The Sterilization Package/Reel is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows: · Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 30 minutes. · Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60℃(140°F). Furthermore, the Sterilization Package/Reel maintains the enclosed devices up until 6 months post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the Sterilization Package/Reel has undergone either a steam or EO sterilization process. The Sterilization Package/Reel is offered 3 types in the following: Self -Seal Sterilization Pouch; Flat Sterilization Reel; Gusseted Sterilization Reel.
Device Story
Sterilization packaging system consisting of self-seal pouches, flat reels, and gusseted reels; used by healthcare workers to enclose medical devices prior to sterilization. Input: medical devices placed inside pouch/reel. Transformation: physical enclosure allows sterilant penetration (Steam or EO) while maintaining post-sterilization barrier. Output: sterile-packaged medical devices with external chemical ink indicators (Type 1 Process Indicator) that change color to verify exposure to sterilization process. Indicators change blue-to-black (Steam) or pink-to-yellow (EO). Device maintains sterility for 6 months post-processing. Healthcare providers use indicators to confirm processing; indicators do not signify sterility, only exposure. Benefits: provides effective method for sterilization and maintenance of sterile barrier for clinical use.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via ISO 11135 (EO) and ISO 17665-1 (Steam) achieving SAL 10^-6. Biocompatibility confirmed per ISO 10993-5, -10, -23. Physical testing included burst strength (ASTM F1140), seal strength (ASTM F88), dye penetration (ASTM F1929), and microbial barrier (DIN 58953-6).
Technological Characteristics
Materials: medical dialysis paper, PET/CPP composite film, adhesive, chemical indicator ink. Sensing: Type 1 chemical process indicators (ISO 11140-1:2014). Form factors: self-seal pouches, flat reels, gusseted reels. Sterilization: Steam (132°C) and EO (735 mg/L). Shelf life: 2 years pre-sterilization; 6 months post-sterilization.
Indications for Use
Indicated for healthcare workers to enclose medical devices for sterilization via Steam or Ethylene Oxide (EO) gas. Maintains sterility of enclosed devices for up to 6 months post-sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
SterileRight Sterilization Pouch and Roll (K212338)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. The logo is simple and professional, and it is easily recognizable.
February 24, 2025
Anqing Clean Dental Instrument Technology Co., Ltd. % Wang Boyle General Manager Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai. 200120 China
Re: K241565
Trade/Device Name: Sterilization Package/Reel Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: January 20, 2025 Received: January 24, 2025
Dear Wang Boyle:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Stephen A. Anisko -S
Digitally signed by Stephen A. Anisko -S Date: 2025.02.24 20:03:28 -05'00'
for: Christopher K. Dugard, M.S. Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241565
Device Name Sterilization Package/Reel
#### Indications for Use (Describe)
The Sterilization Package/Reel is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows: · Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 30 minutes.
· Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60℃(140°F).
Furthermore, the Sterilization Package/Reel maintains the enclosed devices up until 6 months post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the Sterilization Package/Reel has undergone either a steam or EO sterilization process.
The Sterilization Package/Reel is offered 3 types in the following: Self -Seal Sterilization Pouch; Flat Sterilization Reel; Gusseted Sterilization Reel.
The following table (Table 1) lists the model numbers of the Sterilization Package/Reel by type, model, dimensions, and content/max. load (lbs.):
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|--------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | |
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | <div> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| | | Specification Size (WXL) | | Content/Max Load(lbs) | | |
|-------------------------------------------|-------------|--------------------------|-----------------|-----------------------|---------|------------------|
| Product Name | Item | Inch | Dimension | Metal | Plastic | Gauze/Linen<br>s |
| | KLDT-90165 | 3.5"x6.5" | 90mmx165mm | 0.12 | 0.11 | 0.1 |
| | KLDT-305430 | 12"x17" | 305mmx430mm | 1.10 | 1.09 | 0.66 |
| | KLDT-70230 | 2.75"x9" | 70mm×230mm | 0.09 | 0.08 | 0.07 |
| | KLDT-135280 | 3⅛"x11" | 135mmx280mm | 0.50 | 0.47 | 0.46 |
| | KLDT-90135 | 3.5"x5.25 | 90mmx135mm | 0.10 | 0.09 | 0.08 |
| Self<br>Sealing<br>Sterilization<br>Pouch | KLDT-70260 | 2.75"x10" | 70mm×260mm | 0.10 | 0.09 | 0.08 |
| | KLDT-90260 | 3.5"x10" | 90mmx260mm | 0.15 | 0.14 | 0.12 |
| | KLDT-135260 | 3⅛"x10" | 135mmx260mm | 0.48 | 0.46 | 0.45 |
| | KLDT-250370 | 10"x14.5" | 250mmx370mm | 1.06 | 1.05 | 0.62 |
| | KLDT-190360 | 7.5"x14" | 190mmx360mm | 0.60 | 0.58 | 0.55 |
| | KLDT-57130 | 2.25"x5" | 57mm×130mm | 0.04 | 0.03 | 0.005 |
| | KLDT-57100 | 2.25"x4" | 57mm×100mm | 0.03 | 0.02 | 0.004 |
| | KLDT-135330 | 3⅛"x13" | 135mmx330mm | 0.55 | 0.53 | 0.50 |
| | KLDT-90230 | 3.5"x9" | 90mmx230mm | 0.14 | 0.13 | 0.11 |
| Flat<br>Sterilization<br>Reel | KLDT-200200 | 8" | 200mm×200m | 0.60 | 0.58 | 0.55 |
| | KLDT-300200 | 12" | 300mm×200m | 1.04 | 1.03 | 0.60 |
| | KLDT-75200 | 3" | 75mm×200m | 0.09 | 0.05 | 0.004 |
| | KLDT-100200 | 4" | 100mm×200m | 0.14 | 0.13 | 0.11 |
| | KLDT-50200 | 2" | 50mm×200m | 0.02 | 0.01 | 0.003 |
| | KLDT-400200 | 16" | 400mm×200m | 1.08 | 1.07 | 0.64 |
| | KLDT-250200 | 10" | 250mm×200m | 0.80 | 0.78 | 0.58 |
| | KLDT-350200 | 14" | 350mm×200m | 1.06 | 1.05 | 0.62 |
| | KLDT-55200 | 2.25" | 55mm×200m | 0.03 | 0.02 | 0.003 |
| | KLDT-150200 | 6" | 150mm×200m | 0.55 | 0.53 | 0.50 |
| Gusseted<br>Sterilization<br>Reel | KLDT-75100 | 3" | 75mm×30mm×100m | 0.09 | 0.05 | 0.004 |
| | KLDT-350100 | 14" | 350mm×80mm×100m | 1.06 | 1.05 | 0.62 |
| | KLDT-150100 | 6" | 150mm×50mm×100m | 0.55 | 0.53 | 0.50 |
| | KLDT-400100 | 16" | 400mm×80mm×100m | 1.08 | 1.07 | 0.64 |
| | KLDT-250100 | 10" | 250mm×65mm×100m | 0.80 | 0.78 | 0.58 |
| | KLDT-300100 | 12" | 300mm×75mm×100m | 1.04 | 1.03 | 0.60 |
| | KLDT-200100 | 8" | 200mm×55mm×100m | 0.60 | 0.58 | 0.55 |
Table 1. The model numbers of Sterilization Package/Reel (Type, Model and Dimension and Content/Max. Load)
{5}------------------------------------------------
| KLDT-100100 | A 11 | 100mm×50mm×100m | CALI | 0 40<br>V. IV | Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara<br>Vill |
|-------------|------|-----------------|------|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{6}------------------------------------------------
# 510(k) Summary K241565
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
#### 1.0 Submitter's Information
| Name: | Anqing Clean Dental Instrument Technology Co.,LTD. | |
|----------|-------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Address: | Floor 2, D4-A,Jiabao Industrial Park.Yangtze River Bridge<br>Economic Development Zone ,246000 Anqing City, Anhui<br>Province,China | |
| Tel: | | 0086-556-5577869 |
| Contact: | | Wu Yiming |
#### Designated Submission Correspondent
| Contact: | Mr. Boyle Wang |
|----------|-----------------------------------------------------------------------|
| Name: | Shanghai Truthful Information Technology Co., Ltd. |
| Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai<br>200120 China |
| Tel: | +86-21-50313932 |
| Email: | Info@truthful.com.cn |
Date Submitted: Jan.20,2025
#### 2.0 Device Information
| Trade/Device name: | Sterilization Package/Reel |
|------------------------------|----------------------------|
| Common name: | Sterilization Package/Reel |
| Classification name: | Sterilization Wrap; |
| Classification Product Code: | FRG,JOJ |
| Regulation number: | 21 CFR880.6850 |
| Classification: | Class II |
| Panel: | General Hospital |
# 3.0 Predicate Device Information
# Predicate Device:
| Manufacturer: | Sterileright Packaging Mfg., Inc. |
|--------------------|-------------------------------------------|
| Trade/Device Name: | SterileRight Sterilization Pouch and Roll |
{7}------------------------------------------------
510(k) number: K212338
# 4.0 Device Description
The medical devices are inserted into the Sterilization Package/Reel,sealed, and then sterilized. After completion of the sterilization process, the Pouch/Roll main- tains the sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 6 months post Steam or EO gas sterilization.
The Self-seal Sterilization Package permits the sealing of the pouch without the need of heat-sealing equipment, while the Heat-Seal Sterilization Reel are heat-sealed prior to processing in the steam/or EO Sterilization.
The chemical indicators ink printed on the "medical dialysis paper" will exhibit a color change after the Package/Reel is exposed to steam or ethylene oxide gas. The color of the Chemical Indicator changes from blue to black when exposed to Steam. And the color changes from pink to yellow, when exposed to EO gas.
The Chemical Indicator which is Type 1 Process Indicator as categorized by ISO 11140-1:2014 offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to Steam/or EO gas.
## 5.0 Indication for Use Statement
The Sterilization Package/Reel is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
· Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 30 minutes.
• Ethylene Oxide (EO) with a concentration of 735 mg/L at 55° C (131° F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C(140°F).
Furthermore, the Sterilization Package/Reel maintains the enclosed devices up until 6 months post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the Sterilization Package/Reel has undergone either a steam or EO sterilization process.
The Sterilization Package/Reel is offered 3 types in the following: Self -Seal Sterilization Pouch; Flat Sterilization Reel; Gusseted Sterilization Reel.
The following table (Table 1) lists the model numbers of the Sterilization Package/Reel by type, model, dimensions, and content/max. load (Ibs.):
{8}------------------------------------------------
| Product Name | Item | Specification Size (WXL) | | Content/Max Load(lbs) | | |
|-------------------------------------------|-------------|--------------------------|-----------------|-----------------------|---------|------------------|
| | | Inch | Dimension | Metal | Plastic | Gauze/Linen<br>s |
| | KLDT-90165 | 3.5"x6.5" | 90mmx165mm | 0.12 | 0.11 | 0.1 |
| | KLDT-305430 | 12"x17" | 305mmx430mm | 1.10 | 1.09 | 0.66 |
| | KLDT-70230 | 2.75"x9" | 70mm×230mm | 0.09 | 0.08 | 0.07 |
| | KLDT-135280 | 3⅛"x11" | 135mmx280mm | 0.50 | 0.47 | 0.46 |
| | KLDT-90135 | 3.5"x5.25<br>" | 90mmx135mm | 0.10 | 0.09 | 0.08 |
| Self<br>Sealing<br>Sterilization<br>Pouch | KLDT-70260 | 2.75"x10" | 70mm×260mm | 0.10 | 0.09 | 0.08 |
| | KLDT-90260 | 3.5"x10" | 90mmx260mm | 0.15 | 0.14 | 0.12 |
| | KLDT-135260 | 3⅛"×10" | 135mmx260mm | 0.48 | 0.46 | 0.45 |
| | KLDT-250370 | 10"x14.5" | 250mmx370mm | 1.06 | 1.05 | 0.62 |
| | KLDT-190360 | 7.5"x14" | 190mmx360mm | 0.60 | 0.58 | 0.55 |
| | KLDT-57130 | 2.25"x5" | 57mm×130mm | 0.04 | 0.03 | 0.005 |
| | KLDT-57100 | 2.25"x4" | 57mm×100mm | 0.03 | 0.02 | 0.004 |
| | KLDT-135330 | 3⅛"×13" | 135mmx330mm | 0.55 | 0.53 | 0.50 |
| | KLDT-90230 | 3.5"x9" | 90mmx230mm | 0.14 | 0.13 | 0.11 |
| | KLDT-200200 | 8" | 200mm×200m | 0.60 | 0.58 | 0.55 |
| | KLDT-300200 | 12" | 300mm×200m | 1.04 | 1.03 | 0.60 |
| | KLDT-75200 | 3" | 75mm×200m | 0.09 | 0.05 | 0.004 |
| | KLDT-100200 | 4" | 100mm×200m | 0.14 | 0.13 | 0.11 |
| Flat<br>Sterilization<br>Reel | KLDT-50200 | 2" | 50mm×200m | 0.02 | 0.01 | 0.003 |
| | KLDT-400200 | 16" | 400mm×200m | 1.08 | 1.07 | 0.64 |
| | KLDT-250200 | 10" | 250mm×200m | 0.80 | 0.78 | 0.58 |
| | KLDT-350200 | 14" | 350mm×200m | 1.06 | 1.05 | 0.62 |
| | KLDT-55200 | 2.25" | 55mm×200m | 0.03 | 0.02 | 0.003 |
| | KLDT-150200 | 6" | 150mm×200m | 0.55 | 0.53 | 0.50 |
| Gusseted<br>Sterilization<br>Reel | KLDT-75100 | 3" | 75mm×30mm×100m | 0.09 | 0.05 | 0.004 |
| | KLDT-350100 | 14" | 350mm×80mm×100m | 1.06 | 1.05 | 0.62 |
| | KLDT-150100 | 6" | 150mm×50mm×100m | 0.55 | 0.53 | 0.50 |
| | KLDT-400100 | 16" | 400mm×80mm×100m | 1.08 | 1.07 | 0.64 |
| | KLDT-250100 | 10" | 250mm×65mm×100m | 0.80 | 0.78 | 0.58 |
| | KLDT-300100 | 12" | 300mm×75mm×100m | 1.04 | 1.03 | 0.60 |
| | KLDT-200100 | 8" | 200mm×55mm×100m | 0.60 | 0.58 | 0.55 |
| KLDT-100100 | 4" | 100mmx50mmx100m | 0.14 | 0.13 | 0.11 | |
Table 1. The model numbers of Sterilization Package/Reel (Type, Model and Dimension and Content/Max. Load)
{9}------------------------------------------------
# 6.0 Summary of Non-Clinical Testing
Summary of non-clinical and performance testing bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The test results demonstrated the subject device is confirmed to be safe and effective for the intended use.
| Test Item | Test Methodology | Acceptance Criteria or End Point | Test Results |
|-------------------------------------------------------------------------------|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Sterilization<br>Process Validation | ISO 11135:2014 | Ethylene oxide: 60 minutes at 55 °C; relative<br>humidity between 50%- 80%;<br>ethylene oxide concentration<br>is 735 mg/L, 8 hours aeration<br>time at 60°C (140°F).<br>SAL=10-6 | SAL=10-6<br>Pass |
| | ISO 17665-1:2006;<br>ISO TS 17665-2:2009 | Steam; 4 minutes at 132°C;<br>30 minutes dry time.<br>SAL=10-6 | SAL=10-6<br>Pass |
| EO/ECH Residuals | ISO 10993-7:2008 | EO < 4mg/d<br>ECH < 9mg/d | Pass |
| Package Integrity /<br>Material<br>Compatibility<br>/Sterility<br>Maintenance | ASTM F 2251-13 | 52μm±10% | Pass |
| | ASTM<br>F1140/F1140M-13 | Burst value > 3.2 Kpa or No Burst | Pass |
| | ASTM F1929-15 | The dye solution is no any<br>leakage across the seal width<br>of sterile barrier system.<br>(No Infiltration) | No Infiltration<br>Pass |
| | ASTM F88/F88M-21 | Seal strength > 2.5 (N/15mm) | Pass |
| | ISO 1924-2:2008 | Tensile strength:<br>CD $\ge$ 2.2KN/m,MD $\ge$ 4.4KN/m | Pass |
| | ASTM F2096-11 | No Leakage | No Leakage<br>Pass |
| | DIN 58953-6 | CFU = 0 | CFU = 0<br>Pass |
| | ISO 5636-3 | 10-50 mL/min·m² | 20-30 mL/min·m²<br>Pass |
#### Table 2: Performance testing summary – Bench
{10}------------------------------------------------
| Chemical Indicator<br>Efficacy Testing | ISO 11140-1:2014 | Steam: | The real time aging test |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 1. Remain stable<br>before use based on<br>its shelf life.<br><br>2. Maintain the<br>endpoint stability of the<br>color change after<br>being in the presence<br>of the sterilant. | The color of CI changes from<br>blue to black, when exposed<br>to Steam;<br><br>EO:<br>The color of CI changes from<br>pink to yellow, when exposed<br>to EO gas. | was carried out that<br>demonstrates: the test<br>device which exposed to<br>Steam maintain the color<br>of black, the test group<br>which exposed to EO<br>maintain the color of<br>yellow, and the real time<br>aging test was carried out<br>on the empty Sterilization<br>Package/Reel that<br>demonstrates the test<br>device maintain the<br>original color within the<br>stated validity period of 2<br>years. |
| Shelf-Life<br>Validation | / | Shelf Life:2 Years;<br>Shelf Life after Sterilized:<br>6 months | Pass |
# Biocompatibility Testing
The biocompatibility evaluation for the Sterilization Package/Reel was conducted in accordance with the FDA's Biocompatibility Guidance "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process"
Table 3 Summary test of Biocompatibility Testing
| Biocompatibility<br>testing | ISO 10993-5:2009 | Non-cytotoxic | Under conditions of the study, did not<br>show potential toxicity to L-929 cells.<br>Pass |
|-----------------------------|-------------------|-----------------|-------------------------------------------------------------------------------------------|
| | ISO 10993-23:2021 | Non-irritating | Under the conditions of the study, not an<br>irritant.<br>Pass |
| | ISO 10993-10:2021 | Non-sensitizing | Under conditions of the study, not a<br>sensitizer.<br>Pass |
{11}------------------------------------------------
# 7.0 Technological Characteristic Comparison Table
| Item | Subject Device<br>K241565 | Predicate Device<br>K212338 | Comparison<br>Analysis |
|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product Name | Sterilization Package/Reel | SterileRight sterilization pouch and roll | ---- |
| Product Code | FRG,JOJ | FRG,JOJ | Same |
| Regulation No. | 21 CFR 880.6850 | 21 CFR 880.6850<br>21 CFR 880.2800 | Same |
| Class | II | II | Same |
| Indication for<br>Use | The Sterilization Package/Reel is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:<br>Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 30 minutes. Ethylene Oxide (EO) with a concentration of 735 mg/L at 55° C (131 ° F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F). Furthermore, the Sterilization Package/Reel maintains the enclosed devices up until 6 months post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the Sterilization Package/Reel has undergone either a steam or EO sterilization process. | The SterileRight Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:<br>Gravity steam at 121°C (250°F) for 30 minutes; Drying time of 25 minutes. Pre-vacuum steam at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F). The SterileRight provide sterilization pouch and roll made with Paper/Film or Tyvek® /Film. The SterileRight sterilization pouch and roll which are made with Paper maintains the sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of | Similar |
| | | | |
| | The Sterilization Package/Reel is<br>offered 3 types in the following:<br>Self -Seal Sterilization Pouch;<br>Flat Sterilization Reel;<br>Gusseted Sterilization Reel. | manufacture. The pouch's<br>external chemical ink indicators<br>are designed to indicate to the<br>user that the pouch has<br>undergone either a steam or EO<br>sterilization process.<br>The SterileRight sterilization<br>pouch and roll which are made<br>with Tyvek® is for EO gas | |
| | | sterilization only. It also maintains<br>the sterility of the enclosed<br>devices for up to 6 months<br>post EO gas sterilization, and | |
| | | before sterilization has<br>a<br>maximum shelf life of 3 years<br>from the date of manufacture. | |
| Material<br>Composition | Composed of medical dialysis<br>paper and transparent colored<br>PET/CPP composite film , EO and<br>Steam Process Indicator, Print Ink,<br>adhesive | Medical Grade Paper, CPP, PET,<br>adhesive,EO and Steam Process<br>Indicator, Print Ink.<br>Tyvek®, PET, PE, adhesive. | Similar |
| Sterilization<br>Cycles | • Steam Sterilization at 132°C<br>(270°F) for 4 minutes; Drying time<br>of 30 minutes.<br>• Ethylene Oxide (EO) with a<br>concentration of 735 mg/L at 55°<br>C (131 ° F) and 50% to 80%<br>relative humidity for 60 minutes.<br>Aeration time of 8 hours at 60°C<br>(140°F). | • Gravity steam at 121°C (250°F)<br>for 30 minutes; Drying time of 25<br>minutes.<br>• Pre-vacuum steam at 132°C<br>(270°F) for4 minutes; Drying time<br>of 20 minutes.<br>• Ethylene Oxide (EO) with a<br>concentration of 735 mg/L at<br>55°C (131°F) and 50% to 80%<br>relative humidity for 60 minutes.<br>Aeration time of 8hours at 60°C<br>(140°F). | Similar |
| Configuration/<br>Dimension | Various Size, Heat Sealing and<br>Self Sealing | Various Size, Heat Sealing and<br>Self Sealing | Same |
| Sterilant<br>Penetration<br>Efficacy | The test meet the requirement of<br>SAL 10-6 | The test meet the requirement of<br>SAL 10-6 | Same |
| Chemical<br>Indicator<br>(CI)<br>Functionality<br>and Endpoint | The sterilant penetrated through<br>the pouch configuration and<br>affected the CI color change to<br>the endpoint color | The sterilant penetrated through<br>the pouch configuration and<br>affected the CI color change to the<br>endpoint color | Same |
| Device Design<br>of Steam Cl | The color of Chemical Indicator<br>changes from Blue to | The color of Chemical Indicator<br>changes from Pink to | Similar |
| | Black, when exposed to Steam | Brown/Black, when exposed to Steam | |
| Device Design of EO gas CI | The color of Chemical Indicator changes from Pink to Yellow, when exposed to Steam | The color changes from Blue to Yellow/Brown, when exposed to EO gas | Similar |
| Thickness Variations (mm) ASTM F 2251 | Passed | Passed | Same |
| Tensile strength ISO 1924-2 | Passed | Passed | Same |
| Burst Strength (kPa) ASTM F1140 | Passed | Passed | Same |
| Bubble Leak Test ASTM D 3078 or ASTM-F 2096 | Passed | Passed | Same |
| Seal Peel Test (N/15mm) ASTM F88/F88M ; ISO 11607-1 | Passed…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.