br>The DUO Sensors can be used<br>either in combination with<br>special positioning devices to<br>facilitate positioning and<br>alignment with the X-ray beam<br>or they may also be positioned<br>by hand with the assistance of<br>the patient. | Brasseler GEM is a USB<br>driven digital intraoral X-ray<br>sensor which is intended to<br>acquire dental radiographic<br>images.<br>Brasseler GEM must be<br>operated by healthcare<br>professionals who have<br>been trained and are<br>competent using various<br>methods of acquiring<br>radiographic images of<br>dental anatomy. Brasseler<br>GEM can be used with<br>dental positioning devices<br>and holders to assist with<br>aligning an X-ray source<br>beam with the sensor and<br>anatomy. Brasseler GEM can<br>also be aligned by hand with<br>assistance of the patient. | Same |
| Intended Use | The DUO sensors are intended<br>for any dental practice that<br>uses X-ray equipment for<br>intraoral diagnostic purposes.<br>DUO sensors can be used by<br>trained dental professionals<br>for patients receiving intraoral<br>X-ray examinations and<br>procedures for capturing<br>digital X-Ray images. Captured<br>digital X-ray images can be<br>used for examinations and<br>diagnostic purposes with the | Brasseler GEM is intended<br>for any dental practice that<br>uses X-ray equipment for<br>intraoral diagnostic<br>purposes. Brasseler GEM<br>can be used by trained<br>dental professionals for<br>patients receiving intraoral<br>X-ray examinations and<br>produces digital images for<br>patients receiving intraoral<br>X-ray examinations for<br>diagnostic purposes. An<br>image analysis software is | Same |
| | help of optional image analysis<br>software. | not part of this submission. | |
| | The optional image analysis<br>software is not part of this | When Brasseler GEM is to<br>be used in a dental practice, | |
| | submission. DUO sensors can<br>be used with dental | an optional software will be<br>necessary. Brasseler GEM is | |
| | positioning devices and<br>holders to assist with aligning | a USB-driven digital intraoral<br>X-ray sensor which is | |
| | an X-ray source beam with the<br>sensor and anatomy. | intended to acquire dental<br>radiographic images. | |
| | | Brasseler GEM must be | |
| | | operated by healthcare | |
| | | professionals who have | |
| | | been trained and are | |
| | | competent using various | |
| | | methods of acquiring | |
| | | radiographic images of | |
| | | dental anatomy. Brasseler | |
| | | GEM can be used with | |
| | | dental positioning devices | |
| | | and holders to assist with | |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.