The Extremity Medical External Fixation System is indicated for adults and pediatric patients (2 years old and older) for treatment of: Open and closed fractures Osteotomies Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis, infected union, non-union, or malunion of long bones Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction Correction of bony or soft tissue deformity (orthoplastic surgery) Joint arthrodesis Infected fractures Correction of segmental bony or soft tissue defects Management of comminuted intra-articular fractures Bone transport Revision procedures where other treatments or devices have been unsuccessful Bone reconstruction procedures Charcot foot reconstruction Offloading and/or immobilization of ulcers and/or wounds of the foot and ankle Ankle distraction (arthrodiastasis) Septic fusion
Device Story
External fixation system; provides framework for bone/joint stabilization. Components: bridge elements (anodized aluminum rings), connection elements (stainless steel/aluminum threaded rods, pillars, struts), bone anchorage elements (stainless steel wires, half pins). Used in clinical settings by surgeons for orthopedic procedures including fracture fixation, limb lengthening, and deformity correction. Assembly involves connecting circular rings via struts/rods; wires/pins fixate bone to external framework. Walking attachments allow weight-bearing. System provides mechanical stability to bone segments; facilitates healing/correction. Benefits: enables complex reconstruction, stabilization of fractures, and limb lengthening.
Clinical Evidence
No clinical testing was performed. Substantial equivalence established via engineering analysis and static/dynamic construct testing in accordance with ASTM F1541-17.
Indicated for adults and pediatric patients (≥2 years) for treatment of fractures, osteotomies, joint contractures, pseudoarthrosis, non-union/malunion, limb lengthening, deformity correction, joint arthrodesis, bone transport, Charcot foot reconstruction, and offloading/immobilization of foot/ankle ulcers.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Smith & Nephew Ilizarov External Fixator (K201253)
Paragon 28 Monkey Rings External Fixation System (K232838)
Depuy Synthes Maxframe Multi-Axial Correction System (K161417)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below.
February 21, 2025
Extremity Medical, LLC Mary Hoffman Director of Quality Assurance and Regulatory Affairs 300 Interpace Parkway Building A. 2nd Floor Parsippany, New Jersey 07054
Re: K241563
Trade/Device Name: Extremity Medical External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: May 31, 2024 Received: January 22, 2025
Dear Mary Hoffman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Lixin Liu -S
Lixin Liu, Ph.D Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known) K241563
#### Device Name
Extremity Medical External Fixation System
#### Indications for Use (Describe)
The Extremity Medical External Fixation System is indicated for adults and pediatric patients (2 years old and older) for treatment of:
- Open and closed fractures
- · Osteotomies
· Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction
- · Pseudoarthrosis, infected union, non-union, or malunion of long bones
- · Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction
- · Correction of bony or soft tissue deformity (orthoplastic surgery)
- · Joint arthrodesis
- · Infected fractures
- · Correction of segmental bony or soft tissue defects
- · Management of comminuted intra-articular fractures
- · Bone transport
- · Revision procedures where other treatments or devices have been unsuccessful
- Bone reconstruction procedures
- · Charcot foot reconstruction
- · Offloading and/or immobilization of ulcers and/or wounds of the foot and ankle
- · Ankle distraction (arthrodiastasis)
- · Septic fusion
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">✖</span> |
|----------------------------------------------|-------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K241563 (Page 1 of 3)
## Traditional 510(k) Summary:
### External Fixation System
| Submitter | Extremity Medical, LLC<br>300 Interpace Parkway<br>Building A, 2nd Floor<br>Parsippany, NJ 07054 |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Mary Hoffman, MS<br>Director, Quality Assurance and Regulatory Affairs<br>Phone: (973) 588-8980 ext. 502<br>Email: mhoffman@extremitymedical.com |
| Date Prepared | February 20, 2025 |
| Trade Name | Extremity Medical External Fixation System |
| Classification<br>Name and<br>Number | 21 CFR 888.3030 - Single/multiple component metallic bone fixation appliances and<br>accessories |
| Product Code | KTT (Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component) |
| Primary<br>Predicate | K201253- Smith & Nephew Ilizarov External Fixator |
| Additional<br>Predicates | K232838 - Paragon 28 Monkey Rings External Fixation System<br>K161417 - Depuy Synthes Maxframe Multi-Axial Correction System |
| Device<br>Description | The Extremity Medical External Fixation System consists of three basic types of elements:<br>bridge elements, connection elements and bone anchorage elements. The system is intended<br>for use in a variety of applications including but not limited to Charcot foot reconstruction,<br>open fracture fixation, joint arthrodesis, limb lengthening and bone transport. The assembly<br>is made up of two or more circular rings that are connected by struts, rods or pillars to<br>provide a framework for fixation. Pins and tensioned wires are then used to fixate and<br>stabilize bones and joints.<br><br>Bridge elements include rings, manufactured from anodized aluminum, in various types and<br>sizes to accommodate varying patient anatomy. Connection elements, manufactured from<br>stainless steel and anodized aluminum, include threaded rods and threaded pillars available<br>in multiple lengths and struts available in multiple sizes. Bone anchorage elements include<br>wires and half pins, manufactured from stainless steel, available in multiple diameters and<br>lengths.<br><br>All External Fixation components are for single use only. |
| Indications for<br>use | The Extremity Medical External Fixation System is indicated for adults and pediatric<br>patients (2 years old and older) for treatment of:<br>Open and closed fractures Osteotomies Fractures and disease which generally may result in joint contractures or loss of<br>range of motion and fractures requiring distraction Pseudoarthrosis, infected union, non-union, or malunion of long bones Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction Correction of bony or soft tissue deformity (orthoplastic surgery) Joint arthrodesis Infected fractures Correction of segmental bony or soft tissue defects Management of comminuted intra-articular fractures Bone transport Revision procedures where other treatments or devices have been unsuccessful Bone reconstruction procedures Charcot foot reconstruction Offloading and/or immobilization of ulcers and/or wounds of the foot and ankle Ankle distraction (arthrodiastasis) Septic fusion |
| Statement of<br>Technological<br>Comparison | The Extremity Medical External Fixation System and predicate devices are substantially<br>equivalent in terms of design, material, mechanical properties and indications for use.<br>The subject and predicate are based on the following same technological elements:<br>Implants (wires and half pins) are used temporarily to stabilize bones and/or joints<br>and are fixed to external ring components via hardware. Bridge elements are used external to the body and are fixed to bone via implants. Connection elements (threaded rods, threaded pillars and struts) are used to connect<br>rings to other rings. Connection elements (hardware) are used to fix wires to rings, fix half pins to rings,<br>connect struts/rods/pillars to rings, and to connect walking attachments to rings. Walking attachment components are connected to the bottom of foot rings via<br>hardware which are used for weight bearing and walking purposes. All subject and predicate devices are made of the same materials (Stainless Steel,<br>Aluminum 6061-T6, Plastic (Delrin, Radel), Rubber) and thus have equivalent<br>mechanical properties. Key differences in design of the subject device compared to the predicate are as follows: Subject bridge elements (rings) provide a double row of holes versus" tabbed"<br>mounting options, along with olive wire mounting options. Subject bridge elements (foot rings) include tapped holes for the option to directly<br>attach bolts while also allowing for the combination of nuts and bolts to be used. They also include an anatomic option to match the foot shape. |
| Subject accessories provide options for foot rocker or sole style versus “rocker ring” or “foot ring” only. These differences in design do not introduce different issues of safety or effectiveness. | |
| Non-clinical<br>Testing | The Extremity Medical External Fixation System components were compared to the predicates by engineering analysis and static and dynamic construct testing in accordance with ASTM F1541-17.<br><br>The results of this analysis indicate that the Extremity Medical External Fixation System is substantially equivalent to predicate devices and does not introduce different issues of safety or effectiveness. |
| Clinical Testing | No clinical testing was performed. |
| Conclusion | The Extremity Medical External Fixation System is substantially equivalent to its predicate devices. This conclusion is based upon indications for use, principles of operation, design, engineering analysis, and static and dynamic construct testing. |
{5}------------------------------------------------
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.