← Product Code [NAY](/productcode/NAY) · K241638

# 8mm SureForm 30 Curved-Tip Stapler; 8mm SureForm 30 Stapler; 8mm SureForm 30 Reloads (K241638)

_Intuitive Surgical, Inc. · NAY · Sep 10, 2024 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K241638

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [NAY](/productcode/NAY.md)
- **Decision Date:** Sep 10, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Intuitive Surgical 8mm SureForm 30 staplers and reloads and accessories are intended to be used with a compatible da Vinci Surgical system for resection of vasculature and tissue and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.

## Device Story

The 8mm SureForm 30 Staplers are fully wristed, articulating surgical staplers used exclusively with a compatible da Vinci Surgical System. Controlled by a surgeon at the Surgeon Console, the device performs resection, transection, and anastomosis by placing four staggered rows of titanium alloy (Ti3Al2.5V) staples into target tissue, followed by mechanical cutting along the staple line. The system provides the surgeon with precise control over stapling and cutting during minimally invasive procedures. The device benefits patients by enabling secure tissue closure and transection in complex surgical environments. Modifications include a new tungsten material supplier, redesigned driver for improved robustness, and repurposed installation/removal tools (formerly retainer/bottom cover).

## Clinical Evidence

Bench testing only. Evidence includes design verification (mechanical properties, force limits, reliability), transit verification per ASTM D4169-16, biocompatibility per ISO 10993-1:2018, design validation via simulated clinical use, and a summative human factors usability study. No clinical trial data was required.

## Technological Characteristics

Articulating, wristed surgical stapler for da Vinci systems. Materials include titanium alloy (Ti3Al2.5V) staples and tungsten. Complies with ISO 10993-1:2018 for biocompatibility and ASTM D4169-16 for transit testing. Features include four staggered rows of staples with varying leg lengths (2.0mm, 2.5mm, 3.5mm).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- 8mm SureForm 30 Curved-Tip Stapler ([K211997](/device/K211997.md) & [K222839](/device/K222839.md))
- 8mm SureForm 30 Stapler ([K211997](/device/K211997.md) & [K222839](/device/K222839.md))
- 8mm SureForm 30 Reloads ([K211997](/device/K211997.md) & [K222839](/device/K222839.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 10, 2024

Intuitive Surgical, Inc. Kristy Ann Darling Regulatory Affairs Lead 1266 Kifer Road Sunnyvale, California 94086

Re: K241638

Trade/Device Name: 8mm SureForm 30 Curved-Tip Stapler; 8mm SureForm 30 Stapler; 8mm SureForm 30 Reloads Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NAY, GDW, GAG Dated: June 7, 2024 Received: August 13, 2024

Dear Kristy Ann Darling:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Tek Tek N. N. Lamichhane -S Lamichhane -S Date: 2024.09.10 17:11:03 -04'00' Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control

and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{3}------------------------------------------------

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known) K241638

Device Name 8mm SureForm 30 Curved-Tip Stapler 8mm SureForm 30 Stapler 8mm SureForm 30 Reloads

### Indications for Use (Describe)

The Intuitive Surgical 8mm SureForm 30 staplers and reloads and accessories are intended to be used with a compatible da Vinci Surgical system for resection of vasculature and tissue and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D)  |  |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C)   |  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

# 510(k) Summary (21 CFR § 807.92)

### I. Submitter Information

| 510(k) Owner:          | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086                                                  |
|------------------------|---------------------------------------------------------------------------------------------------------------------|
| Contact Person:        | Kristy Ann Darling, RAC<br>Regulatory Affairs Lead<br>kristyann.darling@intusurg.com<br>+1 (408) 528-2644           |
| Date Summary Prepared: | September 05, 2024                                                                                                  |
| II.<br>Subject Device  |                                                                                                                     |
| Trade Name:            | 8mm SureForm 30 Curved-Tip Stapler<br>8mm SureForm 30 Stapler<br>8mm SureForm 30 Reloads                            |
| Common Name:           | Stapler, surgical<br>Stapler, implantable<br>System, surgical, computer controlled instrument                       |
| Classification:        | II                                                                                                                  |
| Regulation:            | 21 CFR 876.1500 Endoscope and Accessories<br>21 CFR 878.4740 Surgical Stapler<br>21 CFR 878.4750 Implantable Staple |
| Product Code:          | NAY, Endoscope and Accessories<br>GAG, Stapler, Surgical<br>GDW, Implantable Staple                                 |

### III. Predicate Device Information

| Predicate Devices: | 8mm SureForm 30 Curved-Tip Stapler, K211997 & K222839<br>8mm SureForm 30 Stapler, K211997 & K222839<br>8mm SureForm 30 Reloads, K211997 & K222839 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|

Image /page/5/Picture/7 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the "I" in the word. The font is sans-serif and the letters are evenly spaced.

{6}------------------------------------------------

### IV. Device Description

The 8mm SureForm 30 Staplers are fully wristed, articulating, surgical staplers which are designed for use exclusively with a compatible da Vinci Surgical System (herein referred to as "system"). The staplers are controlled by the surgeon using the Surgeon Console of the system. They are intended for resection, transection and/or creation of anastomoses in surgery and achieves its intended use by placing multiple staggered rows of implantable staples in the target tissue (stapling) followed by cutting of the target tissue along the middle of the staple line (transection).

The 8mm SureForm 30 Reloads consist of a single-use cartridge that contains four staggered rows of implantable titanium alloy (Ti3Al2.5V) staples. The implantable staples are provided in various staple leg lengths (Gray, White, Blue) to accommodate various tissue types. Table 1 outlines the specifications of the reloads.

| Table 1: Reload Specifications |             |                |                                    |
|--------------------------------|-------------|----------------|------------------------------------|
| Reloads                        | No. of rows | No. of Staples | Unformed Staple Leg<br>Length (mm) |
| 8mm SureForm 30 Gray Reload    | 4           | 34             | 2.0                                |
| 8mm SureForm 30 White Reload   | 4           | 34             | 2.5                                |
| 8mm SureForm 30 Blue Reload    | 4           | 34             | 3.5                                |

### V. Indication for Use

The Intuitive Surgical 8mm SureForm 30 staplers and reloads and accessories are intended to be used with a compatible da Vinci Surgical system for resection of vasculature and tissue and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.

### VI. Technological Characteristics

The subject 8mm SureForm 30 Staplers and 8mm SureForm 30 Reloads are very similar to their predicate devices. They have the same intended use, same indication for use, same fundamental scientific technology, and similar technological characteristics as the predicate device. The modifications consist of a new alternate supplier for Tungsten material and corrective actions related to recall number Z-0340-2023. Results from performance data indicate that the 8mm SureForm 30 Staplers and 8mm SureForm 30 Reloads are substantially equivalent to their predicate devices.

Image /page/6/Picture/10 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "U". The font is sans-serif and the letters are evenly spaced. The image is simple and clean, with a white background.

{7}------------------------------------------------

| 8mm SureForm 30 Staplers       |                                                                                                                                                                        |                                                                                                           |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Attributes                     | Subject Device<br>8mm SureForm 30 Curved-Tip<br>Stapler, 8mm SureForm 30 Stapler                                                                                       | Predicate Device<br>8mm SureForm 30 Curved-Tip<br>Stapler, 8mm SureForm 30 Stapler<br>(K211997 & K222839) |
| Design                         | The subject staplers consist of<br>Distal Jaw Assembly, Middle Shaft<br>and Proximal End.<br>Distal Jaw Assembly - modified<br>driver design to improve<br>robustness. | SIMILAR to subject device.                                                                                |
| Patient Contacting<br>Material | The subject device consists of the<br>same patient contact material, with<br>the exception of the Tungsten<br>material sub-supplier                                    | SIMILAR to subject device                                                                                 |

| 8mm SureForm 30 Reloads                                                                                                                                                                      |                                                                                                                                                                                                                                                      |                                                                                             |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Attributes                                                                                                                                                                                   | Subject Device<br>8mm SureForm 30 Reloads                                                                                                                                                                                                            | Predicate Device<br>8mm SureForm 30 Reloads<br>(K211997 & K222839)                          |
| Retainer and Bottom<br>Cover                                                                                                                                                                 | The Retainer and Bottom Cover<br>were redesigned and repurposed<br>into Installation and Removal tools.                                                                                                                                              | The Retainer and Bottom Cover<br>protect the staples during shipping<br>and transportation. |
| Left: Image of<br>Installation Tool (shown<br>in orange) and Removal<br>Tool (shown in yellow)<br><br>Right: Image of Retainer<br>(shown in orange) and<br>Bottom Cover (shown in<br>yellow) | Image: Installation Tool and Retainer for Subject Device                                                                                                                                                                                             | Image: Installation Tool and Retainer for Predicate Device                                  |
| Reload Design                                                                                                                                                                                | The reload cartridge contains<br>pushers, multiple staggered rows of<br>implantable staples and a stainless-<br>steel component called the switch.<br><br>Modified switch to reduce friction<br>and improve alignment during<br>reload installation. | SIMILAR to subject device.                                                                  |
| Packaging                                                                                                                                                                                    | Modified primary packaging<br>dimensions to accommodate new<br>installation and removal tools.                                                                                                                                                       | SIMILAR to subject device.                                                                  |

# INTUITIVE

{8}------------------------------------------------

| 8mm SureForm 30 Reloads |                                                                                                                                                                              |                                                                    |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Attributes              | Subject Device<br>8mm SureForm 30 Reloads                                                                                                                                    | Predicate Device<br>8mm SureForm 30 Reloads<br>(K211997 & K222839) |
| Instruction For Use     | Added instructions on reload<br>installation and removal.                                                                                                                    | SIMILAR to subject device                                          |
|                         | Updated references to align with<br>redesigned and repurposed<br>components: Installation Tool<br>(previously a Retainer) and<br>Removal Tool (previously a Bottom<br>Cover) |                                                                    |
|                         | New and modified warnings                                                                                                                                                    |                                                                    |

### VII. Performance Data

The subject 8mm SureForm 30 Staplers and 8mm SureForm 30 Reloads underwent a series of tests to evaluate the impact of the modifications made to the predicate devices. Testing was performed with a compatible da Vinci surgical system. Testing included design verification, compatibility verification, transit verification, biocompatibility, design validation, and human factors. The successful completion of testing demonstrated that the subject 8mm SureForm 30 Staplers and 8mm SureForm 30 Reloads design outputs continue to meet design inputs.

## Design Verification

Bench testing was performed to verify functional design outputs met the functional design inputs. The design verification in this section addressed the following:

- Mechanical properties such as critical dimensions, mass, materials, and finish
- Mechanical requirements such as range of motion, friction and offset
- Force limits
- Reliability

## Transit Verification

Transit testing was performed in accordance with ASTM D4169-16, Standard Practice for Performance Testing of Shipping Containers and Systems.

## Biocompatibility

Biocompatibility testing was completed in accordance with the following standards and guidance documents:

- . FDA Guidance: Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process", issued September 2020
- . ISO 10993-1:2018 Biological evaluation of medical devices

{9}------------------------------------------------

## Design Validation

Simulated clinical use testing was performed to validate that the product specifications continue to meet the user's needs and intended use.

## Human Factors

A summative usability study was conducted to assess the use-related risks associated with the device user interface of the 8mm SureForm 30 Staplers and 8mm SureForm 30 Reloads.

### VIII. Conclusion

Based on the intended use, technological characteristics, and performance data, the subject 8mm SureForm 30 Staplers and 8mm SureForm 30 Reloads are substantially equivalent to the predicate devices.

Image /page/9/Picture/8 description: The image shows the word "INTUITIVE" in blue font. There is a small blue circle above the letter "U". The font is sans-serif and the letters are evenly spaced.

---

**Source:** [https://fda-staging.innolitics.com/device/K241638](https://fda-staging.innolitics.com/device/K241638)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
