Guangdong Xinyu Electronic and Technology Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Feb 3, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
Device Story
Arm Blood Pressure Monitor uses oscillometric method to measure systolic/diastolic blood pressure and pulse rate. Device consists of monitor unit (control circuit, air pump, deflation valve, LCD) and arm cuff. User wraps cuff around upper arm; device automatically inflates/deflates via piezoelectric pump. Semiconductor pressure sensor captures oscillations; internal processor calculates blood pressure and pulse rate. Results displayed on LCD; memory stores 2*60 measurements. Used by medical professionals or patients at home. Provides objective data for hypertension monitoring; aids clinical decision-making by tracking blood pressure trends.
Clinical Evidence
Clinical study conducted per ISO 81060-2:2018 with 120 subjects (55 female, 65 male; age 12-85). Primary endpoints: mean error and standard deviation for systolic/diastolic pressure. Results met requirements (mean error ≤ ±5mmHg, SD ≤ 8mmHg). No adverse effects reported.
Technological Characteristics
Oscillometric measurement; semiconductor pressure sensor; piezoelectric pump; automatic rapid deflation valve. Powered by 4xAA or 4xAAA batteries. LCD display. Complies with IEC 60601-1, IEC 60601-1-11, IEC 80601-2-30, IEC 60601-1-2. Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for adult patients to monitor and display systolic/diastolic blood pressure and pulse rate. Contraindicated for use in ambulatory environments and in aircraft.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 3, 2025
Guangdong Xinyu Electronic and Technology Co., Ltd. Rita Long Official Correspondent Number 42, CuiCheng Road, Cuiheng New District Zhongshan, Guangdong 528400 China
Re: K241590
Trade/Device Name: Arm Blood Pressure Monitor (XY-B02, XY-B03, XY-B05, XY-B07, XY-B08. XY-B09) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: January 6, 2025 Received: January 6, 2025
Dear Rita Long:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Robert T. Kazmierski -S for LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241590
Device Name
Arm Blood Pressure Monitor (XY-B01, XY-B02, XY-B05, XY-B07, XY-B07, XY-B08, XY-B09)
Indications for Use (Describe)
Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
#### 1. Summary Prepared Date
20 May 2024
#### 2. Submitter's Information Sponsor
- � Company Name: Guangdong Xinyu Electronic and Technology Co., Ltd.
- � Address: Number 42, CuiCheng Road, Cuiheng New District Zhongshan Guangdong 528400 China
- � Phone: 86-760-88818598
- Contact Person (including title): Ms. Rita Long (Regulatory Manager) �
- E-mail: zcy_cg@163.com �
### Application Correspondent:
- � Company Name: Guangdong Xinyu Electronic and Technology Co., Ltd.
- � Address: Number 42, CuiCheng Road, Cuiheng New District Zhongshan Guangdong 528400 China
- Phone: 86-760-88818598 �
- � Contact Person (including title): Ms. Rita Long (Regulatory Manager)
- � E-mail: zcy_cg@163.com
#### 3. Subject Device Information
- � Type of 510(k) submission: Traditional
- Common Name: Noninvasive blood pressure measurement system �
- Trade Name: Arm Blood Pressure Monitor �
- Model: XY-B01, XY-B02, XY-B03, XY-B05, XY-B07, XY-B08, XY-B09 �
- � Classification Name: System, Measurement, Blood-Pressure, Non-Invasive
- � Review Panel: Cardiovascular
- � Product Code: DXN
- � Regulation Number: 870.1130
- � Regulation Class: II
#### 4. Predicate Device Information
- � 510(k) number: K222926
- � Sponsor: Shenzhen Yuezhongxing Technology Co., Ltd.
- � Trade Name: Arm Blood Pressure Monitor
- � Classification Name: System, Measurement, Blood-Pressure, Non-Invasive
{5}------------------------------------------------
K241590 Summary Guangdong Xinyu Electronic and Technology Co., Ltd.
- Model: 111, B1681, B1682, B1683 �
- � Review Panel: Cardiovascular
- � Product Code: DXN
- Regulation Number: 21 CFR 870.1130 �
- � Requlation Class: II
#### 5. Device Description
The Arm Blood Pressure Monitor, including XY-B01, XY-B03, XY-B05, XY-B07, XY-B08, XY-B09, can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person with arm circumference of 220-320mm through oscillometric method. The initial inflation pressure of the cuff is zero pressure. When start the device, the cuff will be inflated and deflated.
The Arm Blood Pressure Monitor is composed of monitor unit and arm cuff. Of which, the main unit contains main control circuit board, air pump, deflation valve, LCD and shell. The seven models have the same intended use, working principle, measuring range, cuff, and conformance standard, except for appearance and power supply. The devices are powered by 4*AA batteries (XY-B02) or 4*AAA batteries (other models). The Arm Blood Pressure Monitor has a memory function that automatically stores 2*60 sets data of the latest measurements.
#### 6. Indications for Use
Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
#### 7. Comparison to Predicate Device
Compare to the predicate devices, the subject device has same intended use, similar product design, same performance, same safety as the predicate device, the summarized comparison information is listed in the following table:
{6}------------------------------------------------
Guangdong Xinyu Electronic and Technology Co., Ltd.
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Manufacturer | Guangdong Xinyu Electronic and<br>Technology Co., Ltd. | Shenzhen Yuezhongxing<br>Technology Co., Ltd. | - |
| 510 (k) Number | - | K222926 | - |
| Product name | Arm Blood Pressure Monitor | Arm Blood Pressure Monitor | - |
| Models | XY-B01, XY-B02, XY-B03, XY-B05,<br>XY-B07, XY-B08, XY-B09 | 111, B1681, B1682, B1683 | - |
| Product Code | DXN | DXN | Same |
| Regulation No. | 21 CFR 870.1130 | 21 CFR 870.1130 | Same |
| Class | II | II | Same |
| Indication<br>for use | Arm Blood Pressure Monitor is<br>intended for use by medical<br>professionals or at home to monitor<br>and display diastolic, systolic blood<br>pressure and pulse rate on adult each<br>time, with an air cuff buckled around<br>one's arm according to the instruction<br>in the user's guide manual. | Arm Blood Pressure Monitor is<br>intended for use by medical<br>professionals or at home to<br>monitor and display diastolic,<br>systolic blood pressure and<br>pulse rate on adult each time,<br>with an air cuff buckled around<br>one's arm according to the<br>instruction in the user's guide<br>manual. | Same |
| Environment of<br>use | Hospital, Home | Hospital, Home | Same |
| Patient<br>population | Adult | Adult | Same |
| Contraindication | The monitor is contraindicated for<br>use in ambulatory environment. The<br>monitor is contraindicated in aircraft. | The monitor is contraindicated<br>for use in ambulatory<br>environment. The monitor is<br>contraindicated in aircraft. | Same |
| Principle of<br>operation | Cuff oscillometric method | Cuff oscillometric method | Same |
| Measurement site | Upper Arm | Upper Arm | Same |
| Measurement range | Pressure: 0mmHg ~295mmHg(0kPa<br>~39.33kPa)<br>Pulse rate: 40 to 180 beats/min | Pressure:0 to 280 mmHg<br>Pulse rate: 40 to 199<br>beats/min | Different Note 1 |
| Pressure sensor | semiconductor pressure sensor | Semiconductor pressure<br>sensor | Same |
| Arm circumference | 22 cm to 32 cm | 22 cm to 32 cm | Same |
| Accuracy of<br>pressure | ±3mmHg | ±3mmHg | Same |
| Accuracy of<br>pulse rate | ±5% of reading | ±5% of reading | Same |
| Inflation method | Automatic inflation with piezoelectric<br>pump | Automatic inflation with<br>piezoelectric pump | Same |
| Deflation method | Automatic rapid deflation valve | Automatic rapid deflation valve | Same |
| Display | LCD | LCD | Same |
| Power source | 4*AA batteries (XY-B02)<br>4*AAA batteries (other models) | 4 x AAA batteries, or AC<br>adapter (input: 100-<br>240V~,50/60Hz,0.2A output:5V<br>DC, 1.0A,5.0W) | Different<br>Note 2 |
| Operation condition | +10° C ~ +40° C | 5 to 40°C | Different<br>Note 3 |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| | 15% ~ 85% RH<br>860hPa~1060hPa | 700 to 1060hPa | |
| Storage condition | -20 ~ +55 °C<br>10% ~ 93% RH<br>500hPa~1060hPa | -25 to 70°C<br>0 to 93%RH | Different<br>Note 3 |
| Transportation<br>condition | -20 ~ +55°C<br>10% ~ 93% RH<br>500hPa~1060hPa | -25 to 70°C<br>0 to 93%RH | Different<br>Note 3 |
| Irregular heart beat<br>feature | Yes | Yes | Same |
| Hypertension<br>indication | Yes | Yes | Same |
{7}------------------------------------------------
## K241590 Summary
### Guangdong Xinyu Electronic and Technology Co., Ltd.
### Note 1
The Blood Pressure Range of the subject device has minor difference with the blood pressure range of predicate devices, the measuring rang of the subject device was validated according to ISO 80601-2-30. the difference does not raise different questions of safety and effectiveness.
The pulse rate of the subject device is within that of the predicate device. The difference between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
### Note 2
The difference in power source will not affect the safety and effectiveness of the subject device. IEC 60601-1. IEC 60601-1-11 and IEC 60601-1-2 can demonstrate that the subject device can maintain the safety and performance. Thus, this difference does not raise different questions of safety and effectiveness.
### Note 3
All the differences in operation and storage/transportation condition don't affect the safety and effectiveness which is concluded after all the required testing, thus, this difference does not raise different questions of safety and effectiveness.
#### 8. Performance Data
### Non-clinical data
The Arm Blood Pressure Monitor comply with:
1. IEC 60601-1:2005+A1:2012+A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
2. IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
3. IEC 80601-2-30:2018 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers 4. IEC 60601-1-2:2014+A1:2020 Medical electrical equipment-Part1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances-Requirements and tests
5. ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
6. ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
FDA software validation guidance "General Principles of Software Validation; Final Guidance for Industry and FDA Staff, Document issued on: January 11, 2002".
{8}------------------------------------------------
# K241590 Summary
Guangdong Xinyu Electronic and Technology Co., Ltd.
FDA guidance: Content of Premarket Submissions for Device Software Functions, JULY 20, 2023
### Clinical test
The clinical test was conducted to verify that the proposed device met the requirements of the ISO 81060-2 : 2018 . The study population consisted of 120 subjects (55 female and 65 male) with an age range of 12 to 85 years. For systolic and diastolic pressures , the mean error shall be + 5mmHg or less , with a standard deviation of 8mmHg or less, all data is not over the limits. No adverse effect and/or complication is found in this study.
#### 9. Conclusion
Based on the nonclinical and clinical tests, the subject device Arm Blood Pressure Monitor is as safe and as effective as the predicated device(K222926).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.